DIAZYME GLYCATED SERUM PROTEIN ENZYMATIC ASSAY KIT
Applicant
Diazyme Laboratories
Product Code
LCP · Hematology
Decision Date
Oct 29, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.7470
Device Class
Class 2
Indications for Use
Diazyme Glycated Serum Protein Enzymatic Assay Kit in conjunction with Diazyme Glycated Serum Protein single calibrator, are intended for the quantitative determination of glycated serum proteins (fructosamine) in serum. The measurement of glycated serum proteins (fructosamine) is useful for monitoring diabetic patients. Diazyme Glycated Serum Protein Enzymatic Assay Kit contains a single standard. The standard is used in the calculation of glycated serum protein concentrations in unknown serum samples. Diazyme Glycated Serum Protein Enzymatic Assay has controls for normal glycated serum protein level and abnormal glycated serum protein level. The controls are used as reference samples for checking the functionality of the Diazyme Glycated Serum Protein Enzymatic Assay.
Device Story
In vitro diagnostic assay; measures glycated serum proteins (fructosamine) in human serum; used for monitoring diabetic patients. Reagent 1 (Proteinase K) digests glycated serum proteins into glycated protein fragments; Reagent 2 (Fructosaminase) catalyzes oxidation of fragments to release hydrogen peroxide; hydrogen peroxide measured via colorimetric Trinder end-point reaction at 550 nm. Absorbance proportional to GSP concentration. Performed manually or on automated chemistry analyzers (37 °C). Includes lyophilized calibrator and separate bi-level controls. Results assist clinicians in diabetes management.
Clinical Evidence
No clinical studies performed. Evidence consists of analytical bench testing: precision (CV 2.84-2.87%), linearity (R²=0.9986), and interference testing (less than 10% deviation for common substances). Method comparison study against predicate device (n=65) yielded Y=0.9981x + 4.984, R²=0.9789.
Technological Characteristics
Two-reagent (R1/R2) end-point enzymatic assay. Utilizes proprietary fructosyl amine-oxygen oxidoreductase (fructosaminase) and peroxidase. Measurement via colorimetric absorbance at 550 nm. Designed for use on automated clinical chemistry analyzers. No off-line pretreatment required.
Indications for Use
Indicated for the quantitative determination of glycated serum proteins (fructosamine) in serum to monitor diabetic patients.
Regulatory Classification
Identification
A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.
K131307 — ADVIA CHEMISTRY FRUCTOSAMINE ASSAY, ADVIA CHEMISTRY FRUCTOSAMINE CALIBRATOR · Siemens Healthcare Diagnostics, Inc. · Jan 24, 2014
K110188 — DIAZYME GLYCATED SERUM PROTEIN ASSAY / CALIBRATOR KIT / CONTROL KIT · Diazyme Laboratories · May 17, 2011
K133803 — DIAZYME GLYCATED SERUM PROTEIN POC TEST KIT · Diazyme Laboratories · Aug 6, 2014
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
A. 510(k) Number:
k042193
B. Purpose for Submission:
New Device
C. Analyte:
Glycated Serum Proteins (Fructosamine)
D. Type of Test:
Quantitative
E. Applicant:
Diazyme Laboratories
F. Proprietary and Established Names:
Diazyme Glycated Serum Protein Enzymatic Assay Kit
Diazyme Glycated Serum Controls
G. Regulatory Information:
1. Regulation section:
21 CFR 864.7470 Glycosylated Hemoglobin Assay
21 CFR 862.1660 Quality Controls
2. Classification:
Class II - 21 CFR 864.7470 Glycosylated Hemoglobin Assay
Class I (reserved) - 21 CFR 862.1660 Quality Control
3. Product Code:
Diazyme Glycated Serum Protein Enzymatic Assay and Calibrators- LCP
Diazyme Glycated Serum Controls- JJX
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
Diazyme Glycated Serum Protein Enzymatic Assay Kit in conjunction with
Diazyme Glycated Serum Protein single calibrator, are intended for the
quantitative determination of glycated serum proteins (fructosamine) in
{1}
Page 2 of 6
serum. The measurement of glycated serum proteins (fructosamine) is useful for monitoring diabetic patients.
