← Product Code [GKZ](/submissions/HE/subpart-f%E2%80%94automated-and-semi-automated-hematology-devices/GKZ) · K170353

# ABX MICROS ES 60 OT and ABX MICROS ES 60 CT (K170353)

_HORIBA ABX SAS · GKZ · Nov 1, 2017 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/HE/subpart-f%E2%80%94automated-and-semi-automated-hematology-devices/GKZ/K170353

## Device Facts

- **Applicant:** HORIBA ABX SAS
- **Product Code:** [GKZ](/submissions/HE/subpart-f%E2%80%94automated-and-semi-automated-hematology-devices/GKZ.md)
- **Decision Date:** Nov 1, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5220
- **Device Class:** Class 2
- **Review Panel:** Hematology
- **Attributes:** Pediatric

## Indications for Use

The ABX MICROS ES 60 (OT and CT models) is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in clinical laboratories to identify and enumerate the following parameters: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, LYM%, LYM#, MON%, MON#, GRA%, GRA#, in K2EDTA and K3EDTA anticoagulated venous whole blood samples of adult patients and pediatric patients ≥ 1 month of age.

## Device Story

ABX MICROS ES 60 is an automated hematology analyzer; performs quantitative analysis of venous whole blood samples (K2EDTA/K3EDTA). Inputs: venous blood samples; processed via impedance (RBC, PLT, WBC, DIFF) and spectrophotometry (HGB). Outputs: CBC and leukocyte differential parameters (WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, LYM, MON, GRA). Used in clinical laboratories; operated by trained personnel. Manual sample identification via barcode scanner. Provides quantitative data to clinicians for hematological assessment; aids in patient diagnosis and monitoring. Available in Open Tube (OT) and Closed Tube (CT) models; CT features cap-piercing mechanism. Benefits include automated, efficient cell counting and differential analysis for adult and pediatric (≥ 1 month) populations.

## Clinical Evidence

Clinical evaluation included 137 venous whole blood specimens from pediatric patients (1 month to 21 years) across three sites. Samples analyzed in duplicate on ABX MICROS ES 60 and compared against Sysmex XN-10. Bias was estimated at medical decision points; acceptance criteria met for all measurands except RDW (due to calculation differences). Reference interval verification performed on 69 normal pediatric samples confirmed results within published ranges.

## Technological Characteristics

Automated hematology analyzer; dimensions 43x36x36 cm. Measurement principles: Impedance (RBC, PLT, WBC, DIFF) and Spectrophotometry (HGB). Sample volume: 10 µL. Throughput: 60 (OT) or 50 (CT) samples/hour. Connectivity: manual barcode scanner. Software: embedded firmware (no changes from K141161).

## Regulatory Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

## Predicate Devices

- Sysmex XN Series, XN-10 ([K112605](/device/K112605.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 01, 2017

HORIBA ABX SAS Caroline Ferrer Regulatory Affairs Manager Parc Euromedecine Rue du Caducee BP 7290 34184 Montpellier Cedex 4 - France

Re: K170353

Trade/Device Name: ABX MICROS ES 60 OT and ABX MICROS ES 60 CT Regulation Number: 21 CFR 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: Class II Product Code: GKZ Dated: September 28, 2017 Received: October 2, 2017

Dear Caroline Ferrer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Leonthena R. Carrington -S

Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known) k170353

### Device Name

ABX MICROS ES 60 OT (Open Tube model) ABX MICROS ES 60 CT (Close Tube model)

### Indications for Use (Describe)

The ABX MICROS ES 60 (OT and CT models) is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in clinical laboratories to identify and enumerate the following parameters: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, LYM%, LYM#, MON%, MON#, GRA%, GRA#, in K2EDTA and K3EDTA anticoagulated venous whole blood samples of adult patients and pediatric patients ≥ 1 month of age.

| Type of Use (Select one or both, as applicable) | <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)             </span> <span>                 ☐ Over-The-Counter Use (21 CFR 801 Subpart C)             </span> |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
*PRAStaff@fda.hhs.gov*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### Premarket Notification [510(k)] Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is : k170353

