METER TRAX TRILEVEL, LOW LEVEL, MID LEVEL, HIGH LEVEL TRILEVEL MINIPAK
K042336 · Bio-Rad Laboratories, Inc. · JPK · Oct 14, 2004 · Hematology
Device Facts
Record ID
K042336
Device Name
METER TRAX TRILEVEL, LOW LEVEL, MID LEVEL, HIGH LEVEL TRILEVEL MINIPAK
Applicant
Bio-Rad Laboratories, Inc.
Product Code
JPK · Hematology
Decision Date
Oct 14, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.8625
Device Class
Class 2
Indications for Use
Meter Trax is an assayed whole blood control used in monitoring the precision of laboratory testing procedures for hemoglobin, hematocrit and glucose.
Device Story
Meter Trax is a tri-level (low, medium, high) liquid control; consists of stabilized human red blood cell components, animal-derived constituents, chemicals, stabilizers, and preservatives. Used in clinical laboratory settings to monitor precision of testing procedures for hemoglobin, hematocrit, and glucose. Healthcare providers use the control to verify instrument performance and ensure accuracy of patient test results.
Clinical Evidence
No clinical data. Bench testing only. Stability studies confirmed open-vial stability (31 days at 2-8°C for hematocrit; 31 days at 2-8°C or 18-30°C for glucose and hemoglobin) and shelf-life stability (180 days at 2-8°C).
Technological Characteristics
Tri-level liquid suspension; human red blood cell components; includes chemical stabilizers and preservatives. No specific ASTM standards cited. Standalone control product.
Indications for Use
Indicated for use as an assayed whole blood control to monitor the precision of laboratory testing procedures for glucose, hemoglobin, and hematocrit.
Regulatory Classification
Identification
A hematology quality control mixture is a device used to ascertain the accuracy and precision of manual, semiautomated, and automated determinations of cell parameters such as white cell count (WBC), red cell count (RBC), platelet count (PLT), hemoglobin, hematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC).
Special Controls
*Classification.* Class II (special controls). Except when intended for use in blood components, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
K042336
B. Purpose for Submission:
To seek clearance of a new whole blood hematology/glucose control.
C. Measurand:
Hemoglobin, hematocrit and glucose
D. Type of Test:
Assayed whole blood control
E. Applicant:
Bio-Rad Laboratories
F. Proprietary and Established Names:
Meter Trax Control
G. Regulatory Information:
1. Regulation section:
21 CFR 864.8625
21 CFR 862.1660
2. Classification:
Class II
3. Product code:
JPK
JJY
{1}
4. Panel:
81 Hematology
H. Intended Use:
1. Intended use(s):
Meter Trax is an assayed whole blood control used in monitoring the precision of laboratory testing procedures for hemoglobin, hematocrit and glucose.
2. Indication(s) for use:
3. Special conditions for use statement(s):
4. Special instrument requirements:
I. Device Description:
Meter Trax is a tri-level (low, medium, high) liquid suspension of stabilized human red blood cell components with added constituents of animal origin, chemicals, stabilizers, and preservatives.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Meter Trax
Bio-Rad Laboratories (formerly Hematronix, Inc.)
2. Predicate 510(k) number(s):
K904461
3. Comparison with predicate:
{2}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Matrix | Human whole blood based | same |
| Form | Liquid | same |
| Analytes | Hemoglobin, hematocrit, glucose | same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | Meter Trax is an assayed whole blood control intended for use in monitoring the precision of laboratory testing procedures for glucose, hemoglobin, and hematocrit. | Meter Trax™ is intended for use as a whole blood reference control for glucose, hemoglobin, and hematocrit. |
| Stabilizers | Contains stabilizers | Does not contain stabilizers |
# K. Standard/Guidance Document Referenced (if applicable):
# L. Test Principle:
# M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
b. Linearity/assay reportable range:
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
d. Detection limit:
e. Analytical specificity:
f. Assay cut-off:
{3}
2. Comparison studies:
a. Method comparison with predicate device:
b. Matrix comparison:
3. Clinical studies:
a. Clinical Sensitivity:
b. Clinical specificity:
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
5. Expected values/Reference range:
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches