K973696 · Beckman Instruments, Inc. · LCP · Nov 24, 1997 · Hematology
Device Facts
Record ID
K973696
Device Name
SYNCHRON SYSTEM HEMOGLOBIN ALC REAGENT
Applicant
Beckman Instruments, Inc.
Product Code
LCP · Hematology
Decision Date
Nov 24, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.7470
Device Class
Class 2
Indications for Use
The SYNCHRON Systems Hemoglobin A1c (HbA1c) Reagent Kit, in conjunction with SYNCHRON HbA1c Calibrators and SYNCHRON HbA1c Hemolyzing Reagent, is intended for the quantitative determination of Hemoglobin A1c concentrations in whole blood on SYNCHRON Clinical Systems.
Device Story
The SYNCHRON Systems HbA1c Reagent is an in vitro diagnostic kit used on SYNCHRON Clinical Systems to measure HbA1c levels in whole blood. The system processes whole blood samples to determine HbA1c concentration as a percentage of total hemoglobin. It is intended for use in clinical laboratory settings by trained personnel. The device provides quantitative results that assist healthcare providers in assessing glycemic control and adjusting insulin therapy for diabetic patients. The system utilizes specific bar-coded cartridges and pre-programmed assay parameters within the SYNCHRON database to automate the analysis, replacing manual programming required by predicate systems.
Clinical Evidence
Bench testing only. Method comparison studies against the predicate Tina-quant® assay (slope 0.9906, intercept -0.19%, r=0.9698) and Biorad Diamat HPLC method (slope 0.9937, intercept -0.25%, r=0.9678) demonstrate analytical equivalence. Imprecision studies (N=80) showed within-run CVs of 2.9-4.5% and total CVs of 6.0-7.2% across three concentration levels.
Technological Characteristics
In vitro diagnostic reagent kit for use on automated clinical chemistry analyzers. Employs chemical reaction-based assay for HbA1c quantification. Features bar-coded reagent cartridges for automated system recognition and parameter loading. Operates on SYNCHRON Clinical Systems platform.
Indications for Use
Indicated for the quantitative determination of hemoglobin A1c concentration as a percentage of total hemoglobin in whole blood to assess diabetes control and determine insulin dosage in patients.
Regulatory Classification
Identification
A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.
K042459 — SYNCHRON LX SYSTEMS HEMOGLOBIN A1C2 (HBA1C2) REAGENT · Beckman Coulter, Inc. · Nov 17, 2004
K031042 — ATAC HEMOGLOBIN A1C REAGENT KIT · Clinical Data, Inc. · Nov 3, 2003
K091711 — S- TEST HEMOGLOBIN ALC (HBA1C), MODEL RC0023 · Alfa Wassermann Diagnostic Technologies, Inc. · Dec 11, 2009
K031380 — OLYMPUS HBA1C REAGENT, (CATALOG NUMBER OSR6192), OLYMPUS HBA1C CALIBRATOR, (CATOLOG NUMBER ODC3032) · Olympus America, Inc. · Nov 14, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
K973696
Beckman Instruments, Inc., Section 510(k) Notification SYNCHRON Systems Hemoglobin A1c (HbA1c) Reagent Summary of Safety & Effectiveness
NOV
## Summary of Safety & Effectiveness Beckman SYNCHRON® Systems Hemoglobin A1c Reagent
#### 1.0 Submitted Bv:
Lucinda Stockert Senior Regulatory Specialist, Product Submissions Beckman Instruments, Inc. 200 S. Kraemer Blvd., W-337 Brea, California 92822-8000 Telephone: (714) 961-3777 FAX: (714) 961-4457
#### 2.0 Date Submitted:
24 September 1997
#### 3.0 Device Name(s):
## 3.1 Proprietary Names
SYNCHRON® Systems Hemoglobin A1c (HbA1c) Reagent
#### 3.2 Classification Name
Glycosylated hemoglobin assay (21 CFR § 864.7470)
#### 4.0 Predicate Device(s):
| BECKMAN Reagent | Predicate | Predicate Company | Docket Number |
|------------------------------------|----------------------------------------|------------------------------------|---------------|
| SYNCHRON Systems<br>Hemoglobin A1c | Tina-quant®<br>Hemoglobin A1c<br>Assay | Boehringer Mannheim<br>Corporation | K934070 |
#### 5.0 Description:
The SYNCHRON Hemoglobin A1c (HbA1c) Reagent kit contains two reagents, Hb and A1c for determining the Hemoglobin A1c concentrations in whole blood on SYNCHRON Clinical Systems.
#### 6.0 Intended Use:
The SYNCHRON Hemoglobin A1c (HbA1c) Reagent kit, in conjunction with the SYNCHRON Hemoglobin A1c Calibrators, is intended for the quantitative determination of hemoglobin A1c concentration as a percentage of total hemoglobin in whole blood on SYNCHRON Systems.
#### 7.0 Comparison to Predicate(s):
The following tables show similarities and differences between the predicate identified in Section 4.0 of this summary.
