RETICONE SYSTEM FOR EPICS XL FLOW CYTOMETRY SYSTEMS
Applicant
Coulter Corp.
Product Code
GKZ · Hematology
Decision Date
Apr 26, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
The reticONE™ SYSTEM for EPICS® XL™ Flow Cytometry Systems combines a reagent kit consisting of a Coriphosphine-O dye and a Biological Calibrator, and software for automated analysis of reticulocytes in whole blood using EPICS® XL™ Flow Cytometry Systems with SYSTEM II™ Software. The system is intended "For In Vitro Diagnostic Use" and allows identification and enumeration of reticulocyte percentage and absolute count.
Device Story
System uses Coriphosphine-O fluorescent dye to stain whole blood samples; analyzed via EPICS® XL™ Flow Cytometry Systems. Software performs automated gating, data acquisition, and analysis to identify and enumerate reticulocytes. Used in clinical laboratory settings by trained personnel. Output provides reticulocyte percentage and absolute count to clinicians for diagnosing anemias, monitoring erythropoiesis, and assessing bone marrow recovery. Benefits include standardized, automated alternative to manual microscopy.
Clinical Evidence
Bench testing only. Studies included stability (unstained/stained samples), carryover, linearity (0.2% to 12.5% range), and precision (within-run, interlaboratory, site). Accuracy study compared reticONE™ to Retic-COUNT™ using diverse patient populations (n=unspecified, ages 18-85); results demonstrated comparable reticulocyte percentage and absolute count values.
Technological Characteristics
System includes Coriphosphine-O fluorescent dye reagent kit, biological calibrator, and dedicated software. Operates via flow cytometry (fluorescence-based cell staining). Designed for use with EPICS® XL™ Flow Cytometry Systems and SYSTEM II™ Software. Standardized per NCCLS H44-A.
Indications for Use
Indicated for in vitro diagnostic use in adults (18-85 years) to identify and enumerate reticulocyte percentage and absolute count in whole blood. Useful for evaluating erythropoiesis; diagnosing/detecting anemia, hemorrhage, hemoglobinopathies, nutritional/vitamin deficiencies, aplastic anemia, hemolytic anemias, hypersplenism, renal disease, and monitoring bone marrow recovery/transplantation.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
{1}------------------------------------------------
4|26|99
K990327
Image /page/0/Picture/2 description: The image shows the word "COULTER" in a bold, sans-serif font. Above the word is a logo consisting of two overlapping circles with a series of triangles pointing to the right. The logo is black and white, and the word "COULTER" is in black. There is a horizontal line below the word "COULTER".
| COULTER CORPORATION<br>P.O. BOX 169015<br>Miami, Florida 33116-9015 USA | | |
|-----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Customer Service: (800) 526-7694<br>Product Information: (800) 526-6932<br>(800) 327-6531 (305) 327-6531<br>www.coulter.com | Date: | January 29, 1999 |
| | Title: | Summary of Safety and Effectiveness Information for 510(k) Premarket Notification |
| Coulter Corporation<br>Miami, Florida USA | Product: | reticONE™ SYSTEM for EPICS® XL Flow Cytometry Systems |
| Coulter Leasing Corporation<br>Miami, Florida USA | Company: | Coulter Corporation<br>11800 SW 147 Avenue<br>Miami, FL 33196-2500 |
| Coulter Electronics, Pty. Ltd.<br>Sydney, Australia | Contact: | Dr. Marion S. Gaide (M/C: 31-B06)<br>Senior Regulatory Affairs Specialist<br>Premarket Regulatory Affairs |
| Coulter Electronics Ind. & Com., Ltda.<br>Rio de Janeiro, Brazil | | |
| Coulter Electronics of Canada, Ltd.<br>Burlington, Ontario, Canada | Telephone: | 305-380-2594 |
| Coulter Electronics, Ltd.<br>Luton, Bedfordshire, England | Common or Usual or Classification Name:<br>Reticulocyte Analysis System with Reagents and<br>Software for Flow Cytometry | |
