Clinical patient samples from routine surgical/cardiology procedures
The clinical validation study utilized blood samples from 338 patients undergoing cardiovascular surgery, cardiology procedures, or liver transplantation to compare the performance of the new assay cartridge against the predicate device and a laboratory comparator.
338 patients (18 years or older) at increased risk of coagulopathy undergoing liver transplantation, cardiovascular surgery, or cardiology procedures; Sample Size: 338; Number of Sites: 8
TEG 6s with Citrated Multichannel Cartridge (K150041) and STA - Fibrinogen 5 (K840211)
Performance equivalence for parameters R, MA, and LY30
Indications for Use
The TEG 6s Hemostasis System consists of the TEG 6s Hemostasis Analyzer and the Citrated: K, KH, RTH, FFH assay cartridge. The TEG 6s Hemostasis System is intended for in vitro diagnostic use with adult patients where an evaluation of their blood hemostasis properties is desired. The TEG 6s Hemostasis System records the kinetic changes in a sample of 3.2% citrated whole blood as the sample clots and provides semi-quantitative results. The TEG 6s Hemostasis System can be used in the laboratory or at the point-of-care. The Citrated: K, KH, RTH, FFH assay cartridge is intended to be used in patients where heparin/heparinoids may be present and who are at an increased risk of coagulopathy. Hemostasis evaluations are indicated to assess clinical conditions in cardiovascular surgery, cardiovascular procedures (e.g., minimally invasive valve replacement or repairs) and liver transplantation to assess hemorrhage or thrombosis conditions before, during and following the procedure. The Citrated: K, KH, RTH, FFH assay cartridge contains four independent assays (CK, CKH, CRTH and CFFH) and the system output consists of a table of numerical values for parameters R, MA, and LY30. The CK assay monitors the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System. Clotting characteristics are described by the functional parameters R (clotting time) and MA (maximum clot strength). The CKH assay monitors the effects of heparin in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System. CKH is used in conjunction with CK, and heparin influence is determined by comparing Clotting Times (R) between the two tests. LY30 describes fibrinolysis 30 minutes after reaching maximum clot strength. The CRTH assay monitors the hemostasis process after stimulation of both the intrinsic and extrinsic pathways in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System, neutralizing the effect of heparin in the sample. Clotting characteristics are described by the functional parameter MA (maximum clot strength with contributions of both platelets and fibrin). The CFFH assay monitors hemostasis of 3.2% citrated whole blood specimens in the TEG 6s Hemostasis System after blocking platelet contributions to clot strength, neutralizing the effect of heparin in the sample. Clotting characteristics are described by the functional parameter MA (fibrinogen contribution to maximum clot strength). Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests.
Device Story
TEG 6s Hemostasis System uses disposable cartridges with four independent measurement cells to analyze 3.2% citrated whole blood. Device uses piezoelectric actuator to vibrate measurement cells; optical sensors detect meniscus motion. System performs Fast Fourier Transform (FFT) on motion data to calculate resonant frequency and modulus of elasticity (stiffness) as sample clots. Output includes numerical values for R (clotting time), MA (maximum clot strength), and LY30 (fibrinolysis). Used in labs or point-of-care by clinicians to evaluate hemostasis, monitor heparin effects, and assess fibrinogen/platelet contributions. Results assist in managing coagulopathy, hemorrhage, or thrombosis risks during surgery. System provides semi-quantitative data to support clinical decision-making alongside patient history.
Clinical Evidence
Method comparison study across 8 clinical sites (N=617-883 samples depending on parameter) compared subject device to predicate and Clauss Fibrinogen. Primary endpoints (bias at medical decision points, slope, correlation) met acceptance criteria. Pearson correlation coefficients >0.82 for Type 1/2 parameters. Type 3 (CFFH-MA vs Clauss Fibrinogen) showed Spearman correlation of 0.79. Precision studies (repeatability/reproducibility) per CLSI EP05-A3 showed CVs <15% for R and <10% for MA. Interference testing confirmed robustness against common clinical interferents.
Indicated for adult patients at increased risk of coagulopathy where heparin/heparinoids may be present. Used to assess hemorrhage or thrombosis in cardiovascular surgery, cardiovascular procedures (e.g., valve replacement/repair), and liver transplantation.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
TEG 6s with the Citrated Multichannel Cartridge (K150041)
Submission Summary (Full Text)
{1}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K232018
B Applicant
Haemonetics Corporation
C Proprietary and Established Names
Citrated: K, KH, RTH, FFH
D Regulatory Information
| Product Code(s) | | Regulation | |
| --- | --- | --- | --- |
| JPA | Class II | 21 CFR 864.5425 - Multipurpose System For In Vitro Coagulation Studies | HE - Hematology |
## II Submission/Device Overview:
A Purpose for Submission:
New assay
B Measurand:
Citrated Kaolin (CK), Citrated Kaolin with Heparinase (CKH), Citrated RapidTEG with Heparinase (CRTH), and Citrated Functional Fibrinogen with Heparinase (CFFH)
C Type of Test:
Semi-quantitative
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
K232018 - Page 2 of 19
# III Intended Use/Indications for Use:
## A Intended Use(s):
See Indications for Use below.
## B Indication(s) for Use:
The TEG 6s Hemostasis System consists of the TEG 6s Hemostasis Analyzer and the Citrated: K, KH, RTH, FFH assay cartridge. The TEG 6s Hemostasis System is intended for in vitro diagnostic use with adult patients where an evaluation of their blood hemostasis properties is desired. The TEG 6s Hemostasis System records the kinetic changes in a sample of 3.2% citrated whole blood as the sample clots and provides semi-quantitative results. The TEG 6s Hemostasis System can be used in the laboratory or at the point-of-care.
The Citrated: K, KH, RTH, FFH assay cartridge is intended to be used in patients where heparin/heparinoids may be present and who are at an increased risk of coagulopathy. Hemostasis evaluations are indicated to assess clinical conditions in cardiovascular surgery, cardiovascular procedures (e.g., minimally invasive valve replacement or repairs) and liver transplantation to assess hemorrhage or thrombosis conditions before, during and following the procedure.
The Citrated: K, KH, RTH, FFH assay cartridge contains four independent assays (CK, CKH, CRTH and CFFH) and the system output consists of a table of numerical values for parameters R, MA, and LY30.
The CK assay monitors the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System. Clotting characteristics are described by the functional parameters R (clotting time) and MA (maximum clot strength).
The CKH assay monitors the effects of heparin in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System. CKH is used in conjunction with CK, and heparin influence is determined by comparing Clotting Times (R) between the two tests. LY30 describes fibrinolysis 30 minutes after reaching maximum clot strength.
The CRTH assay monitors the hemostasis process after stimulation of both the intrinsic and extrinsic pathways in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System, neutralizing the effect of heparin in the sample. Clotting characteristics are described by the functional parameter MA (maximum clot strength with contributions of both platelets and fibrin).