{1}------------------------------------------------
Beckman Instruments, Inc., Section 510(k) Notification SYNCHRON Systems Hemoglobin A1c (HbA1c) Reagent Summary of Safety & Effectiveness
| Reagent | Aspect/Characteristic | Comments |
|-------------------------------------------------------------|---------------------------|----------------------------------------------------------------------------------------------------------|
| | SIMILARITIES | |
| SYNCHRON<br>Systems Hemoglobin<br>A1c<br>(HbA1c)<br>Reagent | Intended use | Same as the predicate |
| | Formulation | Same source, processing as predicate |
| | Chemical Reaction | Same principle as the predicate |
| | Sample Preparation | Same as the predicate |
| | Calibration | Same as the predicate |
| | Reagent Kit Configuration | Same as the predicate |
| | DIFFERENCES | |
| SYNCHRON Systems<br>Hemoglobin A1c<br>(HbA1c)<br>Reagent | Assay parameters | SYNCHRON parameters reside in the<br>SYNCHRON database, BMC<br>parameters must be operator<br>programmed |
| | Cartridges | Cartridges bar-coded specifically for<br>SYNCHRON HbA1c vs non bar-coded<br>cartridges with BMC reagent |
#### Summary of Performance Data: 8.0
.
The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through method companson, stability, and imprecision experiments that relate results obtained from the SYNCHRON HbA1c Reagent to the Boehringer Mannheim Tina-quant® Hemoglobin A1c assay.
### Method Companison Study Results
| Analyte | Slope | Intercept | r | Predicate Method |
|--------------------------------------------------|--------|-----------|--------|---------------------------------------------------------------------------|
| SYNCHRON<br>Hemoglobin A1c<br>(HbA1c)<br>Reagent | 0.9906 | -0.19% | 0.9698 | Boehringer Mannheim<br>Hemoglobin A1c Reagent on<br>SYNCHRON CX4CE System |
| Reagent | 0.9937 | -0.25% | 0.9678 | Biorad DiamatTM HPLC<br>Hemoglobin A1c Method |
### Stability Study Results
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>ANNUAL AND COLORIAL COLLEGIAN<br>-------------------------------------- | 1 100 11 11 15 15 15 15 15 15 15 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11<br>RECENTRAL II LANDARIES OF LEASTER CONSULTION OF<br>AND CORPARA CORPORATION OF COLLECTION OF |
| Comments of the contraction of the contribution of the commend of the commend of the commend of the many of the many of the many of the many of the many of the many of the ma<br><br>.<br> | <br>11 18 1 1 8 8 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 |
### Estimated SYNCHRON HbA1c Reagent Imprecision
| Sample | Mean (%HbA1c) | S.D. (%) | %C.V. | N |
|------------------------|---------------|----------|-------|----|
| Within-Run Imprecision | | | | |
| Level 1 | 6.05 | 0.27 | 4.5 | 80 |
| Level 2 | 8.75 | 0.26 | 3.0 | 80 |
| Level 3 | 10.73 | 0.31 | 2.9 | 80 |
| Total Imprecision | | | | |
| Level 1 | 6.05 | 0.44 | 7.2 | 80 |
| Level 2 | 8.75 | 0.52 | 6.0 | 80 |
| Level 3 | 10.73 | 0.74 | 6.9 | 80 |
The Summary of Safety and Effectiveness information for the SYNCHRON Systems HbA1c Reagents are found in TAB 1 of this notice and are being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and implementing regulation 21 CFR 807.92.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus or a bird in flight. The logo is presented in black and white.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
NOV 2 4 1997
Lucinda Stockert Senior Regulatory Specialist Beckman Instruments, Inc. 200 S. Kraemer Boulevard, M/S W-337 P.O. Box 800 Brea, California 92822-8000
Re: K973696 Synchron® Systems Hemoglobin A1c (HbA1c) Reagent Regulatory Class :... II . .............................. Product Code: LCP September 24, 1997 Dated: Received: September 26, 1997
Dear Ms. Stockert:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note:
this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{3}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity (CHIA-68), Chila device may require if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as Inis leccer will aremarket notification. The FDA described in your sio\n privalence of your device to a legally marketed predicate device results in a classification for your marketed produces permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Victo diagnoBele develop, productionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general premarket notification" (21 CFR 807.97). information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance obeathed from onember (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Gutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K973696
of
510(k) Number (if known): Not yet assigned
### SYNCHRON® Systems Hemoglobin A1c Reagent Device Name:
Indications for Use:
The SYNCHRON Systems Hemoglobin A1c (HbA1c) Reagent Kit, in conjunction with SYNCHRON HbA1c Calibrators and SYNCHRON HbA1c Hemolyzing Reagent, is intended for the quantitative determination of Hemoglobin A1c concentrations in whole blood on SYNCHRON Clinical Systems.
# 21 CFR 864.7470 Glycosylated hemoglobin assay
(a) Identification. A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a, A1b, A1c) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.
(b) Classification. Class II.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Clinical Laboratory Devices
| 510(k) Number | 2973696 |
|---------------|---------|
|---------------|---------|
| Prescription Use (per 21 CFR 801.109) | Over-the-Counter Use Optional Format 1-2-96 |
|---------------------------------------|---------------------------------------------|
|---------------------------------------|---------------------------------------------|
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.