CELL-DYN Emerald 22 AL System
Device Facts
| Record ID | K190294 |
|---|---|
| Device Name | CELL-DYN Emerald 22 AL System |
| Applicant | Abbott Laboratories |
| Product Code | GKZ · Hematology |
| Decision Date | Mar 15, 2019 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 864.5220 |
| Device Class | Class 2 |
Indications for Use
The CELL-DYN Emerald 22 System is a quantitative, automated hematology analyzer and leukocyte differential counter for in vitro diagnostic use in clinical laboratories.
Device Story
CELL-DYN Emerald 22 System; automated hematology analyzer; performs quantitative blood analysis and leukocyte differential counting. Modification adds sample autoloader for automated sample processing; includes rack transfer, mixing, and unloading. Modifications involve open/close tube operation, sampling module, operating system, and system board. Used in clinical laboratories; operated by laboratory personnel. Output provides hematological data for clinical decision-making regarding patient blood health.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical design verification and analytical performance testing, including background, carryover, linearity, short/long-term imprecision, method comparison, blood tube type study, mode-to-mode equivalency, and clogging rate analysis.
Technological Characteristics
Bench-top analyzer; electrical impedance, absorption spectrophotometry (555nm LED), and optical flow cytometry (455nm LED). Autoloader module for 50-tube capacity; trocar needle for cap piercing. CPU: Phytec Freescale i.MX27, ARM926EJ, 400MHz+. Connectivity: USB ports for external storage/peripherals, LIS integration. Software: OSE Delta operating system. Dimensions: 16.1" x 19.7" x 17.0".
Indications for Use
Indicated for use in clinical laboratories to identify patients with hematologic parameters within and outside of established reference ranges using K2EDTA anticoagulated whole blood.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Predicate Devices
- CELL-DYN Emerald 22 system (K110381)