BAYER ADVIA 2120 HEMATOLOGY ANALYZER WITH AUTOSLIDE SYSTEM
Applicant
Bayer Healthcare, LLC
Product Code
GKZ · Hematology
Decision Date
Sep 17, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
The Bayer ADVIA 2120 Hematology System with Autoslide is a quantitative, automated hematology analyzer that provides a leukocyte differential count and reticulocyte analysis for in vitro diagnostic use in clinical laboratories. In addition, the system provides the added capability to automatically prepare and stain high-quality blood smears on a microscope slide.
Device Story
Modification to ADVIA 120 hematology analyzer; addition of integrated slide maker/stainer. Device detects abnormal samples; automatically reflexes to create slide for visual examination. Used in clinical laboratory settings by trained personnel. Output is prepared slide for microscopic review. Modification maintains fundamental scientific technology of original analyzer.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and performance comparisons to the predicate device.
Technological Characteristics
Integrated slide maker/stainer module added to ADVIA 120 hematology analyzer. System utilizes automated reflex logic based on sample abnormality detection. No changes to fundamental scientific technology.
Indications for Use
Indicated for in vitro diagnostic use in clinical laboratories for quantitative automated hematology analysis, including leukocyte differential counts, reticulocyte analysis, and automated blood smear preparation and staining.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
K971998 — ADVIA 120 HEMATOLOGY SYSTEM, IN VITRO DIAGNOSTIC SYSTEM · Bayer Corp. · Aug 29, 1997
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification
ODE Review Memorandum
To: THE FILE
RE: DOCUMENT NUMBER K042251
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):
1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.)
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials.
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for the purpose of adding a slide maker/stainer integrated to the original ADVIA 120 hematology analyzer so that when the analyzer detects an abnormal sample it can reflex the device to make a slide of that sample for visual examination.
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, technology, and reagents.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
c) A declaration of conformity with design controls. The declaration of conformity should include:
i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and
ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review.
6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices).
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Comments
(Reviewer's Signature)
(Date)
revised:3/27/98
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.