Diazyme Glycated Serum Protein Enzymatic Assay Kit contains a single standard. The standard is used in the calculation of glycated serum protein concentrations in unknown serum samples.
Diazyme Glycated Serum Protein Enzymatic Assay has controls for normal glycated serum protein level and abnormal glycated serum protein level. The controls are used as reference samples for checking the functionality of the Diazyme Glycated Serum Protein Enzymatic Assay.
2. Indication(s) for use:
Diazyme Glycated Serum Protein Enzymatic Assay Kit in conjunction with Diazyme Glycated Serum Protein single calibrator, are intended for the quantitative determination of glycated serum proteins (fructosamine) in serum. The measurement of glycated serum proteins (fructosamine) is useful for monitoring diabetic patients.
Diazyme Glycated Serum Protein Enzymatic Assay Kit contains a single standard. The standard is used in the calculation of glycated serum protein concentrations in unknown serum samples.
Diazyme Glycated Serum Protein Enzymatic Assay has controls for normal glycated serum protein level and abnormal glycated serum protein level. The controls are used as reference samples for checking the functionality of the Diazyme Glycated Serum Protein Enzymatic Assay.
3. Special condition for use statement(s):
For Prescription Use Only
4. Special instrument Requirements:
This assay can be performed manually or in an automated chemistry analyzer capable of maintaining 37 °C and reading absorbance at 550 nm.
I. Device Description:
The Diazyme Glycated Serum Protein (Dia-GSP) Enzymatic Assay consists of Proteinase K reagent 1 (R1) and Fructosaminase reagent 2 (R2). R1 digests Glycated Serum Proteins (GSP) into glycated protein fragment (GPF). Reagent 2 catalyzes the oxidation of GPF into fragments and hydrogen peroxide (H₂O₂). The H₂O₂ released is measured by a colorimetric Trinder end-point reaction. The amount of color developed is measured at 550 nm and is proportional to the concentration of GSP in the sample.
The Diazyme Glycated Serum Protein (Dia-GSP) Enzymatic Assay includes the GSP Calibrator. The GSP Calibrator consists of one level (470 μmol/L) and is supplied in lyophilized. The concentrations of fructosyl propylamine of the calibrators were confirmed by the Randox Fructosamine Assay.
The Diazyme Glycated Serum Protein Controls (GSP Controls) are sold separately and consist of 2 levels (250 μmol/L and 647 μmol/L) and are supplied lyophilized.
{2}
The concentrations of fructosyl propylamine of the controls were confirmed by the Randox Fructosamine Assay.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
Randox Laboratories Limited Fructoasamine
2. Predicate K number(s):
k023763
3. Comparison with predicate:
Diazyme Glycated Serum Protein Enzymatic Assay
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | For the quantitative in vitro determination of glycated protein in human serum. The measurement of glycated serum proteins (fructosamine) is useful for monitoring diabetic patients. | For the quantitative in vitro determination of glycated protein in human serum or plasma. |
| Sample | Serum | Serum, EDTA or Plasma |
| Principle | Glycated Serum Protein (GSP) concentration is determined through a series of enzymatic reactions and the oxidative degradation of glycated serum fragments to form hydrogen peroxide. The hydrogen peroxide is released and measured colorimetrically and the absorbance at 550 nm is proportional to the concentration of GSP. | Glycated Protein fragments are oxidized by ketoamine oxidase to form hydrogen peroxide. The hydrogen peroxide is measured colorimetrically and is proportional to the concentration of glycated protein in a sample. |
| Differences | | |
| Item | Device | Predicate |
| Assay Range | 21 μmol/L to 1354 μmol/L | 11 μmol/L to 1734 mmol/L |
{3}
Page 4 of 6
Diazyme Glycated Serum Controls
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Diazyme Glycated Serum Controls is a bi-level control. The controls are used as reference samples for checking the functionality of the Diazyme Glycated Serum Protein Enzymatic Assay. | Randox Fructoamine Control is bi-level control. The controls are used as reference samples for checking the functionality of the Randox Laboratories Limited Fructosamine Assay. |
K. Standard/Guidance Document Referenced (if applicable):
None Referenced
L. Test Principle:
Glycated Serum Protein (GSP) concentration is determined through a series of enzymatic reactions and the oxidative degradation of glycated protein fragments to release hydrogen peroxide. The hydrogen peroxide released is measured by a colorimetric Trinder end-point reaction. The absorbance at 550 nm is proportional to the concentration of GSP.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The Diazyme GSP Enzymatic Assay was run on two specimens of approximately 280 μmol/L and 586 μmol/L, in duplicates for a total of twenty replicates. Specimen 1 obtained a mean of 290 μmol/L with a CV of 2.87% and specimen 2 obtained a mean of 608 μmol/L with a CV of 2.84%.
