RET-HE PARAMETER ON THE SYSMEX MODEL XE-2100 AUTOMATED HEMATOLOGY ANALYZER

K050589 · Sysmex America, Inc. · GKZ · May 4, 2005 · Hematology

Device Facts

Record IDK050589
Device NameRET-HE PARAMETER ON THE SYSMEX MODEL XE-2100 AUTOMATED HEMATOLOGY ANALYZER
ApplicantSysmex America, Inc.
Product CodeGKZ · Hematology
Decision DateMay 4, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5220
Device ClassClass 2

Indications for Use

The RET-He parameter on the Sysmex® XE-2100, Automated Hematology Analyzer, determines the hemoglobin of reticulocytes for in vitro diagnostic use in clinical laboratories.

Device Story

The RET-He parameter is an analytical feature for the Sysmex XE-2100 Automated Hematology Analyzer. It processes peripheral blood samples collected in EDTA anticoagulant. The device utilizes reticulocyte forward scattered light signals from the analyzer's reticulocyte measurement channel. These signals are processed via a proprietary Sysmex calculation equation to derive the hemoglobin content of reticulocytes. The system requires the XE-Pro and RET Master software modules to function. Results are provided to clinical laboratory professionals to assist in hematological assessment. The device is intended for use in clinical laboratory settings.

Clinical Evidence

Method comparison study performed on 319 clinical samples, including those with high WBC counts, lipemia, bilirubin, platelet clumps, NRBCs, and various hematological abnormalities. The RET-He parameter showed excellent correlation (r=0.98) with the predicate CHr parameter. Analytical precision (total imprecision) ranged from 1.2% to 4.26% CV, meeting the ≤5% specification. Linearity was confirmed for RBC (r=1.0) and Reticulocytes (r=0.9988). Carryover was 0%.

Technological Characteristics

The device is a software-based parameter extension for the existing Sysmex XE-2100 Automated Hematology Analyzer. It utilizes forward scattered light signals from the reticulocyte measurement channel. The system requires XE-Pro and RET Master software modules. It is designed for in vitro diagnostic use in clinical laboratories.

Indications for Use

Indicated for in vitro diagnostic use in clinical laboratories to determine the hemoglobin content of reticulocytes in peripheral blood samples.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K050589 B. Purpose for Submission: Sysmex is adding the RET-He parameter to the XE-2100 Automated Hematology Analyzer (K992875). C. Measurand: The reticulocyte hemoglobin equivalent (RET-He) parameter determines the hemoglobin of reticulocytes. D. Type of Test: The RET-He parameter is derived using the reticulocyte forward scattered light signals from the reticulocyte measurement channel and a proprietary Sysmex calculation equation. E. Applicant: Sysmex America, Inc. F. Proprietary and Established Names: RET-He parameter on the Sysmex XE_2100, Automated Hematology Analyzer G. Regulatory Information: 1. Regulation section: 21 CFR 864.5220, Automated differential cell counter 2. Classification: Class II 3. Product code: GKZ {1} 4. Panel: Hematology (81) ## H. Intended Use: 1. Intended use(s): The RET-He parameter on the Sysmex XE-2100, Automated Hematology Analyzer, determines the hemoglobin of reticulocytes for in vitro diagnostic use in clinical laboratories. 2. Indication(s) for use: N/A 3. Special conditions for use statement(s): N/A 4. Special instrument requirements: Sysmex XE-Pro and RET Master software are required ## I. Device Description: The XE-2100 is an automated hematology analyzer that uses automated fluorescent flow cytometry which in the reticulocyte channel, using a polymethine dye, also measures the mean value of the forward light scatter histogram of mature red blood cells and reticulocytes. ## J. Substantial Equivalence Information: 1. Predicate device name(s): Bayer Advia 120 Hematology system CHr parameter 2. Predicate 510(k) number(s): K971998 {2} 3 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | The RET-He parameter on the XE-2100 determines the hemoglobin of reticulocytes. | The CHr parameter on the Advia 120 determines the hemoglobin of reticulocytes. | | Anticoagulant | EDTA | Same | | Specimen Type | Peripheral Blood | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Methodology | The reticulocyte parameters are derived using the reticulocyte measurement channel and a proprietary Sysmex calculation equation. | The reticulocyte parameters are derived through a combination of laser light scatter and absorption of a nucleic acid dye. | K. Standard/Guidance Document Referenced (if applicable): Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA, 12-4-2001. L. Test Principle: The XE-2100 is an automated hematology analyzer that uses automated fluorescent flow cytometry which in the reticulocyte channel, using a polymethine dye, also measures the mean value of the forward light scatter histogram of mature red blood cells and reticulocytes. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Data was assessed according to EP5 recommendation. Total imprecision {3} ranged between 1.2 and 4.26% CV% on the RET-He parameter and meets manufacturer specifications of ≤5%. b. Linearity/assay reportable range: The data showed linearity for RBC (r=1.0) across the tested range of 0.29-7.48 (x10⁶/μL) and for Reticulocyte (r=0.9988) across the tested range of 0.0135 – 0.6770. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Stability: Samples should be stored from 18 to 26°C (room temperature) or 4-8°C (refrigerated). The variation of RET-He [pg] from 0 hour after blood collection to 24 hours should be within +/-8% (RET#≥0.010x10⁶/μL). d. Detection limit: N/A e. Analytical specificity: N/A f. Assay cut-off: N/A 2. Comparison studies: a. Method comparison with predicate device: Method comparison data was collected on RET-He parameter on the XE-2100 and compared to the CHr parameter on the Bayer Advia 120. The study included 319 samples that included samples with high WBC counts, lipemia, high bilirubin, platelet clumps, NRBC, atypical lymphocytes, immature granulocytes, blast forms, left shift, atypical/abnormal lymphocytes, iron deficiency, thalassemia, etc. Overall, the XE RET-He parameter showed excellent correlation with the CHr parameter. The correlation (r=0.98) concurred with other evaluation studies ranging from r-values of 0.94-0.96. These conclusions point to the fact that the RET-He parameter on the XE-2100 showed excellent correlation the CHr parameter, the predicate method. b. Matrix comparison: N/A {4} 5 3. Clinical studies: a. Clinical Sensitivity: N/A b. Clinical specificity: N/A c. Other clinical supportive data (when a. and b. are not applicable): Carry over results showed zero percent carryover. 4. Clinical cut-off: N/A 5. Expected values/Reference range: Normal Reference Intervals for RET-He Parameter | RET-He (pg) | Children’s Hospital* | Thomas | Johns Hopkins | | --- | --- | --- | --- | | Mean | 32 | 32.0 | 33.7 | | Std. Dev. | 1.5 | 1.9 | 1.4 | | N | 33 | 122 | 31 | | Range | 29-35 | 28.2-35.7 | 30.8-36.6 | * The age range was 1 to 30 years with a mean age of 16. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. P. Other Supportive Device and Instrument Information: {5} N/A 6
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...