K994388 · Streck Laboratories, Inc. · GLQ · Jan 14, 2000 · Hematology
Device Facts
Record ID
K994388
Device Name
XE CHECK
Applicant
Streck Laboratories, Inc.
Product Code
GLQ · Hematology
Decision Date
Jan 14, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.8625
Device Class
Class 2
Indications for Use
XE Check is intended to be used as a control for evaluating complete blood cell count (CBC), white cell five-part differential, and reticulocyte percentage on Sysmex XE – 2100 series hematology instruments. The device will consist of three levels: Abnormal Low (characterized by low CBC, high reticulocyte %), Normal (normal CBC, normal reticulocyte %), and Abnormal High (high CBC, low reticulocyte %).
Device Story
XE Check is an assayed hematology control consisting of stabilized human red blood cells, human white cells, simulated human platelets, and simulated human reticulocytes. It is packaged in 4.6 mL glass vials with polypropylene caps. The device is used in clinical laboratory settings to verify the performance of Sysmex XE-2100 hematology instruments. Laboratory personnel process the control through the instrument to recover CBC, white cell differential, and reticulocyte parameters. By comparing instrument results against established control values, clinicians ensure the accuracy and reliability of patient sample analysis. This verification process supports clinical decision-making by ensuring the hematology analyzer is functioning correctly, thereby reducing the risk of erroneous patient test results.
Clinical Evidence
Bench testing only. Studies included lot-to-lot reproducibility, comparison to whole blood, long-term stability, and open-vial stability. Results confirmed consistent reproducibility and stability for the duration of product dating.
Technological Characteristics
Suspension of stabilized human red blood cells, human white cells, simulated human platelets, and simulated human reticulocytes. Packaged in 4.6 mL glass vials with injection-molded polypropylene screw-top caps. Standalone control material; no energy source or software.
Indications for Use
Indicated for use as a quality control material for complete blood cell count (CBC), white cell 5-part differential, and reticulocyte parameters on Sysmex XE-2100 hematology instruments. No specific patient population or contraindications described.
Regulatory Classification
Identification
A hematology quality control mixture is a device used to ascertain the accuracy and precision of manual, semiautomated, and automated determinations of cell parameters such as white cell count (WBC), red cell count (RBC), platelet count (PLT), hemoglobin, hematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC).
Special Controls
*Classification.* Class II (special controls). Except when intended for use in blood components, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
{0}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
| 510(k) Submitter: | Streck Laboratories, Inc.<br>14124 Industrial Road<br>Omaha, NE 68144 |
|-------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Official Correspondent: | Paul Kittelson<br>Quality Assurance<br>(402) 691-7465 |
| Date Prepared: | December 27, 1999 |
| Names of Device:<br>Trade Name:<br>Common Name:<br>Classification Name: | XE Check<br>Assayed hematology control<br>Hematology quality control mixture (§ 864.8625) |
| Predicate Devices: | SF Check manufactured by Streck Laboratories, Inc.<br>Retic Chex manufactured by Streck Laboratories, Inc. |
Description: XE Check is a suspension of stabilized human red blood cells, human white cells.escription: XE Check is a suspension of stabilized human red blood cells, human white cells, simulated human platelets, and simulated human reticulocytes packaged in glass vials containing 4.6 mL volumes. Closures are injection molded polypropylene screw-top caps. The vials are packaged in polystyrene jars.
Intended Use: XE Check is intended to be used as a control for evaluating complete blood cell count (CBC), white cell five-part differential, and reticulocyte percentage on Sysmex XE – 2100 series hematology instruments. The device will consist of three levels: Abnormal Low (characterized by low CBC, high reticulocyte %), Normal (normal CBC, normal reticulocyte %), and Abnormal High (high CBC, low reticulocyte %).
Comparison with Predicate Devices: Like SF Check is intended to enable the user to verify satisfactory performance of Sysmex XE - 2100 instruments in recovery of CBC and white cell differential parameters on whole blood specimens. Both devices contain stabilized human red blood cells, human white cells, and simulated platelets which properly mimic human whole blood components on Sysmex XE - 2100.
Unlike SF Check, XE Check contains a stabilized human reticulocyte component, similar to Retic Chex. This allows the user to verify proper performance of the on-line reticulocyte analysis system in Sysmex XE -- 2100 instruments with the same device used for CBC and white cell differential analysis verification.
Discussion of Tests and Test Results: Three studies of XE Check were conducted: l) Lot to Lot Reproducibility and Comparison to Whole Blood; II) Long Term Stability; and III) Open Vial Stability. Study results showed XE Check to be consistently reproducible, substantially equivalent to the predicate products, and stable for the entire product dating.
Conclusions Drawn From Tests: XE Check is safe and effective for controlling CBC/Diff/Retic parameters on Sysmex XE - 2100 instruments when used as instructed in the product package insert.
K994388
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized abstract design resembling an eagle or bird in flight. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird symbol.
### Public Health Service
# JAN 1 4 2000
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Paul Kittelson Quality Assurance Streck Laboratories, Inc. 14124 Industrial Road Omaha, Nebraska 68144
Re: K994388 Trade Name: XE Check Regulatory Class: II Product Code: GLO, JPK Dated: December 27, 1999 Received: December 28, 1999
## Dear Mr. Kittelson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Page 1 of 1
510(k) Number (if known): _ H 99 43 88
XE Check Device Name:
Indications For Use:
XE-Check is intended to be used as a control for complete blood cell count (CBC), white cell 5-part differential, and reticulocyte parameters on Sysmex XE - 2100 hematology instrument.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sutule Massi
994388
**Prescription Use** \
(Per 21 CFR 801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
· · · · · ·
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.