The controls are used to verify the performance of the appropriate cartridge on the HMS Plus instrument
Device Story
HDR Control set consists of single-use, non-sterile, in-vitro diagnostic plasma controls; used at point-of-care with HMS Plus instrument. Controls verify functionality of heparinized channels within HDR cartridges by measuring clotting time ratios between heparinized and unheparinized channels. HDR Control 1 verifies channels 1, 2, 5, and 6; HDR Control 2 verifies channels 3, 4, 5, and 6. Healthcare providers use these controls to ensure instrument and cartridge performance prior to clinical use, ensuring accurate heparin dose response monitoring for patients.
Clinical Evidence
Bench testing only. Verification and validation activities performed per design control requirements; results demonstrated that predetermined acceptance criteria were met.
Technological Characteristics
Lyophilized control material (sheep plasma, hexadimethrine bromide). Storage 2-10°C. Rehydration time 3 minutes. Reconstitution stability 1 hour at 15-25°C. Used with HMS Plus instrument and HDR cartridges. Pass/fail criteria based on clot time ratio range.
Indications for Use
Indicated for use as quality control materials to verify the performance of specific cartridges on the HMS Plus coagulation instrument.
Regulatory Classification
Identification
An automated heparin analyzer is a device used to determine the heparin level in a blood sample by mixing the sample with protamine (a heparin-neutralizing substance) and determining photometrically the onset of air-activated clotting. The analyzer also determines the amount of protamine necessary to neutralize the heparin in the patient's circulation.
K051040 — HEPARIN DOSE RESPONSE CONTROLS, MODEL 306-20 POR · Medtronic Perfusion Systems · May 24, 2005
K042206 — PURPLE/BLACK HEPARIN ASSAY CONTROL · Medtronic Vascular · Oct 18, 2004
K042175 — HEPARINASE HR ACT CONTROL, MODEL 550-12 · Medtronic Vascular · Oct 18, 2004
K960234 — TAS HMT CONTROLS · Cardiovascular Diagnostics, Inc. · May 30, 1996
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification
ODE Review Memorandum (Decision Making Document is Attached)
To: THE FILE
RE: DOCUMENT NUMBER K043080
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable:
1. HMS Coagulation System Cartridges and Controls, K042070
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for the addition of a bi-level, Heparin Dose Response (HDR) Control set.
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, sample requirements, methodology, and performance.
Comparison of Predicate to Current Device
| Name | HMS Controls
K894317 | HR ACT Control
K894317 | HDR Controls |
| --- | --- | --- | --- |
| | | | |
| Indication for use | The controls are used to verify the performance of the appropriate cartridge on the HMS Plus instrument | Same | Same |
| Raw Material | Plasma and Heparin | Sheep ROB and Plasma | Sheep Plasma |
| Diluent | DI Water | DI Water | Hexadimethrine
Bromide Solution |
| Instrument | HMS and HMS Plus | Same | Same |
| Cartridge | HPT | HMS HR-ACT | HDR |
| Pass/Fail Criteria | Clot Time Cutoff + channel detect | Clot time range | Clot time ratio range |
| Storage Temp | 2-10°C | Same | Same |
| Rehydration Time | 3 minutes | 2 minutes | 3 minutes |
| Reconstitution stability | 2 hours at 2-25°C | 2 hours at 15-25°C | 1 hour at 15-25°C |
| Storage Type | Lyophilized | Same | Same |
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
{1}
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
c) A declaration of conformity with design controls. The declaration of conformity should include:
i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.
6. A Truthful and Accurate Statement, a 510(k) Summary and the Indications for Use Enclosure (and Class III Summary for Class III devices).
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
| (Reviewer's Signature) | 12-2-2004
(Date) |
| --- | --- |
| Comments | |
Revised: 8/1/03
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches