← Product Code [DAP](/submissions/HE/subpart-h%E2%80%94hematology-kits-and-packages/DAP) · K041438

# ADVANCED D-DIMER ASSAY (K041438)

_Dade Behring, Inc. · DAP · Aug 11, 2004 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HE/subpart-h%E2%80%94hematology-kits-and-packages/DAP/K041438

## Device Facts

- **Applicant:** Dade Behring, Inc.
- **Product Code:** [DAP](/submissions/HE/subpart-h%E2%80%94hematology-kits-and-packages/DAP.md)
- **Decision Date:** Aug 11, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.7320
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

Advanced D-Dimer is a latex-enhanced turbidimetric test for the quantitative determination of cross-linked fibrin degradation products containing D-dimer in human plasma for use with Dade Behring Coagulation Analyzers and Sysmex® Coagulation Systems. The Advanced D-Dimer assay is intended for use as an aid in the diagnosis of venous thromboembolism (VTE) [deep vein thrombosis (DVT) or pulmonary embolism (PE)].

## Device Story

Advanced D-Dimer is a latex-enhanced turbidimetric immunoassay for quantitative measurement of D-Dimer in human plasma. Input: human plasma samples. Principle: polystyrene particles coated with mouse monoclonal antibody (DD5) agglutinate in presence of D-Dimer; agglutination measured via increased turbidity. Output: quantitative D-Dimer concentration (mg/L). Used in clinical laboratories on Dade Behring coagulation analyzers and Sysmex coagulation systems. Automated processing by instrument. Healthcare providers use results as an aid in diagnosing VTE, DVT, or PE. Benefit: assists in clinical assessment of suspected thromboembolic events.

## Clinical Evidence

Retrospective clinical study of 322 patients (BCS system) and 297 patients (Sysmex CA-1500) presenting with suspected VTE. Diagnosis confirmed via objective testing (V/Q scan, angiography, CT, or compression ultrasonography) with 3-month follow-up for negatives. BCS System: 98% sensitivity, 38% specificity. Sysmex CA-1500: 100% sensitivity, 37% specificity. Clinical cut-off validation performed on 110 patients (ages 17-89).

## Technological Characteristics

Latex-enhanced turbidimetric immunoassay. Reagents: polystyrene particles covalently linked to mouse monoclonal antibody (DD5). Instrumentation: Dade Behring coagulation analyzers and Sysmex coagulation systems. Measuring range: 0.4 to 55 mg/L. Sample type: citrate plasma. Automated measurement.

## Predicate Devices

- Dade Behring Advanced D-Dimer ([K992957](/device/K992957.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
K041438

B. Purpose for Submission:
To seek clearance for a modification to their Advanced D-Dimer Assay

C. Analyte:
D-Dimer

D. Type of Test:
Quantitative turbidimetric immunoassay

E. Applicant:
Dade Behring

F. Proprietary and Established Names:
Advanced D-Dimer Assay

G. Regulatory Information:
1. Regulation section:
21 CFR 864.7320, Fibrinogen/Fibrin degradation products assay
2. Classification:
Class II
3. Product Code:
DAP, Fibrinogen and fibrin split products, antigen, antiserum, control
4. Panel:
81 Hematology

H. Intended Use:
1. Intended use(s):
Advanced D-Dimer is a latex-enhanced turbidimetric test for the quantitative determination of cross-linked fibrin degradation products containing D-Dimer in human plasma.
2. Indication(s) for use:
The Advanced D-Dimer is intended for use as an aid in the diagnosis of venous thromboembolism (VTE), deep vein thrombosis (DVT), or pulmonary embolism (PE).
3. Special condition for use statement(s):
None specified

{1}

Page 2 of 5

4. Special instrument Requirements:
The Advanced D-Dimer is intended for use with the Dade Behring coagulation Analyzers, and Sysmex Coagulation Systems.

I. Device Description:
Advanced D-Dimer is a latex-enhanced turbidimetric test for the quantitative determination of cross-linked fibrin degradation products containing D-Dimer in human plasma. Elevated concentrations of d-dimer are indicative of the presence of a clot and have been reported in deep vein thrombosis, pulmonary embolism and disseminated intravascular coagulation.
The assay kit consists of Advanced D-Dimer reagent, Advanced D-Dimer accelerator, and Advanced D-Dimer reconstitution medium. Calibration and quality control material must be provided by the user. All steps necessary for preparing the run and performing the measurements are processed automatically by the instrument.

