MULTISET SOFTWARE FOR AUTOMATED FLOW CYTOMETRIC DATA ANALYSIS
Device Facts
| Record ID | K963963 |
|---|---|
| Device Name | MULTISET SOFTWARE FOR AUTOMATED FLOW CYTOMETRIC DATA ANALYSIS |
| Applicant | Becton Dickinson Immunocytometry Systems |
| Product Code | GKZ · Hematology |
| Decision Date | Dec 12, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 864.5220 |
| Device Class | Class 2 |
| Attributes | Software as a Medical Device |
Indications for Use
MultiSET is an accessory software to in vitro diagnostic reagents employing CD45 gating for identification and enumeration of lymphocyte subsets in human peripheral blood.
Device Story
MultiSET software; accessory to flow cytometry reagents; processes three-color direct immunofluorescence data from erythrocyte-lysed whole blood samples. Input: flow cytometric data using CD45 gating. Operation: automates determination of analysis gates to identify lymphocyte populations based on CD45+ leucocytes with low side scatter. Output: percentages of lymphocyte subsets. Used in clinical laboratories; operated by technicians/clinicians. Benefit: automates manual gating process; provides consistent enumeration of lymphocyte subsets for monitoring immunodeficiency and autoimmune conditions.
Clinical Evidence
Bench testing comparing automated MultiSET to manual CELLQuest. Sample size: 41 donors (17 normal, 24 abnormal). Primary endpoint: accuracy of lymphocyte subset enumeration. Results: MultiSET provided equivalent performance to manual predicate software.
Technological Characteristics
Software accessory for flow cytometry; utilizes CD45 gating for immunophenotyping; automated gate determination algorithm; operates on flow cytometer workstation.
Indications for Use
Indicated for identification and enumeration of lymphocyte subsets in human peripheral blood for monitoring immunodeficiency and autoimmune disease.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Predicate Devices
- CELLQuest (K946055)
Related Devices
- K170974 — BD FACSLyric Flow Cytometer (3-1, 4-2, 4-2-2 and 4-3-3 optical configurations), BD FACSuite Clinical Software, BD Multitest 6-Color Assays, BD Multitest 4-Color Assays, BD Multitest 6-Color TBNK · Becton, Dickinson and Company · Jul 3, 2017
- K141932 — AQUIOS CL FLOW CYTOMETER, AQUIOS TETRA-1 PANEL,AQUIOS TETRA-2+PANEL, AQUIOS IMMUNO-TROL, AQUIOS IMMUNO-TROL LOW, AQUIOS · Beckman Coulter, Inc. · Apr 10, 2015
- K954570 — ORTHO IMMUNOCOUNT FLOW CYTOMETRY SYSTEM · Ortho Diagnostic Systems, Inc. · Apr 30, 1996
- K980858 — MULTITEST CD3/CD16+56/CD45/CD19 REAGENT AND MULTITEST IMK KIT LYSING SOLUTION · Becton Dickinson Immunocytometry Systems · May 22, 1998