STAK-CHEX PLUS RETICS

K090137 · Streck · GLQ · Mar 18, 2009 · Hematology

Device Facts

Record IDK090137
Device NameSTAK-CHEX PLUS RETICS
ApplicantStreck
Product CodeGLQ · Hematology
Decision DateMar 18, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.8625
Device ClassClass 2

Intended Use

STaK-Chex Plus Retics is an assayed whole blood control for evaluating the accuracy and precision of automated, semi-automated and manual procedures that measure blood cell parameters.

Device Story

STaK-Chex Plus Retics is a stabilized whole blood control product used in clinical laboratories to verify the accuracy and precision of automated hematology analyzers. The device consists of a suspension of human red blood cells, a nucleated red blood cell (NRBC) analog, a white blood cell component (human and non-human analogs), and a platelet component (non-human analog) in a preservative medium. It is provided in three levels (low, normal, high) in 4.5ml plastic vials. Laboratory personnel process the control identically to a patient sample on a calibrated hematology instrument. The instrument measures the control, and the resulting values are compared against the expected ranges provided on the product assay sheet. This process allows clinicians to confirm that the instrument is functioning correctly, thereby ensuring the reliability of patient test results.

Clinical Evidence

Bench testing only. Studies included closed vial stability, open vial stability, run-to-run reproducibility, and external site recovery of values. Results demonstrated consistent reproducibility and stability for the claimed shelf life.

Technological Characteristics

Stabilized suspension of human/animal blood components in preservative medium. Vials: 5ml plastic with polypropylene screw caps/polyethylene liners. Storage: 2-10°C. No electronic, software, or connectivity components.

Indications for Use

Indicated for use as an assayed whole blood control to evaluate accuracy and precision of automated, semi-automated, and manual blood cell parameter measurement procedures in clinical laboratory settings.

Regulatory Classification

Identification

A hematology quality control mixture is a device used to ascertain the accuracy and precision of manual, semiautomated, and automated determinations of cell parameters such as white cell count (WBC), red cell count (RBC), platelet count (PLT), hemoglobin, hematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC).

Special Controls

*Classification.* Class II (special controls). Except when intended for use in blood components, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Ko90137 # 510(k) Summary of Safety and Effectiveness | 510(k) Submitter: | Streck<br>7002 South 109th Street<br>Omaha, NE 68128 | MAR 1 8 2009 | |------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|--------------| | | Official Correspondent: Carol Thompson, Quality Assurance Manager<br>(402)-537-5213 | | | Date Prepared: | January 19, 2009 | | | Name of Device:<br>Trade Name:<br>Common Name:<br>Classification Name: | STaK-Chex® Plus Retics<br>Assayed Hematology Control<br>Hematology quality control mixture (864.8625) | | | Predicate Device: | STaK-Chex Plus Retics (K992887) Manufactured by Streck | | ## Description: STaK-Chex Plus Retics is a stabilized suspension of human red blood cells, a nucleated red blood cell analog, a white blood cell component consisting of human analogs and a platelet component consisting of a non-human analog in a preservative medium. The product is packaged in plastic vials containing 5ml. The closures are polypropylene screw caps with polyethylene liners. There are three different levels (low, normal and high). The vials will be packaged in a six (6) or twelve (12) welled vacuum formed "clam-shell" container with the package insert / assay sheet. The product must be stored at 2 - 10° C. ## Intended Use: STaK-Chex Plus Retics is an assayed whole blood control for evaluating the accuracy and precision of automated, semi-automated and manual procedures that measure blood cell parameters. | | STaK-Chex Plus Retics (Predicate Product) | STaK-Chex Plus Retics | |---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use<br>Statement | STak-Chex Plus Retics is intended<br>to be used as a control for complete<br>blood cell count (CBC), white cell 5-<br>part differential, and reticulocyte<br>parameters on Beckman Coulter<br>GenS series hematology<br>instruments. | STaK-Chex Plus Retics is an<br>assayed whole blood control for<br>evaluating the accuracy and<br>precision of automated, semi-<br>automated and manual procedures<br>that measure blood cell<br>parameters. | | Open Vial Stability | 14 days | Same | | Closed Vial Stability | 105 days | 75 days | | Reagents | Stabilized human and animal blood | Low Level - Stabilized human and<br>animal blood, plus a nucleated red<br>blood cell analog<br>Normal and High Levels same as<br>Predicate | | Storage Conditions | 2 - 10°C | Same | # Comparison to Predicate Device: l of 2 {1}------------------------------------------------ # Discussion of Tests and Test Results: Four types of studies were conducted to establish performance of STaK-Chex Plus Retics. The four tests conducted were Closed Vial Stability, Open Vial Stability, Run to Run Reproducibility, and External Site recovery of values. All testing showed that STaK-Chex Plus Retics is consistently reproducible, substantially equivalent to the predicate product and stable for the shelf life claimed. # Conclusions Drawn From Tests: Study results show STak-Chex Plus Retics to be consistently reproducible, substantially equivalent to the products, and stable for the entire product dating. STaK-Chex Plus Retics is a safe and effective product, which fulfills its intended use when used as instructed in the product package insert. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread, clutching a caduceus in its talons. The eagle is facing to the right. #### Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 # MAR 1 8 2009 Streck, Inc. c/o Ms. Kerrie Oetter Quality Assurance Coordinator 7002 South 109th St. Omaha, Nebraska 68128 Re: k090137 Trade/Device Name: STaK-Chex ® Plus Retics. Regulation Number: 21 CFR 864.8625 Regulation Name: Hematology Quality Control mixture Regulatory Class: II Product Code: GLQ Dated: March 4, 2009 Received: March 10, 2009 Dear Ms. Oetter: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the [kit/tray] have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/trav. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not {3}------------------------------------------------ Page 2 - Ms. Kerrie Oetter limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, pennits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240- 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For question regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR), please contact the Division of Surveillance Systems at 240-276-3464. You many obtain other general information on your responsibilities under the Act from the Division of Small Manufactuers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, ia m chan Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indication for Use 510(k) Number (if known): # Device Name: STaK-Chex Plus Retics Indication For Use: STaK-Chex Plus Retics is an assayed whole blood control for evaluating the accuracy and precision of automated procedures that measure blood cell parameters. Prescription Use X (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Stephine Bautista Division Sign-Off Office of In Vito Diagnostic Device Evaluation and Safety 510(k) K090137
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...