← Product Code [JPA](/submissions/HE/subpart-f%E2%80%94automated-and-semi-automated-hematology-devices/JPA) · K993678

# THROMBELASTOGRAPH COAGULATION ANALYZER TEG - 5000 SERIES (K993678)

_Haemoscope Corp. · JPA · May 5, 2000 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HE/subpart-f%E2%80%94automated-and-semi-automated-hematology-devices/JPA/K993678

## Device Facts

- **Applicant:** Haemoscope Corp.
- **Product Code:** [JPA](/submissions/HE/subpart-f%E2%80%94automated-and-semi-automated-hematology-devices/JPA.md)
- **Decision Date:** May 5, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5425
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

The TEG - 5000 Series Analyzer is intended to be used to provide a quantitative and qualitative indication of the coagulation state of a blood sample by monitoring, measuring, and reporting coagulation parameter information. The Thrombelastograph (TEG) Coagulation Analyzer TEG - 5000 Series records the kinetic changes in a sample of whole blood, plasma or platelet rich-plasma as the sample clots, retracts and/or lyses (breaks apart). Results from the TEG analyzer should not be the sole basis for a patient diagnosis; TEG results should be considered along with a clinical assessment of the patient's condition and other coagulation laboratory tests. For Professional Use Only.

## Device Story

TEG-5000 Series Analyzer monitors coagulation state of whole blood, plasma, or platelet-rich plasma. System comprises two-column instrument, computer interface, software, and disposable cups/pins. Principle of operation: blood sample placed in oscillating cylindrical cup; pin suspended by torsion wire immersed in sample. Fibrin-platelet bonding links cup and pin; torque transmitted to pin reflects clot strength. Magnitude of pin motion measured as sample clots, retracts, and lyses. Used by professionals in clinical settings to assist assessment of hemostasis. Output provides quantitative/qualitative coagulation parameters. Results used alongside clinical assessment and other lab tests to inform management of hemorrhage or thrombosis. Benefits include real-time kinetic monitoring of coagulation process.

## Clinical Evidence

No clinical data. Evidence consists of comparative performance testing and software validation testing to demonstrate adherence to design specifications and equivalence to the predicate device.

## Technological Characteristics

Two-column instrument with computer interface. Measures clot physical properties via oscillating cup and torsion wire pin. Includes disposable sample cups and pins. Software-driven menu structure. Connectivity via computer interface module.

## Regulatory Identification

A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.

## Special Controls

*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

## Predicate Devices

- Thrombelastograph® Coagulation Analyzer – 3000S ([K954137](/device/K954137.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary and Certification

[As required by 21 CFR 807.92(c)]

## 1. Submitter's Name / Contact Person

| Eli Cohen, Ph.D.<br>President and CEO<br>Haemoscope Corporation<br>7855 Gross Point Road, Unit G-4<br>Skokie, IL 60077 | Tel: (847) 329-0001, (800) 438-2834<br>Fax: (847) 329-0003 |
|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|

#### 2. General Information

| Trade Name                              | Thrombelastograph® Coagulation Analyzer TEG® - 5000 Series                     |
|-----------------------------------------|--------------------------------------------------------------------------------|
| Common / Usual Name                     | Thrombelastograph Instrument                                                   |
| Classification Name                     | Multipurpose System for In Vitro Coagulation Studies                           |
| Identification of<br>Equivalent Devices | Thrombelastograph® Coagulation Analyzer – 3000S,<br>Haemoscope Corp. (K954137) |

### 3. Device Description

The TEG - 5000 Series Analyzer consists of a two-column TEG instrument, a computer interface module, software, and disposable sample cups and pins. The TEG measures the clot's physical property by the use of a special stationary cylindrical cup that holds the blood and is oscillated. A pin is suspended in the blood by a torsion wire and is monitored for motion. The torque of the rotation cup is transmitted to the immersed pin only after fibrin-platelet bonding has linked the cup and pin together. The strength of these fibrin-platelet bonds affects the magnitude of the pin motion, such that strong clots move the pin directly in phase with the cup motion. The magnitude of the output is, therefore, directly related to the strength of the formed clot. As the clot retracts or breaks apart, these bonds are broken and the transfer of cup motion is diminished.

