← Product Code [GKZ](/submissions/HE/subpart-f%E2%80%94automated-and-semi-automated-hematology-devices/GKZ) · K961701

# MOUSE ANTI-HUMAN CD3, T-CELL, FITC AND CD4, T-CELL, (HELPER/INDUCER) RPE (K961701)

_Dako Corp. · GKZ · Jul 16, 1996 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HE/subpart-f%E2%80%94automated-and-semi-automated-hematology-devices/GKZ/K961701

## Device Facts

- **Applicant:** Dako Corp.
- **Product Code:** [GKZ](/submissions/HE/subpart-f%E2%80%94automated-and-semi-automated-hematology-devices/GKZ.md)
- **Decision Date:** Jul 16, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5220
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

For In Vitro Diagnostic Use Mouse Anti-Human T-cell, CD3/FITC, UCHT1 + Mouse Anti-Human T-cell, CD4/RPE, MT310 (DAKO Anti-CD3/FITC and Anti-CD4/RPE) has been developed for use in flow cytometry for the analysis of CD3+ and CD4+ T-cells. This reagent allows simultaneous detection and quantification of CD3+CD4+ cells (CD4 positive T-lymphocytes) in normal and pathological conditions such as immunodeficiency disorders. It is one component of the suggested monoclonal antibody (MAb) combinations for routine immunophenotyping of lymphocytes in peripheral blood using flow cytometry.

## Device Story

Reagent cocktail containing purified mouse anti-human CD3 (clone UCHT1) conjugated to FITC and mouse anti-human CD4 (clone MT310) conjugated to R-phycoerythrin. Used in clinical laboratories by trained personnel for flow cytometric analysis of peripheral blood specimens. Input: peripheral blood samples with RBCs lysed. Operation: antibodies bind to specific T-cell surface antigens; flow cytometer detects fluorescence signals to identify and quantify CD3+CD4+ T-lymphocyte subpopulations. Output: quantitative data on T-cell subsets. Clinical utility: assists in immunophenotyping for assessment of immunodeficiency disorders and other pathological conditions. Benefits: enables simultaneous detection of two markers in a single assay, streamlining routine lymphocyte analysis.

## Clinical Evidence

Clinical evaluation compared subject reagent to individual predicate reagents using peripheral blood samples from healthy adults. Correlation for CD3+ and CD4+ T-cell measurement was >0.96. Linearity testing using JM and CEM cell lines yielded r=0.999. Reproducibility assessed across two flow cytometers at three antigen concentrations. Cross-reactivity tested against RBCs, monocytes, granulocytes, lymphocytes, and platelets.

## Technological Characteristics

Monoclonal antibody cocktail in 0.05M Tris-HCl buffer (pH 7.2), 15 mM NaN3, 0.1M NaCl, 1% carrier protein. Conjugates: FITC (isomer 1) and R-phycoerythrin. Principle: immunofluorescence staining for flow cytometry. No electronic or software components.

## Regulatory Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

## Predicate Devices

- DAKO Monoclonal Mouse Anti-Human T-cell, CD3/FITC, UCHT1 ([K942797](/device/K942797.md))
- DAKO Monoclonal Mouse Anti-Human helper/inducer T-cell, CD4/RPE, Clone MT310 ([K945212](/device/K945212.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K941701
3

JUL 16 1996

510(k) Summary

Submitter: DAKO Corporation
6392 Via Real
Carpinteria, CA 93013
(805)566-6655

Contact: Gretchen M. Murray, Ph.D., Regulatory Affairs Asst. Manager

Date Summary
Prepared: December 27, 1995

Device Name: Mouse Anti-Human T-cell, CD3/FITC, UCHT1 +
Mouse Anti-Human T-cell, CD4/RPE, MT310

Device
Classification: Class II according to 21 CFR 864.5220, on the basis that monoclonal antibodies are accessories for automated differential cell counters.

Panel: This device classification is under the Hematology and Pathology devices panel, Division of Clinical Laboratory Devices.

