The Stratus® CS Acute Care™ D-dimer assay (DDMR) is an in vitro diagnostic test for the quantitative measurement of cross-linked fibrin degradation products (D-dimer) in human citrated or heparinized plasma on the Stratus® CS analyzer. The Stratus® CS Acute Care™ DDMR assay is intended for use in conjunction with a non-high clinical pretest probability (PTP) assessment model to exclude pulmonary embolism (PE) disease and as an aid in the diagnosis of venous thromboembolism (VTE) [deep vein thrombosis (DVT) or pulmonary embolism (PE)]. This assay is for use by trained health care professionals in the clinical laboratory and point of care (POC) settings.
Device Story
The Stratus® CS Acute Care™ D-Dimer assay is a two-site sandwich Radial Partition Immunoassay (RPIA) performed on the Stratus® CS analyzer. It utilizes dendrimer-linked monoclonal antibodies immobilized on glass fiber paper to capture D-dimer from plasma samples. An enzyme-labeled monoclonal antibody conjugate is added to form a sandwich complex. A substrate wash solution elutes unbound conjugate and initiates an enzymatic reaction proportional to the D-dimer concentration. An optical system measures the reaction rate via front-surface fluorescence. Used in clinical laboratories and point-of-care settings by trained professionals, the device provides quantitative results to assist clinicians in diagnosing VTE and excluding PE when combined with clinical pretest probability assessments. This rapid testing capability supports timely clinical decision-making in emergency departments.
Clinical Evidence
Multi-center studies evaluated the assay in 655 patients (citrated plasma) and 427 patients (heparinized plasma) presenting with suspected PE. Using a 450 ng/mL FEU cutoff, the assay demonstrated high negative predictive value (NPV) for PE exclusion. In citrated plasma, sensitivity was 98.9% and NPV was 99.6%. In heparinized plasma, sensitivity was 98.3% and NPV was 99.1%. Results were compared against imaging (spiral CT/VQ scan) and 3-month follow-up.
Technological Characteristics
Two-site sandwich Radial Partition Immunoassay (RPIA) on glass fiber paper. Uses dendrimer-linked monoclonal antibodies and enzyme-labeled monoclonal antibody conjugate. Detection via front-surface fluorescence optical system. Analyzed on the Stratus® CS analyzer. Quantitative measurement of D-dimer in citrated or heparinized plasma.
Indications for Use
Indicated for use in patients suspected of pulmonary embolism (PE) or venous thromboembolism (VTE) to aid in diagnosis or exclude PE when used with a non-high clinical pretest probability assessment model. For use by trained healthcare professionals in clinical laboratory and point-of-care settings.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k110303
B. Purpose for Submission:
Expanded indications for exclusion of pulmonary embolism (PE) to previously cleared assays (k022976, k063356)
C. Measurand:
D-dimer
D. Type of Test:
Solid Phase Radial Partition Immunoassay (RPIA)
E. Applicant:
Siemens Healthcare Diagnostics, Inc.
F. Proprietary and Established Names:
Stratus® CS Acute Care™ DDMR
G. Regulatory Information:
1. Regulation section:
21 CFR 864.7320; Fibrinogen/Fibrin Degradation Products Assay
2. Classification:
Class II
3. Product code:
DAP; Fibrinogen/Fibrin Split Products, Antigen, Antiserum and Control
4. Panel:
81 Hematology
H. Intended Use:
1. Intended Use(s):
The Stratus® CS Acute Care™ D-dimer assay (DDMR) is an *in vitro* diagnostic test for the quantitative measurement of cross-linked fibrin degradation products (D-dimer) in human citrated or heparinized plasma on the Stratus® CS Analyzer. The Stratus® CS Acute Care™ DDMR assay is intended for use in conjunction with a non-high clinical pretest probability (PTP) assessment model to exclude pulmonary embolism (PE) disease and as an aid in the diagnosis of venous thromboembolism (VTE) [deep vein thrombosis (DVT) or pulmonary embolism (PE)]. This assay is for use by trained health care professionals in the clinical laboratory and point of care (POC) settings.
