K091433 · R&D Systems, Inc. · JPK · Oct 28, 2009 · Hematology
Device Facts
Record ID
K091433
Device Name
R&D SYSTEMS TETIC-I PLUS HEMOTOLOGY CONTROL
Applicant
R&D Systems, Inc.
Product Code
JPK · Hematology
Decision Date
Oct 28, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 864.8625
Device Class
Class 2
Indications for Use
R&D Retic-I Plus Control is a tri-level, assayed hematology control designed to monitor values obtained from automated reticulocyte counting methods.
Device Story
R&D Systems Retic-I Plus Hematology Control is an in vitro diagnostic reagent; composed of human and avian erythrocytes in plasma-like medium with preservatives. Device provides stabilized material to monitor automated reticulocyte counting methods; sampled identically to patient specimens. Available in three levels (Low, Normal, High) in 4 mL tubes. Stored at 2°–8°C. Used by laboratory personnel to verify instrument performance; output is a control value compared against established ranges to ensure accuracy of patient testing.
Clinical Evidence
Bench testing only. Stability studies (open vial 14 days, closed vial 75 days) confirmed performance within established acceptance criteria (mean ± 0.20 for Level 1; mean ± 0.10 for Levels 2 and 3). Precision/reproducibility data provided for three lots across three levels, showing CV% ranging from 0.71% to 7.58%. Value assignment validated using Coulter LH750 and GENS instruments.
Technological Characteristics
Tri-level assayed hematology control; human and avian erythrocytes in plasma-like medium with preservatives. 4 mL tube packaging. Storage 2°–8°C. Analytes: IRF, Retic%, Retic#, RBC, MCVr. No electronic or software components; purely chemical/biological reagent.
Indications for Use
Indicated for use as an in vitro diagnostic hematology control to monitor automated reticulocyte counting methods. No specific patient population or contraindications described.
Regulatory Classification
Identification
A hematology quality control mixture is a device used to ascertain the accuracy and precision of manual, semiautomated, and automated determinations of cell parameters such as white cell count (WBC), red cell count (RBC), platelet count (PLT), hemoglobin, hematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC).
Special Controls
*Classification.* Class II (special controls). Except when intended for use in blood components, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
R&D Systems Advia Retic Plus Hematology Control (K010461)
Related Devices
K130962 — R&D 5D RETIC HEMATOLOGY CONTROL · R&D Systems, Inc. · Jul 24, 2013
K993496 — RETIC-QUINOX · Hematronix, Inc. · Nov 30, 1999
K100050 — R & D SYSTEMS XERET HEMATOLOGY CONTROL · R&D Systems, Inc. · Jan 29, 2010
K042587 — RETIC-CHEX FOR CELL-DYN · Streck Laboratories, Inc. · Oct 19, 2004
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k091433
B. Purpose for Submission:
To obtain clearance for the Retic-I Plus Hematology Control
C. Measurand:
IRF and Absolute Reticulocyte
D. Type of Test:
Assayed controls; automated cell counter
E. Applicant:
R&D Systems, Inc.
F. Proprietary and Established Names:
R&D Systems Retic-I Plus Hematology Control
G. Regulatory Information:
1. Regulation section:
21 CFR 864.8625; Hematology quality control mixture
2. Classification:
Class II
3. Product code:
JPK; Mixture, Hematology Quality Control
4. Panel:
Hematology 81
H. Intended Use:
1. Intended use(s):
R&D Retic-I Plus Control is a tri-level, assayed hematology control designed to monitor values obtained from automated reticulocyte counting methods.
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
Not Applicable
4. Special instrument requirements:
Not Applicable
I. Device Description:
R&D Systems Retic-I Plus Hematology Control is an in vitro diagnostic reagent composed of human and avian erythrocytes in a plasma-like medium with preservatives. It is composed of stabilized materials that provide a means of monitoring automated reticulocyte counting methods. It is sampled in the same manner as a patient specimen. There are three different levels (Low-level 1, Normal-level 2 and High-level 3) available and each control is packaged in a tube containing 4 mL of the control material. The product must be stored at 2° – 8°C.
J. Substantial Equivalence Information:
1. Predicate device name(s):
R&D Systems Advia Retic Plus Hematology Control
2. Predicate K number(s):
{1}
k010461
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | R&D Retic-I Plus Control is a tri-level, assayed hematology control designed to monitor values obtained from automated reticulocyte counting methods. | Same |
| Open Vial Claim | 2° – 8°C (14 days) | 2° – 8°C (14 days) |
| Closed Vial Claim | 2° – 8°C (75 days) | 2° – 8°C (75 days) |
| Analytes | Retic%, MCVr, RBC | Retic%, MRV, RBC |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Analytes | Immature Reticulocyte Fraction (IRF), Retic# (Retic x 10^{6}/uL) | MCVg, CHCMg, CHCMr, CHg, CHr |
*MCVr = MRV, mean corpuscular volume of reticulocytes
K. Standard/Guidance Document Referenced (if applicable):
Points to consider: Guidance for Industry and FDA Staff - Assayed and Unassayed Quality Control Material, February 3, 1999.
L. Test Principle:
Controls are comprised of stabilized materials that provide a means of monitoring the performance of reticulocyte counting methods in accordance with established laboratory practice to monitor the performance of diagnostic tests. It is sampled in the same manner as a patient specimen.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Reproducibility: Reproducibility data was taken from the stability data. Three levels of control material were tested for a minimum of 3 times/day over a 3 to 5 day period. The CV% for each lot and level are reflected below:
| | Level 1 | Level 2 | Level 3 |
| --- | --- | --- | --- |
| Lot 1 | 6.14 | 0.93 | 0.84 |
| Lot 2 | 5.95 | 0.75 | 0.75 |
| Lot 3 | 7.58 | 0.97 | 0.71 |
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Open Vial Stability: To validate open vial stability one vial from each lot and each level were run during a 14 day period. The ranges established for acceptance criteria were Level 1 mean ± 0.20, Level 2 mean ± 0.10 and Level 3 mean ± 0.10. To meet acceptance criteria, all results were to be within the
{2}
range throughout the 14 days of testing. All results were within range and no significant changes in mean or variability were observed.
Value Assignment was determined by analyzing an average of the minimum and maximum values of three lots run on the Coulter LH750 and GENS. This method was utilized because all testing was performed on in-house instruments.
d. Detection limit:
Not Applicable
e. Analytical specificity:
Not Applicable
f. Assay cut-off:
Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
Not Applicable
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
Not Applicable
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.