The CFFH assay monitors hemostasis of 3.2% citrated whole blood specimens in the TEG 6s Hemostasis System after blocking platelet contributions to clot strength, neutralizing the effect of heparin in the sample. Clotting characteristics are described by the functional parameter MA (fibrinogen contribution to maximum clot strength).
Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests.
{3}
For professional use only.
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
## D Special Instrument Requirements:
TEG 6s Hemostasis Analyzer
## IV Device/System Characteristics:
### A Device Description:
The TEG 6s Hemostasis System (TEG 6s Hemostasis Analyzer and the Citrated: K, KH, RTH, FFH assay cartridge) are used to test the hemostasis properties of citrated blood samples using four different assays/reagents simultaneously, one in each of the four cartridge channels. The TEG 6s Hemostasis System is designed to monitor and analyze the coagulation state of blood samples using thromboelastography technology. Thromboelastography has been utilized to understand the hemostatic balance that should exist between the extremes of coagulopathy (impaired clot formation, hypocoagulable) and thrombosis (formation of a blood clot obstructing the flow of blood, hypercoagulable). The TEG 6s analyzer records the kinetic changes in a whole blood sample as the sample clots, retracts and/or lyses. The system output consists of numerical values for each of the four cartridge channels resulting from the hemostasis process over time. For the CK assay, Kaolin-activated test methods are used to reduce variability and to reduce the running time of a native whole blood sample. CKH is similar to CK but with the addition of heparinase, to neutralize the effects of heparin in the blood. If there is heparin present, the R time for CK will be significantly longer than the R time for CKH. The CRTH assay neutralizes the anticoagulant property of heparin and maximally accelerates the clotting process by simultaneously activating the intrinsic and extrinsic coagulation pathways, allowing for the maximum clot strength to be reached more quickly. The CFFH assay neutralizes the anticoagulant property of heparin and inhibits platelet aggregation, excluding its contribution to clot strength, and thereby measures fibrinogen contribution to clot strength.
### B Principle of Operation:
The TEG 6s technology is based on the use of Citrated K, KH, RTH, FFH disposable cartridge containing up to 4 independent measurement cells. Each cell consists of a short vertically oriented injection molded tube (ring). Detection of clotting in the TEG 6s Hemostasis System is performed optically. A piezoelectric actuator vibrates the measurement cell(s) through a motion profile composed of summed sinusoids at different frequencies. The movement of the measurement cells will induce motion in the sample meniscus, which will be detected by a photodiode. The resulting motion of the meniscus is monitored optically and analyzed by the instrument to calculate the resonant frequency and modulus of elasticity (stiffness) of the sample. By performing a Fast Fourier Transform (FFT) on meniscus motion data, the resonant frequencies can be determined. The analyzer monitors the harmonic motion of a hanging drop of blood in response to external vibration. As the sample transitions from a liquid state to a gel-like state during clotting, the modulus of elasticity (stiffness) and therefore resonant frequency
K232018 - Page 3 of 19
{4}
increase. The TEG 6s hemostasis analyzer measures these variations in resonant frequency during clotting and lysis.
Resonance is the tendency of a material or structure to oscillate with greater amplitude at some frequencies than others. The exact frequencies at which resonance occurs will depend on the stiffness and mass of the sample. Stiffness, in turn, is a function of a material's modulus of elasticity and the boundary conditions to which the material is exposed, such as the geometry and materials of a test cell. By holding these boundary conditions and sample mass constant from sample to sample, the TEG 6s Hemostasis System allows direct comparison of elasticity between samples. The output measurements are displayed in a table and on a graphical tracing that reflects the hemostasis profile of the clot formation.
In a typical test, blood that has been delivered to the measurement cell will not clot for several minutes. During this time the sample has no inherent stiffness except that provided by surface tension, and since this remains constant the measured resonant frequencies will not change. Once clotting begins, however, the elastic modulus and thus the resonant frequencies increase rapidly. During fibrinolysis, the process is reversed, with elastic modulus and resonant frequencies decreasing. In tests where clotting does not occur, the resonant frequency of the sample will not change. During coagulation, however, a clot will bind to the ring contained in the cartridge and the resonant frequency will rise with increasing firmness of the clot. The analyzer collects meniscus motion data, tracks changing resonant frequencies and analyzes the frequency data to provide semi-quantitative parameters describing the clot. The System monitors the interaction of platelets within the fibrin mesh of the clot during clot formation and lysis, all in a whole-blood setting. The TEG 6s Hemostasis System uses thromboelastography to provide continuous measurement of clot elasticity.
## CK Principle
Kaolin is a standardized reagent consisting of dry Kaolin and 0.85% Saline solution. Kaolin-activated test methods are used to reduce variability and to reduce the running time of a native whole blood sample. Use of these particles of hydrated aluminum silicate shortens coagulation time because Kaolin acts as a contact surface activator (intrinsic pathway), which activates Factor XII and platelets and stimulates the reserve clotting ability of a blood sample. Kaolin is combined with CaCl₂ to neutralize any sodium citrate in the blood.
## CKH Principle
Heparin is commonly used as an anticoagulant in surgical procedures. Even very low concentrations of heparin, fractions of IU/mL of blood, can noticeably increase the R time and can even completely anticoagulate the blood, making it difficult if not impossible to monitor developing coagulopathies that are masked by high levels of therapeutic heparin. The addition of heparinase in the assay cartridge neutralizes the effects of heparin in the sample. CaCl₂ is included to neutralize any sodium citrate in the blood.
## CRTH Principle
RapidTEG maximally accelerates the clotting process by simultaneously activating the intrinsic and extrinsic coagulation pathways using a high concentration of Kaolin and Tissue Factor (TF). This closely reflects the physiological clotting process and yields results significantly faster than a native, Kaolin, and/or TF activated test. The addition of heparinase in the assay cartridge neutralizes the effects of heparin in the sample. CaCl₂ is included to neutralize any sodium citrate in the blood.
K232018 - Page 4 of 19
{5}
K232018 - Page 5 of 19
# CFFH Principle
The Functional Fibrinogen reagent activates the extrinsic pathway using tissue factor and inhibits platelet aggregation using a platelet inhibitor that binds to GPIIb/IIIa receptors. By excluding the platelet aggregation contribution to clot strength (MA), the reagent measures fibrinogen contribution. The addition of heparinase in the assay cartridge neutralizes the effects of heparin in the sample. Functional Fibrinogen is combined with $\mathrm{CaCl}_2$ to neutralize any sodium citrate in the blood. Determining the independent contribution of platelets and fibrinogen further improves the use of the TEG 6s System in helping to detect and diagnose hemostasis defects.