| Coultronics France, S.A.<br>Margency, France | Product Classification:<br>Product Code: GKZ; C.F.R. Section: 864.5220; Classification<br>Panel: Hematology and Pathology Devices; Device Class: II | |
| Coulter Electronics GmbH<br>Krefeld, Germany | | |
| Coulter Electronics (HK), Ltd.<br>Hong Kong | Intended Use:<br>The reticONE™ SYSTEM for EPICS® XL™ Flow Cytometry Systems<br>combines | |
| Coulter K. K.<br>Tokyo, Japan | a reagent kit consisting of a Coriphosphine-O dye and a Biological Calibrator,<br>and software for automated analysis of reticulocytes in whole blood using<br>EPICS® XL™ Flow Cytometry Systems with SYSTEM II™ Software. The<br>system is intended "For In Vitro Diagnostic Use" and allows identification<br>and enumeration of reticulocyte percentage and absolute count. | |
| Coulter de Mexico S.A., DE C.V.<br>Mexico City, Mexico | | |
| Coulter Electronics, Ltd.<br>Mijdrecht, Netherlands | | |
| Coulter Electronics, Pty. Ltd.<br>Auckland, New Zealand | | |
| Coulter Electronics Sales of PR., Inc.<br>San Juan, Puerto Rico | Substantial Equivalence: 510(k) Premarket Notification: K872166<br>Retic-COUNT™ (Thiazole Orange*) Reagent | |
| Coulter Electronics, Ltd.<br>Johannesburg, South Africa | | 510(k) Premarket Notification: K880636<br>Retic-COUNT™ Reticulocyte Enumeration Software |
| Coulter Electronics Ltd.<br>Istanbul, Turkey | Product Differences:<br>The reticONE™ SYSTEM and Retic-COUNT™ are essentially<br>identical with respect to features and principles of operation. The<br>intended use of the New System and the Predicate System is the same.<br>Further, the New System and the Predicate System use the same, well-<br>established, state-of the-art technologies of cell staining with a<br>fluorescent dye-containing reagent and flow cytometric analysis to<br>measure reticulocytes in whole blood [NCCLS Document H44-A].<br>Also, each system is designed for use with specific reagents,<br>instruments and dedicated software for a) automated instrument set-up,<br>standardization and operation; b) sample gating and data acquisition; c)<br>flow cytometric analysis; and d) results generation and display. Finally,<br>both systems are alternatives to standard microscopy for reticulocyte | |
| Coulter Electronics S.A.<br>Caracas, Venezuela | | |
{2}------------------------------------------------
### Product testing to assess the performance of the reticONE™ SYSTEM is described below. Studies were Product Testing: designed in line with instructions for use given in the reticONE™ SYSTEM Guide, Package Inserts, Product Manuals, and performance specifications. Specimens were assayed with Retic-COUNT™ for comparison purposes. The results of product testing demonstrate that the retic-ONE™ SYSTEM meets all performance specifications and provides reticulocyte percentage and absolute count values comparable to those of Retic-COUNT™.
#### 1. Stability Studies:
Three studies were carried out to demonstrate whole blood specimen and sample stability claims for the reticONE™ SYSTEM.
Stored (Unstained) Whole Blood Specimens Held for 72 Hours at 2-8 ℃: a.
Normal whole blood specimens (n = 3) were collected by venipuncture into K,EDTA and ten replicate samples prepared for each specimen for each of four time points: 0-hours; 48-hours; and 72-hours; and 72-hours after collection. The 0-hours samples were stained and analyzed for reticulocyte percentage within 30 minutes of specimen collection using the reticONE™ SYSTEM and EPICS® XL-MCL™ flow cytometer. The remaining samples were stained and analyzed for reticulocyte percentage after either 24-hours; or 72-hours of storage at 2-8°C. All analyses were carried out at 20-25° using the procedures described in the reticONE™ SYSTEM Guide. The results were analyzed in terms of reticulocyte percentage and standard deviation (± 1 SD) for each donor.
Stored (Unstained) Whole Blood Specimens Held for 6 Hours at 20-25 ℃: b.
