The Hema-Screen TM Specific Gold is for the rapid and qualitative determination of Human Blood fecal samples. It is intended for professional and laboratory use only. It is also useful for determining gastrointestinal bleeding due to a number of gastrointestinal (GI) disorders such as diverticulitis, colitis, polyps and colorectal cancer. This test is recommended for use in routine physical examines, hospital monitoring of bleeding in patients, and for screening for colorectal cancer or gastrointestinal (GI) bleeding from any source.
Device Story
hema-screen™ SPECIFIC Gold; immunochemical fecal occult blood test (iFOBT) for detection of gastrointestinal bleeding; modified version of hema-screen™ SPECIFIC (K060463). Modification involves name change and slimmer test cassette housing design. Device utilizes immunochemical assay methods to detect human hemoglobin in fecal samples. Intended for use by healthcare professionals to aid in identifying gastrointestinal pathology. Performance verified via FMEA risk analysis and comparative testing against predicate; results confirm equivalent safety and effectiveness.
Clinical Evidence
Bench testing only. Comparison of three lots of subject device against predicate device showed 100% agreement, 100% positive percent agreement, and 100% negative percent agreement. Results met acceptance criteria of >98% total agreement.
Technological Characteristics
Lateral flow immunoassay; sandwich dye conjugate; monoclonal antibodies; qualitative visual readout. Housing is a slimmer plastic design compared to predicate. No changes to lateral flow membrane material or chemical composition. Standalone device; no energy source required.
Indications for Use
Indicated for the qualitative detection of fecal occult blood, an indicator of gastrointestinal bleeding, in fecal samples.
Regulatory Classification
Identification
An occult blood test is a device used to detect occult blood in urine or feces. (Occult blood is blood present in such small quantities that it can be detected only by chemical tests of suspected material, or by microscopic or spectroscopic examination.)
Special Controls
*Classification.* Class II (special controls). A control intended for use with an occult blood test is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
K173212 — Instant-view-PLUS immunochemical Fecal Occult Blood Test · Alfa Scientific Designs, Inc. · Feb 15, 2018
K063673 — INNOVACON FLIPCARD FECAL OCCULT BLOOD TEST DEVICE · Innovacon, Inc. · Mar 5, 2007
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification OIR Decision Summary
To: THE FILE
RE: DOCUMENT NUMBER K171484
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):
1. The name and 510(k) number of the SUBMITTER'S previously cleared device: hema-screen™ SPECIFIC, K060463.
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for a product name change from hema-screen™ SPECIFIC to hema-screen™ SPECIFIC Gold and the outer dimensions of the test cassette. The test cassette of the hema-screen™ SPECIFIC Gold has a slimmer housing design while the hema-screen™ SPECIFIC test cassette has a rectangular housing design.
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and assay methods. The only difference between the subject device, hema-screen™ SPECIFIC Gold, and predicate, hema-screen™ SPECIFIC is the housing design of the test cassette.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. A Risk Analysis was used to assess the impact of the modifications on the device and its components. A Risk Analysis was performed following the Failure Modes and Effects Analysis (FMEA) Analysis procedure. A comparison study with fecal test samples was conducted between the predicate (hema-screen™ SPECIFIC) versus the candidate device (hema-screen™ SPECIFIC Gold). Study results showed no impact on device performance. The hema-screen™ SPECIFIC Gold is as safe and effective as the predicate device, hema-screen™ SPECIFIC.
b) Based on the Risk Analysis, an identification of the verification and/or validation required, including methods or tests used and acceptance criteria was applied and found to be satisfactory. No safety risks were introduced. Descriptions of verification and validation activities were provided for the hema-screen™ SPECIFIC Gold. Comparison results between hema-screen™ SPECIFIC Gold and hema-screen™ SPECIFIC were also provided.
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I
{1}
Page 2 of 2
recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.