Eurotrol Hct Control is an assayed hematocrit reference material, to verify the precision and accuracy of the epoc® Blood Analysis System for the measurement of hematocrit. Eurotrol Hct Control is for in vitro diagnostic use only. epoc® Hematocrit Verification Fluids is an assayed hematocrit reference material, to verify the precision and accuracy of the epoc® Blood Analysis System for the measurement of hematocrit. epoc® Hematocrit Verification Fluids is for in vitro diagnostic use only.
Device Story
Eurotrol Hct Control and epoc® Hematocrit Verification Fluids are electrolyte solutions used as quality control materials for the epoc® Blood Analysis System. These fluids simulate clinically relevant hematocrit concentrations via conductivity measurements. The device is used by healthcare professionals to verify the precision and accuracy of the epoc® system's hematocrit channel. The fluids are provided in 2.5 mL glass ampules at multiple concentration levels. The healthcare provider introduces the fluid into the epoc® system; the system measures conductivity and converts it to a hematocrit percentage. This verification process ensures the reliability of patient hematocrit results, supporting clinical decision-making regarding patient blood status.
Clinical Evidence
No clinical data. Bench testing only. Precision/reproducibility studies were conducted across multiple sites and epoc® systems using 3 lots of control material over 20 days. Total variance (CV%) ranged from 1.89% to 4.95% across five levels, meeting the acceptance criterion of ≤9.0% CV. Stability was validated for 3 months (closed-vial) and 15 minutes (open-ampule) using measurand drift analysis.
Technological Characteristics
Electrolyte solutions in a physiologically buffered matrix; no red cells or biological materials. Conductivity-based sensing principle. Dimensions: 2.5 mL glass ampules. Storage: 15-30°C. Connectivity: None (manual sample introduction). Sterilization: Not specified. Software: None (device is a chemical reference material).
Indications for Use
Indicated for use as an assayed hematocrit reference material to verify the precision and accuracy of the epoc® Blood Analysis System for the measurement of hematocrit in a clinical setting. For prescription use only.
Regulatory Classification
Identification
A hematology quality control mixture is a device used to ascertain the accuracy and precision of manual, semiautomated, and automated determinations of cell parameters such as white cell count (WBC), red cell count (RBC), platelet count (PLT), hemoglobin, hematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC).
Special Controls
*Classification.* Class II (special controls). Except when intended for use in blood components, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
RNA Medical QC 900 Hematocrit Control (k955630)
Related Devices
K021924 — MEDICA EASYQC HEMATOCRIT CONTROL · Bionostics, Inc. · Jun 24, 2002
K020618 — AALTO SCIENTIFIC MIRCO HEMATOCRIT CONTROL · Aalto Scientific, Ltd. · May 14, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
K 121467
Eurotrol B.V. Eurotrol Hct Control and epoc Hematocrit Verification Fluids May 10, 2013
### MAY 1 0 2013
### 510(k) Summary
Submitter:
Eurotrol B.V. Keplerlaan 20 6716 BS Ede, The Netherlands +31 318 695777 (Telephone) +31 318 695770 (FAX)
Contact:
Paul B.P. Kooijmans (Official Correspondent) Eurotrol B.V. Keplerlaan 20 6716 BS Ede, The Netherlands +31 318 695777 (Telephone) +31 318 695770 (FAX) pkooijmans@eurotrol.com
Date of Preparation: May 10, 2013 Eurotrol Hct Control and epoc® Hematocrit Verification Fluids Proprietary Names: Hematocrit control, (21 CFR 864.8625, Product Code GLK) Classification Name: Common Name: Eurotrol Hematocrit Control RNA Medical QC 900 Hematocrit Control (K955630) Equivalent to:
#### Substantial Equivalence
Eurotrol Hematocrit Control is substantially equivalent in function, safety and efficacy to currently marketed devices produced by Bionostics.
Comparison of Eurotrol Hematocrit Control to predicate device for substantial equivalency:
| | New Device | Predicate Device |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Eurotrol Hct Control and<br>epoc®Hematocrit Verification<br>Fluids | RNA Medical QC 900 Hematocrit<br>Control |
| 510(k), date | | K955630, 03/01/1996 |
| Number of levels | 5 | 5 |
| Analytes | Hematocrit (Conductivity) | Hematocrit (Conductivity) |
| Container | Clear glass ampules | Clear glass ampules |
| Filling Volume | 2,5 mL | 1,7 mL |
| Color | Clear | Clear |
| Storage<br>temperature | 2 - 30°C/35 - 86°F | 2 - 25°C/35 - 77°F |
| Matrix/Materials | Eurotrol Hematocrit Control is<br>prepared using pure chemicals in a<br>physiologically buffered matrix.<br>Different concentrations provide five<br>distinct Hct levels (conductivity),<br>simulating clinically significant ranges<br>of Hematocrit. | QC 900 is a buffered aqueous solution<br>containing electrolytes and<br>non-conductive ingredients. This<br>product contains no red cells and no<br>human or biological materials. |
{1}------------------------------------------------
510(k) k121467, 510(k) Summary Eurotrol B.V. Eurotrol Hct Control and epoc Hematocrit Verification Fluids May 10, 2013
#### Description of the new device
Eurotrol Het Control and epoc® Hematocrit Verification Fluids are assayed hematocrit reference materials, to verify the precision and accuracy of the epoc@ Blood Analysis System, manufactured by Epocal Inc., Ottawa, ON K1G3P5, Canada, as cleared by FDA, K061597. Eurotrol Hematocrit Control was designed to test the following analytes: Hematocrit.
Eurotrol Hot Control and epoc® Hematocrit Verification Fluids electrolyte solutions with conductivity at five levels appropriate to simulate clinically relevant hematocrit concentrations useful to evaluate the measurement of the epoc Blood Analysis System.
Depending the sales unit, 10 ampules of the same level are packed in a carton box for Eurotrol Hct Control, or 5 ampule per level, are packed in a carton box for epoc®Hematocrit Verification Fluids. A product insert with Intended Use is inserted in each product box.
The assigned values of each batch are printed on a value sheet as available from the Epocal website: http://www.epocal.com/doc_library.html
#### Intended Use
epoc® Hematocrit Verification Fluids is an assayed hematocrit reference material, to verify the precision and accuracy of the epoc® Blood Analysis System for the measurement of hematocrit.
Eurotrol Hot Control is an assayed hematocrit reference material, to verify the precision and accuracy of the epood Blood Analysis System for the measurement of hematocrit.
Eurotrol Hct Control and epoc® Hematocrit Verification Fluids are for professional use only.
#### Technological Characteristics
Eurotrol Hot Control and epoc@ Hematocrit Verification Fluids are filled in 3 mL clear glass ampules. Each ampule contains 2.5 mL product.
This material consists of an Aqueous buffered solution of water and electrolytes in 5 different levels.
The stability of Eurotrol Hematocrit Control is 12 months.
Summary of non-clinical tests submitted with the premarket notification for the device.
Tests were conducted to verify specific performance requirements:
a) Real-time evaluation of the products to support stability.
b) Test precision
Summary of clinical tests submitted with the premarket notification for the device.
N/A
Conclusions drawn from the clinical and non-clinical trials.
Comparison of technological characteristics, formulation and intended use to predicate devices listed in this summary support the claim of substantial equivalence.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
PAUL B. P. KOOIJMANS REGULATORY AFFAIRS MANAGER EUROTROL B.V. KEPLERLAAN 20 6716 BS, EDE THE NETHERLANDS
May 10. 2013
Re: K121467
Trade/Device Name: Eurotrol Het Control and epoc® Hematocrit Verification Fluids Regulation Number: 21 CFR 864.8625 Regulation Name: Hematology quality control mixture Regulatory Class: II Product Code: GLK Dated: April 19, 2013 Received: April 29, 2013
Dear Mr. Kooijmans:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Reena Philip -S
for
Maria M. Chan Director, Division of Immunology and Hematology Devices
Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known):_k121467
Device Name: Eurotrol Hct Control and epoc® Hematocrit Verification Fluids
Indications For Use:
Eurotrol Hct Control
Eurotrol Hct Control is an assayed hematocrit reference material, to verify the precision and accuracy of the epoc® Blood Analysis System for the measurement of hematocrit. Eurotrol Hct Control is for in vitro diagnostic use only.
epoc® Hematocrit Verification Fluids
epoc® Hematocrit Verification Fluids is an assayed hematocrit reference material, to verify the precision and accuracy of the epoc® Blood Analysis System for the measurement of hematocrit.
epoc® Hematocrit Verification Fluids is for in vitro diagnostic use only.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH: Office of In Vitro Diagnostics and Radiological Health (OIR)
## Maria M. @han -S
Division Sign-Off Office of In Vitro Diagnostics and Radiological Health
510(k) k121467
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.