# V Substantial Equivalence Information:
A Predicate Device Name(s):
TEG 6s with the Citrated Multichannel Cartridge
B Predicate 510(k) Number(s):
K150041
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K232018 | K150041 |
| --- | --- | --- |
| Device Trade Name | Citrated: K, KH, RTH, FFH | TEG 6s with the Citrated Multichannel Cartridge |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The TEG 6s Hemostasis System consists of the TEG 6s Hemostasis Analyzer and the Citrated: K, KH, RTH, FFH assay cartridge. The TEG 6s Hemostasis System is intended for in vitro diagnostic use with adult patients where an evaluation of their blood hemostasis properties is desired. The TEG 6s Hemostasis System records the kinetic changes in a sample of 3.2% citrated whole blood as the sample clots and provides semi-quantitative results. The TEG 6s Hemostasis System can be used in the laboratory or at the point-of-care. | The TEG 6s Hemostasis System is intended for in vitro diagnostic use to provide semiquantitative indications of the hemostasis state of a blood sample. The Citrated Multichannel Cartridge, to be used with the TEG 6s analyzer, contains four independent assays (CK, CRT, CKH and CFF), described below.
The CK assay monitors the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the TEG 6s System. Clotting |
{6}
K232018 - Page 6 of 19
| | The Citrated: K, KH, RTH, FFH assay cartridge is intended to be used in patients where heparin/heparinoids may be present and who are at an increased risk of coagulopathy. Hemostasis evaluations are indicated to assess clinical conditions in cardiovascular surgery, cardiovascular procedures (e.g., minimally invasive valve replacement or repairs) and liver transplantation to assess hemorrhage or thrombosis conditions before, during and following the procedure. | characteristics are described by the functional parameters Clotting Time (R), Speed of Clot Formation (K and Alpha angle) and Maximum Clot Strength (MA). The CRT assay monitors the hemostasis process via both the intrinsic and extrinsic pathways in 3.2% citrated whole blood specimens on the TEG 6s System. Clotting characteristics are described by the functional parameter Maximum Clot Strength (MA). The CRT MA parameter is equivalent to the CK MA parameter but the final MA value is reached more quickly using the CRT assay. |
| --- | --- | --- |
| | The Citrated: K, KH, RTH, FFH assay cartridge contains four independent assays (CK, CKH, CRTH and CFFH) and the system output consists of a table of numerical values for parameters R, MA, and LY30. | The CKH assay monitors the effects of heparin in 3.2% citrated whole blood specimens on the TEG 6s System. CKH is used in conjunction with CK, and heparin influence is determined by comparing Clotting Times (R) between the two tests. The CFF assay monitors hemostasis of 3.2% citrated whole blood specimens in the TEG 6s System after blocking platelet contributions to clot strength. Clotting characteristics are described by the functional parameters Maximum Clot Strength (MA) and the Estimated Functional Fibrinogen Level (FLEV). Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be |
| | The CK assay monitors the hemostasis process via the intrinsic pathway in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System. Clotting characteristics are described by the functional parameters R (clotting time) and MA (maximum clot strength). | the CKH assay monitors the effects of heparin in 3.2% citrated whole blood specimens on the TEG 6s System. CKH is used in conjunction with CK, and heparin influence is determined by comparing Clotting Times (R) between the two tests. The CFF assay monitors hemostasis of 3.2% citrated whole blood specimens in the TEG 6s System after blocking platelet contributions to clot strength. Clotting characteristics are described by the functional parameters Maximum Clot Strength (MA) and the Estimated Functional Fibrinogen Level (FLEV). Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be |
{7}
| | The CRTH assay monitors the hemostasis process after stimulation of both the intrinsic and extrinsic pathways in 3.2% citrated whole blood specimens on the TEG 6s Hemostasis System, neutralizing the effect of heparin in the sample. Clotting characteristics are described by the functional parameter MA (maximum clot strength with contributions of both platelets and fibrin).
The CFFH assay monitors hemostasis of 3.2% citrated whole blood specimens in the TEG 6s Hemostasis System after blocking platelet contributions to clot strength, neutralizing the effect of heparin in the sample. Clotting characteristics are described by the functional parameter MA (fibrinogen contribution to maximum clot strength).
Results from the TEG 6s analysis should not be the sole basis for a patient diagnosis, but should be evaluated together with the patient's medical history, the clinical picture and, if necessary, further hemostasis tests.
For professional use only. | evaluated together with the patient’s medical history, the clinical picture and, if necessary, further hemostasis tests. The indication for TEG 6s System use is with adult patients where an evaluation of their blood hemostasis properties is desired.
Hemostasis evaluations are commonly used to assess clinical conditions in cardiovascular surgery and cardiology procedures to assess hemorrhage or thrombosis conditions before, during and following the procedure. |
| --- | --- | --- |
| Measurement Principle | Changes in physical clot elasticity over time; monitoring the physical response of a clot to low levels of applied strain (resonance frequency) | Same |
| Sample Matrix | 3.2% citrated whole blood | Same |
| Measurement Output | Graphical tracings of resonant frequency per reagent type, table of parameters | Same |
K232018 - Page 7 of 19
{8}
| Assays | CK
CKH | Same |
| --- | --- | --- |
| Assay Reagents | CK – kaolin and CaCl2
CKH – kaolin and CaCl2 with heparinase | Same |
| Measuring Chambers | 4 | Same |
| Sample Volume (per chamber) | 20 μL | Same |
| Quality Control | Cartridge Reagent QC - Level 1
Cartridge Reagent QC - Level 2 | Same |
| General Device Characteristic Differences | | |
| Assay Reagents | CRTH – tissue factor, kaolin and CaCl2 with heparinase
CFFH – abciximab, tissue factor and CaCl2 with heparinase | CRT – tissue factor, kaolin and CaCl2
CFF – abciximab, tissue factor and CaCl2 |
| Assay Parameters Reported | CK: R, MA
CKH: R, LY30
CRTH: MA
CFFH: MA | CK: R, K, Angle, MA
CKH: R
CRT: MA
CFF: MA, FLEV |
VI Standards/Guidance Documents Referenced:
- CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition
- CLSI EP28-A3c: Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition
- CLSI EP07-A3: Interference Testing in Clinical Chemistry - 3rd Edition
- CLSI EP25-A Evaluation of Stability of In Vitro Diagnostic Reagents: Diagnostic Reagents; Approved Guideline
- CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline -Second Edition
- IEC 61010-1:2010: Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
- IEC 61010-2-010: 2019: Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-010: Particular requirements for laboratory equipment for the heating of materials
- IEC 61010-2-101: 2018: Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment
- IEC 60601-1-2:2014/AMD1: 2020: Amendment 1 - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
K232018 - Page 8 of 19
{9}
- IEC/ EN61326-1: 2020: Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements
- IEC/ EN61326-2-6: 2020: Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment
## VII Performance Characteristics (if/when applicable):
## A Analytical Performance:
### 1. Precision/Reproducibility:
#### Repeatability – 20-day precision:
A precision study was conducted based on the CLSI EP05-A3 guideline for an assessment of the Citrated: K, KH, RTH, FFH assay cartridge repeatability utilizing the TEG 6s Cartridge Reagent Quality Control (QC) material. Two levels of QC material were tested (Level 1 and 2) with a single lot of Citrated: K, KH, RTH, FFH assay cartridges at a single site with two TEG 6s hemostasis analyzers (one analyzer per QC Level). One trained operator performed two runs per day with two replicates per run for 20 days. To demonstrate precision performance for both levels of QC, the mean, standard deviation (SD) and coefficient of variation (%CV) for each component of variability were calculated including repeatability, between-run, between-day, and within laboratory. The determination of acceptability was based on the SD or %CV.