1.0 Submitted bv : Company: HORIBA ABX SAS HORIBA MEDICAL Parc Euromédecine Rue du Caducée - BP 7290 34184 Montpellier cedex 4 FRANCE

Contact Person: Caroline FERRER Telephone (France): +33 467 141843 Email: caroline.ferrer@horiba.com

Date Submitted : 2.0

28th September 2017

#### 3.0 Device Name and Classification : ABX MICROS ES 60 Trade/Proprietary Name:

### Classification:

Device: Counter, differential cell Panel: 81 Hematology Regulation number: 864.5220 Product Code: GKZ Class: 2

### 4.0 System description : 4.1 Device Description

The ABX MICROS ES 60 is an automated hematology analyzer developed by HORIBA Medical and cleared under k141161.

No modification (hardware or software) to the ABX MICROS ES 60 cleared device (k141161) was made to support the pediatric claim addition.

Performance verification has been done to support the pediatric claim addition.

{4}------------------------------------------------

The corresponding User Instructions for Use was updated to include this additional claim to the intended use.

### 4.2 Principles of Operation

The ABX MICROS ES 60 performs hematology analyses using the following methods:

- RBC / PLT: Impedance ●
- WBC: Impedance ●
- HGB: Spectrophotometry
- DIFF: Impedance ●

The Principles of Operation are unchanged.

### 4.3 Modes of Operation

Two models of ABX MICROS ES 60 are available.

The Closed Tube (CT) instrument: it has a cap-piercing mechanism. The blood collection tube ● is placed directly in the analyzer without removing the cap.

. The Open Tube (OT) instrument: the cap form the blood collection tube must be removed before analyzing the sample.

### 4.4 Specimen Identification

Specimen identification is by manual sample identification with the use of a hand held barcode scanner.

### 4.5 Calibration

Calibration is a procedure that is performed during specific situations such as installation, maintenance or service intervention. It is performed by a HORIBA ABX SAS representative. It ensures that the precision and accuracy of the analyzer are acceptable, so that accurate measurements are performed by the analyzer.

There is no specific calibrator for pediatric addition.

{5}------------------------------------------------

### 4.6 Quality Control

Quality control allows the user to monitor a set of analyses based on known sample values and ranges over a period of several months. Statistical computation performed on these populations allows the extraction of qualitative information related to the stability of the instrument.

The manufacturer's instructions are to be followed for material and frequency of quality control analysis.

There is no specific control for pediatric addition.

### 4.7 Software

FDA has reviewed HORIBA ABX SAS's Hazard Analysis and Software Development processes for this line of product types during the 510(k) file k141161 review.

No change was made to the Hazard Analysis and Software Development processes relative to the pediatric use addition.

#### 5.0 Intended use

### 5.1 Indications for Use :

The ABX MICROS ES 60 (OT and CT models) is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in clinical laboratories to identify and enumerate the following parameters: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, LYM%, LYM#, MON%, MON#, GRA%, GRA#, in K2EDTA and K3EDTA anticoagulated venous whole blood samples of adult patients and pediatric patients > 1 month of age.

