The GSA kit is for the evaluation of human semen. The kit provides the user with a set of reagents used to evaluate semen quality by measuring four parameters, recommended by the World Health Organization (WHO), that are used to help determine whether infertility is caused by abnormalities of one or more of them. The parameters to be tested for include: - . Sperm count - Sperm motility - . Sperm vitality - White blood cell (WBC) count . The GSA kit is an in vitro diagnostic kit intended for use in a clinical laboratory test setting such as a test facility capable of providing laboratory assessment of male infertility, by skilled laboratory technicians, familiar with the handling of semen samples. This product does not contain biological materials of human origin.
Device Story
GSA kit uses reagents to prepare human semen samples for analysis via flow cytometry; replaces manual light microscopy and counting chamber methods. Flow cytometry allows analysis of significantly higher cell counts than manual methods, improving precision and reproducibility. Used in clinical laboratories by skilled technicians; results assist clinicians in diagnosing male infertility based on WHO-recommended parameters. Provides quantitative data on sperm count, motility, vitality, and WBC count. Benefits include higher consistency and reduced technician-dependent variability compared to manual assessment.
Clinical Evidence
Prospective, multi-center clinical study at two sites (Israel and USA) comparing GSA kit (flow cytometry) to manual WHO-recommended method. Evaluated sperm count, motility, vitality, and WBC count. Results showed comparable accuracy to manual method with significantly higher reproducibility. Inter-technician agreement was 91.6% for sperm count and 82.3% for motility in normal samples. Intra- and inter-technician standard deviations were lower for the GSA kit, indicating higher precision.
Technological Characteristics
In vitro diagnostic reagent kit for flow cytometry analysis. Reagents facilitate sample preparation for automated cell counting and characterization. Does not contain human-derived biological materials. Designed for use with standard flow cytometer instrumentation. Shelf life of 6 months under refrigerated storage.
Indications for Use
Indicated for the evaluation of human semen in clinical laboratory settings to assess male infertility by measuring sperm count, motility, vitality, and white blood cell count.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Related Devices
K071737 — GENERAL SEMEN ANALYSIS KIT · Dyn-Bioshaf (2006) , Ltd. · Oct 25, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 03 2003 K024337
# 510(k) Summary of Safety and Effectiveness
# Intended Use
The GSA kit is for the evaluation of human semen. The kit provides the user with a set of reagents used to evaluate semen quality by measuring four parameters, recommended by the World Health Organization (WHO), that are used to help determine whether infertility is caused by abnormalities of one or more of them. The parameters to be tested for include:
- . Sperm count
- Sperm motility
- . Sperm vitality
- White blood cell (WBC) count .
The GSA kit is an in vitro diagnostic kit intended for use in a clinical laboratory test setting such as a test facility capable of providing laboratory assessment of male infertility, by skilled laboratory technicians, familiar with the handling of semen samples. This product does not contain biological materials of human origin.
The performance of the GSA kit was established by comparison to the manual method as recommended by the WHO.
The GSA kit involves specimen acquisition by ejaculation of semen by the patient for testing purposes. This assay presents no more of a safety hazard than the routine tests where semen samples are obtained from subjects.
# Test Performance
The clinical performance of the GSA kit was evaluated via blind, multi-center, prospective clinical studies. The primary objective was to determine whether evaluation of sperm cells using the GSA kit is at least as accurate as the Routine Method. Another objective was to show that using the GSA kit provides more consistent results that are less dependent on the skills of an individual technician.
The study took place at two sites:
- IVF unit and outpatient clinic Lin Center, Carmel Hospital, Israel. .
- Baylor College of Medicine, Houston Medical Center, USA. ●
Each of the above sites has many years of experience performing semen analysis using the Routine Method.
{1}------------------------------------------------
A skilled laboratory technician evaluated each semen sample by the Routine Method and by a different skilled laboratory technician using the Flowcytometry method. The evaluation was based on the parameters recommended by the WHO (see below).
# Sperm Count
Comparison of the GSA kit and the Routine Method showed that using the GSA kit with the Flowcytometer provides comparable accuracy and more precise results. The Flowcytometer allows counting of approximately 100 times more cells as compared to the light microscope and counting chamber.
The inter- and intra-technician experiments showed that in semen samples with approximately normal sperm count there is 91.6% agreement between the two methods while the reproducibility achieved by the Flowcytometry method is significantly higher.
#### Sperm Motilitv
Comparison of the GSA kit with the Routine Method showed that using the GSA kit provides comparable accuracy and more precise results. Again, because of the large amount of sperm cells that can be analyzed and the higher precision in cell count, the determination of the percentage of motile cells is statistically more reproducible using the GSA kit.
