ACL TOP 700 LAS

K091980 · Instrumentation Laboratory CO · GKP · Sep 22, 2009 · Hematology

Device Facts

Record IDK091980
Device NameACL TOP 700 LAS
ApplicantInstrumentation Laboratory CO
Product CodeGKP · Hematology
Decision DateSep 22, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 864.5400
Device ClassClass 2

Indications for Use

The ACL TOP 700 LAS is a bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The system provides results for both direct hemostasis measurements and calculated parameters.

Device Story

ACL TOP 700 modification; addition of hardware arm and interface module for connection to laboratory automation system (LAS). Point-in-Space design allows sample aspiration from primary container on automation tract without removal. Software modified to interface with external computer controlling LAS communications. Used in clinical laboratory settings by trained technicians. Output consists of coagulation test results; facilitates clinical decision-making regarding patient hemostasis status.

Clinical Evidence

Bench testing only; verification and validation activities performed per design control procedures (21 CFR 820.30) to confirm safety and efficacy of LAS interface modification.

Technological Characteristics

Coagulation analyzer with added hardware arm and interface module for LAS connectivity. Point-in-Space sampling mechanism. Software-controlled communication interface. Design controls applied per 21 CFR 820.30.

Indications for Use

Indicated for use in clinical laboratories for the analysis of patient samples to perform coagulation testing.

Regulatory Classification

Identification

A coagulation instrument is an automated or semiautomated device used to determine the onset of clot formation for in vitro coagulation studies.

Special Controls

*Classification.* Class II (special controls). A fibrometer or coagulation timer intended for use with a coagulation instrument is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification Review Memorandum To: THE FILE RE: DOCUMENT NUMBER K091980 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable: 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. ACL TOP (k073377) 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. The modification: Adding an extra arm and hardware to interface with laboratory automation system (LAS) 4. Comparison Information to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics. The difference: A Point-in-Space design solution is utilized where the patient sample remains under the control of the laboratory automated system. For sample analysis, an aliquot is aspirated by the ACL TOP 700 instrument and the primary container from the automation tract is not removed. The instrument software is modified to interface with an interface module computer, which controls the communications to the LAS system 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. FMEA (Failure mode and effects analysis) see section 15 b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied. (See section 15) c) A declaration of conformity (see Section 8) with design controls. The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. 6. A Truthful and Accurate Statement, a 510(k) Summary and the Indications for Use Enclosure. The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...