VIDAS® D-Dimer Exclusion II™ is an automated quantitative test for use on the instruments of the VIDAS family for the immunoenzymatic determination of fibrin degradation products (FbDP) in human plasma (sodium citrate, CTAD) using the ELFA technique (Enzyme Linked Fluorescent Assay). VIDAS D-Dimer Exclusion II is indicated for use in conjunction with a clinical pretest probability assessment model to exclude deep vein thrombosis (DVT) and pulmonary embolism (PE) disease in outpatients suspected of DVT or PE.
Device Story
VIDAS® D-Dimer Exclusion II is an automated, quantitative in vitro diagnostic assay for measuring fibrin degradation products (FbDP) in human plasma. The device utilizes the Enzyme-Linked Fluorescent Assay (ELFA) technique on VIDAS family instruments. Operation involves a two-step sandwich immunoassay: sample is drawn into a Solid Phase Receptacle (SPR) coated with anti-FbDP antibodies; antigen binds to the SPR; unbound components are washed away. A conjugate containing alkaline phosphatase-labeled anti-FbDP monoclonal antibody is added to form a sandwich. After further washing, a substrate (4-Methyl-umbelliferyl phosphate) is introduced; the enzyme catalyzes its conversion to a fluorescent product (4-Methyl-umbelliferone). Fluorescence intensity, measured at 450 nm, is proportional to antigen concentration. The instrument automatically calculates results against a stored calibration curve. Used in clinical laboratories, the assay provides quantitative results to clinicians to help rule out DVT or PE in suspected outpatients, potentially reducing the need for further diagnostic imaging.
Clinical Evidence
Clinical evaluation used 315 banked, frozen plasma samples from patients suspected of VTE. Sensitivity was 100% (95% CI: 95.1%-100%) and specificity was 35.7% (95% CI: 29.6%-42.1%) at the 500 ng/mL FEU cut-off. NPV was 100% (95% CI: 95.8%-100%). Comparison with predicate device showed high concordance. Bench testing included precision (CV <7%), linearity (45-10,000 ng/mL), and interference studies (hemoglobin, lipids, bilirubin, albumin, and 47 drugs).
Indicated for outpatients suspected of deep vein thrombosis (DVT) or pulmonary embolism (PE) to exclude disease when used in conjunction with a clinical pretest probability assessment model.
Predicate Devices
VIDAS® D-dimer Exclusion Assay (k040882)
Related Devices
K020810 — VIDAS D-DIMER NEW (DD2) ASSAY MODEL#30 442 · bioMerieux, Inc. · Oct 3, 2002
K040882 — VIDAS D-DIMER EXCLUSION ASSAY, MODEL 30 442 · bioMerieux, Inc. · Jun 10, 2004
K973819 — VIDAS D-DIMER (DD) ASSAY · Biomerieux Vitek, Inc. · Mar 10, 1998
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k112818
B. Purpose for Submission:
Clearance of a new device
C. Measurand:
D-dimer
D. Type of Test:
Enzyme Linked Fluorescent Assay (ELFA)
E. Applicant:
bioMérieux, Inc.
F. Proprietary and Established Names:
VIDAS® D-dimer Exclusion II Assay (DEX2)
G. Regulatory Information:
1. Regulation section:
21 CFR 864.7320, Fibrinogen/fibrin degradation products assay
2. Classification:
Class II
3. Product code:
DAP, Fibrinogen and Fibrin Split Products, Antigen, Antiserum, Control
4. Panel:
81 Hematology
H. Intended Use:
1. Intended use(s):
VIDAS D-Dimer Exclusion™ II is an automated quantitative test for use on the VIDAS instruments for the immunoenzymatic determination of fibrin degradation products (FbDP) in human plasma (sodium citrate, CTAD) using the ELFA technique (Enzyme Linked Fluorescent Assay). VIDAS D-Dimer Exclusion™ II is indicated for use in conjunction with a clinical pretest probability assessment model to exclude deep vein thrombosis (DVT) and pulmonary embolism (PE) disease in outpatients suspected of DVT or PE.
{1}
2. Indication(s) for use:
Same as intended use.
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
VIDAS (k891385)
miniVIDAS (k923579)
I. Device Description:
The VIDAS® D-dimer Exclusion™ II is a quantitative assay for the immunoenzymatic determination of fibrin degradation products in human citrated plasma using the ELFA technique. Each assay kit contains reagents for 60 tests, and consists of the following:
DEX2 Reagent Strips (STR) - ready to use, polypropylene strip of 10 wells covered with a labeled foil seal. The first well is intended for the sample, and the last well is a cuvette in which the fluorometric reading is performed. The wells in between contain the various reagents required for the assay.