| 20-day Precision – QC Level 1 | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Repeatability | | Between Run | | Between Day | | Within Laboratory | |
| Assay-Parameter | N | Mean | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| CK-R (min) | 80 | 6.10 | 0.54 | 8.9 | 0.00 | 0.0 | 0.27 | 4.4 | 0.61 | 10.0 |
| CKH-R (min) | 80 | 4.46 | 0.30 | 6.7 | 0.09 | 2.0 | 0.10 | 2.3 | 0.33 | 7.3 |
| CK-MA (mm) | 80 | 67.04 | 0.93 | 1.4 | 0.00 | 0.0 | 0.48 | 0.7 | 1.04 | 1.6 |
| CRTH-MA (mm) | 80 | 58.37 | 1.18 | 2.0 | 0.00 | 0.0 | 0.95 | 1.6 | 1.52 | 2.6 |
| CFFH-MA (mm) | 80 | 63.31 | 1.35 | 2.1 | 0.00 | 0.0 | 1.05 | 1.6 | 1.71 | 2.7 |
| CKH LY30 (%) | 80 | Data not collected | | | | | | | | |
| 20-day Precision – QC Level 2 | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Repeatability | | Between Run | | Between Day | | Within Laboratory | |
| Assay-Parameter | N | Mean | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| CK-R (min) | 80 | 1.24 | 0.07 | 5.8 | 0.05 | 3.6 | 0.00 | 0.0 | 0.09 | 6.9 |
| CKH-R (min) | 80 | 1.25 | 0.08 | 6.1 | 0.02 | 1.5 | 0.00 | 0.0 | 0.08 | 6.3 |
| CK-MA (mm) | 79* | 26.71 | 1.21 | 4.5 | 0.00 | 0.0 | 0.84 | 3.1 | 1.47 | 5.5 |
| CRTH-MA (mm) | 80 | 27.69 | 1.29 | 4.7 | 0.00 | 0.0 | 0.89 | 3.2 | 1.57 | 5.7 |
| CFFH-MA (mm) | 80 | 27.83 | 1.26 | 4.5 | 0.00 | 0.0 | 0.91 | 3.3 | 1.55 | 5.6 |
| CKH LY30 (%) | 80 | 92.26 | 0.40 | 0.4 | 0.40 | 0.4 | 0.00 | 0.0 | 0.56 | 0.6 |
*1 outlier removed
K232018 - Page 9 of 19
{10}
# Lot-to-Lot Precision Study
A lot-to-lot precision study was performed at one internal site using three lots of Citrated: K, KH, RTH, FFH assay cartridges and two lots of Cartridge Reagent QC Level 1 and Level 2. Testing was performed by two trained operators using six different TEG 6s hemostasis analyzers over 10 days. To demonstrate precision performance for both levels of QC, the mean, SD and $\% \mathrm{CV}$ for each component of variability were calculated including repeatability, between-day, between-operator, between-QC Lot, between-Cartridge Lot, and within-laboratory. The determination of acceptability was based on the SD or $\% \mathrm{CV}$ .
| Lot-to-Lot Precision - QC Level 1 | | | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Assay-Parameter | N | Mean | Repeatability | | Between Day | | Between Operator | | Between QC Lot | | Between Cartridge Lot | | Within Laboratory | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| CK-R (min) | 120 | 6.22 | 0.50 | 8.1 | 0.11 | 1.7 | 0.00 | 0.0 | 0.12 | 2.0 | 0.02 | 0.4 | 0.53 | 8.5 |
| CKH-R (min) | 120 | 4.76 | 0.41 | 8.5 | 0.12 | 2.6 | 0.00 | 0.0 | 0.25 | 5.2 | 0.12 | 2.5 | 0.50 | 10.6 |
| CK-MA (mm) | 120 | 66.62 | 1.57 | 2.4 | 0.14 | 0.2 | 0.00 | 0.0 | 1.37 | 2.1 | 0.00 | 0.0 | 2.09 | 3.1 |
| CRTH-MA (mm) | 120 | 59.68 | 1.81 | 3.0 | 0.35 | 0.6 | 0.11 | 0.2 | 0.00 | 0.0 | 0.00 | 0.0 | 1.85 | 3.1 |
| CFFH-MA (mm) | 120 | 64.12 | 1.05 | 1.6 | 0.00 | 0.0 | 0.20 | 0.3 | 0.00 | 0.0 | 0.61 | 1.0 | 1.23 | 1.9 |
| CKH LY30 (%) | Data not collected | | | | | | | | | | | | | |
| Lot-to-Lot Precision - QC Level 2 | | | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Assay-Parameter | N | Mean | Repeatability | | Between Day | | Between Operator | | Between QC Lot | | Between Cartridge Lot | | Within Laboratory | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| CK-R (min) | 120 | 1.16 | 0.06 | 5.1 | 0.02 | 1.4 | 0.03 | 2.7 | 0.06 | 5.0 | 0.01 | 0.7 | 0.09 | 7.7 |
| CKH-R (min) | 120 | 1.15 | 0.05 | 4.8 | 0.02 | 1.3 | 0.03 | 2.8 | 0.07 | 5.7 | 0.00 | 0.0 | 0.09 | 8.1 |
| CK-MA (mm) | 118* | 28.02 | 1.51 | 5.4 | 0.26 | 0.9 | 0.47 | 1.7 | 0.72 | 2.6 | 0.23 | 0.8 | 1.78 | 6.3 |
| CRTH-MA (mm) | 120 | 29.40 | 1.52 | 5.2 | 0.74 | 2.5 | 0.81 | 2.8 | 0.83 | 2.8 | 0.30 | 1.0 | 2.07 | 7.0 |
| CFFH-MA (mm) | 120 | 28.84 | 1.08 | 3.8 | 0.32 | 1.1 | 0.68 | 2.4 | 0.95 | 3.3 | 0.13 | 0.5 | 1.63 | 5.7 |
| CKH LY30 (%) | 120 | 92.68 | 0.36 | 0.4 | 0.00 | 0.0 | 0.00 | 0.0 | 0.10 | 0.1 | 0.04 | 0.1 | 0.37 | 0.4 |
* Two outliers removed
# Reproducibility
A reproducibility study was conducted at three sites (two external and one internal) utilizing one lot of TEG 6s Cartridge Reagent QC material. Two levels of QC material were tested (Level 1 and 2) with a single lot of Citrated: K, KH, RTH, FFH assay cartridges by one operator each site per day over a period of five days with five replicates per QC Level per
K232018 - Page 10 of 19
{11}
day. To demonstrate precision performance for both levels of QC, the mean, SD and $\% \mathrm{CV}$ for each component of variability were calculated including repeatability, between-day, between-site (includes operators and analyzer variation) and reproducibility. The determination of acceptability was based on the SD or $\% \mathrm{CV}$ .