2. Indication(s) for Use:
Same as Intended Use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Stratus® CS STAT Fluorometric Analyzer (Stratus® CS Analyzer)
I. Device Description:
The Stratus® CS Acute Care™ D-dimer (DDMR) Testpak:
Reagent Well 1- Alkaline phosphatase conjugated to anti-D-dimer Fab', ACES Buffer, Na azide
{1}
Reagent Well 2- Dendrimer linked D-dimer antibody, Na azide
Reagent Well 3- Tris buffer, Na azide
Reagent Well 4- 4-methylumbelliferyl phosphate, diethanolamine buffer, Na azide
Reagent Well 5- Empty sample well
J. Substantial Equivalence Information:
1. Predicate device name(s): INNOVANCE® D-Dimer
2. Predicate 510(k) number(s): k093626
3. Comparison with predicate:
| | Similarities | |
| --- | --- | --- |
| Item | Device Stratus® CS Acute Care D-Dimer | Predicate INNOVANCE® D-Dimer |
| Intended Use | The Stratus® CS Acute Care™ D-dimer assay (DDMR) is an in vitro diagnostic test for the quantitative measurement of cross-linked fibrin degradation products (D-dimer) in human citrated or heparinized plasma on the Stratus® CS Analyzer. The Stratus® CS Acute Care™ (DDMR) assay is intended for use in conjunction with a non-high clinical pretest probability (PTP) assessment model to exclude pulmonary embolism (PE) disease and as an aid in the diagnosis of venous thromboembolism (VTE)[deep vein thrombosis (DVT) or pulmonary embolism (PE)]. This assay is for use by trained health care professionals in the clinical laboratory and point of care (POC) settings. | For the quantitative determination of cross-linked fibrin degradation products (D-dimers) in human plasma on Siemens Healthcare Diagnostics and Sysmex® Coagulation Systems. The INNOVANCE® D-Dimer Assay is intended for use in conjunction with a non-high clinical pretest probability (PTP) assessment model to exclude deep vein thrombosis (DVT) and pulmonary embolism (PE). |
| Technology | Immunochemical reaction | Same |
| Reagents | Liquid; no preparation | Same |
| Antibody | Monoclonal from mouse | Same |
| | Differences | |
| --- | --- | --- |
| Item | Device Stratus® CS Acute Care D-dimer | Predicate INNOVANCE® D-Dimer |
| Intended Use | Exclusion of PE. The Stratus® CS Acute Care™ D-dimer assay is for use by trained health care professionals in the clinical laboratory and point of care (POC) settings | Cleared for exclusion of DVT as well as PE. |
| Technology | Fluorometry | Turbidimetry |
| Assay Cutoff | 450 ng/mL[μg/L] FEU | 0.50 mg/L FEU |
{2}
# K. Standard/Guidance Document Referenced (if applicable):
Not applicable
# L. Test Principle:
The Stratus® CS Acute Care™ D-dimer method is a two-site sandwich assay based upon solid phase Radial Partition Immunoassay (RPIA) technology. In this procedure, dendrimer linked monoclonal antibody is added to the center portion of a square piece of glass fiber paper in the DDMR TestPak. This antibody recognizes a distinct antigenic site on the D-dimer molecule. Sample is then added onto the paper where it reacts with the immobilized antibody. After a short incubation, a conjugate consisting of enzyme-labeled monoclonal antibody directed against a second distinct antigenic site on the D-dimer molecule is pipetted onto the reaction zone of the paper. During this second incubation period, enzyme-labeled antibody reacts with the bound D-dimer, forming an antibody-antigen-labeled antibody sandwich. The unbound labeled antibody is later eluted from the field of view of the Stratus® CS analyzer by applying a substrate wash solution to the center of the reaction zone. By including substrate for the enzyme within the wash solution, initiation of enzyme activity occurs simultaneously with the wash. The enzymatic rate of the bound fraction increases directly with the concentration of D-dimer in the sample. The reaction rate can be measured by an optical system that monitors the reaction rate via front surface fluorescence. All data analysis functions are performed by the microprocessor within the analyzer.
# M. Performance Characteristics:
1. Analytical performance: Refer to original $510(\mathrm{k})$ submissions k022976, k051597 and k063356 for analytical performance data (precision/reproducibility, linearity/assay reportable range, detection limit, analytical specificity, assay cutoff, traceability, stability, expected values).
2. Comparison studies:
a. Method comparison was determined in a total of 396 citrated samples from patients suspected of PE using the Stratus® CS Acute Care™ D-dimer assay and the INNOVANCE® D-Dimer assay. The range of D-dimer values in the correlation studies was 54 to $4506\mathrm{ng / mL}$ $[\mu \mathrm{g} / \mathrm{L}]$ FEU.
| Comparative Method n = 396 | Slope | Regression Statistics | |
| --- | --- | --- | --- |
| | | Intercept (mg/L FEU) | Correlation Coefficient |
| INNOVANCE® D-Dimer | 0.950 | -12.32 | 0.938 |
b. Matrix comparison:
Refer to original 510(k) submission (k022976) for Stratus® DDMR matrix data.
3. Clinical studies:
a. Clinical Sensitivity and Specificity: Citrated Plasma
The Stratus® CS Acute Care™ D-dimer assay was evaluated on the Stratus CS® Analyzer in a multi-center study to validate the exclusion of PE using fresh citrated plasma specimens collected from 730 consecutive patients presenting to the emergency department with suspected PE. Of these 730 patients, 75 were excluded for a total of 655 patients available for final analysis.