The inter- and intra- technician experiments showed that in semen samples characterized by approximately normal sperm count, there is close agreement (82.3%) between the two methods while the reproducibility achieved by the GSA kit is significantly higher.
#### Sperm Vitality
Comparison of the GSA kit with the Routine Method showed that using the GSA kit provides comparable accuracy and significantly more precise results.
The inter- and intra-technician experiments showed that in semen samples characterized by approximately normal sperm count, there is close agreement between the two methods while the reproducibility achieved by the GSA kit is significantly higher.
#### White Blood Cells (WBC)
Comparison of the GSA kit with the Routine Method showed that using the GSA kit provides comparable accuracy and more precise results.
#### Intra-Technician Precision
In an attempt to determine and compare the intra-technician precision of the two methods, five semen samples were repeatedly tested by both conventional (light microscope and counting chamber) analysis and Flowcytometry analysis according to the WHO manual. For each analysis method the same technician repeatedly evaluated the
{2}------------------------------------------------
semen samples in order to evaluate the intra-technician variability obtained by each methodology. The results show that by using the GSA kit the typical intra-technician standard deviation for each parameter is smaller then by using the Routine Method indicating a high intra-technician precision for Flowcytometry assay.
# Inter-Technician Precision
To compare the inter-technician precision of the two methods six semen samples were tested by both conventional analysis and Flowcytometry analysis according to the WHO manual method. For each analysis method three technicians evaluated all semen samples and the results obtained for each sample by each technician were compared in order to evaluate the inter-technician variability obtained by each methodology. The results show that by using the GSA kit the typical inter-technician standard deviation for each parameter is significantly smaller then by using the Routine Method indicating a high inter-technician precision for Flowcytometry assay.
### Interfering Substances
The following table summarizes data obtained using human semen samples characterized by certain abnormalities.
| Interfering Substance | Amount | Result |
|------------------------------------|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| High level of WBC | $\geq 4 \times 10^7$ /ml | May cause high staining background and by<br>this false WBC count result. A new sample<br>must be obtained after the infection is over. |
| Fungi infection | Visible<br>cloudiness | May cause auto-fluorescence which can result<br>in erroneous results. Thus, contaminated<br>semen samples should not be tested. |
| Extremely viscous<br>semen sample. | - | In rare cases dilution of the sample is<br>practically impossible and dilution treatment<br>using Bromolin may be required. |
# Product Stability
The stability of the kit was tested in real time as the components were stored refrigerated and the performance tested periodically. The shelf life of the kit has been determined to be 6 months at the specified storage conditions.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circle around the eagle. The eagle is depicted in a simple, line-art style, and the text is in a sans-serif font. The overall design is clean and professional.
10903 New Hampshire Avenue Silver Spring, MD 20993
DEC 21 2012
Devices & Diagnostics Consulting Group, Inc. c/o Mr. Thomas M. Tsakeris President 16809 Briardale Road Rockville, Maryland 20855
Re: k024337
Trade/Device Name: BioShaf Limited General Semen Analysis (GSA) Kit Regulation Number: 21 CFR § 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: Class II Product Code: GKZ Date: April 28, 2003 Received: April 28, 2003
Dear Mr. Tsakeris:
This letter corrects our substantially equivalent letter of July 3, 2003.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
Page 2 - Mr. Thomas Tsakeris
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Sincerely yours,
marie m chan
Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
510(k) Number: K024337
Device Name: General Semen Analysis (GSA) Kit
Indications for Use:
The GSA kit is for the evaluation of human semen. The kit provides the user with a set of reagents used to evaluate semen quality by measuring four (4) parameters recommended by the World Health Organization (WHO) that are used to determine whether infertility is caused by abnormalities of one or more of them. The parameters include:
- . Sperm count
- Sperm motility .
- Sperm vitality t
- . White blood cell (WBC) count
In most fertility test laboratorics these parameters are manually evaluated on a routine basis using a light microscope and a counting chamber. The performance characteristics of the GSA kit is based on a comparison between the GSA kit and the routine manual method performed according to the latest edition of the Manual for the Examination of Human Semen and Sperm published by the WHO.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of In Vitro Diagnostic Evaluation and Safety (OIVDES)
Professional use:
Prescription Use:
(Per 21 CFR 801.109)
OR
Over-the-counter Use: (Optional Format 1-2-96)
:
:
Caroline Bartoli
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K024337
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.