| Reproducibility - QC Level 1 | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Repeatability | | Between Day | | Between Site* | | Reproducibility | |
| Assay-Parameter | N | Mean | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| CK-R (min) | 75 | 6.18 | 0.40 | 6.4 | 0.04 | 0.6 | 0.39 | 6.3 | 0.56 | 9.0 |
| CKH-R (min) | 75 | 4.62 | 0.33 | 7.2 | 0.09 | 1.9 | 0.26 | 5.7 | 0.43 | 9.4 |
| CK-MA (mm) | 75 | 66.06 | 1.33 | 2.0 | 0.00 | 0.0 | 0.56 | 0.8 | 1.44 | 2.2 |
| CRTH-MA (mm) | 75 | 60.34 | 1.52 | 2.5 | 0.37 | 0.6 | 1.76 | 2.9 | 2.36 | 3.9 |
| CFFH-MA (mm) | 75 | 63.97 | 1.03 | 1.6 | 0.00 | 0.0 | 2.11 | 3.3 | 2.34 | 3.7 |
| CKH LY30 (%) | 75 | Data not collected | | | | | | | | |
*Between site includes operator and analyzer variation.
| Reproducibility - QC Level 2 | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Repeatability | | Between Day | | Between Site* | | Reproducibility | |
| Assay-Parameter | N | Mean | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| CK-R (min) | 75 | 1.13 | 0.22 | 0.2 | 0.12 | 0.1 | 0.00 | 0.0 | 0.25 | 0.3 |
| CKH-R (min) | 75 | 1.13 | 0.05 | 4.7 | 0.00 | 0.0 | 0.01 | 1.1 | 0.05 | 4.8 |
| CK-MA (mm) | 75 | 28.18 | 0.77 | 2.8 | 0.24 | 0.8 | 0.51 | 1.8 | 0.96 | 3.4 |
| CRTH-MA (mm) | 75 | 30.20 | 1.12 | 3.7 | 0.44 | 1.5 | 0.38 | 1.3 | 1.26 | 4.2 |
| CFFH-MA (mm) | 75 | 29.53 | 0.81 | 2.7 | 0.31 | 0.1 | 0.27 | 0.9 | 0.91 | 3.1 |
| CKH LY30 (%) | 75 | 92.30 | 0.22 | 0.2 | 0.12 | 0.1 | 0.00 | 0.0 | 0.25 | 0.3 |
*Between site includes operator and analyzer variation.
# Whole Blood Repeatability
A whole blood repeatability study was conducted using whole blood samples under the following conditions: normal, contrived hypocoagulable (low CFFH MA, low CRTH MA), hypercoagulable (high CRTH MA and High CKH LY30) and hyperfibrinolytic (High CKH LY30). All sample conditions were analyzed by two trained operators using three lots of Citrated: K, KH, RTH, FFH assay cartridges with two replicates per run. Normal whole blood samples were tested in one day with two operators testing three cartridge lots on two different TEG 6s hemostasis analyzers with two replicates per run $(n = 24)$ . Due to sample stability limitations for contrived samples (hypocoagulable, hypercoagulable, and hyperfibrinolytic), testing was completed in one day by two operators using three cartridge lots on four analyzers for one replicate $(n = 24)$ . To demonstrate precision performance for all sample types, the mean, SD and $\% \mathrm{CV}$ for each component of variability were calculated including repeatability, between-operator, between-lot, between device (normal sample type only), and within-laboratory. The determination of acceptability was based on the SD or $\% \mathrm{CV}$ .
K232018 - Page 11 of 19
{12}
| Whole Blood Repeatability – Normal Sample | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Repeatability | | Between Operator | | Between Lot | | Between Device/Day | | Within Laboratory | |
| Assay-Parameter | N | Mean | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| CK-R (min) | 24 | 7.50 | 0.4 | 5.9 | 0.00 | 0.0 | 0.65 | 8.7 | 0.28 | 3.8 | 0.83 | 11.2 |
| CKH-R (min) | 24 | 7.60 | 0.75 | 10.0 | 0.00 | 0.0 | 0.00 | 0.0 | 0.14 | 1.8 | 0.77 | 10.1 |
| CK-MA (mm) | 24 | 54.80 | 0.96 | 1.7 | 0.32 | 0.6 | 0.54 | 1.0 | 0.25 | 0.5 | 1.17 | 2.1 |
| CRTH-MA (mm) | 24 | 17.00 | 0.27 | 1.6 | 0.46 | 2.7 | 0.09 | 0.6 | 0.60 | 3.5 | 0.81 | 4.8 |
| CFFH-MA (mm) | 24 | 57.00 | 1.43 | 2.5 | 0.65 | 1.1 | 0.20 | 0.4 | 0.00 | 0.0 | 1.58 | 2.8 |
| CKH LY30 (%) | 24 | 0.30 | 0.13 | 51.0 | 0.00 | 0.0 | 0.10 | 36.5 | 0.08 | 32.1 | 0.18 | 70.4 |
| Whole Blood Repeatability – Hypocoagulable (Sample 2) Citrated whole blood was spiked with Abciximab and Dabigatran | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Repeatability* | | Between Operator | | Between Lot | | Within Laboratory | | |
| Assay-Parameter | N | Mean | SD | %CV | SD | %CV | SD | %CV | SD | %CV | |
| CK-R (min) | 24 | 13.29 | 1.47 | 11.01 | 0.00 | 0.0 | 0.00 | 0.0 | 1.47 | 11.1 | |
| CKH-R (min) | 24 | 13.09 | 1.52 | 11.6 | 0.61 | 4.6 | 0.39 | 3.0 | 1.68 | 12.8 | |
| CK-MA (mm) | 24 | 52.78 | 1.04 | 2.0 | 0.00 | 0.0 | 0.89 | 1.7 | 1.37 | 2.6 | |
| CRTH-MA (mm) | 24 | 58.55 | 0.51 | 0.9 | 0.00 | 0.0 | 0.35 | 0.6 | 0.62 | 1.1 | |
| CFFH-MA (mm) | 24 | 18.02 | 0.17 | 1.0 | 0.03 | 0.2 | 0.12 | 0.7 | 0.21 | 1.2 | |
| CKH LY30 (%) | 24 | 1.18 | 0.19 | 16.1 | 0.00 | 0.0 | 0.00 | 0.0 | 0.19 | 16.1 | |
* Repeatability includes between analyzer variance.
