CRYO CHECK FACTOR XI DEFICIENT PLASMA, MODEL FDP11-10 (1.OML) AND FDP11-15 (1.5ML)
K981173 · Precision Biologicals, Inc. · GJT · Jan 13, 1999 · Hematology
Device Facts
Record ID
K981173
Device Name
CRYO CHECK FACTOR XI DEFICIENT PLASMA, MODEL FDP11-10 (1.OML) AND FDP11-15 (1.5ML)
Applicant
Precision Biologicals, Inc.
Product Code
GJT · Hematology
Decision Date
Jan 13, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.7290
Device Class
Class 2
Indications for Use
Cryov Check™ Factor XI Deficient Plasma is recommended for use as a substrate in clot-based Factor XI assays using the activated partial thromboplastin time (APTT).
Device Story
Cryo Check™ Factor XI Deficient Plasma is a frozen human plasma product used as a substrate in clinical laboratory clot-based Factor XI assays. It is prepared from citrated pooled normal human plasma depleted of Factor XI via immunoadsorption; Factor XI activity is <1% of normal levels, while other coagulation factors remain within normal ranges. The device is used by laboratory technicians in clinical settings to perform APTT-based coagulation testing. By providing a standardized, Factor XI-deficient environment, the product allows for the accurate quantification of Factor XI in patient samples. This aids clinicians in diagnosing coagulation factor deficiencies, supporting the assessment of hemostatic function.
Clinical Evidence
Bench testing only. Open vial stability study performed on 5 samples. Factor XI assays conducted at 0 and 24 hours showed mean recovery of 97.2% and 96.2% respectively, with a total average recovery of 96.7% and C.V. < 5%. Results support an 8-hour open vial stability claim.
Technological Characteristics
Frozen human plasma; Factor XI depleted via immunoadsorption; Factor XI activity <1%; other coagulation factors >normal range. Standalone in vitro diagnostic reagent.
Indications for Use
Indicated for use as a substrate in clot-based Factor XI assays using APTT. Intended for laboratory use in the diagnosis of Factor XI deficiency, which may have congenital or acquired etiologies.
K981174 — CRYO CHECK FACTOR XII DEFICIENT PLASMA MODEL FDP12-10 (1.0ML) AND FDP12-15 (1.5ML) · Precision Biologicals, Inc. · Jan 13, 1999
K972286 — FACTOR DEFICIENT PLASMAS - FACTOR XI (11) · Universal Reagents, Inc. · Jul 14, 1997
K030287 — HEMOSIL FACTOR XI DEFICIENT PLASMA · Instrumentation Laboratory CO · Mar 19, 2003
K971227 — CRYO CHECK FACTOR X DEFICIENT PLASMA · Precision Biologicals, Inc. · Jul 18, 1997
K971225 — CRYO CHECK FACTOR VIII DEFICIENT PLASMA · Precision Biologicals, Inc. · Jul 17, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
LG81173
JAN 1 3 1999
## 510(k) Summary for Cryov Check™ Factor XI Deficient Plasma
- Submitter's Address and Contact Information 1.
- Precision Biologicals Incorporated a) 900 Windmill Rd. Unit # 100 Dartmouth, Nova Scotia Canada B3B 1P7
- b) Contact
Mr. Sandy Morrison Manager, Technical Operations Phone: (902) 468-6422 (902) 468-6421 Fax: E-mail: pbi@fox.nstn.ca
- Date Prepared: March 09, 1998 c)
- Device Name 2.
| a) | Proprietary (trade) name: | Cryo✓Check™ Factor XI Deficient Plasma |
|----|-----------------------------|------------------------------------------------------------------|
| b) | Common name: | Factor XI Deficient Plasma (human) |
| c) | Classification name: | Coagulation Factor Deficient Plasma |
| d) | Classification information: | Regulatory Class II<br>Hematology Panel<br>Product Code - 81 GJT |
## 3. Device Description:
Cryo Check™ Factor XI Deficient Plasma is frozen human plasma deficient in the Factor XI coagulation factor. It is prepared from citrated pooled normal human plasma which has been depleted of Factor XI by immunoadsorption. Activity levels of Factor XI are assayed at less than 1% normal levels while all other coagulation factors are within normal ranges.