Normal whole blood specimens (n = 10) were collected by venipuncture into K,EDTA and ten replicate samples prepared for each specimen for each of two time points: 0-hours after collection. The 0hours samples were stained and analyzed for reticulocyte percentage within 30 minutes of specimen collection using the reticONE™ SYSTEM and EPICS® XL-MCL™ flow cytometer. The remaining samples were stained and analyzed for reticulocyte percentage after 6-hours of storage at 20-25°C. All analyses were carried out at 20-25° using the procedures described in the reticONE™ SYSTEM. The results were analyzed in terms of reticulocyte percentage and standard deviation (± 1 SD) for each donor.
#### Stored Prepared (Stained) Samples Held for 6 Hours at 20-25 ℃: C.
Normal whole blood specimens (n = 5) were collected by venipuncture into K.EDTA and ten replicate samples prepared for each specimen. The samples (0-hours) were stained and analyzed for reticulocyte percentage within 30 minutes of specimen collection using the reticONE™ SYSTEM and EPICS® XL-MCL™ flow cytometer. These same stained samples were reanalyzed for reticulocyte percentage after 6-hours of storage at 20-25°C. All analyses were carried out at 20-25° using the procedures described in the reticONE™ SYSTEM Guide. The results were analyzed in terms of reticulocyte percentage and standard deviation (± 1 SD) for each donor.
The results for the three stability studies clearly demonstrated that the reticONE™ SYSTEM meets stability claims for both stored (unstained) whole blood specimens and prepared (stained) samples under the storage times and temperature conditions studied.
#### 2. Carryover Study:
Carryover percent was evaluated on two EPICS® XL-MCL™ flow cvtometers by the approved NCCLS procedure for high-to-low carryover testing in a reticulocyte flow cytometric analysis system [NCCLS Document H44-A]. The High Level and the Low Level of the Streck Laboratories, Inc® multiple-level Reticulocyte Control product, Retic-Chex, were used in place of a whole blood specimen with an elevated RBC count and a whole blood specimen with a low RBC count. A single sample was prepared for each Level and stained with the COULTER® reticONE™ Reagent Kit. The stained High Level samples were then analyzed for reticulocyte percentage in three successive runs (Runs: 1), I, 1, 1, using the reticONE™ SYSTEM and EPICS® XL-MCL™ flow cytometer. The stained Low Level samples were next analyzed in the same manner (Runs: J, J, J.). All sample staining and analyses were carried out at 20-25° using the procedures described in the reticONE™ SYSTEM Guide. Carryover percent was calculated using the following equation: Carryover Percent (%) == [{J ] - J x 100] == [(1} - J}). The results for the carryover study clearly demonstrated that the reticONE™ SYSTEM exhibits minimal carryover compared to the Flow Cytometer Specification.
{3}------------------------------------------------
#### 3. Linearity Study:
Two (n = 2) whole blood specimens were collected by venipuncture into K,EDTA and serially diluted (n = 10 data points: low, high and 8 serial dilutions in between) to achieve a defined range of 0.2% to 12.5% reticulocytes. A sample was prepared for each data point and stained with the COULTER® reticONE™ Reagent Kit. Three replicate measurements of reticulocyte percentage were then made for each stained sample using the reticONE™ SYSTEM and EPICS® XL-MCL™ flow cytometer. All sample staining and analyses were carried out at 20-25° using the procedures described in the reticONE™ SYSTEM Guide. The results were analyzed in terms of regression and correlation analyses for recovered versus expected reticulocyte percentage. The results for the linearity study clearly demonstrated linearity over the defined (reportable) range for the reticONE™ SYSTEM.
#### Precision Study: 4.
Three studies were carried out to demonstrate the reproducibility of the reticONE™ SYSTEM.
- Within Run (Intralaboratory) Precision: a.
Measurements were made on sample [normal or abnormal] whole blood specimens for each of five levels of reticulocyte percentage using one EPICS® XL-MCL™ flow cytometer. All measurements were made in replicates of ten determinations. Results were analyzed in terms of mean reticulocyte percentage, standard deviation (± 1 SD) and coefficient of variation (%CV).