| Whole Blood Repeatability – Hypocoagulable (Sample 3) Citrated Whole blood sample was diluted to 50% fibrinogen with fibrinogen deficient plasma. | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Repeatability* | | Between Operator | | Between Lot | | Within Laboratory | |
| Assay-Parameter | N | Mean | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| CK-R (min) | 24 | 4.91 | 0.31 | 6.3 | 0.03 | 0.7 | 0.12 | 2.4 | 0.33 | 6.8 |
| CKH-R (min) | 24 | 4.85 | 0.20 | 4.2 | 0.00 | 0.0 | 0.11 | 2.4 | 0.23 | 4.8 |
| CK-MA (mm) | 24 | 44.71 | 0.69 | 1.6 | 0.49 | 1.1 | 0.39 | 0.9 | 0.94 | 2.1 |
| CRTH-MA (mm) | 24 | 43.69 | 0.83 | 1.9 | 0.00 | 0.0 | 0.00 | 0.0 | 1.19 | 2.7 |
| CFFH-MA (mm) | 24 | 10.18 | 0.83 | 8.1 | 0.20 | 2.3 | 0.30 | 3.3 | 0.92 | 9.2 |
| CKH LY30 (%) | 24 | Data not collected | | | | | | | | |
| Whole Blood Repeatability – Hypercoagulable (Sample 4) Citrated whole blood sample was collected and prepared using platelet rich plasma (PRP) and concentrated fibrinogen. | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Repeatability* | | Between Operator | | Between Lot | | Within Laboratory | |
| Assay-Parameter | N | Mean | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| CK-R (min) | 23* | 5.47 | 0.61 | 10.7 | 0.37 | 6.5 | 0.05 | 0.8 | 0.71 | 12.5 |
| CKH-R (min) | 23* | 5.91 | 0.72 | 12.2 | 0.00 | 0.0 | 0.46 | 7.8 | 0.85 | 14.5 |
| CK-MA (mm) | 23* | 70.69 | 0.35 | 0.5 | 0.27 | 0.4 | 0.27 | 0.4 | 0.51 | 0.7 |
| CRTH-MA (mm) | 24 | 71.59 | 0.39 | 0.5 | 0.00 | 0.0 | 0.00 | 0.0 | 0.39 | 0.5 |
| CFFH-MA (mm) | 24 | 51.38 | 0.92 | 1.8 | 0.18 | 1.5 | 2.81 | 5.5 | 3.06 | 6.0 |
| CKH LY30 (%) | 24 | Data not collected | | | | | | | | |
*One outlier removed
K232018 - Page 12 of 19
{13}
K232018 - Page 13 of 19
| Whole Blood Repeatability – Hypercoagulable (Sample 5)
Citrated whole blood was spiked with Fibrinogen using cryo-precipitate | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Repeatability* | | Between Operator | | Between Lot | | Within Laboratory | | |
| Assay-Parameter | N | Mean | SD | %CV | SD | %CV | SD | %CV | SD | %CV | |
| CK-R (min) | 24 | 6.60 | 0.51 | 7.7 | 0.10 | 1.5 | 0.24 | 3.6 | 0.57 | 8.6 | |
| CKH-R (min) | 24 | 6.28 | 0.66 | 10.6 | 0.00 | 0.0 | 0.24 | 3.8 | 0.71 | 11.2 | |
| CK-MA (mm) | 24 | 63.85 | 0.75 | 1.2 | 0.00 | 0.0 | 0.19 | 0.3 | 0.77 | 1.2 | |
| CRTH-MA (mm) | 24 | 66.59 | 0.26 | 0.4 | 0.00 | 0.0 | 0.13 | 0.2 | 0.29 | 0.4 | |
| CFFH-MA (mm) | 24 | 32.92 | 0.66 | 2.0 | 0.00 | 0.0 | 0.53 | 1.6 | 0.85 | 2.6 | |
| CKH LY30 (%) | 24 | 0.45 | 0.44 | 97.1 | 0.00 | 0.0 | 0.14 | 31.0 | 0.46 | 101.9 | |
* Repeatability includes between analyzer variance.
| Whole Blood Repeatability – Hyperfibrinolytic (Sample 6)
Whole blood was spiked with tPA (tissue factor) | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Repeatability* | | Between Operator | | Between Lot | | Within Laboratory | | |
| Assay-Parameter | N | Mean | SD | %CV | SD | %CV | SD | %CV | SD | %CV | |
| CK-R (min) | 24 | 6.38 | 0.45 | 7.1 | 0.00 | 0.0 | 0.51 | 8.0 | 0.68 | 10.7 | |
| CKH-R (min) | 24 | 6.53 | 0.54 | 8.3 | 0.00 | 0.0 | 0.00 | 0.0 | 0.54 | 8.3 | |
| CK-MA (mm) | 24 | 58.69 | 0.96 | 1.6 | 0.00 | 0.0 | 0.10 | 0.2 | 0.96 | 1.6 | |
| CRTH-MA (mm) | 24 | 61.56 | 0.68 | 1.1 | 0.38 | 0.6 | 0.00 | 0.0 | 0.78 | 1.3 | |
| CFFH-MA (mm) | 24 | 21.70 | 1.01 | 4.6 | 0.28 | 1.3 | 1.25 | 5.7 | 1.63 | 7.5 | |
| CKH LY30 (%) | 24 | 15.87 | 1.59 | 9.7 | 0.00 | 0.0 | 0.00 | 0.0 | 1.59 | 9.7 | |
* Repeatability includes between analyzer variance.
| Whole Blood Repeatability – Hypercoagulable (Sample 7)
Citrated whole blood sample was spiked with diluted tissue factor | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Repeatability* | | Between Operator | | Between Lot | | Within Laboratory | | |
| Assay-Parameter | N | Mean | SD | %CV | SD | %CV | SD | %CV | SD | %CV | |
| CK-R (min) | 24 | 4.41 | 0.17 | 3.8 | 0.00 | 0.0 | 0.13 | 2.9 | 0.21 | 4.8 | |
| CKH-R (min) | 24 | 4.31 | 0.19 | 4.3 | 0.00 | 0.0 | 0.08 | 1.9 | 0.20 | 4.7 | |
| CK-MA (mm) | 24 | 64.38 | 0.44 | 0.7 | 0.11 | 0.2 | 0.28 | 0.4 | 0.54 | 0.8 | |
| CRTH-MA (mm) | 24 | 65.65 | 0.40 | 0.6 | 0.16 | 0.2 | 0.00 | 0.0 | 0.43 | 0.6 | |
| CFFH-MA (mm) | 24 | 25.18 | 0.43 | 1.7 | 0.17 | 0.7 | 0.52 | 2.1 | 0.69 | 2.7 | |
| CKH LY30 (%) | 24 | 0.67 | 0.15 | 23.1 | 0.03 | 4.2 | 0.13 | 20.0 | 0.21 | 30.8 | |
* Repeatability includes between analyzer variance.
2. Linearity:
Not applicable.
