AMAX CS-190 COAGULATION ANALYZER SYSTEM
Device Facts
| Record ID | K955720 |
|---|---|
| Device Name | AMAX CS-190 COAGULATION ANALYZER SYSTEM |
| Applicant | Sigma Diagnostics, Inc. |
| Product Code | JPA · Hematology |
| Decision Date | Mar 13, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 864.5425 |
| Device Class | Class 2 |
Indications for Use
The Amelung AMAX CS-190 Coagulation Analyzer is an automated random access multipurpose analyzer. The AMAX CS-190 Coagulation Analyzer can be used for the detection of fibrin formation utilizing either mechanical principles (ball method) or photo-optical principles to perform clot based tests such as prothrombin time(PT), activated partial thromboplastin time (APTT), fibrinogen, factor assays, and other clotting tests. In addition, the AMAX CS-190 Coagulation Analyzer can be used for chromogenic assays such as antithrombin III (AT III).
Device Story
Automated random access multipurpose coagulation analyzer; utilizes mechanical (ball method) or photo-optical principles to detect fibrin formation. Performs clot-based assays (PT, APTT, fibrinogen, factor assays) and chromogenic assays (AT III). Used in clinical laboratory settings by trained personnel. Provides quantitative results for coagulation parameters; assists clinicians in diagnosing and monitoring coagulation disorders. Benefits patients through rapid, automated, and precise coagulation testing.
Clinical Evidence
Bench testing only. Comparison studies performed against TOA CA-5000 (mechanical), BBL Fibrometer, and manual spectrophotometric methods. PT (optical) r=0.989; APTT (optical) r=0.907; Factor IX (optical) r=0.954; Factor X (optical) r=0.980; Fibrinogen (optical) r=0.979; AT III r=0.848. Precision studies showed CV <4% for PT/APTT, <9% for factor assays, <10% for fibrinogen, and <6% for AT III.
Technological Characteristics
Automated random access analyzer; dual-mode detection: mechanical (ball method) and photo-optical. Performs clot-based and chromogenic assays. No specific materials, connectivity, or software architecture details provided.
Indications for Use
Indicated for clinical laboratory use to perform coagulation testing, including PT, APTT, fibrinogen, factor assays, and chromogenic assays like AT III, for patients requiring coagulation status assessment.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Reference Devices
- TOA Medical Electronics CA-5000
- BBL Fibrometer
Related Devices
- K021162 — AMAX DENSITY COAGULATION ANALYZER, AMAX DENSTINY COAGULATION ANALYZER-EU, AMAX DESTINY COAGULATION ANALYZER-REFURBISHED · Sigma Diagnostics, Inc. · Aug 30, 2002
- K960875 — KC 40 MICRO COAGULATION ANALYZER SYSTEM (K3632) · Sigma Diagnostics, Inc. · Sep 10, 1996
- K962664 — ELECTRA 1800C AUTOMATIC COAGULATION ANALYZER · Medical Laboratory Automation Systems, Inc. · Sep 10, 1996
- K972260 — CS-400 COAGULATION ANALYZER SYSTEM (A1208) · Sigma Diagnostics, Inc. · Oct 20, 1997
- K970431 — BEHRING COAGULATION SYSTEM · Behring Diagnostics, Inc. · May 5, 1997