V - 1
{1}------------------------------------------------
## Intended Use 4.
b)
Cryov Check™ Factor XI Deficient Plasma is recommended for use as a substrate in clot-based Factor XI assays using the activated partial thromboplastin time (APTT).
- Substantially Equivalent Device న్.
- a) 510(k) number: K900411
- Factor XI Deficient Plasma Trade Name: c)
- d) Manufacturer: Sigma
- Substantial Equivalence Comparison e)
Cryo Check™ Factor XI Deficient Plasma is similar to the predicate device in that they both have the same "indications for use"; target population; and are both made from human plasma.
Cryo Check™ Factor XI Deficient Plasma differs from the predicate device in that it is a frozen liquid preparation and not a lyophillized Additionally, CryovCheck™ Factor XI Deficient Plasma is product. prepared from normal human plasma from which Factor XI has been immunoadsorbed, while the predicate device is derived from human donors with a congenital Factor XI deficiency.
To our knowledge, these differences do not affect the intended use or performance of the device.
- Non-Clinical Performance Data 24 Hour Open Vial Stability : 6.
- a) Testing Performed:
- i) Factor XI assays were performed on a known reference plasma using vials of Cryo Check™ Factor XI Deficient Plasma as a substrate. Recovered factor XI values were measured at 0 hours and 24 hours. (see table for results)
- b) Conclusions:
Test results indicate that a claim of 8 hours open vial stability is acceptable.
{2}------------------------------------------------
| Table S1 | | | |
|------------------------------------|---------|----------|---------|
| Open Vial Stability of Cryo✓Check™ | | | |
| Factor XI Deficient Plasma | | | |
| Summary Statistics (% Recovery) | | | |
| | 0 Hours | 24 Hours | Average |
| MEAN | 97.2 | 96.2 | 96.7 |
| MAXIMUM | 99 | 99 | 99 |
| MINIMUM | 95 | 93 | 93 |
| S.D. | 1.48 | 2.28 | 1.89 |
| 2 S.D. | 2.97 | 4.56 | 3.78 |
| SAMPLE SIZE | 5 | 5 | 10 |
| C.V.% | 1.53 | 2.37 | 1.95 |
Note: Reference Value =
.
Acceptable values are: Mean (+/-) 5% of reference value; and %C.V. < 5%
{3}------------------------------------------------
JAN 1 3 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Sandy Morrison Manager, Technical Operations Precision Biologicals Incorporated 900 Windmill Road Unit # 100 Dartmouth, Nova Scotia CANADA B3B 1P7
Re: K981173
Trade Name: Cryo Check™ Factor XI Deficient Plasma Regulatory Class: II Product Code: GJT Dated: March 27, 1998 Received: April 1, 1998
Dear Mr. Morrison:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number: KG81173
Device Name: CryovCheck™ Factor XII Deficient Plasma
## Indications for Use
Deficiencies in coagulation factors may have congenital or acquired etiologies and can compromise in vivo hemostasis. Factor XII (also known as Hageman factor) is a serine protease located in the "intrinsic coagulation pathway". Patients with Factor XII deficiency generally do not have significant bleeding tendencies. Factor XII deficiency is commonly diagnosed in vitro through the use of the activated partial thromboplastin time (APTT).
CryovCheck™ Factor XII Deficient Plasma is human plasma deficient in the Factor XII coagulation protein while having all other coagulation factors greater than 50%. It is recommended for use as a substrate in clot-based Factor XII assays using the activated partial thromboplastin time (APTT).
Peter E. Marki
(Division Sign-Off)
Division of Clinical Laboratory Devices K451173
510(k) Number
Recyphoi
Use ✓
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.