3. Analytical Specificity/Interference:
Interference testing was conducted using Citrated: K, KH, RTH, FFH assay cartridge with 3.2% sodium citrated whole blood in accordance with CLSI EP07-A3. The study included 13 potential interferents and the impact of four potential interferent factors: lack of discard tube, excess hemolysis, hemodilution, and short draw. Each potential interferent or interferent factor was evaluated in three replicates on normal, hyperfibrinolytic, and
{14}
hypocoagulable conditions. A control sample was tested with each condition in which the blood was not treated with the potential interferent or interferent factor.
The following interfering substances were evaluated and showed no significant interference up to the specified concentration in whole blood samples collected in 3.2% sodium citrate anticoagulant collection tubes.
| Interferent | Highest Concentration with No Interference |
| --- | --- |
| Endogenous Substances | |
| lipemia | 300 mg/dL |
| Exogenous Substances | |
| alcohol | 320 mg/dL |
| ticagrelor | 1.8 μg/mL |
| tranexamic acid (TXA) | 200 μg/mL |
| epsilon aminocaproic acid (EACA) | 600 μg/mL |
| Mycophenolic Acid | 42 μg/mL |
| Tacrolimus | 144 ng/mL |
| Prednisone | 99 ng/mL |
| Rifaximin | 40.5 ng/mL |
| Lactulose | 12 μg/mL |
| Aspirin | Normal – 6.5 mg/dL
Hyperfibrinolytic – 6.5 mg/dL (CK-R, CK-MA, CKH-R, CRTH-MA, CFFH-MA)
0 mg/dL (CKH-LY30) |
| Dabigatran | Normal – 0 ng/mL (CK-R, CKH-R), 90 ng/mL (CKH-LY30), 180 ng/mL (CK-MA, CRTH-MA, CFFH-MA)
Hyperfibrinolytic – 0 ng/mL (CK-R, CKH-R, CKH-LY30), 90 ng/mL (CK-MA), 180 ng/mL (CRTH-MA, CFFH-MA) |
| Rivaroxaban | Normal – 50 ng/mL (CK-R, CKH-R) 400 ng/mL (CK-MA, CKH-LY30, CRTH-MA, CFFH-MA)
Hyperfibrinolytic – 50 ng/mL (CK-R, CKH-R), 200 ng/mL (CK-MA), 400 ng/mL (CKH-LY30, CRTH-MA, CFFH-MA) |
| Interferent Factors | Specimen Condition | Highest Concentration with No Interference |
| --- | --- | --- |
| No Discard Tube | Normal | No interference |
| | Hypocoagulable | No interference |
| | Hyperfibrinolytic | CKH-LY30 |
| Hemolysis | Normal | CK-R, CK-MA, CKH-R, CRTH-MA, CFFH-MA – 400 mg/dl
CKH-LY30 – 200 mg/dL |
| | Hypocoagulable | All parameters – 400 mg/dl |
| | Hyperfibrinolytic | CK-R, CK-MA, CFFH-MA – 400 mg/dl
CKH-R – 200 mg/dL
CRTH-MA – 100 mg/dL
CKH-LY30 – 0 mg/dL |
| Hemodilution | Normal | CK-R, CK-MA, CKH-R, CKH-LY30 – 50%
CRTH-MA – 20%
CFFH-MA – 40% |
K232018 - Page 14 of 19
{15}
| Interferent Factors | Specimen Condition | Highest Concentration with No Interference |
| --- | --- | --- |
| | Hypocoagulable | CK-R, CKH-R, CKH-LY30 – 50%
CK-MA, CRTH-MA – 0%
CFFH-MA – 40% |
| | Hyperfibrinolytic | CK-R, CKH-R – 50%
CK-MA, CKH-LY30, CRTH-MA – 0%
CFFH-MA – 20% |
| Short Draw | Normal | CK-R, CK-MA, CKH-R, CKH-LY30, CFFH-MA – 60%
CRTH-MA – 70% |
| | Hypocoagulable | CK-R, CK-MA, CKH-LY30, CRTH-MA, CFFH-MA – 60%
CKH-MA – 70% |
| | Hyperfibrinolytic | CK-R, CK-MA, KH-R, CRTH-MA, CFFH-MA – 60%
CKH-LY30 – 70% |
## 4. Assay Reportable Range:
Validation that the Citrated: K, KH, RTH, FFH assay cartridge parameters demonstrate functional performance across the claimed reportable ranges for each parameter was supported by data from the clinical validation study.
| Parameter | Reportable Range |
| --- | --- |
| CK-R (min) | 0.4–17 |
| CK-MA (mm) | 40–75 |
| CKH-R (min) | 0.3–17 |
| CRTH-MA (mm) | 20–78 |
| CFFH-MA (mm) | 6–61 |
| CKH-LY30 (%) | 0–30 |
## 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
### Shelf-Life Stability Study
To establish the shelf-life stability of the Citrated: K, KH, RTH, FFH assay cartridge, a real-time stability study is designed following CLSI EP25-A. The shelf-life stability study includes three cartridge lots stored at $2 - 8^{\circ}\mathrm{C}$ and cartridges are taken out for evaluation at 0, 12, 18, 24 and 26.5 months. Cartridges are tested using $3.2\%$ sodium citrate samples using 16 cartridges at each sample condition. Sample conditions include unspiked donor whole blood and donor whole blood spiked with $5\mathrm{IU / mL}$ Heparin. The cartridges were tested at time zero (T0) and lot one was tested after the transport stability study (day 93). Cartridge stability will be assessed through regression analysis of the individual value difference from T0 compared to each sample type and assay parameter in accordance with CLSI EP25-A. Shelf-life cartridge stability studies are on-going.
### Transport Stability Study
The transport simulation study was performed with one lot of Citrated: K, KH, RTH, FFH assay cartridge to simulate transport under stressed conditions. To create the stressed condition, cartridges were exposed to an extended seven-day International Safe Transit Association (ISTA) 7-day shipping conditions ($5^{\circ}\mathrm{C}$ for 24 hours (Preconditioning), $22^{\circ}\mathrm{C}$ for 4 hours, $35^{\circ}\mathrm{C}$ for 6 hours, $30^{\circ}\mathrm{C}$ for 152 hours, and $35^{\circ}\mathrm{C}$ for 6 hours). The measured drift for
K232018 - Page 15 of 19
{16}
each testing time point and condition was compared to the mean level obtained at T0 (time 0). All results were within the acceptable allowable drift. The cartridges passed transport stability testing and demonstrate the cartridges are robust. The one lot of cartridges used in the Shelf-Life Stability Study and Cartridge In-Use Stability Study were the lot tested in the Transport Stability Study to demonstrate the real-world use scenario.
## Cartridge In-Use Stability Study
To support a 2 hour in-use stability of the Citrated: K, KH, RTH, FFH assay cartridge, three lots of cartridges were tested at T0 and 2.5 hours after removal from packaging. Sodium citrated whole blood collected from normal donors were tested on 10 cartridges and 1 IU/mL Heparin Spiked Blood and 7 IU/mL Heparin Spiked Blood were tested on 18 cartridges each. The results from each assay were compared to the control measurements at T0. The data support the cartridge can be stable after removal from the sealed cartridge pouch for up to 2 hours at ambient room temperature (20–22°C) and humidity (<50%).
