K070660 · Alfa Scientific Designs, Inc. · KHE · Jun 21, 2007 · Hematology
Device Facts
Record ID
K070660
Device Name
INSTANT-VIEW FECAL OCCULT BLOOD (FOB) RAPID TEST
Applicant
Alfa Scientific Designs, Inc.
Product Code
KHE · Hematology
Decision Date
Jun 21, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.6550
Device Class
Class 2
Indications for Use
The INSTANT-VIEW® Fecal Occult Blood (FOB) Rapid Test is a qualitative immunoassay for the detection of Fecal Occult Blood by non-professional, lay individuals as an over-the-counter product for home use. Measurement of FOB is a useful as an aid to detect blood in stool as found in a number of gastrointestinal disorders, e.g. diverticulitis, colitis, polyps, and colorectal cancer. It is intended for over-the-counter use. The Fecal Occult Blood (FOB) Rapid Test is a qualitative immunoassay for the detection of Fecal Occult Blood by non-professional, lay individuals as an over-the-counter product for home use. Measurement of FOB is a useful as an aid to detect blood in stool as found in a number of gastrointestinal disorders, e.g. diverticulitis, colitis, polyps, and colorectal cancer. It is intended for over-the-counter use.
Device Story
The INSTANT-VIEW FOB Self-Test is a one-step lateral flow chromatographic immunoassay; detects human hemoglobin in fecal specimens. Device consists of a test strip in a plastic cassette housing; strip contains a conjugate pad with colloidal gold-labeled mouse anti-human hemoglobin monoclonal antibodies and a nitrocellulose membrane with test (T) and control (C) lines. User applies fecal sample; if human hemoglobin is present, it binds to the conjugate and is captured at the T line. Results are visually interpreted by the user. Intended for over-the-counter, home use by lay individuals. Provides a qualitative result to aid in detecting gastrointestinal bleeding; enables early identification of potential conditions like colorectal cancer. Offers higher specificity than traditional chemical-based tests by targeting human hemoglobin specifically, avoiding false positives from dietary red meat consumption.
Clinical Evidence
Bench testing only. Analytical performance established via spiked stool samples (0–500 ng/ml hHb). Precision/reproducibility confirmed across three lots. Analytical specificity verified against various dietary substances (meats, vegetables, iron, Vitamin C) with no interference. Method comparison study (n=40) showed 100% agreement with expected results. Multi-site evaluation (n=100) at physician office and medical laboratories showed 98% overall agreement. Consumer study (n=120) using spiked PBS samples showed 95% agreement with expected results compared to 100% for professional testers.
Technological Characteristics
Lateral flow chromatographic immunoassay. Components: colloidal gold-conjugated mouse anti-human hemoglobin monoclonal antibodies; nitrocellulose membrane coated with anti-human hemoglobin (test line) and goat anti-mouse IgG (control line). Detection limit: 50 ng/ml hHb in buffer or 50 µg hHb/g in stool. Standalone, single-use cassette format. No external energy source required.
Indications for Use
Indicated for non-professional, lay individuals for the qualitative detection of human blood in stool as an aid in identifying gastrointestinal disorders such as diverticulitis, colitis, polyps, and colorectal cancer.
Regulatory Classification
Identification
An occult blood test is a device used to detect occult blood in urine or feces. (Occult blood is blood present in such small quantities that it can be detected only by chemical tests of suspected material, or by microscopic or spectroscopic examination.)
Special Controls
*Classification.* Class II (special controls). A control intended for use with an occult blood test is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
Hemoccult® test by Beckman Coulter, Inc. (K880499)
Related Devices
K173212 — Instant-view-PLUS immunochemical Fecal Occult Blood Test · Alfa Scientific Designs, Inc. · Feb 15, 2018
K162333 — Wondfo One Step Fecal Occult Blood (FOB) Test · Guangzhou Wondfo Biotech Co., Ltd. · May 14, 2017
Submission Summary (Full Text)
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE
A. 510(k) Number:
K070660
B. Purpose for Submission:
Marketing in the U.S.
C. Analyte:
Human Hemoglobin
D. Type of Test:
Fecal Occult Blood
E. Applicant:
Alfa Scientific Designs, Inc.
F. Proprietary and Established Names:
INSTANT-VIEW® Fecal Occult Blood (FOB) Rapid Test
Fecal Occult Blood (FOB) Rapid Test
G. Regulatory Information:
1. Regulation section:
864.6550 Occult blood test
2. Classification:
Class II
3. Product Code:
KHE
4. Panel:
Hematology
H. Intended Use:
1. Intended use(s):
The INSTANT-VIEW® Fecal Occult Blood (FOB) Rapid Test is a qualitative immunoassay for the detection of Fecal Occult Blood by non-professional, lay individuals as an over-the-counter product for home use. Measurement of FOB is a useful as an aid to detect blood in stool as found in a number of gastrointestinal disorders, e.g. diverticulitis, colitis, polyps, and colorectal cancer. It is intended for over-the-counter use.
