UA-Cellular™ for IQ is an assayed cellular urine control for evaluating the accuracy and precision of automated procedures that measure urinary sediment parameters.
Device Story
UA-Cellular™ for IQ is a quality control product for the IRIS IQ 200 urinalysis analyzer. It consists of stabilized human red blood cells, white blood cells, and inert particles (non-squamous epithelial cells and crystals) in a preservative medium. The device is provided in two levels, packaged in 120ml plastic bottles. Laboratory personnel use the control by processing it like a patient sample on the IRIS IQ 200 analyzer. The instrument measures the cellular parameters, and the resulting values are compared against the expected ranges provided in the assay sheet. This process confirms the precision and accuracy of the instrument's operation, ensuring reliable clinical results for patient samples.
Clinical Evidence
Bench testing only. Studies included run-to-run reproducibility, open vial stability (30 days), and closed vial stability (60 days). Results confirmed lot-to-lot consistency and performance stability for the claimed shelf life.
Technological Characteristics
Stabilized human red/white blood cells, non-squamous simulated epithelial cells, and crystal components in preservative medium. Packaged in 120ml plastic bottles with polypropylene screw caps and polyethylene liners. Storage: 2-10°C. No electronic or software components.
Indications for Use
Indicated for use as an assayed cellular urine control to evaluate the accuracy and precision of automated procedures measuring urinary sediment parameters in clinical laboratory settings.
Regulatory Classification
Identification
A hematology quality control mixture is a device used to ascertain the accuracy and precision of manual, semiautomated, and automated determinations of cell parameters such as white cell count (WBC), red cell count (RBC), platelet count (PLT), hemoglobin, hematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC).
Special Controls
*Classification.* Class II (special controls). Except when intended for use in blood components, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
K170091 — UA-Cellular Complete · Streck · Feb 9, 2017
K131444 — UA-CELLULAR COMPLETE · Streck · Jan 29, 2014
K051706 — IQ BODY FLUIDS CONTROL · Streck · Aug 17, 2005
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
K090201
B. Purpose for Submission:
New Device
C. Measurand:
White Blood Cells (WBC), Red Blood Cells (RBC), Non-squamous Epithelial Cells, Crystals
D. Type of Test:
Quantitative
E. Applicant:
Streck, Inc.
F. Proprietary and Established Names:
UA-Cellular™ for IQ
G. Regulatory Information:
1. Regulation section:
21 CFR 864.8625, Hematology Quality Control Mixture
2. Classification:
Class II
3. Product code:
JPK, Mixture Hematology Quality Control
{1}
4. Panel:
81 (Hematology)
H. Intended Use:
1. Intended use(s):
UA-Cellular™ for IQ is an assayed cellular urine control for evaluating the accuracy and precision of automated procedures that measure urinary sediment parameters.
2. Indication(s) for use:
UA-Cellular™ for IQ is an assayed cellular urine control for evaluating the accuracy and precision of automated procedures that measure urinary sediment parameters.
3. Special conditions for use statement(s):
Not applicable.
4. Special instrument requirements:
For use on the IRIS IQ 200 analyzer.
I. Device Description:
UA-Cellular™ for IQ is a urinalysis control which contains stabilized human red and white blood cells and other inert particles in a preservative medium. UA-Cellular™ for IQ is a urine control for the IRIS IQ 200 analyzer. The product is packaged in plastic bottles containing 120ml. The closures are propropylene screw caps with polyethylene liners. There are two different levels; level 1 and level 2. The bottles are packaged in a box with the package insert/ assay sheet. The product is stored at 2-10° C.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Cell-Chex Auto
2. Predicate K number(s):
K053362
{2}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | UA-CellularTM for IQ | Cell-Chex Auto |
| Intended use | Used as an assayed cellular urine control for evaluating the accuracy and precision of automated procedures that measure urinary sediment parameters. | Used as an assayed whole blood control for evaluating the accuracy and precision of hematology instruments that measure blood cell counts in patient body fluid samples. |
| Reagent composition | Stabilized human red and white blood cells combined with non-squamous simulated epithelial and crystal components. | Stabilized human red and white blood cells. |
| Open vial stability | 30 days | Same |
| Closed vial stability | 60 days | 75 days |
| Storage temperature | 2 – 10°C | Same |
K. Standard/Guidance Document Referenced (if applicable):
Not Applicable.
L. Test Principle:
Daily use of this whole blood control provides quality control data for confirming the precision and accuracy of instrument operation. Use of stabilized cell preparations for controlling laboratory testing protocols is an established procedure. When handled like a patient sample and assayed on a properly calibrated and functioning instrument the whole blood control will provide values within the expected range indicated on the assay sheet.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Run-to-Run reproducibility studies were performed on three lots of each level. Each value was calculated from 10 consecutive analyses performed on a single vial of each. Reproducibility is expressed as a CV%, which is calculated by dividing the standard deviation by the mean value and multiplying by 100.
{3}
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
**Value assignment:** Assay values are assigned to each lot by replicate analysis. A minimum of three vials per level are tested on the IRIS IQ 200. Each vial is tested for a minimum of three test events performed on different dates. Off-site (referee) laboratories are used to supplement data collection for in-house instrumentation or provide unique system specific data. Referees are requested to run each vial in duplicate.
A ten run reproducibility event is run using a vial from each level on the IRIS IQ 200 in addition to the test listed above.
The data is entered into the validated QC link database program which calculates mean, standard deviation, and coefficient of variation for each parameter analyzed. Final assay assignment values are determined using the data collection and established product performance characteristics.
**Open vial stability:** Level 1 and Level 2 from three lots were run 5 times during the 30 day period to establish open vial stability. On the week days the vials were not run they were brought to room temperature, mixed and opened. Each vial was returned to refrigerated storage (2-10°C) between analysis periods. All lots performed as expected, with parameter recovery within the established assay ranges.
**Closed vial stability:** Level 1 and Level 2 from three lots were run once per week from the ship date to the expiration date. The vials were stored in a monitored refrigerator between 2-10°C between analysis periods. No significant trends occurred and there was consistent recovery of values within the indicated assay range.
d. Detection limit:
Not applicable.
e. Analytical specificity:
Not applicable.
{4}
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
The mean and expected assay values provided for each parameter are derived from replicate analyses on calibrated instruments. The assay values are obtained using reagents recommended by instrument manufactures and are to be used for instrument control (they are not absolute assay for calibration). Upon receipt of a new control lot, it is recommended that an individual laboratory establish its own mean and limits of each parameter. However, the control mean established by the laboratory should fall within the expected range specified for the control. The expected ranges listed represent estimates of variation due to different laboratories instrument calibration, maintenance, and operator technique.
{5}
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
6
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.