## Whole Blood Sample Stability Study
A whole blood sample stability study was conducted to support the stability of whole blood samples collected in 3.2% sodium citrate tubes after 10 minutes of incubation and up to 2 hours of storage at ambient room temperature (20–22°C). Samples were tested at T0 (as soon as practical post draw), 10 (incubation time point prior to assay start), 66, 99, and 132 minutes. A total of six donor samples per sample type were tested in six replicates per time point. The sample types used included normal, hypocoagurable, hypercoagulable, and hyperfibrinolytic whole blood. Replicates were executed simultaneously using six TEG 6s hemostasis Analyzers. All Citrated: K, KH, RTH, FFH reported parameters (CK-R, CK-MA, CKH-R, CKH-LY30, CRTH-MA, and CFFH-MA) were evaluated. Results of the study support the use of whole blood collected in 3.2% sodium citrate tubes up to two hours after collection and following a 10-minute incubation after collection.
6. Detection Limit:
Not applicable.
7. Assay Cut-Off:
Not applicable.
## B Comparison Studies:
1. Method Comparison with Predicate Device:
The method comparison study was performed at eight clinical sites in the United States, comparing to the TEG 6s hemostasis analyzers with the Citrated Multichannel Cartridge (K150041) and STA - Fibrinogen 5 (K840211). The study enrolled 338 patients (145 females and 192 males), 18 years of age or older who are at an increased risk of coagulopathy. Enrolled subjects were patients undergoing liver transplantation, cardiovascular surgery, or cardiology procedures. Blood samples were drawn before, during, and after the procedures and analyzed by 24 trained operators.
To support the use of the CKH-MA and CKH-LY30 parameters in the Citrated Multichannel Cartridge (K150041), additional analytical validation studies were performed. The results of
K232018 - Page 16 of 19
{17}
the method comparison study comparing the candidate device, Citrated: K, KH, RTH, FFH assay cartridge to the predicate, Citrated Multichannel Cartridge (K150041) and STA - Fibrinogen 5 (K840211) demonstrate substantial equivalence for all reportable parameters.
| Parameter | N | Intercept [95% CI] | Slope [95% CI] | Pearson Coefficient [95% CI] |
| --- | --- | --- | --- | --- |
| CK-R | 617 | -0.45 [-0.769; -0.125] | 1.08 [1.041; 1.126] | 0.90 [0.878; 0.910] |
| CK - MA | 539 | -3.43 [-5.127; -1.731] | 1.05 [1.022; 1.082] | 0.95 [0.939; 0.956] |
| CKH - R | 829 | -0.17 [-0.538; 0.193] | 1.01 [0.960; 1.055] | 0.82 [0.800; 0.844] |
| CKH - LY30 | 828 | -0.01 [-0.095; 0.071] | 1.00 [0.988; 1.010] | 0.99 [0.985; 0.989] |
| CRTH - MA | 870 | -5.54 [-6.4626; -4.459] | 1.11 [1.094; 1.133] | 0.97 [0.962; 0.910] |
| CFFH - MA * | 883 | -78.71 [-98.9; -58.53] | 1.32 [1.25; 1.40] | -- |
*Clauss Fibrinogen comparator
2. Matrix Comparison:
Not applicable.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
CK Sensitivity and Specificity:
To define the sensitivity and specificity for the CK channel, blood samples were collected from four normal donors to establish hypocoagulable, hypercoagulable and hyperfibrinolytic conditions. Four Levels – including baseline of citrated whole blood samples spiked with increasing concentration of heparin (0, 0.1, 0.15, and 0.20 IU/mL) were tested with five TEG 6s Citrated: K, KH, RTH, FFH assay cartridges for a total of 20 cartridges per run.
All hypo-coagulable, hyper-coagulable and hyper-fibrinolytic samples without heparin showed 100% specificity and all samples spiked at 0.2 IU/ml showed 100% sensitivity for the detection of heparin.
K232018 - Page 17 of 19
{18}
K232018 - Page 18 of 19
# Heparin Neutralization Requirement Verification
A study was performed to validate the use of heparinase to neutralize the effect of heparin in blood samples. Tests were performed using Unfractionated Heparin (UFH) and Low Molecular Weight Heparin (LMWH) spiked samples and un-spiked samples. A total of 200 cartridges were tested across whole blood samples collected in 3.2% sodium citrate tubes from four donors in 10 replicates. The mean result per test specimen condition fell within the respective normal reference range for CRTH-MA, CFFH-MA, CKH-R, and CKH-LY30 for all donors. The study verified the defined heparin neutralization performance requirements:
- The CKH, CRTH, and CFFH assays shall be able to neutralize up to 5.0 IU/mL (+/- 0.1 IU/mL) of heparin.
- The CKH, CRTH, and CFFH assays shall be able to neutralize up to 0.013 mg/mL (+/- 0.001 mg/mL) of LMWH.
The tolerance interval (95% confidence for 90% of the population) calculated for CFFH-MA, and CRTH-MA for all specimen types fell within the respective normal reference range. The study acceptance criteria were met and has verified the following heparin neutralization performance requirements:
- Blood treated with 5.0 IU/ml (+/- 0.1 IU/mL) of heparin, measured CFFH-MA, shall remain within the normal reference range, 90% of the time.
- Blood treated with 5.0 IU/ml (+/- 0.1 IU/mL), measured CRTH-MA, shall remain within the normal reference range, 90% of the time.
For all 0.008 mg/mL LMWH contrived samples, the CK-R value was greater than the CKH-R value. The study acceptance criteria were met and has verified the following heparin neutralization performance requirement for the LMWH sample condition:
- CK-R shall be greater than CKH-R when at least 0.2 IU/mL of heparin or at least 0.008 mg/mL LMWH is present in the blood sample.
# D Clinical Cut-Off:
Not applicable.
# E Expected Values/Reference Range:
The reference ranges for the Citrated: K, KH, RTH, FFH assay cartridge parameters were established according to CLSI EP28-A3c. The study included 164 apparently healthy individuals (84 males and 80 female), greater than 18 years old with no known coagulopathies and not taking any drugs that would potentially affect patient hemostasis. The non-parametric method for analysis was used to determine the reference range for each assay parameter. The following table contains the reference range data for each parameter.
| Parameter | N | Range |
| --- | --- | --- |
| CK-R (min) | 157 | 4.6–9.1 |
| CK-MA (mm) | 151 | 52–69 |
| CKH-R (min) | 155 | 4.3–8.3 |
| CKH-LY30 (%) | 148 | 0–3.2 |
{19}
| Parameter | N | Range |
| --- | --- | --- |
| CRTH-MA (mm) | 162 | 53–69 |
| CFFH-MA (mm) | 162 | 15–34 |
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K232018 - Page 19 of 19
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.