ORTHO COUNT CALIBRATION KIT FOR ORTHO CYTORONABSOLUTE LASER FLOW CYTOMETRY SYSTEM
K935720 · Ortho Diagnostic Systems, Inc. · GKZ · May 9, 1996 · Hematology
Device Facts
Record ID
K935720
Device Name
ORTHO COUNT CALIBRATION KIT FOR ORTHO CYTORONABSOLUTE LASER FLOW CYTOMETRY SYSTEM
Applicant
Ortho Diagnostic Systems, Inc.
Product Code
GKZ · Hematology
Decision Date
May 9, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
Ortho-Count Calibration Kit for ORTHO CYTORONABSOLUTE Laser Flow Cytometry System is intended to be used to calibrate and verify calibration of the absolute lymphocyte counting function of the ORTHO CYTORONABSOLUTE.
Device Story
Ortho-Count Calibration Kit consists of a Calibrator Bead Suspension and three Verifier Bead Suspensions (I, II, III) used with the ORTHO CYTORONABSOLUTE Laser Flow Cytometry System. The Calibrator contains microparticle beads at a known concentration (events/unit volume) in a proprietary buffer, sized to emulate leukocytes. The system uses precision stepper motors to deliver diluted samples to a flow cell at a constant rate; calibration involves setting the analysis time so that the instrument counts a volume equivalent to 1 uL of undiluted sample. Verifier suspensions provide low, normal, and high concentration targets to verify calibration accuracy and serve as daily quality control. A properly calibrated instrument yields absolute counts within ± 10% of labeled values. This process enables the direct determination of absolute lymphocyte subset counts on a single instrument, replacing the traditional requirement for separate flow cytometry and hematology instruments.
Clinical Evidence
Bench testing only. Performance compared against Coulter JT2 (calibrated with S-CAL) and whole blood methods. Study across four sites demonstrated inter-laboratory reproducibility and accuracy using Verifier Bead Suspensions. Direct absolute lymphocyte subset counts (CD19, CD8+, CD4+, CD3+) on the ORTHO CYTORONABSOLUTE showed strong correlation with combined hematology/flow cytometry calculations (r=0.96, slope=0.89). Quality control study over 60 days confirmed stability and reliability of calibration.
Technological Characteristics
Kit contains microparticle bead suspensions in a proprietary buffer. Beads are sized to emulate leukocytes. System uses laser flow cytometry with precision stepper motor-controlled sample delivery. Calibration is based on setting analysis time to correspond to a fixed volume (1 uL).
Indications for Use
Indicated for use in calibrating and verifying the absolute lymphocyte counting function of the ORTHO CYTORONABSOLUTE Laser Flow Cytometry System for immune status monitoring.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
{0}
ITEM 7. 807.87(h). A 510(k) SUMMARY AS DESCRIBED IN 807.92 OR A 510(k) STATEMENT AS DESCRIBED IN 807.93.
K935720
MAY - 9 1996
Ortho Diagnostic Systems Inc.
Ortho-Count Calibration Kit for
ORTHO CYTORONABSOLUTE™ Laser Flow Cytometry System
Page 56
{1}
510(k) SUMMARY
SUBMITTER: Ortho Diagnostic Systems Inc.
1001 U.S. Hwy 202
Raritan, NJ 08869-0606
CONTACT: Ms. Joanne Harris
Tel.: (908) 218-8404
Fax.: (908) 218-8168
DEVICE NAME: Ortho-Count Calibration Kit
for ORTHO CYTORONABSOLUTE™
Laser Flow Cytometry System
[864.8185 Calibrator for red
cell and white cell counting]
PREDICT: S-CAL® Kit
[K813375]
Coulter Diagnostics
DATE: November 29, 1993
## DEVICE DESCRIPTION:
Ortho-Count Calibration Kit for ORTHO CYTORONABSOLUTE Laser Flow Cytometry System, contains a Calibrator Bead Suspension for absolute lymphocyte count calibration and Verifier Bead Suspensions (I, II and III) to verify accuracy of absolute count calibration.