### 5.2 Special Conditions for Use Statements:

For prescription use only

{6}------------------------------------------------

#### 6.0 Comparison with the original device :

# Similarities

| Item                                      | ABX MICROS ES 60<br>Device with additional pediatric<br>intended use                                              | ABX MICROS ES 60 (K141161)<br>Previous Device |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| Manufacturer                              | HORIBA ABX SAS                                                                                                    | HORIBA ABX SAS                                |
| Type of product                           | Automated blood cell counter                                                                                      | Automated blood cell counter                  |
| Parameters                                | Complete Blood Cell Count (CBC)<br>WBC, RBC, HGB, HCT, MCV,<br>MCH, MCHC, RDW, PLT, MPV                           | Complete Blood Cell Count (CBC)<br>Same       |
|                                           | Leukocyte Differential (DIFF):<br>LYM (# and %), MON (# and %),<br>GRA (# and %)                                  | Leukocyte Differential (DIFF):<br>Same        |
| Sample types                              | K2 and K3EDTA anti-coagulated<br>whole blood                                                                      | Same                                          |
| Principles of<br>measurements             | RBC / PLT: Impedance<br>WBC / DIFF: Impedance<br>HGB: Spectrophotometry<br>RDW, MCV, MCH, MCHC:<br>Calculation    | Same                                          |
| Reagents                                  | ABX Minidil LMG<br>ABX Minilyse LMG or ABX<br>Alphalyse 360<br>ABX Miniclean<br>ABX Minipack LMG<br>ABX Minoclair | Same                                          |
| Calibrators                               | ABX Minocal                                                                                                       | Same                                          |
| Controls                                  | ABX Minotrol 16 (3 levels)                                                                                        | Same                                          |
| Throughput                                | OT / CT models: 60 / 50 samples per<br>hour                                                                       | Same                                          |
| Dimensions<br>(Height x Width x<br>Depth) | 43 cm x 36 cm x 36 cm                                                                                             | Same                                          |

{7}------------------------------------------------

## Differences

| Item         | ABX MICROS ES 60<br>Device with additional pediatric<br>intended use                                                                                                                                                                                                                                                                                                                                                                                                    | ABX MICROS ES 60 (K141161)<br>Previous Device                                                      |
|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Intended Use | The ABX MICROS ES 60 (OT and CT<br>models) is a quantitative multi-<br>parameter, automated hematology<br>analyzer for in vitro diagnostic use in<br>clinical laboratories to identify and<br>enumerate the following parameters:<br>WBC, RBC, HGB, HCT, MCV, MCH,<br>MCHC, RDW, PLT, MPV, LYM%,<br>LYM#, MON%, MON#, GRA%,<br>GRA#, in K2EDTA and K3EDTA<br>anticoagulated venous whole blood<br>samples of adult patients and pediatric<br>patients ≥ 1 month of age. | Same except:<br>- sample type was venous only<br>- intended population was adult<br>patients only. |

{8}------------------------------------------------

#### 7.0 Substantial Equivalence Information :

The following tables show the similarities and differences between the candidate device and its predicate device identified below.

### 7.1 Predicate Device Name and 510(k) number:

For the validation of the pediatric use on the ABX MICROS ES 60 the following predicate device was used during the clinical evaluation:

| Candidate device | Predicate device        | Manufacturer | Predicate 510(k) number |
|------------------|-------------------------|--------------|-------------------------|
| ABX MICROS ES 60 | Sysmex XN Series, XN-10 | SYSMEX       | K112605                 |

