HEMOSIL VON WILLEBRAND ACTIVITY

K040843 · Instrumentation Laboratory CO · GGP · Apr 30, 2004 · Hematology

Device Facts

Record IDK040843
Device NameHEMOSIL VON WILLEBRAND ACTIVITY
ApplicantInstrumentation Laboratory CO
Product CodeGGP · Hematology
Decision DateApr 30, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7290
Device ClassClass 2

Indications for Use

The IL HemosIL vWF Activity Kit is an in vitro diagnostic automated latex enhanced immunoturbidometric assay for the quantitative determination of vWF activity in human citrated plasma on IL Coagulation Systems.

Device Story

The HemosIL vWF Activity Kit is an in vitro diagnostic assay used on IL Coagulation Systems (ACL 8000, 9000, 10000) to quantify vWF activity in human citrated plasma. The device utilizes polystyrene latex particles coated with anti-vWF monoclonal antibodies directed against the glycoprotein Ib receptor binding site. When mixed with patient plasma, the vWF in the sample causes latex particle agglutination. The degree of agglutination is proportional to vWF activity and is measured by the decrease in transmitted light (immunoturbidometry). The assay is performed in clinical laboratory settings by trained technicians. The resulting quantitative output allows clinicians to assess vWF activity levels, aiding in the diagnosis or monitoring of vWF-related deficiencies. The system provides automated analysis, replacing manual or ELISA-based methods with a faster, automated turbidimetric approach.

Clinical Evidence

Bench testing only. Precision evaluated per NCCLS EP5-A (N=84, CVs 1.7–9.0%). Linearity confirmed (R2 0.9845–0.9941) across 21–200% range. Method comparison (N=108) against Shield Diagnostics ELISA showed r=0.9723. Comparison (N=114) against Dade Behring BC vWF:RCo reagent showed r=0.9497. Reference ranges established using 248 healthy donors stratified by blood type.

Technological Characteristics

Latex particle enhanced immunoturbidometric assay. Components: lyophilized polystyrene latex particles coated with anti-vWF monoclonal antibody (hybridoma RFF-VIII:R/1) and Tris buffer. Energy source: optical detection (transmitted light) via IL Coagulation Systems. Calibration traceable to NIBSC WHO/ECBS 4th International Standard (Code: 97/586).

Indications for Use

Indicated for the quantitative determination of von Willebrand Factor Activity in human citrated plasma for patients requiring assessment of VWF activity levels. For professional in vitro diagnostic use on IL Coagulation Systems.