{3}
All patients were evaluated using the Wells' rules to estimate a high, moderate or low pre-test probability (PTP) of PE. Patient specimens were tested with the Stratus® CS Acute Care™ D-dimer assay and results were compared to a cut-off value of 450 ng/mL [μg/L] (FEU). A D-dimer result < 450 ng/mL [μg/L] (FEU) was considered negative and a D-dimer result ≥ 450 ng/mL (FEU) [μg/L] was considered positive.
Patients with a positive D-dimer result and/or high PTP were evaluated by imaging methods, e.g. spiral CT and/or VQ scan. Patients with a negative D-dimer result and a low or moderate PTP underwent imaging at the physician's discretion, and patients with negative imaging results were followed for three months to evaluate potential development of PE.
The overall prevalence of PE in those patients available for final analysis was 14.0% (92/655). The following instrument-specific sensitivity, specificity and negative predictive value (NPV) with upper and lower 95% confidence limits (CL) were obtained with the Stratus® CS Acute Care™ D-Dimer clinical cutoff of 450 ng/mL [μg/L] (FEU).
## Citrated Plasma: All Patients
| Instrument | PE Patients (n) | Cutoff ng/mL[μg/L] FEU | Sensitivity (CL) % | Specificity (CL) % | NPV (CL) % |
| --- | --- | --- | --- | --- | --- |
| Stratus® CS System | 655 | 450 | 98.9 (94.1 – 100.0) | 42.5 (38.3-46.7) | 99.6 (97.7-100.0) |
## Citrated Plasma: Patients with low and moderate pre-test probability
| Instrument | PE Patients (n) | Cutoff ng/mL[μg/L] FEU | Sensitivity (CL) % | Specificity (CL) % | NPV (CL) % |
| --- | --- | --- | --- | --- | --- |
| Stratus® CS System | 625 | 450 | 98.7 (92.9 – 100.0) | 43.0 (38.8-47.3) | 99.6 (97.7-100.0) |
## Heparinized Plasma
The Stratus® CS Acute Care™ D-dimer assay was evaluated on the Stratus CS® Analyzer in a multi-center study to validate the exclusion of PE using fresh heparinized plasma specimens collected from 468 consecutive patients presenting to the emergency department with suspected PE. Of these 468 patients, 41 were excluded for a total of 427 patients available for final analysis.
All patients were evaluated using the Wells' rules to estimate a high, moderate or low pre-test probability (PTP) of PE. Patient specimens were tested with the Stratus® CS Acute Care™ D-dimer assay and results were compared to a cut-off value of 450 ng/mL [μg/L] (FEU). A D-dimer result < 450 ng/mL [μg/L] (FEU) was considered negative and a D-dimer result ≥ 450 ng/mL [μg/L] (FEU) was considered positive.
Patients with a positive D-dimer result and/or high PTP were evaluated by imaging methods, e.g. spiral CT and/or VQ scan. Patients with a negative D-dimer result and a low or moderate PTP underwent imaging at the physician's discretion, and patients with negative imaging results were followed for three months to evaluate potential development of PE.
{4}
The overall prevalence of PE in those patients available for final analysis was 14.1% (60/427). The following instrument-specific sensitivity, specificity and negative predictive value (NPV) with upper and lower 95% confidence limits (CL) were obtained with the Stratus® CS Acute Care™ D-Dimer clinical cut-off of 450 ng/mL [μg/L] (FEU).
## Heparinized Plasma: All Patients
| Instrument | PE Patients (n) | Cutoff ng/mL[μg/L] FEU | Sensitivity (CL) % | Specificity (CL) % | NPV (CL) % |
| --- | --- | --- | --- | --- | --- |
| Stratus® CS System | 427 | 450 | 98.3 (91.1 – 100.0) | 29.7 (25.1-34.7) | 99.1 (95.04-100.0) |
## Heparinized Plasma: Patients with low and moderate pre-test probability
| Instrument | PE Patients (n) | Cutoff ng/mL[μg/L] FEU | Sensitivity (CL) % | Specificity (CL) % | NPV (CL) % |
| --- | --- | --- | --- | --- | --- |
| Stratus® CS System | 401 | 450 | 97.9 (88.7 – 100.0) | 29.9 (25.2-35.0) | 99.1 (94.9-100.0) |
b. Other clinical supportive data:
Not applicable
4. Clinical cut-off:
450 ng/mL [μg/L] FEU; cutoff previously established and evaluated in k051597.
5. Expected values:
Less than 552 ng/mL [μg/L] FEU (citrated plasma)
Citrated plasma samples from apparently healthy individuals ranged from 38 to 804 ng/mL [μg/L] FEU with a mean of 258 ng/mL [μg/L] FEU. Samples consisted of 81 males and 50 females with ages ranging from 18 to 59 years.
Less than 682 ng/mL[μg/L] FEU (lithium heparin plasma)
Lithium heparin plasma samples from apparently healthy individuals range from 37 to 971 ng/mL [μg/L] FEU with a mean of 304 ng/mL [μg/L] FEU. Samples consisted of 81 males and 50 females with ages ranging from 18 to 59 years.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.