K965076 · Organon Teknika Corp. · KFF · Mar 18, 1997 · Hematology
Device Facts
Record ID
K965076
Device Name
MDA HEPARIN ANTI-X ASSAY
Applicant
Organon Teknika Corp.
Product Code
KFF · Hematology
Decision Date
Mar 18, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.7525
Device Class
Class 2
Indications for Use
MDA™ Heparin anti-Xₐ is achromogenic assay for the quantitative determination of heparin anti-Xₐ activity in human plasma.
Device Story
MDA™ Heparin anti-Xₐ is a chromogenic assay for measuring heparin activity in citrated human plasma. The device utilizes a two-step amidolytic reaction based on the Teien and Lee method. In the first step, plasma containing heparin is incubated with antithrombin III (AT-III) and Factor Xa (FXa); heparin accelerates the AT-III-mediated inhibition of FXa. In the second step, residual active FXa cleaves a chromogenic peptide substrate, releasing a yellow chromophore. The device measures absorbance at 405 nm; color intensity is inversely proportional to heparin concentration. The assay is designed for use on the MDA 180 instrument. Healthcare providers use the resulting quantitative heparin activity levels to monitor anticoagulant therapy, helping to prevent thrombotic complications or manage existing thrombi. The assay is particularly useful for monitoring low molecular weight heparins, which are not readily measured by traditional APTT assays.
Clinical Evidence
Accuracy evaluated using 125 specimens (90 standard heparin, 35 LMW heparin) compared to a reference chromogenic method per NCCLS EP9-A; results showed high correlation (r=0.952 overall). Precision evaluated per NCCLS EP5-T2 over 19-21 days with 76 measurements per control level; within-run CVs were 4.5-6.1% and total CVs were 6.7-9.6%.
Technological Characteristics
Chromogenic assay; reagents include Buffer Concentrate, FXa reagent, Substrate reagent, and AT-III reagent. Principle: two-step amidolytic reaction measuring absorbance at 405 nm. Designed for automated use on the MDA 180 instrument. Sample type: citrated plasma.
Indications for Use
Indicated for the quantitative determination of heparin anti-Xa activity in human plasma for patients undergoing anticoagulant therapy with unfractionated or low molecular weight heparins.
Regulatory Classification
Identification
A heparin assay is a device used to determine the level of the anticoagulant heparin in the patient's circulation. These assays are quantitative clotting time procedures using the effect of heparin on activated coagulation factor X (Stuart factor) or procedures based on the neutralization of heparin by protamine sulfate (a protein that neutralizes heparin).
K983178 — COAMATIC HEPARIN · Instrumentation Laboratory CO · Nov 3, 1998
Submission Summary (Full Text)
{0}
K96507
510(k) SUMMARY
MDA™ Heparin anti-Xₐ
MAR 18 1997
This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and the final rule under 21 CFR 807.92 published December 14, 1994.
(a) (1) The submitter's name, address, telephone number, a contact person, and the date the summary was prepared;
Submitter's Name: Organon Teknika Corporation
Submitter's Address: 100 Akzo Avenue, Durham, North Carolina 27712, USA
Submitter's Telephone: (919) 620-2373
Submitter's Contact: Ron Sanyal, M. Pharm, CQE, RAC Ron Sanyal
Date 510(k) Summary Prepared: December 17, 1996
(a) (2) The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known;
Trade/Proprietary Name: MDA™ Heparin anti-Xₐ
Common/Usual Name: Heparin assay
Classification Name: Heparin assay
(a) (3) An identification of the legally marketed device to which the submitter claims substantial equivalence.
Device Equivalent to: Stachrom® Heparin (K925433)
(a) (4) A description of the device(System)
## Description of MDA Heparin anti-Xa
Heparin is administered to prevent thrombotic complications (for example, to prevent thrombosis following surgery) and to reduce or prevent the extension of existing thrombi (for example, in treatment of deep venous thrombosis). Heparin acts by binding to antithrombin III (AT-III), a plasma protein that inhibits coagulation enzymes including thrombin and FXa. Inhibition of these enzymes is greatly accelerated when AT-III is bound to heparin.
