The CoaguChek XS System is intended for properly selected and suitably trained users or their caregivers on the prescription or other order of the treating doctor. Users should be stabilized on anticoagulation medications such as Coumadin® or warfarin prior to self-testing with the CoaguChek XS System. The system uses capillary blood.
Device Story
CoaguChek XS System is a point-of-care diagnostic device for home monitoring of prothrombin time (PT) in patients on anticoagulation therapy. System consists of a handheld meter and disposable test strips containing human recombinant tissue factor. User applies capillary blood sample to test strip; electrochemical sensor detects thrombin activity via amperometric measurement. Meter converts electrical signal to INR value. Device includes on-board quality control (OBC) to verify strip integrity; if QC fails, no result is displayed. Meter stores up to 100 results with timestamps. Patients or caregivers operate device at home; results assist in managing anticoagulation therapy. Data transfer via infrared communication. System provides patient-side monitoring, reducing need for frequent clinical visits.
Clinical Evidence
Clinical study compared self-trained patient results (N=258) against healthcare professional results using the same meter. Results showed high correlation (Slope=1.00, Intercept=0.0, r=0.974). Precision study (N=222 patients, N=257 professionals) demonstrated comparable CVs (5.19% vs 5.38%) and SDs (0.132 vs 0.135), confirming self-testing accuracy and precision are equivalent to professional use.
Technological Characteristics
Electrochemical system with amperometric detection of thrombin activity. Test strips utilize human recombinant tissue factor calibrated to ISI 1.0. Meter features infrared communication, memory for 100 results, and bi-level on-board quality control (OBC) integrated into test strips. OBC monitors for reagent defects, capillary compression, electrode defects, and environmental damage (humidity/light/temperature). Code chip provides lot-specific calibration data.
Indications for Use
Indicated for patients on anticoagulation therapy (e.g., Coumadin, warfarin) requiring prothrombin time (INR) monitoring. Intended for use by properly selected, trained patients or caregivers in home settings upon physician prescription. Patients must be stabilized on medication prior to use.
Regulatory Classification
Identification
A prothrombin time test is a device used as a general screening procedure for the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway, which involves the reaction between coagulation factors III and VII, and to monitor patients receiving coumarin therapy (the administration of one of the coumarin anticoagulants in the treatment of venous thrombosis or pulmonary embolism).
Predicate Devices
Roche Diagnostics CoaguChek XS System (professional) (K060978)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K062925
B. Purpose for Submission:
Clearance of a new instrument and test strip
C. Measurand:
Prothrombin Time
D. Type of Test:
Electrochemical
E. Applicant:
Roche Diagnostics
F. Proprietary and Established Names:
CoaguChek® XS System (Patient Self-testing)
G. Regulatory Information:
1. Regulation section:
21 CFR 864.7750
2. Classification:
Class II
3. Product code:
GJS
4. Panel:
81 Hematology
{1}
2
H. Intended Use:
1. Intended use(s):
The CoaguChek XS System is intended for properly selected and suitably trained users or their caregivers on the prescription or other order of the treating doctor. Users should be stabilized on anticoagulation medications such as Coumadin® or warfarin prior to self-testing with the CoaguChek XS System. The system uses capillary blood.
2. Indication(s) for use:
3. Special conditions for use statement(s):
Intended for home use
4. Special instrument requirements:
I. Device Description:
The CoaguChek® XS System includes a meter and CoaguChek® XS PT test strips. The test strip contains a human recombinant tissue factor, and is calibrated to an ISI of 1.0.
The test strip incorporates quality control material that accesses strip integrity.
The CoaguChek® XS meter automatically stores up to 100 test results along with their dates and times in memory.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Roche Diagnostics CoaguChek XS System
2. Predicate 510(k) number(s):
K060978
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Measure prothrombin time | same |
| Technology | Electrochemical with amperometric detection of thrombin activity | same |
| Dosing | Top and side dosing | same |
{2}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Indications | Home users | Professional use |
## K. Standard/Guidance Document Referenced (if applicable):
## L. Test Principle:
When a blood sample is applied to the test strip, thromboplastin activates the coagulation cascade which leads to the formation of thrombin. Thrombin cleaves the thrombin substrate creating an electrochemically active peptide, which generates an electrical signal. The signal is converted to an INR value and displayed by the CoaguChek XS System.
The on-board quality control is a bi-level control that accesses test strip integrity. The PT test and QC testing are performed simultaneously. The test system determines whether the quality control is within preset limits. If it is, the meter displays a short term “QC√”, and then the PT test result. If the QC is not within limits, the meter displays “error QC”, and no PT test result will be displayed.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Precision of duplicates for capillary blood results was calculated for both trained users and healthcare professional. The following results were obtained:
| | User results | Professional Results |
| --- | --- | --- |
| N | 214 | 249 |
| Mean | 2.57 | 2.52 |
| SD | 0.13 | 0.13 |
| CV | 5.13 | 5.36 |
#### b. Linearity/assay reportable range:
Established under K060978
#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Established under K060978
#### d. Detection limit:
{3}
e. Analytical specificity:
Established under K060978
f. Assay cut-off:
2. Comparison studies:
a. Method comparison with predicate device:
Study compared test results obtained by trained users to results obtained by healthcare professionals using the CoaguChek XS meter. The patient tested once a week at home for 8 weeks, with 3 scheduled visits to their study site. At the site visit the patient performed two tests on the CoaguChek XS meter, and the healthcare professional performed two tests on the subject using the subject's meter and supplies. A venous reference sample (3.2% sodium citrate) was collected on site visits 2 and 4 and the healthcare professional checked the setup of the meter. 107 patients were enrolled, and a total of 91 patients completed all four visits. Comparing test results obtained by patients with results obtained by the healthcare professional demonstration good correlation (y = 1.00x-0, r= 0.977, n=463). Test results obtained by the patient was also compared to a laboratory-based plasma method with good results (y = 1.154x - 0.2, r = 0.934, n=297).
b. Matrix comparison:
3. Clinical studies:
a. Clinical Sensitivity:
b. Clinical specificity:
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
5. Expected values/Reference range:
Established under K060978
N. Instrument Name:
CoaguChek XS meter
O. System Descriptions:
1. Modes of Operation:
{4}
Manual, closed system
2. Software:
The CoaguChek XS software controls the user interface, buttons used to navigate the user interface and to configure the device, storage of patient results, transfer of measurement results using serial infrared communication in an open mode, transfer of production and calibration results using serial infrared communication in a protected mode, reading and storage of specific information for strip LOT from code key, calculation of PT time based on data received from measurement cycle, and checks for failsafe in order to recognize malfunctions of the measurement electronics or malfunctions within the strip used for testing.
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☑ or No ☐
3. Specimen Identification:
Date and time of testing is recorded by the CoaguChek XS meter
4. Specimen Sampling and Handling:
Whole blood is manually applied to the target area of the test strip either from the top or side of the strip.
5. Calibration:
The CoaguChek XS Test strips are calibrated to a master reagent lot which has in turn been calibrated to a WHO International Reference Preparations (rTF/95) using the manual tilt tube method.
6. Quality Control:
The CoaguChek XS System incorporates a bi-level on-board quality control (OBC) within the CoaguChek XS test strip that monitors test strip integrity.
Level 1 OBC detects strip defects such as reagent defects, capillary compression and electrode defects. Level 2 OBC directly measures strip damage due to such things as exposure to increased humidity, light, and temperature.
The pre-determined OBC ranges are programmed into the lot specific code chip that is packaged with the matching test strip lot.
Acceptable data was presented validating the OBC.
5
{5}
P. Other Supportive Instrument Performance Characteristics Data Not Covered In the "Performance Characteristics" Section above:
An instrument failsafe checklist was presented outlining the QC check made by the CoaguChek XS meter.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
6