The device is a modification of the Sponsor's currently cleared Advanced D-Dimer (K992957). The Sponsor modified the intended use statement to include use as an aid in the diagnosis of venous thromboembolism and revised the performance section to include a cutoff value

J. Substantial Equivalence Information:
1. Predicate device name(s):
Dade Behring Advanced D-Dimer
2. Predicate K number(s):
K992957
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Principle | Latex-enhanced turbidimetric test | Same  |
|  Antibody | Mouse monoclonal | Same  |
|  Measuring Range | 0.4 to 55 mg/L | Same  |
|  Sample Requirement | Citrate plasma | Same  |
|  Instrumentation | Dade Behring coagulation Analyzers and Sysmex Coagulation Systems | Same  |
|  Differences  |   |   |
|  Item | Device | Predicate  |
|  Intended Use | Quantitative determination of cross-linked fibrin degradation products | Quantitative determination of cross-linked fibrin degradation products  |

{2}

Page 3 of 5

|   | containing D-Dimer in human plasma and as an aid in the diagnosis of VTE. | containing D-Dimer in human plasma.  |
| --- | --- | --- |
|  Instrument specific cut-off | BCS System: 1.6 mg/L
Sysmex CA-1500: 1.0 mg/L | none  |

## K. Standard/Guidance Document Referenced (if applicable):

## L. Test Principle:

Polystyrene particles covalently linked to a monoclonal antibody (DD5) to the cross-linkage region of D-Dimer agglutinate when mixed with samples containing D-Dimer. The sterosymmetrical structure is detected turbidimetrically via the increase in turbidity.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

a. Precision/Reproducibility:

2 levels of controls and two levels of human plasma were tested following guidelines outlined in NCCLS Guideline EP5-A. Intra-assay precision &lt;3.0%, inter assay precision &lt;4.0%

b. Linearity/assay reportable range:

c. Traceability (controls, calibrators, or method):

d. Detection limit:

e. Analytical specificity:

No interference with the Advanced D-dimer assay on the BCS System by:

|   | Up to  |
| --- | --- |
|  Bilirubin | 24 mg/dL  |
|  Hemoglobin | 100 mg/dL  |
|  Lipids | 120 mg/dL  |
|  Heparin | 2.5 IU/mL  |
|  Rheumatoid Factor | 98 IU/mL  |

Users should refer to the respective application sheet for additional interference information. Testing for interference from higher levels of lipids or turbid samples can lead to falsely elevated or decreased values. Human Anti-mouse Antibodies (HAMA) testing has not been performed with the Advanced D-Dimer assay. However plasma samples from patients containing heterophilic antibodies and from patients who have received preparation of mouse monoclonal antibodies for diagnosis or therapy may show either falsely elevated or falsely decreased values when tested with assays kits that use

{3}

Page 4 of 5

mouse monoclonal antibodies. Dilution of samples may lead to discordant results in certain cases resulting in non-specific reactions (non-specific binding) independent of the concentration of D-Dimer fragment.

f. Assay cut-off:

2. Comparison studies:
a. Method comparison with predicate device:
b. Matrix comparison:

3. Clinical studies:
Frozen samples were collected from out-patients at three sites, and retrospectively evaluated with the Advanced D-Dimer assay on both the BCS System and CA-1500 Analyzer. 322 patients were tested on the BCS system (198 female, 122 male), and 297 on the Sysmex CA-1500 (184 female, 111 male). Patients presenting to the emergency room with clinically suspected VTE were evaluated using the Wells Pre-Test probability models to estimate the probability (high, moderate, or low) of DVT or PE. Patients were diagnosed as DVT or PE positive by standard objective tests as appropriate and patients initially diagnosed as negative were followed for three months.

a. Clinical sensitivity:
|  BCS System | 98%  |
| --- | --- |
|  Sysmex CA-1500 | 100%  |

c. Clinical specificity:
|  BCS System | 38%  |
| --- | --- |
|  Sysmex CA-1500 | 37%  |

c. Other clinical supportive data (when a and b are not applicable):

4. Clinical cut-off:
The Advanced D-Dimer was tested at one site, on 110 patients using both the BCS System and Sysmex CA-1500 Analyzers (78 females, 32 males, between the ages of 17-89). PE was ruled out for confirmed by ventilation-perfusion (V/Q) lung scan, pulmonary angiography scintigraphy and/or spiral CT. DVT was ruled out or confirmed by compression ultrasonography. Results:
|  BCS System | 1.6  |
| --- | --- |
|  Sysmex CA-1500 | 1.0  |

5. Expected values/Reference range:

{4}

Page 5 of 5

In a study of 136 health subjects using the BCS system the following data were obtained: Mean 1.20 mg/L, median 1.07 mg/L, 90% interval 0.54-2.09 mg/L.

**N. Conclusion:**
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/HE/subpart-h%E2%80%94hematology-kits-and-packages/DAP/K041438](https://fda.innolitics.com/submissions/HE/subpart-h%E2%80%94hematology-kits-and-packages/DAP/K041438)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