# 7.2 Comparison with predicate Device : Similarities

| Item                          | ABX MICROS ES 60 (k141161)<br>Candidate device                                                                                                                                                                                                                                                                                                                                                                                     | Sysmex XN Series, XN-10<br>(k112605)<br>Predicate Device                                                                                                                                                                                                                                                                                                                                                                              |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                  | HORIBA ABX SAS                                                                                                                                                                                                                                                                                                                                                                                                                     | Sysmex                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Type of product               | Automated blood cell counter                                                                                                                                                                                                                                                                                                                                                                                                       | Automated blood cell counter                                                                                                                                                                                                                                                                                                                                                                                                          |
| Parameters                    | Complete Blood Cell Count (CBC)<br>WBC, RBC, HGB, HCT, MCV,<br>MCH, MCHC, RDW, PLT, MPV<br>Leukocyte Differential (DIFF):<br>LYM (% and #)                                                                                                                                                                                                                                                                                         | Complete Blood Cell Count (CBC)<br>WBC, RBC, HGB, HCT, MCV,<br>MCH, MCHC, RDW-CV, PLT,<br>MPV<br>Leukocyte Differential (DIFF):<br>LYM (# and %),                                                                                                                                                                                                                                                                                     |
| Sample types                  | K2 and K3EDTA anti-coagulated<br>venous samples                                                                                                                                                                                                                                                                                                                                                                                    | K2 and K3EDTA anti-coagulated<br>whole blood                                                                                                                                                                                                                                                                                                                                                                                          |
| Principles of<br>measurements | RBC / PLT: Impedance<br>HGB: Spectrophotometry                                                                                                                                                                                                                                                                                                                                                                                     | RBC: Impedance (Sheath Flow)<br>HGB: Spectrophotometry                                                                                                                                                                                                                                                                                                                                                                                |
| Item                          | ABX MICROS ES 60 (k141161) Candidate device                                                                                                                                                                                                                                                                                                                                                                                        | Sysmex XN Series, XN-10 (k112605) Predicate Device                                                                                                                                                                                                                                                                                                                                                                                    |
| Intended Use                  | The ABX MICROS ES 60 (OT and CT models) is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in clinical laboratories to identify and enumerate the following parameters: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, LYM%, LYM#, MON%, MON#, GRA%, GRA#, in K2EDTA and K3EDTA anticoagulated venous whole blood samples of adult patients and pediatric patients ≥ 1 month of age. | The Sysmex® XN-10 and XN-20 modules are quantitative multi-parameter automated hematology analyzers intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XN-Series modules classify and enumerate the following parameters: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, NRBC%/#, IG%/#, , MPV, RET%/#, IRF, IPF, RET-He |
| Parameters                    | Leukocyte Differential (DIFF):<br>MON (% and #)<br>GRA (% and #)<br><br>RDW-SD, NRBC%/#, IG%/#, , MPV, RET%/#, IRF, IPF, RET-He: parameters not available                                                                                                                                                                                                                                                                          | Leukocyte Differential (DIFF):<br>NEUT%/#<br>ΜΟΝΟ%/#,<br>EO%/#,<br>BASO%/#<br><br>RDW-SD, NRBC%/#, IG%/#, , MPV, RET%/#, IRF, IPF, RET-He                                                                                                                                                                                                                                                                                             |
| Minimum<br>Sample<br>Volume   | 50 µL                                                                                                                                                                                                                                                                                                                                                                                                                              | Sampler Mode – 88 µL<br>Manual (Closed Cap) Mode – 88 µL<br>Manual (Open Cap) Mode – 88 µL<br>Dilution Mode – 70 µL                                                                                                                                                                                                                                                                                                                   |
| Specimen<br>sample volume     | 10 µL                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Dimensions                    | OT / CT models:<br>43 x 36 x 36 cm                                                                                                                                                                                                                                                                                                                                                                                                 | 85.5 cm x 64.5 cm x 75.5 cm                                                                                                                                                                                                                                                                                                                                                                                                           |
| Weight                        | OT model: 14 kg<br>CT model: 17 kg                                                                                                                                                                                                                                                                                                                                                                                                 | 57 kg                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Throughput                    | OT / CT models: 60 / 50 samples per hour                                                                                                                                                                                                                                                                                                                                                                                           | 100 samples/hour maximum depending on mode used.                                                                                                                                                                                                                                                                                                                                                                                      |
| Modes of<br>Operation         | Open Tube Mode<br>Closed Tube Mode                                                                                                                                                                                                                                                                                                                                                                                                 | Sampler Analysis Mode<br>Manual Closed Analysis Mode<br>Pre-dilute Analysis Mode Dilute sample 1:7                                                                                                                                                                                                                                                                                                                                    |
| Principles of Measurement     |                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| WBC, WBC<br>Differential      | Impedance                                                                                                                                                                                                                                                                                                                                                                                                                          | WBC: Radio-frequency (RF) / Direct-current (DC) Detection Method Flow                                                                                                                                                                                                                                                                                                                                                                 |
| Item                          | ABX MICROS ES 60 (k141161)<br>Candidate device                                                                                                                                                                                                                                                                                                                                                                                     | Sysmex XN Series, XN-10 (k112605)<br>Predicate Device                                                                                                                                                                                                                                                                                                                                                                                 |
| (LYM, MON,<br>GRA)            |                                                                                                                                                                                                                                                                                                                                                                                                                                    | Cytometry Methods using a Semiconductor Laser<br>Particle characterization and identification is<br>based on detection of forward scatter,<br>fluorescence and adaptive cluster analysis.<br>The XN-Series analyzers automatically<br>classify cells from whole blood                                                                                                                                                                 |
| Calibrator                    | ABX Minocal                                                                                                                                                                                                                                                                                                                                                                                                                        | XN CAL (XN-10/XN-20 Calibrator)<br>XN CAL PF (Platelet F Calibrator)                                                                                                                                                                                                                                                                                                                                                                  |
| Quality<br>Controls           | ABX Minotrol 16<br>(3 levels)                                                                                                                                                                                                                                                                                                                                                                                                      | XN-Check - 3 Levels                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Reagents                      | ABX Minidil LMG (Diluent)<br>ABX Minilyse LMG or ABX Alphalyse<br>360 (Lyse)<br>ABX Miniclean (Cleaner)<br>ABX Minoclair (Concentrated cleaning<br>reagent)<br>ABX Minipack LMG (Reagent Pack)                                                                                                                                                                                                                                     | CELLPACKTM DCL (Diluent)<br>CELLPACKTM DFL (Diluent)<br>LYSERCELL WNR (Lyse) LYSERCELL<br>WDF (Lyse) LYSERCELL WPC (Lyse)<br>FLUOROCELL WNR (Stain)<br>FLUOROCELL WDF (Stain)<br>FLUOROCELL RET (Stain)<br>FLUOROCELL PLT (Stain)<br>FLUOROCELL WPC (Stain)                                                                                                                                                                           |

