STA - LIATEST VWF TEST KIT
Device Facts
| Record ID | K962675 |
|---|---|
| Device Name | STA - LIATEST VWF TEST KIT |
| Applicant | American Bioproducts Co. |
| Product Code | GJT · Hematology |
| Decision Date | Oct 25, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 864.7290 |
| Device Class | Class 2 |
Indications for Use
The STA®-Liatest® vWF test kit is intended for use with the STA® analyzer to perform quantitative assays of von Willebrand Factor (vWF) antigen in citrated plasma by the immuno-turbidimetric method based on the measurement of light absorbance produced by a suspension of microlatex particles coated with specific anti-vWF antibodies.
Device Story
In vitro diagnostic test kit; performs quantitative assay of vWF antigen in citrated plasma. Uses immuno-turbidimetric method; input is citrated plasma sample; reagent consists of microlatex particles coated with rabbit anti-human vWF antibodies. Device operates on STA® analyzer; measures light absorbance of latex particle suspension. Output is quantitative vWF antigen level. Used in clinical laboratory settings by trained technicians. Results assist clinicians in diagnosing or monitoring conditions related to vWF levels.
Clinical Evidence
No clinical data provided; substantial equivalence based on performance characteristics of the immuno-turbidimetric assay compared to the predicate electro-immunodiffusion method.
Technological Characteristics
Immuno-turbidimetric assay; utilizes microlatex particles coated with rabbit anti-human vWF antibodies. Reagents include glycine buffer and diluent. Designed for use on STA® analyzer; measures light absorbance. Storage 2-8°C; 18-month shelf life; 15-day on-board stability.
Indications for Use
Indicated for quantitative determination of von Willebrand Factor (vWF) antigen levels in citrated plasma for patients requiring assessment of vWF levels.
Predicate Devices
- ASSERA®-PLATE vWF kit (K861417)
Related Devices
- K992704 — IL TEST VON WILLEBRAND FACTOR · Instrumentation Laboratory CO · Nov 1, 1999
- K220728 — vWF Ag · Siemens Healthcare Diagnostics Products GmbH · Jun 2, 2023
- K133005 — HEMOSIL VON WILLEBRAND ACTIVITY ASSAY · Instrumentation Laboratory CO · Mar 13, 2014