DEX2 Solid Phase Receptors (SPR) - The SPR is coated with alkaline phosphatase labeled mouse anti-D-Dimer monoclonal antibody.
C1 and C2 controls – bi-level lyophilized FbDP controls obtained from human plasma
S1 calibrator – lyophilized FbDP calibrator obtained from human plasma
R1 diluent- Buffer containing preservatives
Master Lot Entry (MLE) Card-the MLE is printed with barcode readable data needed for establishing the Master Curve for that assay.
J. Substantial Equivalence Information:
1. Predicate device name(s):
VIDAS® D-dimer Exclusion Assay
2. Predicate 510(k) number(s):
k040882
3. Comparison with predicate:
{2}
| Similarities | | |
| --- | --- | --- |
| Item | DeviceVIDAS® D-dimerExclusion™ II Assay | PredicateVIDAS® D-dimer ExclusionAssay |
| Intended use | VIDAS D-Dimer Exclusion™II is an automated quantitative test for use on the VIDAS instruments for the immunoenzymatic determination of fibrin degradation products (FbDP) in human plasma (sodium citrate and CTAD) using the ELFA technique (Enzyme Linked Fluorescent Assay).VIDAS D-Dimer Exclusion II is indicated for use in conjunction with a clinical pretest probability assessment model to exclude deep vein thrombosis (DVT) and pulmonary embolism (PE) disease in outpatients suspected of DVT or PE. | VIDAS D-Dimer Exclusionis an automated quantitative test for use on the VIDAS instruments for the Immunoenzymatic determination of fibrin degradation products (FbDP) in human plasma (sodium citrate) using the ELFA technique (Enzyme Linked Fluorescent Assay). VIDAS D-Dimer Exclusion is indicated for use in conjunction with a clinical pretest probability assessment model to exclude deep vein thrombosis (DVT) and pulmonary embolism (PE) disease in outpatients suspected of DVT or PE. |
| Principle of operation | Enzyme-linked fluorescent assay | Same |
| Automated | Yes | Same |
| Sample type | Plasma containing 3.2% or 3.8% sodium citrate or CTAD | Same |
| Enzyme conjugate | Alkaline phosphatase | Same |
| Substrate | 4-Methyl-ombelliferyl phosphate | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Assay Duration | 19 minutes, 57 seconds | 35 minutes, 35 seconds |
| Solid Phase Receptacle (SPR) | SPR is coated with mouse anti-D-Dimer monoclonal antibody | SPR is coated with mouse anti-D-Dimer monoclonal antibody along with an azide preservative |
| Controls (C) and Calibrators (S) | S1: 3000-3500 ng/mL FEUC1: 5200-6040 ng/mL FEUC2: 250-380 ng/mL FEUReconstituted with water | S1: 3700-4400 ng/mL FEUS2: 420-580 ng/mL FEUC1: 4200-4800 ng/mL FEUC2: 380-520 ng/mL FEUReconstituted with the R1 diluent included in the kit |
| Calibration | Every 28 days | Every 14 days |
| Assay Range | 45-10 000 ng/mL | 45-5 000 ng/mL |
{3}
4
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods, Approved Guideline- Second Edition
CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures
CLSI EP7-A, Interference Testing in Clinical Chemistry; Approved Guideline
CLSI EP9-A2, Method comparison and Bias Estimation Using Patient Samples
CLSI EP12-A2, User Protocol for Evaluation of Qualitative Test Performance, Approved Guideline- Second Edition
CLSI EP17-A, Protocols for Determination of Limits of Detection and Limits of Quantitation
L. Test Principle:
The assay combines a two-step enzyme immunoassay sandwich method with a final fluorescent detection step (ELFA).
All assay steps are performed automatically by the instrument. The sample is transferred into the well of the Solid Phase Receptacle (SPR) containing an alkaline-phosphatase labeled anti-FbDP monoclonal antibody. The sample/conjugate mixture is cycled in and out of the SPR several times to increase the reaction speed. The antigen in the sample binds to the antibodies coated on the SPR and to the conjugate.
The remaining free antigen sites are then saturated by cycling the conjugate in the fifth well of the reagent strip in and out of the SPR. Unbound components are eliminated during the washing steps.
