Sysmex XW-100 Automated Hematology Analyzer

K210346 · Sysmex America, Inc. · GKZ · Nov 8, 2022 · Hematology

Device Facts

Record IDK210346
Device NameSysmex XW-100 Automated Hematology Analyzer
ApplicantSysmex America, Inc.
Product CodeGKZ · Hematology
Decision DateNov 8, 2022
DecisionSESE
Submission TypeDual Track
Regulation21 CFR 864.5220
Device ClassClass 2
AttributesPediatric

Indications for Use

The XW-100 Automated Hematology Analyzer (XW-100) is a quantitative automated hematology analyzer intended for in vitro diagnostic use to classify and enumerate the following parameters for venous whole blood anticoagulated with K2/K3 EDTA: WBC, RBC, HGB, HCT, MCV, PLT, LYM%, Other WBC%, NEUT%, LYM#, Other WBC#, and NEUT#. It is not for use in diagnosing or monitoring patients with primary or secondary chronic hematologic diseases/disorders, oncology patients, critically ill patients, or children under the age of 2.

Device Story

XW-100 is an automated hematology analyzer for CLIA-waived settings; processes venous whole blood samples (K2/K3 EDTA). Uses impedance technology (direct current) with hydrodynamic focusing for cell counting/sizing; photometric method for hemoglobin. Operator manually mixes and loads samples. Device generates histograms from pulse heights to identify cell populations and calculate 12 hematology parameters (CBC, 3-part WBC differential, MCV). Software version 1.14 includes modified print suppression rules to simplify flagging for non-specialist operators. Results assist clinicians in basic hematologic assessment; suppression rules prevent reporting of potentially unreliable data in complex clinical cases. Benefits include rapid, automated blood analysis in point-of-care environments by non-laboratory personnel.

Clinical Evidence

No clinical trials; bench testing only. Internal study (N=139) and prospective sample challenge study (N=173) compared software v1.14 against v1.03 and v1.13. Samples included normal and abnormal blood with various interferences (e.g., cold agglutinins, lipemia, nucleated RBCs). Deming regression showed high correlation (r > 0.90) for all parameters. Confirmatory methods included blood smear analysis (CLSI H20-A2) and Sysmex XN-10 analyzer. Results confirmed that v1.14 maintains appropriate suppression of unreliable parameters.

Technological Characteristics

Impedance-based (DC) cell counter with hydrodynamic focusing; photometric hemoglobin measurement. Dimensions: 7"x14"x18". Standalone system. Factory calibrated. Uses XW Pack L (lyse) and XW Pack D (diluent). Software-based result suppression logic. Complies with IEC 61010-1, IEC 61326-1, and IEC 61326-2-6.