MDA Heparin anti-FXa measures the ability of heparin to accelerate AT-III inhibition of FXa using a chromogenic substrate. This substrate is a small peptide linked to a chromophore which is specifically cleaved by FXa to produce yellow color. The assay principle is based on the method of Teien and Lee (1,2) and uses a two-step reaction:
1) Plasma containing heparin is mixed with buffer and AT-III, and incubated with FXa reagent for a fixed period of time. Some of the FXa is inactivated:
$$
\mathrm{FXa} + \mathrm{AT-III-Heparin} (\text{complex}) \longrightarrow \mathrm{FX-AT-III} (\text{inactive}) + \text{Heparin}
$$
residual FXa
2) Remaining active FXa cleaves chromogenic substrate to produce yellow color, monitored by measuring absorbance at 405 nm. The amount of color produced is proportional to the amount of FXa remaining and inversely proportional to the amount of heparin in the specimen plasma:
$$
\mathrm{FXa} + \text{peptide-pNA} \longrightarrow \text{peptide} + \text{p-NO}_2\text{-aniline} (\text{yellow})
$$
Reagents included in the assay kit include Buffer Concentrate, FXa reagent, Substrate reagent and AT-III reagent.
Until recently, the heparins used for anticoagulant therapy were high-molecular weight (~15,000-30,000 daltons) sulfated polysaccharides that could be monitored using Activated Partial Thromboplastin Time (APTT, PTT) assay, chromogenic assays, or other methods (3). Recently, low molecular weight heparins (Mr < 15,000 daltons) have become available for clinical use. These LMW heparins can be measured using chromogenic assays, but are not readily measured by APTT.
{1}
(a) (5) A statement of the intended use of the device.
Device Intended Use: MDA™ Heparin anti-Xₐ is achromogenic assay for the quantitative determination of heparin anti-Xₐ activity in human plasma.
(a) (6) A summary of the technological characteristics of the new device in comparison to those of the predicate device.
The technological characteristics of the new device in comparison to those of the device [Stachrom® Heparin (K925433)] are given in the Table 1 below.
TABLE 1
| PARAMETERS | ORGNON TEKNIKA's MDA™ Heparin anti-Xₐ | DIAGNOSTICA STAGO's Stachrom® Heparin |
| --- | --- | --- |
| | | |
| CATEGORY | MDA™ Heparin anti-Xₐ is a chromogenic assay. | Stachrom® Heparin assay is a chromogenic assay |
| | | |
| INTENDED USE | MDA™ Heparin anti-Xₐ is a chromogenic assay for the quantitative determination of heparin anti-Xa activity in human plasma. | Stachrom® Heparin assay is a chromogenic assay for quantitative determination of the plasma level of unfractionated heparins (UFH) or low molecular weight heparins (LMWH) by the measurement of their anti-Xa activity using the amidolytic method with chromogenic substrate. |
| | | |
| SAMPLE | Citrated Plasma | Citrated Plasma |
| | | |
| AUTOMATION | Yes (MDA 180 Instrument) | Yes (Adaptation protocol is available on request from Diagnotica Stago) |
| | | |
| TEIEN AND LEE METHOD | Yes (Two step chromogenic measurement based on Teien and Lee method) | Yes (Two step chromogenic measurement based on Teien and Lee method) |
| | | |
{2}
(b) (1) A brief discussion of the nonclinical tests submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalency.
Not Applicable
(b)(2) A brief discussion of the clinical tests submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalency.
## Accuracy
Results from MDA Heparin anti-FX₄ reagents obtained on an MDA 180 were compared with another commercial chromogenic reagent and system for specimens tested in duplicate according to NCCLS Tentative Guideline EP9-A. The following results for slope, intercept and correlation were observed for linear regression comparing MDA Heparin anti Xₐ (y-axis) and the reference method (x-axis):
| Type of heparin therapy | n | Slope | Intercept | r |
| --- | --- | --- | --- | --- |
| | | | | |
| Standard heparin | 90 | 1.06 | 0.00 | 0.954 |
| Low molecular weight heparin | 35 | 0.95 | -0.02 | 0.954 |
| All specimens | 125 | 0.98 | 0.00 | 0.952 |
## Precision
Total precision and within-run precision for the MDA Heparin anti-Xa assay were determined in accordance with NCCLS guideline EP5-T2. Controls were tested in duplicate on two MDA instruments twice daily. Instruments were calibrated using duplicate determinations at the beginning of each week. For each instrument, data were collected for 19-21 days, with a minimum of 38 runs and 76 measurements at each control level per instrument. The following precision was observed:
| Sample | Mean (IU/ml) | SD(within-run) (IU/ml) | CV(within-run) % | SD(total) (IU/ml) | CV(total) % |
| --- | --- | --- | --- | --- | --- |
| | | | | | |
| 1 | 0.193 | 0.012 | 6.1 | 0.018 | 9.6 |
| 2 | 0.389 | 0.018 | 4.5 | 0.026 | 6.7 |
(b) (3) The conclusion drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a) (3).
In conclusion, the MDA™ Heparin anti-Xₐ has successfully met all aspects of clinical testing and have demonstrated that the device is safe and effective and has performed well and is substantially equivalent to the legally marketed device [Stachrom® Heparin (K925433)].
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.