Substrate is then cycled in and out of the SPR. The conjugate catalyzes the hydrolysis of the substrate into a fluorescent product that is measured at 450 nm. The intensity of the fluorescence is proportional to the concentration of antigen present in the sample. Results are automatically calculated in relation to the calibration curve stored in memory.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The precision study was performed per CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods, Approved Guideline-Second Edition. Three samples prepared from pooled human citrated plasma were tested at three sites using two VIDAS and one miniVIDAS instruments. Analyte concentrations included all medically relevant levels with target concentrations of 277.97 ng/mL FEU, 544.14 ng/mL FEU and 7788.88 ng/mL FEU. Testing was performed twice a day, in duplicate over 40 runs (n=240) at three sites. Two lots of reagent were included in the studies.
{4}
The predetermined acceptance criteria of $7\%$ CV for repeatability and reproducibility were met.
| Sample Name/Mean concentration | Sample 1 | | Sample 2 | | Sample 3 | |
| --- | --- | --- | --- | --- | --- | --- |
| | 277.97 ng/mL FEU | | 544.14 ng/mL FEU | | 7788.88 ng/mL FEU | |
| | SD (ng/mL) | CV (%) | SD (ng/mL) | CV (%) | SD (ng/mL) | CV (%) |
| Repeatability | 6.88 | 2.5 | 11.05 | 2.0 | 113.83 | 1.5 |
| Between-run | 2.24 | 0.8 | 4.63 | 0.9 | 83.94 | 1.1 |
| Between-day | 2.46 | 0.9 | 3.42 | 0.6 | 19.58 | 0.3 |
| Between-calibration | 6.54 | 2.4 | 12.79 | 2.3 | 246.22 | 3.2 |
| Between-lot | 8.49 | 3.1 | 9.79 | 1.8 | 340.79 | 4.4 |
| Between-instrument | 12.65 | 4.6 | 24.85 | 4.6 | 148.47 | 1.9 |
| Total within-instrument within-lot | 10.06 | 3.6 | 17.85 | 3.3 | 284.62 | 3.7 |
| Total within-instrument between-lot | 13.16 | 4.7 | 20.36 | 3.7 | 444.01 | 5.7 |
| Total between instrument | 18.25 | 6.6 | 32.13 | 5.9 | 468.18 | 6.0 |
# b. Linearity/assay reportable range:
A linearity study was performed per CLSI EP6-A guideline, April 2003, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach. Two VIDAS D-Dimer Exclusion II assay lots were tested on two VIDAS instruments. Nine samples were prepared by mixing a high concentration FbDP (above $10,000\mathrm{ng / mL}$ FEU) of human citrated plasma with a low concentration FbDP (below $45\mathrm{ng / mL}$ FEU) sample in varying ratios.
Each sample was tested in triplicate in random order in the same single run. Results showed that deviation from linearity was within $\pm 12\%$ at each analyte level which met the established acceptance criteria. The VIDAS D-Dimer Exclusion II assay was found to be linear from 45 to $10,000~\mathrm{ng / mL}$ FEU.
# c. Traceability, Stability, Expected values (controls, calibrators, or methods):
VIDAS D-Dimer Exclusion II Assay was standardized using the VIDAS D-Dimer Exclusion Assay to provide metrological traceability of the calibration and controls. A reference standard was created by measuring samples with the VIDAS D-Dimer Exclusion Assay. A panel of working standards and adjustment panel traceable to the reference standards were consequently used
{5}
to assign values to the final calibrator and control material.
Real time shelf life stability was determined for three lots of assay reagents and controls and calibrators. Thermal shocks were performed during the first 3 months of stability testing to mimic differences in temperature the kit could encounter during transport. The results are within acceptance criteria based on the inter kit-lot precision profile (+/-13.3% to 20.7%) and support a shelf-life of 12 months at 2-8°C.
d. Detection limit:
Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ) were determined according to CLSI EP17-A, Guideline, October 2004; Protocols for Determination of Limits of Detection and Limits of Quantitation. Two lots of VIDAS D-Dimer Exclusion II were tested on two VIDAS instruments using 6 human citrated plasma samples (1/90 dilution of patient sample in D-Dimer Exclusion II diluent, five low-level samples).
With each lot, the blank sample was tested in 12 replicates in a single run per day for 5 days on one VIDAS system. Each low-level sample was tested in 4 replicates in a single run per day for 5 days on one VIDAS instrument. The LoD was determined to be 16.471 ng/mL FEU, LoB 7.406 ng/mL FEU, and LoQ 29 ng/mL FEU.
e. Analytical specificity:
Interference studies were conducted according to CLSI EP7-A2. Varying concentrations of hemoglobin (5 g/L), lipids (30 g/L), bilirubin (0.35 g/L), and albumin (60 g/L) were spiked into aliquots of 3 human plasma pools corresponding to 3 medical decision levels of D-Dimer (7500 ng/mL FEU, 500 ng/mL FEU, and 250 ng/mL FEU) and tested with 2 assay lots on one VIDAS instrument.