The Fecal Occult Blood (FOB) Rapid Test is a qualitative immunoassay for the detection of Fecal Occult Blood by non-professional, lay individuals as an over-the-counter product for home use. Measurement of FOB is a useful as an aid to detect blood in stool as found in a number of gastrointestinal disorders, e.g. diverticulitis, colitis, polyps, and colorectal cancer. It is intended for over-the-counter use.
2. Indication(s) for use:
N/A
3. Special condition for use statement(s):
{1}
Page 2 of 5
N/A
4. Special instrument Requirements:
N/A
I. Device Description:
The INSTANT-VIEW® Fecal Occult Blood (FOB) Rapid Test is a one-step lateral flow chromatographic immunoassay. The test strip is contained in a cassette and consists of a burgundy-colored conjugate pad containing colloidal gold coupled with mouse anti-human hemoglobin monoclonal antibodies and a nitrocellulose membrane containing a test line (T line) and a control line (C line). The T line is coated with anti-human hemoglobin antibodies, and the C line is coated with goat anti-mouse IgG antibodies.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Beckman Coulter, Inc. Hemoccult® Fecal Occult Blood Test
2. Predicate K number(s):
K880499
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | Qualitative detection of occult blood in feces | Same |
| Sample Type | Feces | Same |
| Differences | | |
| Item | Device | Predicate |
| Detection method | Immunochromatographic assay | Chemical guaiac based assay |
| Analyte | Human hemoglobin | Hemoglobin |
| Users | Healthcare professionals and lay persons (OTC) | Healthcare professional use |
K. Standard/Guidance Document Referenced (if applicable):
N/A
L. Test Principle:
This test kit is one-step lateral flow chromatographic immunoassay. The test consists of three burgundy colored conjugate pads containing mouse anti-hHb antibodies conjugated with colloidal gold and a nitrocellulose membrane containing the test line. The human hemoglobin line appears as a visible burgundy line if the concentration of hHb in the specimen at or above 50 ng/ml.
A fecal sample is collected and prepared for testing in a sample collection tube containing extraction buffer. The extracted sample is added directly to the test device and migrates by capillary action through the test strip. If the level of hemoglobin in the fecal specimen is at or above the cutoff concentration, the T line appears as a visible burgundy line. If the level of hemoglobin in the fecal specimen is below the cutoff, no T line develops.
{2}
Page 3 of 5
The control line is a internal procedural control. Goat anti-mouse IgG is used for the "Control Line" coating. When the gold conjugate reagent flows to the "Control Line" area, the goat anti-mouse IgG reacts to form a burgundy colored band regardless of the presence of hemoglobin.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
a. Precision/Reproducibility:
Human stool extraction samples were spiked with hHb at the following concentrations: 0, 37.5, 50, 62.5, and 500 ng/ml. 4 samples were tested at each concentration with three lots of the Alfa-Scientific FOBT. Samples with concentrations below 50 ng/ml hHb tested negative with the candidate device while samples at and above 50 ng/ml tested positive.
b. Linearity/assay reportable range: N/A
c. Traceability (controls, calibrators, or method):
Internal Control: Procedural controls are included in the test device. A magenta line appearing the control region in considered as internal procedural control. It confirms sufficient specimen volume, adequate membrane wicking and correct procedural technique.
External Control: Controls are not provided with this kit. It is recommended that positive and negative controls be performed to verify proper test performance.
d. Detection limit:
The minimal detection limit is 50 ng/ml of hHb in buffer or 50 µg hHb/g in stool.
e. Analytical specificity:
Positive and negative stool samples were spiked with the following substances: Beef, chicken, fish (meat extract), horse, goat, pig, rabbit, and sheep hemoglobin; horseradish peroxidase; red radish; raw turnip; cauliflower; broccoli; parsnip; cantaloupe; Vitamin C; and iron. Addition of these substances had no effect on the test results.
f. Assay cut-off:
### 2. Comparison studies:
a. Method comparison with predicate device:
{3}
Page 4 of 5
a. Forty human stool extraction samples were spiked with hHb at the following concentrations: 0, 37.5, 50, 62.5, and 500 ng/ml. The candidate device was compared to the predicate at each of the concentrations and agreed 100% of the time with the predicate and the expected results.
b. The candidate device was evaluated at three physician office laboratories and one medical laboratory by personnel with diverse educational backgrounds and work experience. One hundred human stool extraction samples were spiked with hHb at the following concentrations: 0, 37.5, 50, 62.5, and 500 ng/ml. (20 at each concentration) Results obtained from three POL sites agreed 97.7% with the expected results and results from the medical laboratory agree 99% with the expected results. Overall agreement was 98%.
c. The candidate device was evaluated in a consumer study of 120 lay individuals with various educational levels. 120 hHB spiked phosphate buffer saline (PBS) samples at the following concentrations: 0, 37.5, 62.5 and 500 ng/ml, were tested. Consumer results agreed 95% with the expected results, while professional results agreed 100% with the expected results.
b. Matrix comparison:
3. Clinical studies: N/A
a. Clinical sensitivity:
b. Clinical specificity:
c. Other clinical supportive data (when a and b are not applicable):
4. Clinical cut-off: N/A
5. Expected values/Reference range:
Negative
N. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
{4}
Page 5 of 5
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.