{9}------------------------------------------------

# 7.3 Comparison with predicate Device : Differences

{10}------------------------------------------------

#### 8.0 Special Control/Guidance Document Referenced :

### 8.1 Standards Followed

- CLSI EP09-A3: Measurement Procedure Comparison and Bias Estimation Using ● Patient Samples - 2013
- CLSI EP28-A3: Defining, Establishing and Verifying Reference Intervals in the ● Clinical Laboratory - 2008

### 8.2 FDA Guidances Followed

- Guidance for Industry and FDA Staff : Format for Traditional and Abbreviated 510(k)s ● - 2005
- Final Guidance for Industry and FDA: Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells - 2001

{11}------------------------------------------------

#### 9.0 Summary of Performance Data :

### 9.1 Comparability with Predicate Device

A total of 137 venous whole blood specimens (collected in K2EDTA) from pediatric patients between 1 month and 21 years of age were collected and analyzed at two test sites in the US and one test site in France. Different instruments and operators were used at each site.

Each of the samples was analyzed in duplicate on the ABX MICROS ES 60 and once on the predicate XN Series, XN-10 from SYSMEX.

Bias was estimated at various data points, particularly around medical decision points, for each parameter. Bias was estimated for one replicate. Acceptance criteria were met for all measurands at all levels, except for RDW, due to the difference of calculation methods on the two analyzers. These findings support the claim that the ABX MICROS ES 60 candidate device and the XN Series, XN-10 predicate device are substantially equivalent, and demonstrate acceptable levels of bias for pediatric samples.

### 9.2 Reference Interval

69 (37 female and 32 male) normal pediatric samples (whole blood samples collected in K2EDTA) were analyzed, when possible in duplicate, on the ABX MICROS ES 60 at one test site in the US. Results were compared to a pediatric reference range obtained from several US hospitals published ranges. Results were contained in the pediatric reference range, confirming the transference of the pediatric reference range below for the ABX MICROS ES 60:

