CELL-DYN 1200 SYSTEM

K991142 · Abbott Laboratories · GKZ · Jun 23, 1999 · Hematology

Device Facts

Record IDK991142
Device NameCELL-DYN 1200 SYSTEM
ApplicantAbbott Laboratories
Product CodeGKZ · Hematology
Decision DateJun 23, 1999
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 864.5220
Device ClassClass 2

Indications for Use

The CELL-DYN® 1200 System is a multi-parameter hematology analyzer intended to classify the following formed elements of EDTA anti-coagulated blood: White Blood Cell Parameters: WBC -- White Blood Cell or leukocyte count, GRAN -- Granulocyte absolute count, %GRAN -- Granulocyte percent, LYM - Lymphocyte absolute count, %L - Lymphocyte percent, MID - Mid-range absolute count, %M - Mid-range percent; Red Blood Cell Parameters: RBC -- Red Blood Cell or erythrocyte count, HCT -- Hematocrit, MCV -- Mean Corpuscular Volume, RDW -- Red Cell Distribution Width; Platelet Parameters: PLT - Platelet Count, MPV -- Mean Platelet Volume, *PDW - Platelet Distribution Width, *PCT - Plateletcrit; Hemoglobin Parameters: HGB -- Hemoglobin Concentration, MCH -- Mean Cell Hemoglobin, MCHC -- Mean Cell Hemoglobin Concentration. * Clinical significance has not been established for these parameters. Therefore, they are not reportable in the US.

Device Story

Automated hematology analyzer for clinical laboratory use; processes EDTA-anticoagulated whole blood specimens. System aspirates, dilutes, and mixes samples. Uses focused flow impedance transducer (two non-corrosive electrodes) to count/size WBCs, RBCs, and PLTs; uses LED/photodiode optical absorbance for HGB measurement. Microprocessor-controlled system; operator interface via keypad and monochrome LCD. Outputs include hemogram parameters (WBC, RBC, HGB, PLT, etc.) displayed on screen, printed via optional printer, or sent to LIS. Provides distributional flags for data review. Benefits include automated hematological analysis, reducing manual differential requirements. Healthcare providers use output for clinical decision-making regarding patient hematological status.

Clinical Evidence

Bench testing only. Performance evaluated via background, carryover, precision, correlation, and linearity studies. Data demonstrates performance to manufacturer's specifications and substantial equivalence to the CELL-DYN® 1700 predicate for all reported parameters (WBC, WBC differential, RBC, HGB, MCV, RDW, PLT, MPV).

Technological Characteristics

Focused flow impedance transducer for cell counting/sizing; optical absorbance for HGB (LED/photodiode). Microprocessor-based control, data acquisition, and analysis. Interface: keypad, monochrome LCD, RS232 LIS port, diskette drive. Cyanide-free HGB reagent. Standalone unit with optional printer.

