← Product Code [GJS](/submissions/HE/subpart-h%E2%80%94hematology-kits-and-packages/GJS) · K020679

# INRATIO (K020679)

_Hemosense, Inc. · GJS · May 6, 2002 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HE/subpart-h%E2%80%94hematology-kits-and-packages/GJS/K020679

## Device Facts

- **Applicant:** Hemosense, Inc.
- **Product Code:** [GJS](/submissions/HE/subpart-h%E2%80%94hematology-kits-and-packages/GJS.md)
- **Decision Date:** May 6, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.7750
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

The HemoSense INRatio is an in vitro diagnostic system that provides a quantitative prothrombin time results, expressed in seconds and as an International Normalized Ratio (INR). It uses fresh capillary whole blood and is intended for use by health care professionals in monitoring patients who are on warfarin-type (coumarin) anticoagulation therapy. This device is not intended to be used for screening purposes.

## Device Story

INRatio system monitors PT/INR in patients on warfarin therapy; used by healthcare professionals in point-of-care settings. Input: drop of fresh capillary whole blood applied to test strip. Operation: blood drawn into three-channel test strip; one channel performs PT test via reagent-induced coagulation; two channels perform internal control tests. Detection: electrodes monitor reaction; change in sample impedance indicates fibrin formation. Output: quantitative PT (seconds) and INR displayed on meter. Internal controls verify test integrity; error message displayed if controls fail. Benefits: provides rapid, point-of-care anticoagulation monitoring without external quality control requirements.

## Clinical Evidence

Field studies compared INRatio accuracy against CoaguChek S and a reference method; correlation coefficient r>0.90. Precision and linearity evaluations performed with acceptable results. Additional testing conducted on interfering substances and hematocrit effects.

## Technological Characteristics

In vitro diagnostic system; uses test strips with three channels containing reagents and electrodes. Sensing principle: impedance change detection of fibrin formation. Standalone meter with digital display. No external quality control required.

## Regulatory Identification

A prothrombin time test is a device used as a general screening procedure for the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway, which involves the reaction between coagulation factors III and VII, and to monitor patients receiving coumarin therapy (the administration of one of the coumarin anticoagulants in the treatment of venous thrombosis or pulmonary embolism).

## Predicate Devices

- Roche CoaguChek S ([K013488](/device/K013488.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K020679

MAY 0 6 2002

Appendix 10. Summary

## PREMARKET NOTIFICATION 510(k) Summary

Introduction:

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

- 1. Submitter's name, HemoSense. Inc. 600 Valley Way address, contact: Milpitas, CA 95035 (408) 719-1393 (phone) (408) 719-1184 (fax)
Contact person: Judith Blunt Date Prepared: February 28, 2002

- 2. Device name: Common or usual name: Prothrombin Time Test
Classification Name: Prothrombin Time Test

Trade or Proprietary Name: INRatio

- The Roche CoaguChek STM system: device for testing 3. Predicate device Prothrombin Time and INR in whole blood.
- 4. Device description: The INRatio Prothrombin Time Monitoring System uses a modified version of the one-stage prothrombin time test. After a drop of blood is applied to the test strip, it is drawn into the test area and mixed with reagents that cause coagulation to begin. The test area on the test strip is separated into three channels, each of which contains electrodes for detection of the blood clot. One channel contains the reagents to perform the prothrombin time test on the blood sample. The other two channels contain the reagents to run control tests. No external quality control tests are required for the INRatio System. The meter monitors the reactions, and calculates the PT and INR for the blood sample, which are reported on the display. If the control results are not within a set

{1}------------------------------------------------

range, it indicates a problem with the test and an error message is reported on the display instead of a result.

- The INRatio is an in vitro diagnostic system that 5. Intended use: provides a quantitative prothrombin time result, expressed in seconds and as an International normalized ratio (INR). It is intended for use by healthcare professionals in monitoring patients who are on warfarin-like (coumarin) anticoagulation therapy. This device is not intended to be used for screening purposes.
- The INRatio is substantially equivalent in 6. Comparison to materials, design and intended use to other products predicate device: that measure Prothrombin Time in human blood. Most notably, it is substantially equivalent to the CoaguChek S™, manufactured by Roche Diagnostics. Both products are prothrombin time devices, have the same intended use and serve the same professional, point-of-care market. The HemoSense INRatio and the CoaguChek both measure the reaction of the extrinsic coagulation pathway. Both products measure the elapsed time between the start of the reaction and the formation of fibrin from fibrinogen; the CoaguChek measure fibrin formation via the motion of magnetic particles, while the INRatio measures it via a change in the impedance of the sample.
- The accuracy of the INRatio was compared 7. Summary of to the CoaguChek S and a reference method in field Performance data: studies and found to be equivalent (r>0.90). Precision and linearity evaluations were performed on the INRatio and the results were found acceptable. Additional testing on interfering substances, and hematocrit were performed and the results are reflected in the product labeling.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or movement.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Judith Blunt Director of Clinical Affairs & Customer Support HemoSense, Inc. 600 Valley Way Milpitas, California 95035

**MAY 06 2002**

k020679 Re:

> Trade/Device Name: INRatioTM Regulation Number: 21 CFR § 864.7750 Regulation Name: Prothrombin Time Test Regulatory Class: II Product Code: GJS Dated: February 28, 2002 Received: March 4, 2002

Dear Ms. Blunt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Appendix 11. Indications for Use Statement

Feb 28, 2002

## PREMARKET NOTIFICATION INDICATIONS FOR USE STATEMENT (As required by 21 CFR 807.87(j))

510(K) Number:

Device Name: HemoSense INRatio PT Monitoring System

The HemoSense INRatio is an in vitro diagnostic system that provides a quantitative prothrombin time results, expressed in seconds and as an International Normalized Ratio (INR). It uses fresh capillary whole blood and is intended for use by health care professionals in monitoring patients who are on warfarin-type (coumarin) anticoagulation therapy. This device is not intended to be used for screening purposes.

PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED

| Signed:   | Judith Blunt                                     |
|-----------|--------------------------------------------------|
| Name:     | Judith Blunt                                     |
| Position: | Director of Clinical Affairs and Product Support |
| Date:     | Feb. 28, 2002                                    |

on Bign-Off

510(K) Number

Division of Clinical Laboratory Devices

000564

---

**Source:** [https://fda.innolitics.com/submissions/HE/subpart-h%E2%80%94hematology-kits-and-packages/GJS/K020679](https://fda.innolitics.com/submissions/HE/subpart-h%E2%80%94hematology-kits-and-packages/GJS/K020679)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