Interference studies were also conducted using the same human plasma pools for 47 drugs at the concentrations indicated in the table below.
6
{6}
| Drug | Concentration tested (mg/dL-unless otherwise noted) | Drug | Concentration tested (mg/dL-unless otherwise noted) | Drug | Concentration tested (mg/dL-unless otherwise noted) |
| --- | --- | --- | --- | --- | --- |
| Acetaminophen | 20 | Creatinine | 30 | Lithium chloride | 14 |
| Acetylsalicyclic acid= aspirin | 65 | Cyclosporine A | 0.4 | L-Thyroxine | 0.06 |
| Allopurinol | 4 | Dextran 75 | 2500 | Nicotine | 0.1 |
| Amikacin sulfate | 10.4 | Diazepam | 0.5 | Nifedipine | 0.04 |
| Ampicillin | 5 | Digoxin | 0.0006 | Penicillin G sodium salt | 2500U/dL |
| Ascorbic Acid (vitamin C) | 6 | D-L methyl dopa hydrochloride | 1.8 | Pentobarbital | 9 |
| Atenolol | 1 | Dopamine hydrochloride | 0.1 | Phenobarbital | 10 |
| Caffeine | 6 | Erythromycin | 6 | Phenytoine | 5 |
| Captopril | 0.5 | Ethanol | 400 | Primidone | 4 |
| Carbamazepine | 3 | Ethosuximide | 25 | Propranolol hydrochloride | 0.2 |
| Chloramphenicol | 5 | Furosemide | 6 | Theophylline | 4 |
| Chlordiazepoxide hydrochloride | 1.1 | Gentamicin sulfate | 1 | Urea | 500 |
| Chlorpromazine hydrochloride | 0.2 | Heparin lithium salt | 300U/dL | Uric acid | 24 |
| Cimetidine | 2 | Heparin sodium salt | 300u/dL | Valproic acid sodium salt | 60 |
| Cinnarizine | 3 | Ibuprofen | 50 | Verapamil hydrochloride | 0.2 |
| | | Lidocaine | 1.2 | Warfarin | 1 |
All observed interferences were within the pre-established acceptance criteria for each D-Dimer level sample summarized in the table below.
| D-Dimer level (ng/mL) | Acceptance criteria |
| --- | --- |
| 250 | <12% |
| 500 | <10% |
| 7500 | <10% |
f. Assay cut-off:
The assay cut-off of 500 ng/mL FEU was transferred from the predicate device (VIDAS® D-dimer Exclusion Assay (k040882)). The cut-off was validated in frozen specimens from previous clinical management studies.
2. Comparison studies:
a. Method comparison with predicate device:
Method comparison was performed on 326 leftover citrated plasma from
{7}
blood samples submitted to the laboratory for routine D-Dimer measurements. Samples were from hospitalized patients or outpatients at three clinical sites. Samples were tested in singlicate with both VIDAS D-Dimer Exclusion II ref 30455 or VIDAS D-Dimer Exclusion ref 30442 over a minimum of 5 days on two VIDAS and one miniVIDAS instruments. Two lots of reagent were used at each participating site. The bias between the VIDAS D-Dimer Exclusion II ref 30455 and the VIDAS D-Dimer Exclusion ref 30442 at different analyte levels was determined according to CLSI EP9-A2. Linear regression analysis results by Passing- Bablok analysis are summarized in the table below.
