DADE BEHRING HEPARIN CALIBRATOR AND CONTROLS

K042941 · Dade Behring, Inc. · GIZ · Dec 2, 2004 · Hematology

Device Facts

Record IDK042941
Device NameDADE BEHRING HEPARIN CALIBRATOR AND CONTROLS
ApplicantDade Behring, Inc.
Product CodeGIZ · Hematology
Decision DateDec 2, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5425
Device ClassClass 2

Indications for Use

The Dade Behring Heparin Calibrator is an in vitro diagnostic product used to calibrate the Berichrom® Heparin assay for the measurement of unfractionated heparin. Dade Behring Heparin Controls are in vitro diagnostic products intended to be used as assayed, unfractionated heparin quality control materials for the Berichrom® Heparin assay.

Device Story

Lyophilized calibrator and quality control materials derived from citrated human plasma and porcine-sourced unfractionated heparin; used to calibrate and monitor performance of Berichrom® Heparin assay on photo-optical coagulation analyzers; provides reference values for heparin measurement; assists laboratory personnel in ensuring assay accuracy and quality control; supports clinical monitoring of heparin therapy.

Clinical Evidence

Bench testing only. Traceability established using USP Heparin Sodium Reference Standard. Method comparison performed on Dade Behring BCS® and Sysmex® CA-1500 analyzers showed high correlation (r=0.9985 and r=0.9996 respectively) against USP standard-spiked plasma samples. Reconstitution stability confirmed for 8 hours at 2-8°C.

Technological Characteristics

Lyophilized product; citrated human plasma base; porcine-sourced unfractionated heparin. Used as an accessory for coagulation analyzers. No electronic or software components.

Indications for Use

Indicated for use as a calibrator and quality control material for the Berichrom® Heparin assay to measure unfractionated heparin in human plasma samples.

Regulatory Classification

Identification

A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.

Special Controls

*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K042941 B. Purpose for Submission: New Device C. Measurand: Calibrator and Quality control material for heparin D. Type of Test: Quantitative E. Applicant: Dade Behring Inc. F. Proprietary and Established Names: Dade Behring Heparin Calibrator and Controls G. Regulatory Information: 1. Regulation section: 21 CFR 864.5425, Multipurpose system for in vitro coagulation studies 2. Classification: Class II 3. Product code: GIZ, Plasma, Control, Normal 4. Panel: Hematology (81) {1} H. Intended Use: 1. Intended use(s): Dade Behring Heparin Calibrator is an *in vitro* diagnostic product used to calibrate the Berichrom® Heparin assay for the measurement of unfractionated heparin. Dade Behring Heparin Controls are *in vitro* diagnostic products intended to be used as assayed, unfractionated heparin quality control materials for the Berichrom® Heparin assay. 2. Indication(s) for use: Dade Behring Heparin Calibrator is an *in vitro* diagnostic product used to calibrate the Berichrom® Heparin assay for the measurement of unfractionated heparin. Dade Behring Heparin Controls are *in vitro* diagnostic products intended to be used as assayed, unfractionated heparin quality control materials for the Berichrom® Heparin assay. 3. Special conditions for use statement(s): Not applicable. 4. Special instrument requirements: Not applicable. I. Device Description: Dade Behring Heparin Calibrator and Controls are lyophilized products prepared from citrated human plasma and contains unfractionated heparin form a porcine source. The kit consists of one calibrator and two levels of assayed controls intended to monitor the performance of Berichrom® Heparin reagent when testing for unfractionated heparin using coagulation analyzers. J. Substantial Equivalence Information: 1. Predicate device name(s): aca® Heparin Calibrator Ci-Trol® Heparin Controls, Low and High 2. Predicate 510(k) number(s): K843202, K812424 {2} 3. Comparison with predicate: # CALIBRATOR Similarities | Item | Device | Predicate | | --- | --- | --- | | | Heparin Calibrator | aca® Heparin Calibrator | | Intended Use | Used to calibrate the Berichrom® Heparin assay for the measurement of unfractionated heparin. | Used to calibrate the aca® discrete clinical analyzer for the Heparin method. This product was designed to meet the needs of users to assure accurate results over the assay range of this method. | | Composition | Citrated human plasma and heparin from porcine intestine. | Same | | Analytes | Unfractionated heparin | Same | | Form | Lyophilized | Same | Differences | Item | Device | Predicate | | --- | --- | --- | | Instrumentation | Photo-optical coagulation analyzers | aca® discrete clinical analyzer | CONTROLS | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | Heparin Controls, Level 1 and Level 2 | Ci-Trol® Heparin Controls, Low and High | | Intended Use | Used as assayed, unfractionated heparin quality control materials for Berichrom® Heparin assay. | Used as a control for heparin assay procedures. | | Composition | Citrated human plasm and heparin from porcine intestine. | Citrated human plasma with sodium heparin, buffer and stabilizers. | | Analytes | Unfractionated heparin | Same | | Form | Lyophilized | Same | {3} | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Instrumentation | Photo-optical coagulation analyzers | Photo-optical and mechanical coagulation analyzes. | **K. Standard/Guidance Document Referenced (if applicable):** USP Heparin Sodium Reference Standard **L. Test Principle:** Not Applicable. **M. Performance Characteristics (if/when applicable):** 1. **Analytical performance:** a. **Precision/Reproducibility:** Not Applicable. b. **Linearity/assay reportable range:** Not Applicable. c. **Traceability, Stability, Expected values (controls, calibrators, or methods):** Using at least duplicate determinations, reconstitution stability data met the acceptance criteria of recovering within the assigned values when stored for eight (8) hours at 2 to 8°C. Different dilutions of USP Heparin Sodium Reference Standard were spiked into citrated plasma samples and the heparin recovery evaluated using a calibration curve established using the USP Heparin Sodium Reference Standard. The same samples were then measured using a calibration curve established with the new Heparin calibrator. Results were compared using regression analysis and the following statistics were obtained: Dade Behring BCS® Analyzer: Y = 1.07 x -0.01, r = 0.9985 Sysmex® CA-1500 Analyzer: Y = 0.97 x +0.01, r = 0.9996 d. **Detection limit:** Not Applicable. {4} e. Analytical specificity: Not Applicable. f. Assay cut-off: Not Applicable. 2. Comparison studies: a. Method comparison with predicate device: Not Applicable. b. Matrix comparison: Not Applicable. 3. Clinical studies: a. Clinical Sensitivity: Not Applicable. b. Clinical specificity: Not Applicable. c. Other clinical supportive data (when a. and b. are not applicable): 4. Clinical cut-off: Not Applicable. 5. Expected values/Reference range: The product insert contains mean values derived from multiple analyzers and reagent lots, and are specific for the lot of product. It is recommended that each laboratory establish its own means and acceptable ranges and use those provided only as guides. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. {5} O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 6
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...