K992959 · Dade Behring, Inc. · JPA · Jan 11, 2000 · Hematology
Device Facts
Record ID
K992959
Device Name
BEHRING COAGULATION SYSTEM (BCS)
Applicant
Dade Behring, Inc.
Product Code
JPA · Hematology
Decision Date
Jan 11, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5425
Device Class
Class 2
Indications for Use
The Behring Coagulation System performs quantitative assays of specific parameters in human citrated plasma.
Device Story
The Behring Coagulation System (BCS) is a fully automated photometric coagulation analyzer used in clinical laboratories. It accepts human citrated plasma samples to perform coagulometric, chromogenic, and immunochemical assays. The system utilizes photometric technology at various wavelengths to measure immuno complexes and coagulation reactions. Results are provided to clinicians to assist in the diagnosis and monitoring of coagulation-related conditions. The modification adds immunochemical measuring capabilities, specifically for latex-enhanced turbidimetric tests like Advanced D-Dimer. The device automates the analysis process, reducing manual labor and providing quantitative data for clinical decision-making.
Clinical Evidence
Bench testing only. Correlation study of the Advanced D-Dimer assay against Asserachrom® D-Di using 316 samples (range 0.43–85.9 mg/l) showed a correlation coefficient of 0.91, y-intercept 0.54, and slope 0.98. Precision studies conducted per NCCLS EP5-A using two control levels and two human plasma pools showed inter-assay precision of 0.8–3.8% and intra-assay precision of 1.3–3.0%.
Technological Characteristics
Automated photometric coagulation analyzer. Employs coagulometric, chromogenic, and immunochemical (latex-enhanced turbidimetric) measuring methods. Photometric sensing principle. Designed for clinical laboratory use. Complies with 21 CFR 820 (Quality System Regulation).
Indications for Use
Indicated for in vitro diagnostic use in clinical laboratories for the quantitative assay of coagulation parameters including PT, APTT, Antithrombin IIIa, Batroxibin, D-dimer, Deficient Plasmas, Fibrinogen, Heparin, Plasminogen, Protein C (clotting and chromogenic), Thrombin Time, and Von Willebrand factor in human citrated plasma.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
K970431 — BEHRING COAGULATION SYSTEM · Behring Diagnostics, Inc. · May 5, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
Dade Behring Inc. Behring Coagulation System 510(k)-Modification
JAN 1 1 2000
# 510(k) Summary for Behring Coagulation System
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
1992959 The assigned 510(k) number is:
### 1. Manufacture's Name, Address, Telephone, and Contact Person, Date of Preparation:
Manufacturer:
Dade Behring Marburg GmbH Emil-von-Behring Str. 76 Marburg/Germany
Dade Behring Inc. Glasgow Site P.O. Box 6101 Newark, Delaware 19714 Attn: Rebecca S. Ayash
Preparation date:
Contact Information:
August 31, 1999
### 2. Device Name/ Classification:
Behring Coagulation System:
Multipurpose svstem for in vitro Coaqulation studies
Class II (864.5425)
Classification Number:
Identification of the Legally Marketed Device: 3.
Bio-Tek EL309 microplate reader (K842085)
### 4. Device Description:
The current BCS was determined to be substantially equivalent as a fully automated photometric coagulation analyzer in 510(k) notification K970431. The current BCS was cleared to perform coagulometric and chromogenic tests, such as the routine tests prothrombin time, partial thromboplastin time, and fibrinogen, as well as the special tests, single factor determination antithrombin IIIa, protein C and plasminogen. The modified BCS has the ability to perform coagulometric, chromogenic and immunochemical assays (e.g., latex enhanced turbidimetric test Advanced D-Dimer). The addition of the immunochemical measuring method is the subject of this modification.
#### ક. Device Intended Use:
The Behring Coagulation System performs quantitative assays of specific parameters in human citrated plasma.
000036
{1}------------------------------------------------
Dade Behring Inc. Behring Coagulation System 510(k)-Modification
- 6. Medical device to which equivalence is claimed and comparison information: There are a number of photometric analyzers for the measurement of immunochemical assays. One such product is the Bio-Tek EL309 microplate reader. This instrument is a photometric analyzer for measuring immuno complexes. The BCS is substantially equivalent in intended use and results obtained to the Bio-Tek EL309 microplate reader, which was the subject of 510(k) K842085. Both instruments use photometric technology at various wavelengths for the measurement of immuno complexes.
#### 7. Device Performance Characteristics:
### Correlation:
The Advanced D-Dimer assay was compared to the Asserachrom® D-Di by evaluating 316 samples ranging from 0.43 to 85.9 mg/l. A correlation coefficient of 0.91 was obtained, with a y-intercept value of 0.54 and a slope of 0.98.
### Precision:
Precision studies were performed by the evaluation of two levels of control material and two levels of human plasma pools in a manner consistent with NCCLS Guideline EP5-A. The inter-assay precision ranged from 0.8 to 3.8%, while the intra-assay precision ranged from 1.3 to 3.0%.
# 000037
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized caduceus-like symbol with three intertwined strands. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
# JAN 1 1 2000
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Rebecca S. Ayash Manager, Regulatory Affairs, Biology Dade Behring, Inc. P.O. Box 6101 Newark, Delaware 19714
Re: K992959
> Trade Name: Behring Coagulation System Regulatory Class: II Product Code: JPA Dated: December 6, 1999 Received: December 7, 1999
Dear Ms. Ayash:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
## Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrl/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Dade Behring Inc. Behring Coagulation System Additional Information 510(k) - K992959
# Indications Statement
Behring Coagulation System Device Name:
### Indications for Use:
The Behring Coagulation System (BCS) is an automated coagulation analyzer for in vitro diagnostic use in clinical laboratories. The instrument performs the following parameters:
- · Prothrombin Time (PT)
- ·Activated Pattial Thromboplastin Time (APTT)
- ·Antithrombin Illa
- •Batroxibin
- ·D-dimer
- ·Deficient Plasmas
- ·Fibrinogen
- ·Heparin
- ·Plasminogen
- ·Protein C-clotting
- ·Protein C-chromogenic
- · Thrombin Time
- .Von Willebrand factor
Rita E. Mader
(Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number --
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
Over-The-Counter-Use_ (Optional Format 1-2-96)
000009
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.