The Diazyme GSP Controls were tested for reproducibility at two laboratories (on the Hitachi 717 and Randox Daytona) on low and high control serum samples. Site one tested two controls (244 μmol/L and 611 μmol/L glycated serum proteins) in duplicates and obtained an average of 246 μmol/L and 615 μmol/L glycated serum proteins. Site two tested two controls (235 μmol/L and 560 μmol/L glycated serum proteins) in duplicates and obtained an average of 233 μmol/L and 557 μmol/L.
b. Linearity/assay reportable range:
A linearity study was conducted on the Hitachi 717 in duplicates on eight dilutions of glycated serum proteins purchased from Sigma-Aldrich and yielded an equation of Y=0.9557x + 5.3377 with a R² of 0.9986.
{4}
Page 5 of 6
The assay has a linear range of 21 to 1354 µmol/L.
c. Traceability (controls, calibrators, or method):
The Diazyme GSP Controls are sold separately from the assay and the Calibrators are included with the assay. The Diazyme GSP Controls and Calibrators are gravimetrically prepared from commercially available reagents. The concentrations of fructosyl propylamine were confirmed by the Randox Fructosamine Assay. Two lots of the Diazyme GSP Enzymatic Assay kits were used in a real-time and an accelerated stability studies. The results of the stability studies supported a claim for reconstituted reagents at 2-8 °C of 28 days and for the lyophilized form at 2-8 °C of 12 months. The Diazyme GSP Calibrators and controls were both studied in real time at 4 °C and the results supported a stability of 4 weeks when reconstituted. The Diazyme GSP calibrators and control in the lyophilized form were evaluated in an accelerated study and the results supported a 12 weeks stability claim at 25 °C.
d. Detection limit:
See Linearity section above.
e. Analytical specificity:
Interference was tested on substances normally present in the serum with Diazyme GSP Enzymatic Assay by spiking 160 µmol/L glycated protein serum samples. The study found less than 10% deviation when tested with the following substance levels: 2 mM Uric Acid, 200 mg/dl Haemoglobin, 133 mM Glucose, 2000 mg/dl Triglycerides, 0.230 mM Ascorbic Acid and 20 mg/dl Bilirubin.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Sixty-five real patient serum samples were tested with both the Diazyme GSP Enzymatic Assay and the Randox Fructosamine Assay. On the Hitachi 717. The Diazyme GSP calibrator and controls were gravimetrically prepared with commercially available reagents and was verified by the Randox Fructosamine Assay. The equation in the correlation study was Y=0.9981x + 4.984 with a R² of 0.9789.
b. Matrix comparison:
Not applicable
{5}
Page 6 of 6
3. Clinical studies:
a. Clinical sensitivity: Not applicable
b. Clinical specificity: Not applicable
c. Other clinical supportive data (when a and b are not applicable): Not applicable
4. Clinical cut-off: Not applicable
5. Expected values/Reference range: A normal glycated serum proteins in adults (20 to 60 years) range from 100-285 μmol/L as obtained from the literature was provided.
N. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.