Product Code: GKZ

Predicate Device(s): DAKO Monoclonal Mouse Anti-Human T-cell, CD3/FITC, UCHT1 (DAKO Code Number F0818, FDA K942797) and
DAKO Monoclonal Mouse Anti-Human helper/inducer T-cell, CD4/RPE,
Clone MT310 (DAKO Code Number R0805, FDA K945212)

Device Description: Purified mouse anti-human CD3, Clone UCHT1, conjugated with fluorescein isothiocyanate, isomer 1 (FITC) + purified mouse anti-human CD4, Clone MT310, conjugated with R-phycoerythrin, present in 0.05M Tris-HCl buffer, pH 7.2, 15 mM NaN₃, 0.1M NaCl, stabilized with 1% carrier protein

Subpopulations of lymphocytes may be stained with fluorochrome-conjugated antibody and evaluated in peripheral blood specimens when contaminating red blood cells (RBC's) are lysed prior to flow cytometric analysis. A subpopulation of WBC's are selected for assessment based upon cell morphology.

Intended Use: For In Vitro Diagnostic Use

Mouse Anti-Human T-cell, CD3/FITC, UCHT1 + Mouse Anti-Human T-cell, CD4/RPE, MT310 (DAKO Anti-CD3/FITC and Anti-CD4/RPE) has been developed for use in flow cytometry for the analysis of CD3⁺ and CD4⁺ T-cells. This reagent allows simultaneous detection and quantification of CD3⁺CD4⁺ cells (CD4 positive T-lymphocytes) in normal and pathological conditions such as immunodeficiency disorders. It is one component of the suggested monoclonal antibody (MAb) combinations for routine immunophenotyping of lymphocytes in peripheral blood using flow cytometry.

Comparison of
Technological
Characteristics

Performance characteristics have been established by clinical evaluation of DAKO Anti-CD3/FITC and Anti-CD4/RPE compared to the individual single reagent predicate devices that quantitatively measure CD3⁺ and CD4⁺ T-cells that have been previously cleared by FDA (DAKO CD3/FITC, Code No. F0818 and DAKO CD4/RPE, Code No. R0805). When flow cytometric tests of peripheral blood samples obtained from apparently healthy adults were

1

{1}

completed, correlation of Anti-CD3/FITC, UCHT1 with DAKO Anti-CD3/FITC and Anti-CD4/RPE approached a direct 1 : 1 comparison for measurement of CD3 + cells. Correlation of Anti-CD4/RPE, MT310 with DAKO Anti-CD3/FITC and Anti-CD4/RPE approached a direct 1 : 1 comparison for measurement of CD4 + cells. Data for the measurement of CD3 + T-cells by DAKO Anti-CD3/FITC and Anti-CD4/RPE reagent compared to DAKO CD3/FITC gave a correlation greater than 0.96 using the whole blood method for flow cytometry. Data for the measurement of CD4 + T-cells by DAKO Anti-CD3/FITC and Anti-CD4/RPE reagent compared to DAKO CD4/RPE gave a correlation greater than 0.96 using the whole blood method for flow cytometry.

The CD3 antibody clone, UCHT1, was clustered at the First Leukocyte Typing Workshop, Paris, France, 1982. The CD4 antibody clone, MT310, was clustered at the Second Leukocyte Typing Workshop, Boston, 1984.

Linearity testing of DAKO CD3/FITC using JM cells gave the following linear equation:

$$
y = 0.02 + 0.98x; \quad r = 0.999
$$

Linearity testing of DAKO CD4/RPE using CEM cells gave the following linear equation:

$$
y = -0.01 + 1.03x; \quad r = 0.999
$$

In addition, reproducibility of DAKO reagents using replicates (from peripheral blood) run on two different flow cytometers was measured at three concentrations of each antigen. Cross-reactivity of Anti-CD3/FITC, plus Anti-CD4/RPE with peripheral blood cells (red blood cells, monocytes, granulocytes, lymphocytes, and platelets) was measured.

## Conclusions:

Results of the above testing as well as the information provided by the First and Second Leukocyte Typing Workshops indicate that the DAKO Anti-CD3/FITC plus Anti-CD4/RPE reagent performs as well as DAKO CD3/FITC in the detection and enumeration of CD3⁺ lymphocytes while the DAKO Anti-CD3/FITC plus Anti-CD4/RPE reagent performs as well as DAKO CD4/RPE in the detection and enumeration of CD4⁺ lymphocytes using flow cytometry. Safety of the DAKO Anti-CD3/FITC plus Anti-CD4/RPE reagent and its individual predicate devices is high as all reagents are used for in vitro testing.

---

**Source:** [https://fda.innolitics.com/submissions/HE/subpart-f%E2%80%94automated-and-semi-automated-hematology-devices/GKZ/K961701](https://fda.innolitics.com/submissions/HE/subpart-f%E2%80%94automated-and-semi-automated-hematology-devices/GKZ/K961701)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