- b. Interlaboratory Precision:
Measurements were made on samples from a single [normal] whole blood specimen for one level of reticulocyte percentage on the same day by three laboratories using different EPICS® XL-MCL™ flow cvtometers. All measurements were made in replicates of ten determinations. Results were analyzed in terms of mean reticulocyte percentage, standard deviation (± 1 SD) and coefficient of variation (%CV).
- C. Site Precision:
Measurements were made on sample [normal or abnormal] whole blood specimens for each of five levels of reticulocyte percentage using one EPICS® XL-MCL™ flow cytometer. All measurements were made in replicates of ten determinations. Results were analyzed in terms of mean reticulocyte percentage, standard deviation (± 1 SD) and coefficient of variation (%CV).
The results for the three precision studies clearly demonstrated the reproducibility of the reticONE™ SYSTEM.
#### ડ. Accuracv Study:
Normal and abnormal whole blood specimens were collected from geographically diverse populations of males and females unselected as to race and ranging in age from 18 to 85 years. Specimens were divided, processed and assayed in parallel with reticONE™ SYSTEM and Retic-COUNT™. Reticulocyte percentage (%) and absolute count (cells/uL) values were determined with EPICS® XL-MCL or FACScan™ flow cytometers (as applicable). Red blood cell counts were obtained using a COULTER® STKS™ Analyzer. The results were analyzed in terms of minimum, maximum, mean, standard deviation (± 1 SD), confidence intervals with 95% limits, regression and correlation analyses, and analyses of variance. The results for the accuracy study clearly demonstrated that the reticONE™ SYSTEM and Retic-COUNT™ identify and enumerate comparable numbers of reticulocytes in whole blood specimens.
Marion S. Gaiser
Marion S. Gaide, Ph.D. Senior Regulatory Affairs Specialist Premarket Regulatory Affairs
January 29, 1999
Date
{4}------------------------------------------------
Public Health Service
Marion S. Gaide, Ph.D. Senior Regulatory Affairs Specialist Premarket Regulatory Affairs Coulter Corporation 11800 SW 147 Avenue Miami. Florida 33196-2500
APR 2 6 1999
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Re: K990327
Trade Name: reticONE™ SYSTEM for EPICS® XL™ Flow Cytometry Systems Regulatory Class: II Product Code: GKZ Dated: January 29, 1999 Received: February 2, 1999
Dear Dr. Gaide:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Dutman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## INDICATIONS FOR USE
# 199032
## 510(k) Number (if known): Not Yet Assigne
#### reticONE™ SYSTEM for EPICS® XL™ Flow Cytometry Systems Device Name:
## Indications For Use:
The reticONE™ SYSTEM for EPICS® XL™ Flow Cytometry Systems combines a reagent kit consisting of a Coriphosphine-O dye and a Biological Calibrator, and software for automated analysis of reticulocytes in whole blood using EPICS® XL™ Flow Cytometry Systems with SYSTEM II™ Software. The system is intended "For In Vitro Diagnostic Use" and allows identification and enumeration of reticulocyte percentage and absolute count.
The reticulocyte count is diagnostically useful in discriminating between normal erythropoiesis. It can be useful in the diagnosis or detection of anemal hemorthaging, hemoglobinopathies and certain nutritional or vitamin deficiencies. Decreased or defective red cell production may result in a lower reticulocyte count such as in aplastic anemias. Elevated reticulocyte counts may be observed in clinical conditions where red cell destruction is increased (for example, hemolytic anemias and hypersplenism) or where there is increased red cell production (for example, erythropoietin drug therapy and a response to treated anemia).
Since the kidneys are a primary source of erythropoietin (a hormone that regulates erythroid development), the reticulosyte count is also affected in individuals with renal disease. In cases of renal atrophy, the reticulocyte count will be decreased. In cases of malignancy or hypersplenism, the reticulocyte count will be elevated.
Reticulocyte counts are also used as an indicator of bone marrow recovery following intensive or a bone marrow transplantation. Increased reticulocyte counts in these patients are indicative of bone marrow regeneration.
Peter E. Mackin
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number. K970321
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
1 Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
retI4Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.