Predicate Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: #K040843 B. Purpose for Submission: New device C. Analyte: Von Willebrand Factor (vWF) D. Type of Test: Quantitative E. Applicant: Instrumentation Laboratory Co. F. Proprietary and Established Names: IL HemosIL vWF Activity Kit; vWF Deficiency Test G. Regulatory Information: 1. Regulation section: CFR Section 864.7290 – Factor Deficiency Test 2. Classification: Class II 3. Product Code: GGP – Quantitative Factor Deficiency Test 4. Panel: Hematology (81) H. Intended Use: 1. Intended use(s): The IL HemosIL vWF Activity Kit is an *in vitro* diagnostic automated latex enhanced immunoturbidometric assay for the quantitative determination of vWF activity in human citrated plasma on IL Coagulation Systems. 2. Indication(s) for use: Same as above 3. Special condition for use statement(s): N/A 4. Special instrument Requirements: IL Coagulation Systems (ACL 8000, 9000, and 10000 Analyzers) I. Device Description: The HemosIL vWF Activity Kit is a latex particle enhanced immunoturbidometric assay to quantify vWF activity in plasma. It consists of (1) a lyophilized suspension of polystyrene latex particles coated with purified anti-vWF mouse monoclonal antibody directed against a functional epitope of vWF, containing bovine serum albumin (BSA), stabilizers and preservative; and (2) Tris buffer containing BSA, stabilizers and preservative. {1} Page 2 of 4 # J. Substantial Equivalence Information: 1. Predicate device name(s): Shield Diagnostics, Ltd. vWF Activity ELISA; Dade Behring BC vWF:RCofactor Activity Reagent 2. Predicate K number(s): #K000398; #K972116 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Methodology | Immunoassay | Same | | Monoclonal antibody origin | Hybridoma cell line, RFF-VIII:R/1 | Same | | Calibration standard traceability | NIBSC WHO/ECBS 4^{th} International Std. for FVIII/vWF Activity in Plasma (Code: 97/586) | Same | | Sample type | Citrated plasma | Same | | Stability | 2° - 8° C. | Same | | Differences | | | | Item | Device | Predicate | | Test principle | Latex agglutination enhanced immunoturbidometry (ITA) | Enzyme-linked immunosorbency (ELISA) | | Detection limit | 12.0% | 1.6% | # K. Standard/Guidance Document Referenced (if applicable): IFCC Approved Recommendation (1987) on the Theory of Reference Values; NIBSC WHO/ECBS 4th International Standard for Factor VIII and vWF in Plasma (Code: 97/586); NCCLS EP5-A: Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline (1999); and NCCLS H21-A3: Collection, Transport and Processing of Blood Specimens for Testing Plasma-Based Coagulation Assays; Approved Guideline-4th Edition (2003). {2} Page 3 of 4 L. Test Principle: The HemosIL vWF Activity Kit uses a specific anti-vWF monoclonal adsorbed onto the latex reagent and directed against the platelet binding site of vWF (glycoprotein Ib receptor). When it reacts with the vWF in patient plasma, the degree of agglutination is directly proportional to the activity of vWF activity in the sample, and is determined by measuring the decrease in transmitted light caused by the aggregates. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Testing was performed on the ACL 9000, using duplicates of (3) levels of controls (N=84), each, according to NCCLS EP5-A Protocol. The HemosIL Normal Control had a mean of 79.6%; the Level 1, a mean of 49.8%; and the Level 2, a mean of 27.7%. They all met the acceptance criteria with %CV's ranging from 1.7 – 9.0%. b. Linearity/assay reportable range: Reagent, (3) lots and (5) replicates of Calibration Plasma/Factor Diluent were used in dilutions to obtain the assay range. Reported values vs expected values yielded R2 values ranging 0.9845 – 0.9941. The activity (%) ranged 19.4 – 275.5 to support the claim of 21 – 200%. c. Traceability (controls, calibrators, or method): Value assignment was made from multiple runs against a house standard traced to the NIBSC WHO/ECBS 4th International Standard for FVIII/vWF Activity in Plasma (Code :97/586). d. Detection limit: Replicates (10), of (3) lots of reagent, diluted with Diluent, yielded 12.2% (the greater of the mean + 2SD). e. Analytical specificity: Pooled normal and abnormal plasmas, (2) each, were spiked with varying levels of interferents, bilirubin and lipids. Bilirubin, at 3.8 mg/dL, and lipids, at 265 mg/dL, met the acceptance criteria of ±15% recovery of the unspiked value. f. Assay cut-off: N/A {3} Page 4 of 4 2. Comparison studies: a. Method comparison with predicate device: The company provided an in-house comparison study, performed on the ACL 9000, using citrated samples (N=108) obtained from (3) hospitals. Samples ranged from 0 – 192% vWF activity. HemosIL vs Shield Diagnostics vWF Activity ELISA yielded these values: $$ \mathrm{r} = 0.9723; \mathrm{S} = 0.832; \mathrm{I} = 4.782 $$ Duplicate vs singlet testing was also performed, and yielded values, as follows, to demonstrate no significant difference between the two: $$ \mathrm{r} = 0.9710; \mathrm{S} = 0.849; \mathrm{I} = 4.207 $$ b. Matrix comparison: N/A 3. Clinical studies: a. Clinical sensitivity: N/A b. Clinical specificity: N/A c. Other clinical supportive data (when a and b are not applicable): The company also provided a clinical comparison of the HemosIL Reagent vs a reference assay, Dade Behring BC vWF:RCo Reagent. Duplicate testing of samples (N = 114) was performed on the ACL 9000 at the Coagulation Lab in Hamburg, Germany. Samples ranged from 0 – 213% vWF activity. They yielded values of: $$ \mathrm{r} = 0.9497; \mathrm{S} = 0.8406; \mathrm{I} = 9.7557 $$ 4. Clinical cut-off: N/A 5. Expected values/Reference range: Using IFCC guidelines for determining reference values, the company tested healthy blood donors (N = 248) from the Hospital Vall d’Hebron in Barcelona, Spain. Type “O” (N = 122), with M = 51 and F = 71; and Types “A”, “B” and “AB” (N = 126), with M = 59 and F = 67 yielded the following results: | | Type “O” | | Other types | | | --- | --- | --- | --- | --- | | | 95% Limit | 90% C.I. | 95% Limit | 90% C.I. | | Lower | 38.03 | 34.65 – 41.05 | 49.19 | 47.55 – 57.85 | | Upper | 125.24 | 116.00 - 142.50 | 169.71 | 156.50 – 182.50 | N. Conclusion: The submitted material in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...