Indications for Use

Indicated for clinical laboratory use to analyze EDTA-anticoagulated human whole blood specimens to count, size, and classify white blood cells, red blood cells, and platelets, and to measure hemoglobin concentration.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ CELL-DYN® 1200 System JUN 23 1999 II. 510(k) Summary per 21 CFR807.92 Abbreviated 510(k) Notification K991142 | Submitted By: | Abbott Laboratories<br>5440 Patrick Henry Drive<br>Santa Clara, CA 95054 | |----------------------|--------------------------------------------------------------------------| | Contact Person: | Janice E. Brown<br>(408) 567-3521 | | Date Prepared: | April 2, 1999 | | Proprietary Name: | CELL-DYN® 1200 System | | Common Name: | Automated Hematology Analyzer | | Classification Name: | Automated Differential Cell Counter<br>21 CFR 864.5220 | | Predicate Devices: | CELL-DYN® 1700 System, K870233<br>Manual Differential | ### Description of Device: The CELL-DYN® 1200 System is a single unit device with a stand alone printer used in the clinical laboratory setting. The Analyzer aspirates, dilutes, mixes and analyzes each whole blood specimen. The computer in the Analyzer controls all system processing, provides the primary operator interface with the system, and controls the optional printer, which generates reports. The CELL-DYN® 1200 System is a multi-parameter, automated hematology analyzer designed to use EDTA-anticoagulated whole blood specimens to count, size and classify cells by using an focused flow impedance transducer, equipped with two non-corrosive electrodes to measure WBCs, and PLTs. The HGB is measured colorimetrically in the HGB transducer by an LED light source and a light sensitive diode. The analyzer is menu-driven and controlled by a microprocessor. {1}------------------------------------------------ # II. 510(k) Summary per 21 CFR807.92 (cont.): # Intended Use: The CELL-DYN® 1200 System is a multi-parameter hematology analyzer intended to classify the following formed elements of EDTA anti-coagulated blood: | White Blood Cell Parameters: | Red Blood Cell Parameters: | |-------------------------------------------------------------------------------------------------------------------------|--------------------------------------------| | WBC -- White Blood Cell or leukocyte count | RBC -- Red Blood Cell or erythrocyte count | | GRAN -- Granulocyte absolute count | HCT -- Hematocrit | | %GRAN -- Granulocyte percent | MCV -- Mean Corpuscular Volume | | LYM - Lymphocyte absolute count | RDW -- Red Cell Distribution Width | | %L - Lymphocyte percent | | | MID - Mid-range absolute count | Hemoglobin Parameters: | | %M - Mid-range percent | HGB -- Hemoglobin Concentration | | Platelet Parameters: | MCH -- Mean Cell Hemoglobin | | PLT - Platelet Count | MCHC -- Mean Cell Hemoglobin Concentration | | MPV - Mean Platelet Volume | | | *PDW - Platelet Distribution Width | | | *PCT - Plateletcrit | | | * Clinical significance has not been established for these<br>parameters. Therefore, they are not reportable in the US. | | # Technological Characteristics: The CELL-DYN® 1200 System utilizes the following technologies: optical absorbance methods; focused flow impedance; microprocessors for systems control, data acquisition, and data analysis; video keypads, liquid crystal display (LCD), optional printer and an on-line LIS interface. # Similarities and Differences: The following table describes the various parameters, subsystems and interfering substances of the CELL-DYN® 1200 System as compared to the predicate device the CELL-DYN® 1700 System. The two systems are similar in that: a) Both systems provide quantitation of the hemogram parameters in EDTAanticoagulated human whole blood specimens. b) Both systems will accept specimens presented manually by the operator. c) Both systems automatically aspirate the specimen and present it for automated processing. {2}------------------------------------------------ # II. 510(k) Summary per 21 CFR807.92 (cont.): # Similarities and Differences (cont'd): The two systems are similar in that: d) Both systems utilize optical absorbance methods. e) Both systems use microprocessors for systems control, data acquisition, and data analysis. f) Both systems accept input from keypads and output data to a video screen, diskette, optional printer, and on-line LIS. g) Both systems provide distributional flags to assist in data review. The two systems are different in that: a) The CELL-DYN® 1200 System utilizes a focused flow impedance measurement, while the CELL-DYN® 1700 System utilizes impedance. b) The CELL-DYN® 1700 utilizes both a keyboard and a hard drive for data operations, while the CELL-DYN® 1200 utilizes keypads and a diskette. c) The CELL-DYN® 1200 uses a cyanide free hemoglobin reagent while the CELL-DYN® 1700 utilizes a cyanide containing hemoglobin reagent. d) The CELL-DYN® 1200 utilizes a liquid crystal monochrome display while the CELL-DYN® 1700 utilizes high resolution color monitor. {3}------------------------------------------------ : : : - #### II.