The Calibrator Bead Suspension is manufactured as known "events"/unit volume (i.e. the Calibration Number) in the form of microparticle beads suspended in a proprietary buffer. The size of the microparticles has been selected to emulate the range of size of leukocytes detected by the ORTHO CYTORONABSOLUTE under specified gating conditions.
The Verifier Bead Suspensions (I, II and III), are composed of materials identical to the Calibrator Bead Suspension and collectively provide a range of particle bead concentrations. The target values chosen approximate values expected from patients with low (Verifier I), normal (Verifier II) and high (Verifier III) white blood cell counts, respectively. The Verifiers are used to verify accuracy of the absolute lymphocyte count calibration immediately following calibration and are also used as daily quality control verification that the ORTHO CYTORONABSOLUTE is in calibration. A properly calibrated instrument will provide absolute counts within ± 10% of the labeled events/uL for each Verifier suspension.
In the context of immune status monitoring, absolute cell count is defined as the number of cells per unit volume. The ORTHO CYTORONABSOLUTE uses precision stepper motors to deliver diluted sample to the instrument's flow cell at a constant rate. Since constant flow rate means movement of a fixed volume per unit time, analysis time determines the volume of sample analyzed. Calibration of the ORTHO CYTORONABSOLUTE consists of setting analysis time such that a volume corresponding to 1 uL of undiluted sample is examined.
Ortho Diagnostic Systems Inc.
Ortho-Count Calibration Kit for
ORTHO CYTORONABSOLUTE™ Laser Flow Cytometry System
Page 57
{2}
Ortho-Count Calibrator Bead Suspension, representing a known number of particles per unit volume (i.e. events/uL) at the same dilution as used for testing samples, is counted until its "Calibration Number" is reached; the instrument is set to this analysis time. In subsequent analyses events are counted for this predetermined period of time, now calibrated to correspond to passage of the equivalent of 1 uL of undiluted sample through the instrument flowcell. As a result of this calibration, sample results can be read in absolute count, i.e. "events"/uL.
## INTENDED USE:
Ortho-Count Calibration Kit for ORTHO CYTORONABSOLUTE Laser Flow Cytometry System is intended to be used to calibrate and verify calibration of the absolute lymphocyte counting function of the ORTHO CYTORONABSOLUTE. This intended use is substantially equivalent to the intended use of S-CAL Kit, the predicate device, for lymphocyte counting. The intended use is different than that for S-CAL Kit in that Ortho-Count Calibration Kit may not be used to calibrate the ORTHO CYTORONABSOLUTE for parameters other than lymphocyte counting. This difference does not affect the safety or effectiveness of Ortho-Count Calibration Kit for its intended use since the ORTHO CYTORONABSOLUTE is not intended for all of the same uses as the Coulter Systems with which S-CAL Kit is intended to be used.
## TECHNOLOGICAL CHARACTERISTICS:
Ortho-Count Calibration Kit and S-CAL Kit both contain components which are recognized by cell counters as leukocytes and are used to convert the electronic signal obtained from such instruments into accurate leukocyte count results expressed in clinical terms. Both kits consist of suspensions of particles, whose size, shape, concentration and other characteristics have been accurately determined. Ortho-Count Calibration Kit employees microparticle beads, while S-CAL Kit uses fixed erythrocytes in suspension, to simulate leukocytes.
## PERFORMANCE DATA:
Performance of an ORTHO CYTORONABSOLUTE calibrated using Ortho-Count Calibration Kit (Meth 2) was compared to performance of the same instrument calibrated using whole blood whose total lymphocyte count was determined with a Coulter JT2 calibrated using S-CAL Kit (Meth 1), and to performance of the Coulter JT2 calibrated with S-CAL Kit (JT2). As shown in Table 1, there were no significant differences in absolute lymphocyte counts that could be attributed to either the instrument or the method of calibration.
Ortho Diagnostic Systems Inc.
Ortho-Count Calibration Kit for
ORTHO CYTORONABSOLUTE™ Laser Flow Cytometry System
Page 58
{3}
TABLE 1
LYMPHOCYTE COUNTS FROM A HEMATOLOGY ANALYZER VERSUS
LYMPHOCYTE COUNTS FROM ORTHO CYTORONABSOLUTE CALIBRATED BY
DIFFERENT METHODS
| Donor Number | Mean JT2 | Std. Dev. | Mean Method 1 | Std. Dev. | Mean Method 2 | Std. Dev. | Ratio Meth1 JT2 | Ratio Meth2 JT2 | Ratio Meth2 Meth1 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 1 | 1983 | 75 | 1776 | 23 | 1940 | 90 | 0.90 | 0.98 | 1.09 |
| 2 | 2383 | 41 | 2189 | 94 | 2147 | 88 | 0.92 | 0.90 | 0.98 |
| 3 | 1767 | 137 | 1736 | 71 | 1668 | 93 | 0.98 | 0.94 | 0.96 |
| 4 | 2217 | 75 | 2076 | 81 | 2129 | 90 | 0.94 | 0.96 | 1.03 |
| 5 | 1800 | 47 | 1805 | 52 | 1807 | 110 | 1.00 | 1.00 | 1.00 |
| 6 | 2250 | 55 | 2265 | 29 | 2132 | 56 | 1.01 | 0.95 | 0.94 |
| 7 | 1600 | 122 | 1557 | 25 | 1476 | 98 | 0.97 | 0.92 | 0.95 |
| 8 | 2300 | 71 | 2323 | 138 | 2271 | 61 | 1.01 | 0.99 | 0.98 |
| 9 | 3820 | 164 | 3730 | 81 | 3855 | 94 | 0.98 | 1.01 | 1.03 |
| 10 | 2517 | 117 | 2481 | 73 | 2413 | 141 | 0.99 | 0.96 | 0.97 |
| | | | | Mean | | | 0.97 | 0.96 | 0.99 |
| | | | | Standard Deviation | | | 0.04 | 0.04 | 0.05 |
Four ORTHO CYTORONABSOLUTE instruments located in different regions of the United States, were calibrated using Ortho-Count Calibration Kit to assess inter-laboratory reproducibility as evidenced by Ortho-Count Calibration Kit Verifier Bead Suspensions. Table 2 shows the accuracy and precision of the absolute counts at each site. Overall precision of the absolute counts between laboratories was equivalent to the within laboratory precision.
Table 2. Across Site Accuracy and Precision of Absolute Counts on Four Microparticle Calibrated CytoronAbsolutes
| Location | Site One | Site Two | Site Three | Site Four | Overall Inter-Lab |
| --- | --- | --- | --- | --- | --- |
| Verifier I | | | | | |
| Mean | 1736 | 1699 | 1669 | 1670 | 1693 |
| Std. Dev. | 61 | 43 | 43 | 44 | 55 |
| % CV | 3.5 | 2.5 | 2.6 | 2.6 | 3.2 |
| Verifier II | | | | | |
| Mean | 7851 | 7726 | 7805 | 7614 | 7748 |
| Std. Dev. | 174 | 225 | 101 | 173 | 193 |
| % CV | 2.2 | 2.9 | 1.3 | 2.3 | 2.5 |
| Verifier III | | | | | |
| Mean | 15434 | 15169 | 15209 | 14755 | 15140 |
| Std. Dev. | 257 | 492 | 225 | 288 | 409 |
| % CV | 1.7 | 3.2 | 1.5 | 2.0 | 2.7 |
Consistency of calibration of ORTHO CYTORONABSOLUTE instruments was demonstrated in a study comparing absolute lymphocyte counts obtained on a panel of donors at each of the four study sites.
Ortho Diagnostic Systems Inc.
Ortho-Count Calibration Kit for
ORTHO CYTORONABSOLUTE™ Laser Flow Cytometry System
Page 59
{4}
Mean absolute lymphocyte counts from eight determinations on each donor per site were plotted against the overall mean lymphocyte count for each donor across sites (Figure 1).