| Combined Pediatric         |           | MALE         |              |              | FEMALE       |              |              |
|----------------------------|-----------|--------------|--------------|--------------|--------------|--------------|--------------|
| Reference Range parameters |           | 1 m - 2 y    | >2y - 12 y   | >12 y - 21 y | 1 m - 2 y    | >2y - 12 y   | >12 y - 21 y |
| WBC                        | (10³/mm³) | 5.0 - 20.0   | 3.4 - 17.0   | 3.5 - 13.0   | 5.0 - 20.0   | 3.4 - 17.0   | 3.5 - 13.0   |
| RBC                        | (10⁶/mm³) | 2.7 - 6.0    | 3.9 - 5.5    | 4.0 - 5.7    | 2.7 - 6.0    | 3.8 - 5.5    | 3.7 - 5.2    |
| HGB                        | g/dL      | 8.9 - 18.0   | 10.2 - 15.5  | 11.0 - 18.0  | 9.0 - 18.0   | 10.2 - 15.5  | 10.6 - 16.0  |
| HCT                        | %         | 28.0 - 55.0  | 31.0 - 45.0  | 33.9 - 49.0  | 27.7 - 55.0  | 31.0 - 45.0  | 32.9 - 49.0  |
| MCV                        | μm³       | 70 - 123     | 70 - 95      | 77 - 102     | 70 - 123     | 70 - 95      | 76.9 - 102   |
| RDW                        | %         | 11.0 - 16.5  | 11.0 - 16.0  | 11.0 - 15.6  | 11.0 - 16.5  | 11.0 - 16.0  | 11.0 - 15.5  |
| MCH                        | pg        | 22.7 - 36.0  | 23.0 - 33.0  | 25.0 - 35.0  | 24.4 - 35    | 23.7 - 33.0  | 24.8 - 35.0  |
| MCHC                       | %         | 28 - 37      | 30 - 37      | 31 - 37      | 28 - 36      | 30 - 37      | 31 - 37      |
| PLT                        | (10³/mm³) | 140 - 562    | 140 - 450    | 140 - 450    | 140 - 597    | 140 - 450    | 140 - 450    |
| MPV                        | μm³       | 6.5 - 12.4   | 6.5 - 12.4   | 6.5 - 12.4   | 6.5 - 12.4   | 6.5 - 12.4   | 6.5 - 12.4   |
| LYM#                       | (10³/mm³) | 1.50 - 16.50 | 0.97 - 10.50 | 0.85 - 6.50  | 1.50 - 16.50 | 1.16 - 10.50 | 0.90 - 6.50  |
| LYM%                       | %         | 26 - 86      | 15 - 75      | 12 - 53      | 27 - 87      | 17 - 75      | 13 - 50      |
| MON#                       | (10³/mm³) | 0.00 - 1.40  | 0.00 - 1.00  | 0.00 - 0.90  | 0.00 - 1.40  | 0.00 - 1.00  | 0.00 - 0.90  |
| MON%                       | %         | 0 - 14       | 0 - 12       | 0 - 13       | 0 - 13       | 0 - 12       | 0 - 13       |
| GRA#                       | (10³/mm³) | 0.84 - 10.1  | 1.50 - 9.45  | 1.50 - 8.80  | 1.01 - 10.45 | 1.50 - 9.45  | 1.50 - 8.80  |
| GRA%                       | %         | 10 - 80      | 12 - 85      | 31 - 87      | 9 - 86       | 12 - 81      | 31 - 87      |

These intervals are given in the labeling of the ABX MICROS ES 60.

However, expected values will vary with sample population and/or geographical location. HORIBA highly recommends that each laboratory establishes its own normal ranges based upon its local population.

{12}------------------------------------------------

### 10.0 Proposed Labeling :

The labeling is written as per the recommendations given in standard EN18113-2. It takes into account the requirements of 21 CFR Part 809.10.

### 11.0 Conclusion :

As per 21CFR Part §807.92(b)(3), the nonclinical tests demonstrate that the ABX MICROS ES 60 device is as safe, as effective, and performs as well as or better than the predicate device for pediatric population (≥1 month of age).

The submitted information in this premarket notification is complete and supports a substantial equivalence decision on pediatric use addition.

---

**Source:** [https://fda.innolitics.com/submissions/HE/subpart-f%E2%80%94automated-and-semi-automated-hematology-devices/GKZ/K170353](https://fda.innolitics.com/submissions/HE/subpart-f%E2%80%94automated-and-semi-automated-hematology-devices/GKZ/K170353)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
