VIDAS D-DIMER EXCLUSION ASSAY, MODEL 30 442

K040882 · bioMerieux, Inc. · DAP · Jun 10, 2004 · Hematology

Device Facts

Record IDK040882
Device NameVIDAS D-DIMER EXCLUSION ASSAY, MODEL 30 442
ApplicantbioMerieux, Inc.
Product CodeDAP · Hematology
Decision DateJun 10, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7320
Device ClassClass 2

Indications for Use

The VIDAS® D-dimer Exclusion Assay is an automated quantitative test for the immunoenzymatic determination of fibrin degradation products (FbDP) in citrated human plasma using the ELFA (Enzyme Linked Fluorescent Assay). The VIDAS® D-dimer Exclusion Assay is indicated for use in conjunction with a clinical Pre-test Probability Assessment (PTP) assessment model to exclude deep venous thrombosis (DVT) and pulmonary embolism (PE) in outpatients suspected of DVT or PE.

Device Story

The VIDAS D-Dimer Exclusion Assay is an automated, quantitative, enzyme-linked fluorescent immunoassay (ELFA) performed on the VIDAS analyzer. It measures fibrin degradation products (FbDP) in citrated human plasma. The system utilizes a Solid Phase Receptacle (SPR) that functions as both a solid phase and a pipettor for the assay. Reagents are pre-dispensed in sealed strips. The device is intended for use in clinical settings to assist healthcare providers in excluding DVT and PE in symptomatic outpatients. By integrating assay results with a clinical Pre-test Probability (PTP) assessment, the device helps clinicians rule out these conditions, potentially reducing the need for further diagnostic imaging or invasive procedures. The instrument automates all assay steps and temperature controls.

Clinical Evidence

Clinical evidence includes two multi-center prospective cohort studies. DVT study (n=556) and PE study (n=965). Patients evaluated using Wells (DVT) or WICKIE (PE) pre-test probability models. Primary endpoint: negative predictive value (NPV) for VTE exclusion. Results: NPV for DVT exclusion was 100% (95% CI: 97.8-100.0). NPV for PE exclusion was 100% (95% CI: 98.7-100.0). Sensitivity was 100% across all probability groups.

Technological Characteristics

ELFA-based quantitative immunoassay. Components: polypropylene reagent strips, anti-D-Dimer mouse monoclonal immunoglobulin-coated Solid Phase Receptors (SPR), lyophilized FbDP controls/calibrators. Energy source: VIDAS analyzer-integrated. Detection: fluorometric measurement at 450 nm. Connectivity: Master Lot Entry (MLE) barcode for curve calibration. Automated processing.