Figure 1. Absolute Lymphocyte Counts Measured on Cross-Calibrated ORTHO CYTORONABSOLUTE Instruments
Figure 1. Absolute lymphocyte counts were determined using cross-calibrated ORTHO CYTORONABSOLUTE instruments for a whole blood panel kept overnight at room temperature or shipped for next day delivery to each site. The average lymphocyte count obtained at each site is plotted versus the mean lymphocyte count across sites for each of 41 samples.
The determination of an absolute lymphocyte subset count requires both a lymphocyte count and the percentage of lymphocytes positive for the marker. By convention, these measurements have required separate flow cytometry and hematology instruments. A study was performed to demonstrate that when calibrated using Ortho-Count Calibration Kit it is possible to obtain equivalent absolute lymphocyte subset counts directly from the ORTHO CYTORONABSOLUTE. Figure 2 shows strong correlation between absolute count values determined directly on the ORTHO CYTORONABSOLUTE and those calculated from combined hematology and other flow cytometer/reagent results. Regression analysis of the data yields a slope of 0.89, a y-intercept of $-31 \pm 13$ and a correlation coefficient $(r) = 0.96$.
Ortho Diagnostic Systems Inc.
Ortho-Count Calibration Kit for
ORTHO CYTORONABSOLUTE™ Laser Flow Cytometry System
Page 60
{5}

Figure 2. Direct Determination of Absolute Lymphocyte Counts on the ORTHO CYTORONABSOLUTE Versus Hematology Plus Other Flow Cytometer Calculations
Figure 2. Absolute lymphocyte subset values were determined directly using a single ORTHO CYTORONABSOLUTE. These data are plotted versus absolute lymphocyte counts determined using separate flow cytometers and automated hematology analyzers. Absolute lymphocyte subsets plotted are: CD19, CD8+ cells that are also CD3+, CD4+ cells that are also CD3+ and total CD3+ cells
The utility of the Ortho-Count Calibrator Kit Verifier Bead Suspensions as quality control indicators of proper calibration of the ORTHO CYTORONABSOLUTE was demonstrated in a study in which counts/volume (absolute counts) were measured on the three verifier suspensions twice per week over a period of 60 days. Figure 3 shows the absolute counts obtained at each laboratory over the two months. The tight agreement in measurements made using each of the Verifier Bead Suspensions confirms the suitability of Ortho-Count Calibration Kit as an accurate quality control measure of the status of calibration of the ORTHO CYTORONABSOLUTE.
Ortho Diagnostic Systems Inc.
Ortho-Count Calibration Kit for
ORTHO CYTORONABSOLUTE™ Laser Flow Cytometry System
Page 61
{6}

Figure 3. Ortho-Count Calibration Kit Verifier Bead Suspensions as Quality Control Indicators of Status of Calibration of the ORTHO CYTORONABSOLUTE
Figure 3. Four ORTHO CYTORONABSOLUTE instruments were calibrated using Ortho-Count Calibration Kit. The absolute count of each Verifier Bead Suspension is plotted for each laboratory on each day throughout the study.
## CONCLUSIONS:
1. Ortho-Count Calibration Kit is substantially equivalent to S-CAL Kit used to calibrate Coulter hematology instruments, for the intended use of calibration and verification of the calibration of the absolute lymphocyte counting function of the ORTHO CYTORONABSOLUTE.
2. Ortho-Count Calibration Kit functions as a suitable cross-calibration standard to ensure inter-laboratory reproducibility.
3. Ortho-Count Calibration Kit provides accurate calibration and verification of calibration of the ORTHO CYTORONABSOLUTE, permitting direct determination of absolute lymphocyte subset values equivalent to those obtained using values calculated from separate flow cytometers and hematology analyzers.
4. Ortho-Count Calibration Kit provides reliable quality control of the calibration status of the ORTHO CYTORONABSOLUTE.
Ortho Diagnostic Systems Inc.
Ortho-Count Calibration Kit for
ORTHO CYTORONABSOLUTE™ Laser Flow Cytometry System
Page 62
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.