BEHRING COAGULATION SYSTEM
Device Facts
| Record ID | K970431 |
|---|---|
| Device Name | BEHRING COAGULATION SYSTEM |
| Applicant | Behring Diagnostics, Inc. |
| Product Code | JPA · Hematology |
| Decision Date | May 5, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 864.5425 |
| Device Class | Class 2 |
Indications for Use
The Behring Coagulation System is an instrument intended to perform quantitative assays of specific parameters in plasma.
Device Story
Fully automated coagulation instrument; processes plasma samples using Behring Diagnostics test kits. Performs coagulometric and chromogenic coagulation tests. Used in clinical laboratory settings by trained laboratory personnel. Instrument automates assay execution; provides quantitative results for coagulation parameters. Healthcare providers use output to assess patient coagulation status and guide clinical management of hemostatic disorders.
Clinical Evidence
Bench testing only. Comparative analysis performed against predicate devices (BCT, BFA) using various coagulometric and chromogenic assays; correlation coefficients ranged from 0.97 to 0.996. Precision studies conducted per NCCLS EP5 guidelines; within-run precision 0.5-6.3%, total precision 0.90-8.1%.
Technological Characteristics
Fully automated coagulation analyzer. Utilizes coagulometric and chromogenic detection principles. Designed for in vitro diagnostic use. System operates as a standalone laboratory instrument.
Indications for Use
Indicated for use in clinical laboratories to perform quantitative coagulometric and chromogenic coagulation assays on plasma samples.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
Related Devices
- K960875 — KC 40 MICRO COAGULATION ANALYZER SYSTEM (K3632) · Sigma Diagnostics, Inc. · Sep 10, 1996
- K962664 — ELECTRA 1800C AUTOMATIC COAGULATION ANALYZER · Medical Laboratory Automation Systems, Inc. · Sep 10, 1996
- K955720 — AMAX CS-190 COAGULATION ANALYZER SYSTEM · Sigma Diagnostics, Inc. · Mar 13, 1996
- K013114 — BCS SYSTEM · Dade Behring, Inc. · Nov 9, 2001
- K031377 — SYSMEX AUTOMATED COAGULATION ANALYZER CA500 SERIES (CA-510, -520, -530, -540. -550, AND -560 · Dade Behring, Inc. · Sep 8, 2003