#### 4. Intended Use

The TEG - 5000 Series Analyzer is intended to be used to provide a quantitative and qualitative indication of the coagulation state of a blood sample by monitoring, measuring, and reporting coagulation parameter information. The Thrombelastograph (TEG) Coagulation Analyzer TEG - 5000 Series records the kinetic changes in a sample of whole blood, plasma or platelet rich-plasma as the sample clots, retracts and/or lyses (breaks apart).

Results from the TEG analyzer should not be the sole basis for a patient diagnosis; TEG results should be considered along with a clinical assessment of the patient's condition and other coagulation laboratory tests. For Professional Use Only.

{1}------------------------------------------------

## 5. Technological Characteristic Comparisons

The Thrombelastograph Coagulation Analyzer TEG - 5000 Series is substantially equivalent to the Haemoscope Corporation Thrombelastograph Coagulation Analyzer -- 3000S (K954137). Compared to the predicate device, the TEG - 5000 Series Analyzer has the same intended use, principles of operation and function as the TEG - 3000S Analyzer. There are no technological differences between the two TEG analyzer families. The TEG - 5000 Series Analyzer, a modified version of the Haemoscope TEG 3000S Analyzer, offers instructions and software for determining the functional fibrinogen level of a clot sample. In addition, the TEG - 5000 Series offers an improved sample cup placement and retrieval system, and an improved software menu structure, enhancing ease of use and user safety.

#### 6. Summary of Studies

Comparative performance testing and software validation testing was performed to demonstrate that the TEG - 5000 Series analyzer meets established design specifications and performance requirements. The test results confirmed that, despite differences, the TEG - 5000 Series Analyzer meets established design specifications and performs as well as the predicate device. No new questions of safety or effectiveness were raised.

#### 7. Conclusion (statement of equivalence)

The data and information provided in this submission support a substantial equivalence determination and, therefore, clearance of the 510(k) premarket notification for the Thrombelastograph Coagulation Analyzer TEG - 5000 Series.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, which is a traditional symbol of medicine and healthcare. The logo is rendered in black and white.

MAY - 5 2000

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Eli Cohen, Ph.D. President and CEO Haemoscope Corporation 7855 Gross Point Road, Unit G-4 Skokie, Illinois 60077

Re: K993678

> Trade Name: Thrombelastograph® Coagulation Analyzer TEG® 5000 Series Regulatory Class: II Product Code: JPA Dated: April 3, 2000 Received: April 4, 2000

Dear Dr. Cohen:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nonication. The I Dr initially of succion for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire spoonly an invitro diagnostic devices), please contact the Office of Compliance at additionally 607.10 for in the arguestions on the promotion and advertising of your device, (2017) 591-1566. Traditional (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## INDICATIONS FOR USE STATEMENT

Page 1 of 1

510(k) Number (if known): K993678

Device Name:

Thrombelastograph® (TEG®) Coagulation Analyzer TEG® – 5000 Series

Indications for Use:

The Thrombelastograph (TEG) Coagulation Analyzer TEG - 5000 Series is a noninvasive diagnostic instrument designed to monitor and analyze the coagulation state of a blood sample in order to assist in the assessment of patient clinical hemostasis conditions. The TEG is indicated for use with adult patients where an evaluation of their blood coagulation properties is desired. Coagulations are commonly used to assess clinical conditions such as post-operative hemorrhage and / or thrombosis during and following cardiovascular surgery, organ transplantation, trauma, and cardiology procedures.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Prescription-✓

Concurrence of CDRH, Office of Device Evaluation (ODE)

Stur E. Madim

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K993678

---

**Source:** [https://fda.innolitics.com/submissions/HE/subpart-f%E2%80%94automated-and-semi-automated-hematology-devices/JPA/K993678](https://fda.innolitics.com/submissions/HE/subpart-f%E2%80%94automated-and-semi-automated-hematology-devices/JPA/K993678)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