| Sample Level | Sample Level Value (ng/mL) | Bias(%) | Confidence Interval (95%) |
| --- | --- | --- | --- |
| Low | 250 | 5.0% | [2.0%; 8.1%] |
| Medical Decision Level | 500 | 12.2% | [8.5%; 16.0%] |
| High | 3000 | 18.2% | [14.5%; 22.0%] |
b. Matrix comparison:
i. Samples collected in sodium citrate vs. CTAD
Two blood draws from 124 donors were collected into one tube containing sodium citrate and one containing CTAD (0.109M sodium citrate, theophyline, adenosine, dipyridamole) each. Eighteen (18) contrived samples with higher D-Dimer concentrations were also included in the study. The paired samples were tested with the VIDAS D-Dimer Exclusion Assay II assay in singlicate. The D-Dimer concentrations observed ranged from 54.5 ng/mL – 7053.5 ng/mL for the CTAD tube collected samples and from 47.7 ng/mL-6730.2 ng/mL for the sodium citrate tubes. The slope and intercept Passing and Bablok regression was used to calculate the relationship between the two sample matrices. The following bias estimates were derived:
| Sample Level | Sample Level Value (ng/mL) | Bias (%) | Confidence Interval (95%) |
| --- | --- | --- | --- |
| Low | 250 | -0.7% | [-1.4%; 2.8%] |
| Medical Decision Level | 500 | -0.8% | [-4.2%; 2.6%] |
| High | 3000 | -2.0% | [-5.8%; 1.7%] |
ii. Samples collected in 3.2% vs. 3.8% sodium citrate
Blood was collected from 91 donors into two blood collection tubes, one containing 3.2% sodium citrate, and the other containing 3.8% sodium citrate. Samples from each tube were analyzed using the DEX2 assay. Eighteen (18) additional donor samples were spiked to create high D-Dimer concentrations. The resulting 109 samples were tested in singlet.
{8}
9
| | | 1 Outlier Removed | | 1 Outlier Removed | |
| --- | --- | --- | --- | --- | --- |
| Sample Level | Sample Level value (ng/mL) | Bias (%) | Confidence Interval (95%) | Bias (%) | Confidence Interval (95%) |
| Low | 250 | -1.9% | [-4.9%; 1.1%] | -1.8% | [-4.8%; 1.2%] |
| Medical Decision Level | 500 | -1.7% | [-8.9%; 5.7%] | -1.5% | [-9.1%; 6.1%] |
| High | 3000 | -1.4% | [-2.7%; -0.2%] | -1.3% | [-2.6%; -0.1%] |
3. Clinical studies:
a. Clinical Sensitivity:
Clinical studies were performed on banked, frozen samples collected from patients suspected of VTE from previous VTE monitoring studies. Sample stability was assessed with the D-Dimer Exclusion assay ref. 30442 by comparison with results obtained prior to freezing (fresh specimen). The results of the stability study were found to be acceptable.
Confirmation of DVT was done by objective testing (e.g. serial compression ultrasound, CT scans, etc. according to physician request) and a 3-month individual follow-up. Overall prevalence of VTE in the total study population was 23.5% (74/315). Two lots each of VIDAS D-Dimer Exclusion II and VIDAS D-Dimer Exclusion reagents were included in the study.
The sensitivity, specificity, negative and positive predictive values (NPV and PPV) were determined using a clinical cut-off of 500 ng/mL FEU.
| | Patients with suspected VTE | | |
| --- | --- | --- | --- |
| | | Low & intermediate pretest probability N=303 | All probabilities N=315 |
| % Sensitivity (95%CI) | VIDAS D-Dimer Exclusion II ref. 30 455 | 100% (94.2%-100%) | 100% (95.1%-100%) |
| | VIDAS D-Dimer Exclusion ref 30 442 | 100% (94.2%-100%) | 100% (95.1%-100%) |
| % Specificity (95% CI) | VIDAS D-Dimer Exclusion II ref. 30 455 | 35.7% (29.6%-42.1%) | 35.7% (29.6%-42.1%) |
| | VIDAS D-Dimer Exclusion ref 30 442 | 37.8% (31.6%-44.2%) | 37.8% (31.6%-44.2%) |
| % NPV (95% CI) | VIDAS D-Dimer Exclusion II ref. 30 455 | 100% (95.8%-100%) | 100% (95.8%-100%) |
| | VIDAS D-Dimer Exclusion ref 30 442 | 100% (95.8%-100%) | 100% (96.0%-100%) |
| %PPV (95% CI) | VIDAS D-Dimer Exclusion II ref. 30 455 | 28.6% (22.7%-35.1%) | 32.3% (26.3%-38.8%) |
| | VIDAS D-Dimer Exclusion ref 30 442 | 29.2% (23.2%-35.9%) | 33.0% (26.9%-39.6%) |
{9}
b. Clinical specificity: See section 3a.
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off: 500 ng/mL FEU
5. Expected values: 215 fresh citrated plasma samples collected from blood donors were tested on four VIDAS instruments using three VIDAS D-Dimer Exclusion II assay lots. After assay recalibration, each sample was tested in singlicate.
| Total number of blood donors | 215 |
| --- | --- |
| Number of donors with D-Dimer value <500 ng/mL | 193 |
| Percentage of donors with D-Dimer value <500 ng/mL | 90% |
N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.