___________________________________________________________________________________________________________________________________________________________________________ 510(k) Summary per 21 CFR807.92 cont.: # Table: Comparison of the CELL-DYN® 1200 to the CELL-DYN® 1700 | | CELL-DYN® 1200 | CELL-DYN® 1700 K870233 | |----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------| | Intended Use | The CELL-DYN® 1200 System is a<br>multi-parameter, automated hematology<br>analyzer intended to classify the<br>following formed elements of EDTA<br>anti-coagulated blood: | Same | | Measurement(s)<br>and (Parameters) | Focused Flow Impedance | Impedance | | | • White Blood Cell (WBC) Count | • White Blood Cell (WBC) Count | | | • Lymphocyte # and % | • Lymphocyte # and % | | | • Mid Cells # and % | • Mid Cells # and % | | | • Granulocyte # and % | • Granulocyte # and % | | | • Red Blood Cell (RBC) Count | • Red Blood Cell (RBC) Count | | | • Red Cell Distribution Width (RDW) | • Red Cell Distribution Width (RDW) | | | • Mean Corpuscular Volume (MCV) | • Mean Corpuscular Volume (MCV) | | | • Platelet (PLT) Count | • Platelet (PLT) Count | | | • Mean Platelet Volume (MPV) | • Mean Platelet Volume (MPV) | | | • Plateletcrit (PCT)** | • Plateletcrit (PCT)** | | | • Platelet Distribution Width (PDW)** | • Platelet Distribution Width (PDW)** | | | ** for laboratory use only; not reportable<br>in US | ** for laboratory use only; not reportable<br>in US | | | Optical Absorbance | Optical Absorbance | | | • Hemoglobin (HGB) | • Hemoglobin (HGB) | | | Calculated values: | Using Modified Cyanmethemoglobin<br>Calculated values: | | | • Hematocrit (HCT) | • Same | | | • Mean Cell Hemoglobin (MCH) | | | | • Mean Cell Hemoglobin<br>Concentration (MCHC) | | | Device Description | Main Components: | Same | | | • Analyzer w/integrated display/keypad | | | Interface<br>(Data Output/<br>Data Input) | Data Output:<br>• ¼ VGA LCD Display, High Resolution<br>Monochrome<br>• Printer (optional)<br>• RS232 LIS Interface Port<br><br>Data Input:<br>• Keypad | • High Resolution Color Monitor | | | | • Keypad | | | | • Keyboard | | Specimen Type | EDTA (K3, K2) Anticoagulated Human<br>Whole Blood | Same | | | CELL-DYN® 1200 | CELL-DYN® 1700 K870233 | | Sampling | • Direct sampling of a well mixed whole<br>blood specimen from an open or closed<br>collection tube that has been identified<br>and presented manually by the operator. | Same | | Dilution | • Automatic dilution of the aspirated<br>sample and presentation of each<br>dilution for measurement. | Same | | Operating<br>Principles | • Enumeration and sizing of WBCs,<br>RBCs, and PLTs by impedance | Same | | | • HGB is measured by optical<br>absorbance. | Same | | Potential<br>Interfering<br>Substances | • WBC | Same | | | • NRBCs | | | | • Lyse Resistant RBCs | | | | • PLT clumps | | | | • Cryoglobulin and cryofibrinogen | | | | RBC | Same | | | • Elevated WBCs | | | | • Increased # giant PLTs | | | | • Autoagglutination | | | | • In vitro hemolysis | | | | HGB | Same | | | • Elevated WBC | | | | • Increased plasma substances<br>(triglycerides, bilirubin, in vivo<br>hemolysis) | | | | • Lyse Resistant RBCs | | | | MCV | Same | | | • Elevated WBCs | | | | • Increased # giant PLTs | | | | • Hyperglycemia | | | | • In vitro hemolysis | | | | CELL-DYN® 1200 | CELL-DYN® 1700 K870233 | | Potential<br>Interfering<br>Substances Cont. | PLT<br>• WBC fragments<br>• Increased # giant PLTs<br>• Microcytic RBCs<br>• PLT clumping<br>• Cryoglobulins<br>• In vitro hemolysis | Same | | | • MPV<br>• None Stated | Same | {4}------------------------------------------------ , - # II. # Table: Comparison of the CELL-DYN® 1200 to the CELL-DYN® 1700 {5}------------------------------------------------ #### 510(k) Summary per 21 CFR807.92 cont .: II. # Table: Comparison of the CELL-DYN® 1200 to the CELL-DYN® 1700 # Equivalency Data: The data compiled to support the claim that the CELL-DYN® 1200 System is substantially equivalent to the CELL-DYN® 1700 and the manual differential includes background, carryover, precision, correlation, linearity, and flagging data. Equivalence is demonstrated between the CELL-DYN® 1200 System and the CELL-DYN® 1700 System for the following measured parameters: White Blood Cell (WBC), three part WBC Differential sub-populations, Red Blood Count (RBC), Hemoglobin concentration (HGB), Mean Corpuscular Volume (MCV), Red Cell Distribution Width (RDW), Platelet Count (PLT), Mean Platelet Volume (MPV). Background, carryover, precision, correlation, and linearity data show performance to manufacturer's specifications. #### Conclusion: The conclusion of the testing performed on the CELL-DYN® 1200 System at Abbott Laboratories, Santa Clara, CA demonstrates the device is as safe and effective and performs as well as or better than the CELL-DYN® 1700 System. 12 {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle. JUN 23 1999 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Janice E. Brown Regulatory Affairs Manager Abbott Laboratories 5440 Patrick Henry Drive Santa Clara, California 95054 Re: K991142 Trade Name: CELL-DYN® 1200 System Regulatory Class: II Product Code: GKZ Dated: April 2, 1999 Received: April 5, 1999 Dear Mr. Brown: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {7}------------------------------------------------ Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in yitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Putman Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure : - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - {8}------------------------------------------------ CELL-DYN® 1200 System # VI. INDICATIONS FOR USE STATEMENT 510(k) Number (if known):_ Kgy 1147 Device Name: CELL-DYN® 1200 System Indications For Use: The CELL-DYN® 1200 is a multi-parameter hematology analyzer intended to classify the following formed elements of EDTA anti-coagulated blood: White Blood. Cell Parameters: WBC - White Blood Cell or leukocyte count GRAN -- Granulocyte absolute count %GRAN -- Granulocyte percent LYM - Lymphocyte absolute count %L - Lymphocyte percent MID – Mid-range absolute count %M - Mid-range percent Platelet Parameters: PLT - Platelet Count MPV – Mean Platelet Volume *PDW - Platelet Distribution Width *PCT - Plateletcrit Red Blood Cell Parameters: RBC -- Red Blood Cell or erythrocyte count HCT - Hematocrit MCV -- Mean Corpuscular Volume RDW -- Red Cell Distribution Width Hemoglobin Parameters: HGB - Hemoglobin concentration MCH -- Mean Cell Hemoglobin MCHC - Mean Cell Hemoglobin Concentration * Clinical significance has not been established for these parameters. Therefore, they are not reportable in the US Pitu E. Malvin prescription ✓
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...