REAGENT, RUSSEL VIPER VENOM - LA SCREEN

K132130 · Dsrv, Inc. · GIR · Jan 10, 2014 · Hematology

Device Facts

Record IDK132130
Device NameREAGENT, RUSSEL VIPER VENOM - LA SCREEN
ApplicantDsrv, Inc.
Product CodeGIR · Hematology
Decision DateJan 10, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 864.8950
Device ClassClass 1

Indications for Use

Frozen Format LA Screen (FFLAS) and Frozen Format LA Confirm (FFLAC) are simplified DRVVT reagents for detection of Lupus Anticoagulants (LA) in one-stage clotting tests. Frozen Format LA Screen is a simplified DRVVT reagent to screen for the presence of Lupus Anticoagulants. This product is for in-vitro diagnostic use only.

Device Story

Lupo-Test Reagent; in vitro diagnostic reagent. Modification involves transition from lyophilized format to frozen format stored at -80°C. Manufacturing and formulation processes remain identical to previously cleared version, excluding the removal of the reconstitution step. Used in clinical laboratory settings for diagnostic testing. Output provides diagnostic results based on reagent interaction with patient samples. Modification verified via FMEA per ISO 14971 to ensure performance characteristics remain equivalent to predicate.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

In-vitro diagnostic reagent; Dilute Russell Viper Venom Test (DRVVT) chemistry; frozen liquid format; intended for use in one-stage clotting assays.

Indications for Use

Indicated for in-vitro diagnostic screening of Lupus Anticoagulants (LA) in patients using one-stage clotting tests (DRVVT).

Regulatory Classification

Identification

Russell viper venom reagent is a device used to determine the cause of an increase in the prothrombin time.

Related Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification ODE Review Memorandum (Decision Making Document is Attached) To: THE FILE RE: DOCUMENT NUMBER k132130 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable: 1. The Name and 510(k) number of the SUBMITTER'S previously cleared device: Lupo-Test Reagent (k922326). 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling: Proposed labeling i.e., vial and box labels were provided as Attachment 1 and Attachment 2 in k132130/S002 and package insert was provided in k132130/S003. 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed: This change was for modification of Lupo-Test Reagent from lyophilized to a frozen format. The Submitter confirmed (k132130/S001, p.1) that there is no difference in the manufacturing and formulation process between the lyophilized and frozen format reagent except the additional step of reconstituting the lyophilized reagent, aliquoting and storage at -80°C. 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including labeling, intended use, stability and performance characteristics (k132130, p.4-5 of 10). 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis: Risk analysis followed harmonized standard ISO 14971. Failure Mode and Risk Analysis (FMEA) method was applied (k132130, p.8 of 10). The Risk Management Plan, Risk Management Process and Preliminary Risk Analysis was provided in k132130/S001, pp.14-21. b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied. (k132130 p.8 of 10 and S001 pp.15-21). c) A declaration of conformity with design controls. The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met. (k132130, p.7 of 10). ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review (k132130/S002). 6. A Truthful and Accurate Statement (k132130 p.10 of 10), a 510(k) Statement (k132130, p.9 of 10) and the Indications for Use Enclosure (provided by email). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...