STA R MAX

K151867 · Diagnostica Stago S.A.S · JPA · Aug 7, 2015 · Hematology

Device Facts

Record IDK151867
Device NameSTA R MAX
ApplicantDiagnostica Stago S.A.S
Product CodeJPA · Hematology
Decision DateAug 7, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 864.5425
Device ClassClass 2

Indications for Use

The STA R Max® is a fully automatic clinical instrument designed to perform tests on human plasmas, the results of which aid in the diagnosis of coagulation abnormalities or in monitoring anticoagulant therapy.

Device Story

STA R Max is an automated hemostasis analyzer for clinical laboratory use. It processes patient samples to perform coagulation testing using chronometric and photometric detection methods. Modifications from the predicate (STA-R Evolution Expert Series) include a new graphical user interface (GUI), updated external design (color, shape, ergonomics), a new computer table, and enhanced communication protocols for Laboratory Information System (LIS) or middleware integration. The device utilizes an integrated PC with updated microprocessor, memory, and hard disk specifications. It maintains the same fundamental scientific technology as the predicate, including coagulation detection methods, barcode identification for samples/reagents, and disposables. The system is operated by laboratory personnel to provide diagnostic data for clinical decision-making.

Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing, risk assessment, and verification/validation activities confirming that modifications to the predicate device do not affect safety or effectiveness.

Technological Characteristics

Fully automatic coagulation analyzer. Detection: chronometric (inductive sensors measuring metal ball oscillation) and photometric (tungsten-halogen lamp, 405/540 nm filters, optical fibers). Materials: standard laboratory cuvettes. Connectivity: RS232, USB, RJ45. Electrical safety: IEC 61010-2-101:2003. Software: embedded system with GUI and LIS communication protocols.

Indications for Use

Indicated for use in hospital or healthcare laboratories to perform in vitro coagulation tests on human plasma to aid in the diagnosis of coagulation abnormalities or monitoring of anticoagulant therapy in patients.

Regulatory Classification

Identification

A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.

Special Controls

*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification OIR Decision Memorandum To: THE FILE RE: DOCUMENT NUMBER K151867 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable: 1. The name and 510(k) number of the SUBMITTER'S previously cleared device: STA-R Evolution® Expert Series Hemostasis System, K093001. 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials. 3. A description of the device MODIFICATIONS, including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for the following modifications: (1) new device name (STA R Max®), (2) implementation of a new graphical user interface (GUI) to use the graphical chart of Max line of STA Analyzers, (3) changes to the external design (color and shape) of the analyzer which include modifications of the openings and the addition of a new computer table for improved ergonomics, (4) improvement of the communication protocol with a Laboratory Information System (LIS) or a middleware so more information can be exchanged. (Attachment 9) 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, target population, anatomical sites, point of use, electrical safety, chronometric and photometric method of coagulation detection, barcode ID of samples and reagents, and disposables are similar. Differences include dimensions, integrated PC microprocessor, memory and hard disk, weight, connections, computer table, and touch screen. (Attachment 9) 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis utilizing the FMEA model in compliance with ISO 14971:2007. (p 27-29/44 and Attachment 11) b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied. (Attachment 18) c) Declaration of conformity with design controls. The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual and the results demonstrated that the predetermined acceptance criteria were met. (Attachment 10) ii) A statement signed by the individual responsible, the manufacturing facility is in conformance with the design control procedure requirement specified in 21 CFR 820.30 and the records are available for review. (Attachment 10). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modifications and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...