Indications for Use

Indicated for quantitative hematology analysis of venous whole blood (K2/K3 EDTA) in patients aged 2 years and older. Not for use in oncology, critically ill, or patients with chronic hematologic disorders.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY ## I Background Information: A 510(k) Number K210346 B Applicant Sysmex America, INC. C Proprietary and Established Names Sysmex XW-100 Automated Hematology Analyzer D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | GKZ | Class II | 21 CFR 864.5220 - Automated Differential Cell Counter | HE - Hematology | ## II Submission/Device Overview: A Purpose for Submission: This submission is a Dual 510(k) and CLIA Waiver by Application (Dual Submission) tracked as K210346 and CW170012. K210346 was submitted for clearance of a device modification. B Type of Test: Complete blood count (WBC, RBC, HGB, HCT, MCV, PLT) and leukocyte 3-part differential (LYM%, Other WBC%, NEUT%, LYM#, Other WBC#, NEUT#). ## III Intended Use/Indications for Use: A Intended Use(s): See Indications for Use below. Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} B Indication(s) for Use: The XW-100 Automated Hematology Analyzer (XW-100) is a quantitative automated hematology analyzer intended for in vitro diagnostic use to classify and enumerate the following parameters for venous whole blood anticoagulated with K2/K3 EDTA: WBC, RBC, HGB, HCT, MCV, PLT, LYM%, Other WBC%, NEUT%, LYM#, Other WBC#, and NEUT#. It is not for use in diagnosing or monitoring patients with primary or secondary chronic hematologic diseases/disorders, oncology patients, critically ill patients, or children under the age of 2. C Special Conditions for Use Statement(s): Rx - For Prescription Use Only The XW-100 is intended to be used by operators with a minimum of an earned high school diploma or equivalent. IV Device/System Characteristics: A Device Description: The XW-100 Automated Hematology Analyzer (XW-100) for CLIA Waived Use is an electrical resistance blood cell counter. This technology may also be referred to as Direct Current (DC) or impedance. The XW-100 analyzes human whole blood specimens anticoagulated with K2EDTA or K3EDTA and reports results for 12 hematology parameters, including the basic complete blood count (CBC), 3-part white blood cell (WBC) differential, and MCV. Principles of Operation The XW-100 uses direct current with hydrodynamic focusing for all parameters except hemoglobin, which is measured photometrically. The patient sample is aspirated, measured, diluted with diluent (and lysed for WBC measurement), then directed into a transducer chamber by a hydrodynamic focusing nozzle. The transducer chamber has a minute hole, or aperture. Electrodes are mounted on both sides of the aperture chamber, through which the direct current flows. Blood cells suspended in the diluted sample are injected through the aperture by the hydrodynamic focusing nozzle. The hydrodynamic focusing nozzle is positioned in front of the aperture and in line with the aperture's center. All blood cells are separated from each other and pass through the aperture in one direction, one cell at a time. When a cell passes through the aperture, it causes a change in the direct current resistance, which is proportional to the cell's size. These resistance changes are captured as electric pulses. The various blood cell counts are calculated by counting the pulses that occur in each cell size category. The analyzer then determines blood cell volume and identifies cells by creating and analyzing histograms of the various cell populations using their respective pulse heights. Hemoglobin is measured photometrically using a non-cyanide method. B Instrument Description Information: 1. Instrument Name: XW-100 Automated Hematology Analyzer for CLIA Waived Use K210346 - Page 2 of 11 {2} K210346 - Page 3 of 11 2. Specimen Identification: Specimen identification input is manual (by operator). 3. Specimen Sampling and Handling: The XW-100 processes anticoagulated venous whole blood collected in K2EDTA or K3EDTA collection tubes. Samples are manually mixed by inversion and loaded into an onboard sample adapter one at a time. 4. Calibration: The XW-100 is factory calibrated. 5. Quality Control: The XW-100 system performance is evaluated using XW QC CHECK, a stabilized whole blood matrix quality control material designed for statistical process control of the analyzer. Assayed parameters include: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, LYM%, OTHER WBC%, NEUT%, LYM#, OTHER WBC#, NEUT#, RDW-SD, RDW-CV and MPV. V Substantial Equivalence Information: A Predicate Device Name(s): XW-100 Automated Hematology Analyzer for CLIA Waived Use B Predicate 510(k) Number(s): K172604 C Comparison with Predicate(s): | Device & Predicate Device(s): | K210346 | K172604 | | --- | --- | --- | | Device Trade Name | XW-100 Automated Hematology Analyzer for CLIA Waived Use | XW-100 Automated Hematology Analyzer for CLIA Waived Use | | General Device Characteristic Similarities | | | | Intended Use/Indications For Use | The XW-100 Automated Hematology Analyzer (XW-100) is a quantitative automated hematology analyzer intended for in vitro diagnostic CLIA waived use to classify and enumerate the following | Same | {3} | | parameters for venous whole blood anticoagulated with K2/K3 EDTA: WBC, RBC, HGB, HCT, MCV, PLT, LYM%, Other WBC%, NEUT%, LYM#, Other WBC#, and NEUT#. It is not for use in diagnosing or monitoring patients with primary or secondary chronic hematologic diseases/disorders, oncology patients, critically ill patients, or children under the age of 2. | | | --- | --- | --- | | Test Principle | Impedance technology (direct current detection) with hydrodynamic focusing for all parameters except hemoglobin, which is measured photometrically. | Same | | Measuring Channel | Single hydrodynamic focused impedance chamber | Same | | Sample Type | Anticoagulated (K2EDTA or K3EDTA) venous whole blood | Same | | Sample Aspiration Volume | 15 μL | Same | | Analysis Reagents | XW Pack L (Lyse) XW Pack D (Diluent) | Same | | System Throughput | 20 cycles per hour | Same | | Test System Dimensions | Width: 7 inches Height: 14 inches Depth: 18 inches | Same | | Mode of Operation | Whole blood mode | Same | | Calibration and Quality Control | XW QC CHECK (K143577) SCSTM-1000 Calibrator | Same | K210346 - Page 4 of 11 {4} | | (K943268) | | | --- | --- | --- | | General Device Characteristic Differences | | | | CLIA Waiver Software Version | Version 1.14 with modification to print suppression rule | Version 1.03 | VI Standards/Guidance Documents Referenced: CLSI H20-A2 – Reference Leukocyte (WBC) Differential Count (Proportional) and Evaluation of Instrumental Methods; Approved Standard – Second Edition IEC 61010-1, Safety requirements for electrical equipment for measurement, control and laboratory use. Part 1: General requirements IEC 61326-1, Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements IEC 61326-2-6, Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: Refer to K172604 2. Linearity: Refer to K172604 3. Analytical Specificity/Interference: Refer to K172604 4. Accuracy (Instrument): Refer to K172604 and CW170012 5. Carry-Over: Refer to K172604 K210346 - Page 5 of 11 {5} B Other Supportive Instrument Performance Characteristics Data: 1. Internal study to evaluate new version of software v1.14 with modification to print suppression rule An internal study was performed to evaluate: - The effectiveness of the new software version with suppression rule modification to appropriately suppress parameters according to the XW-100 algorithm suppression rules (used to simplify flagging for CLIA waived use). - The results unsuppressed by the new software and to assess the accuracy of unsuppressed results for all reported parameters. A total of 139 residual whole blood samples (mix of normal and abnormal) collected in K2EDTA anticoagulant were run on two XW-100 CLIA waived analyzers using the on-market software version (v1.13) and within a target of one hour all samples were run on two additional XW-100 CLIA waived analyzers using the proposed modified software version (v1.14). The results using both versions of software were compared to assess the frequency of suppression for all parameters, as well as evaluate the comparability of the reported parameters. Deming linear regression analysis was performed on all unsuppressed results for all reported parameters. The results of this regression analysis are shown in the table below. K210346 - Page 6 of 11 {6} | Internal Study Deming Regression Analysis | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | Measurand | N | Result Range (N=139) | Correlation Coefficient | Slope | 95% (CI) | Intercept | 95% (CI) | | WBC (x 10³/μL) | 70 | 3.0–22.8 | 0.9981 | 0.9985 | 0.9792, 1.0178 | 0.0279 | -0.1043, 0.1601 | | RBC (x 10⁶/μL) | 129 | 2.28–5.86 | 0.9967 | 1.0086 | 0.9937, 1.0236 | -0.0111 | -0.0695, 0.0473 | | HGB (g/dL) | 57 | 10.2–15.6 | 0.9874 | 0.9977 | 0.9580, 1.0373 | 0.0145 | -0.4867, 0.5157 | | HCT (%) | 58 | 28.0–47.1 | 0.9927 | 1.0030 | 0.9759, 1.0300 | 0.1220 | -0.9107, 1.1548 | | PLT (x 10³/μL) | 121 | 107–570 | 0.9945 | 0.9885 | 0.9657, 1.0113 | 3.4479 | -2.5141, 9.4099 | | MCV (fL) | 58 | 80.6–108.2 | 0.9971 | 1.0024 | 0.9854, 1.0195 | -0.3465 | -1.9205, 1.2275 | | NEUT# (x 10³/μL) | 70 | 1.1–16.8 | 0.9978 | 1.0022 | 0.9816, 1.0228 | -0.0062 | -0.1141, 0.1017 | | NEUT% | 70 | 21.8–91.9 | 0.9821 | 1.0126 | 0.9701, 1.0552 | -0.9453 | -4.1267, 2.2362 | | LYMPH# (x 10³/μL) | 70 | 0.2–8.2 | 0.9956 | 0.9740 | 0.9392, 1.0087 | 0.04297 | -0.0077, 0.0936 | | LYMPH% | 70 | 1.2–47.5 | 0.9951 | 0.9958 | 0.9717, 1.0200 | 0.02710 | -0.4449, 0.4991 | | OTHER WBC# (x 10³/μL) | 61 | 0.1–2.5 | 0.9098 | 0.9164 | 0.6813, 1.1515 | 0.0648 | -0.0962, 0.2258 | | OTHER WBC % | 62 | 1.2–30.7 | 0.9129 | 1.0580 | 0.9417 to 1.1743 | -0.4484 | -1.5124 to 0.6156 | Conclusion: The results from this study demonstrated that the new version of software is comparable to the prior version of software. Appropriate suppression of results is maintained with the new version of software. # 2. Sample challenge study A prospective study was performed to confirm the appropriate suppression of results with v1.14 software for samples associated with interferences, artifacts, or associated conditions for which results from certain parameters should be suppressed in a CLIA Waived environment. In this study, 173 pre-selected, de-identified leftover K2EDTA venous whole K210346 - Page 7 of 11 {7} blood (normal and abnormal) samples were tested on XW-100 CLIA Waived analyzers using software versions v1.03 (K172604 clearance), v1.13 (on-market), or v1.14 (candidate device) by trained operators. The samples were also run on the Sysmex XN-10 analyzer (K112605). Blood smear analysis and WBC differential counts were performed on all samples as a confirmatory method by qualified examiners in accordance with CLSI H20-A2 – Reference Leukocyte (WBC) Differential Count (Proportional) and Evaluation of Instrumental Methods; Approved Standard – Second Edition. Samples with interfering substances, rare cells, pathologies, or morphologies with emphasis on samples with known conditions to affect RBC indices were specifically targeted for challenge samples. Samples with parameter flags (Alert High and Alert Low) were also evaluated in the study. A list of abnormalities targeted for challenge samples is provided below. | Challenge Samples | | --- | | Cold Agglutinins | | Fragmented RBC’s | | High Lipids | | High WBC Count | | Hyperglycemia | | Hypernatremia | | Hypochromic Anemia | | Hyponatremia | | Immunoglobulin | | Immunosuppressive Drugs | | In Vivo Hemolysis | | Large and Giant Platelets | | Microcytes | | Microorganisms (bacterial aggregates, parasites, fungi) | | Nucleated RBCs | | Platelet Agglutination | | Platelet Aggregates | | Warm Agglutinins | | Other (primarily High MCV, Immature Granulocytes, Atypical Lymphocytes) | Deming linear regression analysis was performed on all unsuppressed results for all reported parameters. The results of this regression analysis are shown in the table below. K210346 - Page 8 of 11 {8} | Sample Challenge Study Software version v1.03 (K172604) compared to v1.14 (candidate) | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | Measurand | N | Result Range (N=173) | Correlation Coefficient | Slope | 95% (CI) | Intercept | 95% (CI) | | WBC (x 10³/μL) | 87 | 3.0–45.9 | 0.9995 | 1.0070 | 0.9922, 1.0218 | -0.02695 | -0.1549, 0.1010 | | RBC (x 10⁶/μL) | 133 | 2.37–5.71 | 0.9985 | 1.0116 | 1.0013, 1.0218 | -0.0118 | -0.0497, 0.0261 | | HGB (g/dL) | 91 | 10.2–18.3 | 0.9971 | 1.0053 | 0.9900, 1.0207 | -0.1249 | -0.3184, 0.0686 | | HCT (%) | 112 | 25.0–51.1 | 0.9982 | 1.0115 | 1.0001, 1.0229 | 0.0794 | -0.3247, 0.4834 | | PLT (x 10³/μL) | 102 | 104–804 | 0.9977 | 1.0045 | 0.9913, 1.0178 | -0.7271 | -4.4504, 2.9962 | | MCV (fL) | 112 | 71.3–111.0 | 0.9990 | 1.0188 | 1.0098, 1.0278 | -1.3206 | -2.1296, -0.5116 | | NEUT# (x 10³/μL) | 87 | 1.7–40.5 | 0.9993 | 1.0156 | 1.0015, 1.0296 | -0.0528 | -0.1457, 0.0400 | | NEUT% | 87 | 29.6–97.1 | 0.9926 | 0.9761 | 0.9524, 0.9999 | 2.0669 | 0.2159, 3.9179 | | LYMPH# (x 10³/μL) | 87 | 0.2–9.4 | 0.9974 | 1.0167 | 0.9876, 1.0459 | -0.0029 | -0.0472, 0.0412 | | LYMPH% | 87 | 2.0–66.7 | 0.9968 | 0.9938 | 0.9764, 1.0111 | 0.2754 | -0.0460, 0.5968 | | OTHER WBC# (x 10³/μL) | 79 | 0.0–2.2 | 0.9188 | 0.9753 | 0.8721, 1.0785 | -0.0272 | -0.0975, 0.0430 | | OTHER WBC % | 78 | 0.7–29.3 | 0.9523 | 0.9545 | 0.8906, 1.0184 | -0.1008 | -0.6642, 0.4626 | K210346 - Page 9 of 11 {9} | Sample Challenge Study Software version v1.13 (on-market) compared to v1.14 (candidate) | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | Measurand | N | Result Range (N=173) | Correlation Coefficient | Slope | 95% (CI) | Intercept | 95% (CI) | | WBC (x 10³/μL) | 82 | 3.2–54.4 | 0.9997 | 1.0074 | 0.9979, 1.0169 | -0.0206 | -0.1118, 0.0705 | | RBC (x 10⁶/μL) | 131 | 2.33–5.73 | 0.9988 | 1.0320 | 1.0228, 1.0412 | -0.0655 | -0.1006, -0.0303 | | HGB (g/dL) | 83 | 10.0–17.9 | 0.9980 | 1.0228 | 1.0104, 1.0351 | -0.0069 | -0.1708, 0.1570 | | HCT (%) | 100 | 25.3–50.7 | 0.9982 | 1.0353 | 1.0210, 1.0496 | -0.4120 | -0.9451, 0.1211 | | PLT (x 10³/μL) | 102 | 116–821 | 0.9968 | 1.0111 | 0.9896, 1.0326 | 0.7837 | -4.7345, 6.3020 | | MCV (fL) | 100 | 72.2–110.1 | 0.9983 | 1.0485 | 1.0373, 1.0598 | -3.6551 | -4.6946, -2.6155 | | NEUT# (x 10³/μL) | 82 | 1.6–48.5 | 0.9995 | 0.9986 | 0.9836, 1.0135 | 0.0603 | -0.0437, 0.1643 | | NEUT% | 82 | 29.3–96.9 | 0.9911 | 0.9636 | 0.9372, 0.9899 | 3.0143 | 0.9764, 5.0522 | | LYMPH# (x 10³/μL) | 82 | 0.2–9.4 | 0.9961 | 1.0062 | 0.9509, 1.0615 | -0.0089 | -0.0904, 0.0726 | | LYMPH% | 82 | 2.1–66.9 | 0.9963 | 0.9945 | 0.9697, 1.0192 | -0.0445 | -0.4965, 0.4074 | | OTHER WBC# (x 10³/μL) | 73 | 0.0–2.1 | 0.9186 | 1.0611 | 0.9086, 1.2136 | -0.0526 | -0.1476, 0.0424 | | OTHER WBC % | 75 | 0.4–28.9 | 0.9462 | 0.9654 | 0.8319, 1.0990 | 0.0311 | -0.9067, 0.9690 | The results of this study demonstrated that appropriate and expected suppression of results occurred for the XW-100 CLIA Waived analyzer with v1.14 software for challenge samples, as confirmed by blood smear analysis with WBC differential counts and results from the XN-10 analyzer. Regression analysis indicated that the v1.14 software performed similar to the previously cleared software (v1.03, K172604 and CW170012) and software currently on the market (v1.13). K210346 - Page 10 of 11 {10} Conclusion: The software change does not impact the appropriate suppression of results for challenging samples and unaffected parameters maintain similar performance to the previously cleared XW-100 CLIA Waived analyzer. ## VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. ## IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K210346 - Page 11 of 11
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...