Indications for Use

Indicated for the quantitative determination of fibrin degradation products (FbDP) in citrated human plasma to aid in the exclusion of deep venous thrombosis (DVT) and pulmonary embolism (PE) in outpatients suspected of DVT or PE, when used in conjunction with a clinical Pre-test Probability Assessment (PTP) model.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: K040882 B. Purpose for Submission: The purpose of this submission is to obtain clearance for a modification to the indications for use of this device to include an exclusionary claim for pulmonary embolism. C. Analyte: D-dimer D. Type of Test: ELISA E. Applicant: bioMerieux F. Proprietary and Established Names: VIDAS® D-dimer Exclusion Assay G. Regulatory Information: 1. Regulation section: 21 CFR 864.7320 2. Classification: II 3. Product Code: DAP 4. Panel: 81 Hematology H. Intended Use: 1. Intended use(s): The VIDAS® D-dimer Exclusion Assay is an automated quantitative test for the immunoenzymatic determination of fibrin degradation products (FbDP) in citrated human plasma using the ELFA (Enzyme Linked Fluorescent Assay). 2. Indication(s) for use: The VIDAS® D-dimer Exclusion Assay is indicated for use in conjunction with a clinical Pre-test Probability Assessment (PTP) assessment model to exclude deep venous thrombosis (DVT) and pulmonary embolism (PE) in outpatients suspected of DVT or PE. 3. Special condition for use statement(s): 4. Special instrument Requirements: VIDAS analyzers I. Device Description: The VIDAS® D-dimer Exclusion Assay is a quantitative for the immunoenzymatic determination of fibrin degradation products in human citrated plasma using the {1} Page 2 of 6 ELFA technique. The assay is intended for use either on the VIDAS (K981385), or the miniVIDAS (K923579). Each assay kit contains 60 tests, and consists of the following; DD2 Reagent Strips- ready to use, polypropylene strip of 10 wells covered with a labeled foil seal. The first well is intended for the sample, and the last well is a cuvette in which the fluorometric reading is performed. The wells in between contain the various reagents required for the assay. DD2 Solid Phase Receptors (SPR) - The SPR is coated during production with anti-D-Dimer mouse monoclonal immunoglobulin. C1 and C2 controls – bilevel lyophilized FbDP controls obtained from human plasma. S1 and S2 calibrators – bilevel lyophilized FbDP calibrators Master Lot Entry (MLE) Card-the MLE is printed with bar code readable data needed for establishing the Master Curve for that assay. J. Substantial Equivalence Information: 1. Predicate device name(s): bioMerieux VIDAS D-Dimer (DD) New Assay-Exclusion DVT Claim 2. Predicate K number(s): K030328 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Quantitative determination of FbDP in citrated human plasma | Same | | Methodology | Enzyme Linked Fluorescent Assay (ELFA) | Same | | Differences | | | | Item | Device | Predicate | | Indications for use | Along with a Pre-Test Probability Assessment Model (PTP), for the exclusion of DVT and PE | Along with a PTP for the exclusion on DVT | K. Standard/Guidance Document Referenced (if applicable): {2} Page 3 of 6 L. Test Principle: The assay combines a two-step enzyme immunoassay sandwich method with a final fluorescent detection (ELFA). All of the assay steps are performed automatically by the instrument. The sample is transferred into the well of the SPR containing an alkaline-phosphatase labeled anti-FbDP monoclonal antibody. The sample /conjugate mixture is cycled in and out of the SPR several times to increase the reaction speed. The antigen in the sample binds to the antibodies coated on the SPR and to the conjugate. The remaining free antigen sites are then saturated by cycling the conjugate in the fifth well of the reagent strip in and out of the SPR. Unbound components are eliminated during the washing steps. Substrate is then cycled in and out of the SP. The conjugate catalyzes the hydrolysis of the substrate into a fluorescent product that is measured at 450 nm. The intensity of the fluorescence is proportional to the concentration of antigen and present in the sample. Results are calculated in relation to two calibration curves stored in memory. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: | | | | Within-run reproducibility | Reproducibility | | --- | --- | --- | --- | --- | | Plasma | N | Concentration (ng/ml) (FEU) | %CV | %CV | | Level 1 | 80 | 264 | 5.0 | 5.7 | | Level 2 | 80 | 549 | 3.9 | 5.8 | | Level 3 | 80 | 7283 | 5.3 | 7.1 | b. Linearity/assay reportable range: 45 to 10,000 ng/ml (FEU) c. Traceability (controls, calibrators, or method): d. Detection limit: The smallest concentration of D-Dimer significantly different from the zero concentration with a risk α of 5%: ≤ 45 ng/ml (FEU) e. Analytical specificity: Fibrinogen: the results of the dilution tests for two plasmas with titers of 2076 ng/ml (FEU) and 8093 ng/ml (FEU) are not significantly different when the kit diluent for plasma with a low concentration do D-Dimer is used. {3} Page 4 of 6 Fibrinogen degradation products: the excess concentration of plasma with a titer of 63 ng/ml (FEU) of D-Dimer, with a mixture of fibrinogen degradation products, does not modify the result up to a concentration ratio of 1 to 100. The specificity of the two antibodies used in this assay was not tested against FDPE, therefore cross-reactivity cannot be ruled out. f. Assay cut-off: 500 ng/ml (FEU) 2. Comparison studies: a. Method comparison with predicate device: b. Matrix comparison: 3. Clinical studies: a. Clinical sensitivity: **DVT EXCLUSION** Frozen samples collected from patients enrolled in a multi-center prospective cohort study were used. Consecutive eligible outpatients (n= 556) with a first suspected DVT from 3 sites were evaluated. Patients were classified as having a high, moderate, or low pretest probability of DVT, using the Wells model to estimate probability. A d-dimer result ≥ 500 ng FEU/ml was considered positive and a result of < 500 ng FEU/ml was considered negative. Those patients having a negative D-dimer result and a low or moderate PTP underwent no further diagnostic testing, and were followed for 3 months for development of DVT. Patients with a positive d-dimer test result and/or high PTP underwent serial compression ultrasound. Total Population | Patients With Suspected DVT | N | %Sensitivity (95% CI) | % Specificity (95% CI) | % Negative Predictive Value (95%) CI | | --- | --- | --- | --- | --- | | All probabilities | 555 | 100.0 (93.6-100.0) | 32.9 (28.8 – 37.2) | 100.0 (97.8-100.0) | {4} Page 5 of 6 ## Patients with low PTP | Patients With Suspected DVT | N | %Sensitivity (95% CI) | % Specificity (95% CI) | % Negative Predictive Value (95%) CI | | --- | --- | --- | --- | --- | | Low probability | 295 | 100.0 (81.5-100.0) | 39.7 (33.9 – 45.7) | 100.0 (96.7-100.0) | ## Patients with Moderate PTP | Patients With Suspected DVT | N | %Sensitivity (95% CI) | % Specificity (95% CI) | % Negative Predictive Value (95%) CI | | --- | --- | --- | --- | --- | | Mod probability | 189 | 100.0 (80.5-100.0) | 26.7 (20.3 – 34.0) | 100.0 (92.3-100.0) | ## Patients with High PTP | Patients With Suspected DVT | N | %Sensitivity (95% CI) | % Specificity (95% CI) | % Negative Predictive Value (95%) CI | | --- | --- | --- | --- | --- | | High probability | 71 | 100.0 (83.9-100.0) | 16.0 (7.2 – 29.1) | 100.0 (63.1-100.0) | ## PE EXCLUSION In a multi-center (3) prospective study, fresh samples were collected from patients presenting to the ER with suspected PE (n=965). Patients were classified as having a high, moderate, or low pretest probability of PE, using the WICKIE model to estimate probability. A d-dimer result ≥ 500 ng FEU/ml was considered positive and a result of < 500 ng FEU/ml was considered negative. Those patients having a negative D-dimer result received no treatment, and underwent no further diagnostic testing. Patients with a positive d-dimer test result were further evaluated using ultrasound, Helical CT scans, and/or angiography. These patients were followed up for 3 months to evaluate any possible VTE and episodes of bleeding. {5} Page 6 of 6 Total Population | All Patients | %Sensitivity (95% CI) | % Specificity (95% CI) | % Negative Predictive Value (95% CI) | %Positive Predictive Value (95% CI) | | --- | --- | --- | --- | --- | | 965 | 100.0 (98.4-100.0) | 37.7 (34.2 – 41.3) | 100.0 (98.7-100.0) | 32.4 (28.9-36.1) | Patients with low and moderate PTP | Low & Intermediate | %Sensitivity (95% CI) | % Specificity (95% CI) | % Negative Predictive Value (95%) CI | %Positive Predictive Value (95% CI) | | --- | --- | --- | --- | --- | | 891 | 100% (97.7-100%) | 37.6 (34.0-41.2) | 100.0 (98.7 – 100.0) | 25.8 (22.4-29.5) | Patients with High PTP | High | %Sensitivity (95% CI) | % Specificity (95% CI) | % Negative Predictive Value (95%) CI | %Positive Predictive Value (95% CI) | | --- | --- | --- | --- | --- | | 74 | 100.0 (94.3-100.0) | 45.5 (16.7 – 76.6) | 100.0 (47.8-100.0) | 91.3 (82.0-96.7) | b. Clinical specificity: See above c. Other clinical supportive data (when a and b are not applicable): 4. Clinical cut-off: 500 ng/ml (FEU) 5. Expected values/Reference range: In a study involving 200 blood donors, 96% of values were below 500 ng/ml (FEU). N. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...