ACCUCLOT THROMBIN TIME REAGENT MODEL A8713/A4589

K982209 · Sigma Diagnostics, Inc. · GJA · Oct 15, 1998 · Hematology

Device Facts

Record IDK982209
Device NameACCUCLOT THROMBIN TIME REAGENT MODEL A8713/A4589
ApplicantSigma Diagnostics, Inc.
Product CodeGJA · Hematology
Decision DateOct 15, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7875
Device ClassClass 2

Indications for Use

Sigma Diagnostics ACCUCLOT™ Thrombin Time Reagent (A8713/A4589) is a device intended to measure fibrinogen concentration and to detect fibrin or fibrinogen split products for the evaluation of bleeding disorders.

Device Story

ACCUCLOT™ Thrombin Time Reagent is an in vitro diagnostic reagent used to measure thrombin clotting time (TT). It provides thrombin to catalyze the conversion of fibrinogen to fibrin. The reagent is used in clinical laboratory settings by trained personnel on coagulation analyzers (e.g., Amelung KC-4A, CS-190). The device detects functional fibrinogen and inhibitors like heparin, fibrin/fibrinogen degradation products, or myeloma proteins by measuring the prolongation of clotting time. Results assist clinicians in diagnosing bleeding disorders, hypofibrinogenemia, afibrinogenemia, and dysfibrinogenemia. It serves as a screening tool for hemostasis and thrombosis, helping to identify coagulation abnormalities.

Clinical Evidence

Bench testing only. Correlation studies compared ACCUCLOT™ to the predicate using 101 samples on AMAX analyzers and 49 samples on KC4a. Correlation coefficients (r2) ranged from 0.959 to 0.978. Precision studies showed within-run CV <7% and total precision <6%. Sensitivity demonstrated for heparin levels 0.2–0.8 U/mL and low fibrinogen levels starting at 98 mg/dL.

Technological Characteristics

In vitro diagnostic reagent containing thrombin for fibrinogen hydrolysis. Formulated for use on photo-optical and mechanical coagulation analyzers. No specific materials of construction or software algorithms described; functions as a biochemical assay reagent.

Indications for Use

Indicated for use as a screening procedure to detect functional fibrinogen and thrombin inhibitors (e.g., heparin) in citrated plasma for the evaluation of bleeding disorders, including disorders of thrombosis and hemostasis.

Regulatory Classification

Identification

A thrombin time test is a device used to measure fibrinogen concentration and detect fibrin or fibrinogen split products for the evaluation of bleeding disorders.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ OCT 1 5 1998 K982209 ## Summarv of Safety and Effectiveness ACCUCLOT™ Thrombin Time Reagent (Product No. A8713, A4589) Sigma Diagnostics' ACCUCLOT™ Thrombin Time Reagent is intended for use as a screening procedure to detect functional fibrinogen and thrombin inhibitors such as heparin, in citrated plasma. The thrombin clotting time (TT) is an important screening procedure for disorders of thrombosis and hemostasis as well as the presence of heparin.1-3 Thrombin Time testing is a rapid assay procedure that measures the polymerization of fibrinogen to fibrin. Any interference with this conversion will be reflected in a prolongation in the clotting time of the test procedure. The ACCUCLOT™ Thrombin Time Reagent provides thrombin for fibrinogen hydrolysis and is able to detect normal fibrinogen function even in the presence of severe coagulation abnormalities such as hemophilia (Factor VIII or Factor IX deficiency) or a decrease in factor levels associated with vitamin K deficiency. Abnormal thrombin clotting times occur in cases of hypofibrinogenemia. afibrinogenemia, or dysfibrinogenemia. The TT is also prolonged in the presence of inhibitors such as heparin, myeloma proteins and fibrin/fibrinogen degradation products, which block either thrombin cleavage of fibrinopeptides or fibrin monomer polymerization. 3-5 Sigma Diagnostics' ATROXIN® assay (Catalog # 845-2) can provide data to rule out the presence of heparin in samples with elongated thrombin clotting times. The safety and effectiveness of the Sigma Diagnostics' ACCUCLOT™ Thrombin Time Reagent has been demonstrated by showing its substantial equivalence to Organon's Thromboquik™ Thrombin Time Reagent. Both Thrombin Time reagents were compared using normal and abnormal plasma samples. Sigma's ACCUCLOT™ Thrombin Time Reagent (y) was used on an Amelung KC-4A, an CS-190 (optically and mechanically) and Organon Teknika's AMAX Thromboquik™ (x) was used on a photo-optical analyzer, the Coag-A-Mate X2. Results of these correlation studies are presented in the table below. | Instrument | Equation | r2 | N | |-----------------|---------------|-------|-----| | AMAX Optical | Y=2.11x-13.33 | 0.978 | 101 | | AMAX Mechanical | Y=2.87x-18.10 | 0.972 | 101 | | KC4a | Y=1.95x-7.26 | 0.959 | 49 | Precision studies demonstrated a within run CV of less than 7% and a total {1}------------------------------------------------ precision of less than 6%, using normal and abnormal control plasmas. Sigma Diagnostics' ACCUCLOT™ Thrombin Time Reagent is sensitive to heparin levels corresponding to 0.2 heparin U/mL and up to at least 0.8 heparin U/mL. The reagent also, demonstrated sensitivity for abnormal low fibrinogen levels starting at 98 mg/dL fibrinogen. - 1. Coleman RW, Hirsh J. Marder VJ, Salzman EW: Hemostasis and Thrombosis_2nd ed. J B Lippincott Co., Philadelphia, 1987 - 2. Jarret L. Sonnenworth AC: Gradwohl's Clinical Laboratory Methods and Diagnosis, Vol 1, 2nd ed. C V Mosby Co., St. Louis (MO), 1977 - 3. Miale JB: Laboratory Medicine Hematology, 5th ed. C V Mosby Co., St. Louis (MO), 1977 - 4. Williams WT, Beutler E, Erslev AJ, Rundles RW: Hematology, 2nd ed. _McGraw-Hill, New York, 1977 - 5. Wintrobe MM, Lee GR, Boggs DR, Bitchell JC, Foerster J, Athens JW, Lukens JN: Clinical Hematology, 8th ed. Lea & Febiger, Philadelphia, 1981 {2}------------------------------------------------ Image /page/2/Picture/2 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. OCT 1 5 1998 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 William R. Gilbert, Ph.D. Manager, Scientific Affairs Siqma Diagnostics 545 South Ewing Avenue St. Louis, Missouri 63103 Re: K982209/S1 Trade Name: ACCUCLOT Thrombin Time Reagent Regulatory Class: II Product Code: GJA Dated: September 17, 1998 Received: September 18, 1998 Dear Dr. Gilbert: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Putman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page ! of I ## 510(k) Number (if known): ___ Device Name: _ Sigma Diagnostics ACCUCLOT™ Thrombin Time Reagent (A8713/A4589) ## Indications For Use: Sigma Diagnostics ACCUCLOT™ Thrombin Time Reagent (A8713/A4589) is a device intended to measure fibrinogen concentration and to detect fibrin or fibrinogen split products for the evaluation of bleeding disorders. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Teter E. Marlini (Division Sign-Off) Division of Clinical Laboratory Devices / 4982209 510(k) Number **Prescription Use** √ (Per 21 CFR 801.109) - ) Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...