STA COMPACT AUTOMATED MULTI-PARAMETRIC ANALYZER
Device Facts
| Record ID | K961579 |
|---|---|
| Device Name | STA COMPACT AUTOMATED MULTI-PARAMETRIC ANALYZER |
| Applicant | American Bioproducts Co. |
| Product Code | GKP · Hematology |
| Decision Date | Jul 10, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 864.5400 |
| Device Class | Class 2 |
Indications for Use
The STA Compact® instrument is a fully automated laboratory analyzer designed to perform tests which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.
Device Story
Fully automated laboratory analyzer; performs clot-based and photometric coagulation assays. Input: patient samples and reagents. Operation: random access testing; automated sample handling, reagent delivery, analysis, and reporting. CPU manages instrument functions, patient results, quality control, and maintenance. Barcode-enabled reagent management; supports third-party reagents. Used in clinical laboratory settings by trained laboratory personnel. Output: coagulation test results. Benefit: walkaway system efficiency; aids diagnosis of coagulation disorders and anticoagulant therapy monitoring.
Clinical Evidence
Bench testing only; sample-to-sample carryover demonstrated at 0.05%.
Technological Characteristics
Automated multi-parametric analyzer; clot-based and photometric detection; random access; barcode-enabled reagent management; CPU-controlled; supports Diagnostica Stago and third-party reagents.
Indications for Use
Indicated for use in clinical laboratories to perform coagulation tests for diagnosing coagulation abnormalities or monitoring anticoagulant therapy in patients requiring such assessment.
Regulatory Classification
Identification
A coagulation instrument is an automated or semiautomated device used to determine the onset of clot formation for in vitro coagulation studies.
Special Controls
*Classification.* Class II (special controls). A fibrometer or coagulation timer intended for use with a coagulation instrument is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
- STA® analyzer (K942117)
Related Devices
- K983460 — STAR AUTOMATED MULTI-PARAMETRIC ANALYZER · American Bioproducts Co. · Nov 12, 1998
- K093001 — STA-R EVOLUTION EXPERT SERIES HEMOSTASIS SYSTEM AUTOMATED MULTI-PARAMETRIC ANALYZER · Diagnostica Stago, Incorporated · Dec 18, 2009
- K031377 — SYSMEX AUTOMATED COAGULATION ANALYZER CA500 SERIES (CA-510, -520, -530, -540. -550, AND -560 · Dade Behring, Inc. · Sep 8, 2003
- K250965 — Automated Blood Coagulation Analyzer CN-Series (CN-6000) · Sysmex America, Inc. · Jun 2, 2025
- K151867 — STA R MAX · Diagnostica Stago S.A.S · Aug 7, 2015