MICROS 60

K030799 · Abx Diagnostics · GKZ · Apr 8, 2003 · Hematology

Device Facts

Record IDK030799
Device NameMICROS 60
ApplicantAbx Diagnostics
Product CodeGKZ · Hematology
Decision DateApr 8, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 864.5220
Device ClassClass 2

Indications for Use

The ABX MICROS 60 is a fully automated (microprocessor controlled) multi-parameter hematology analyzer intended for in in-vitro diagnostic use in the clinical laboratory environment.

Device Story

Benchtop hematology analyzer; processes whole blood specimens; performs complete blood count (CBC) and 3-part differential leukocyte count (LMG). Operates as closed tube/closed system or open tube/open system. Microprocessor-controlled; evaluates 5, 8, 16, or 18 hematology parameters based on configuration. Used in clinical laboratory settings by trained personnel. Output provides quantitative hematology data to assist clinicians in patient diagnosis and monitoring. Benefits include automated, precise, and efficient blood analysis.

Clinical Evidence

Bench testing and inter-procedural correlation study. Precision evaluated per NCCLS EPS-A guidelines. Accuracy compared against Abbott CD 4000; correlation R > 0.95 for all parameters. Linearity ranges established for WBC, RBC, HGB, HCT, and PLT. No significant bias observed.

Technological Characteristics

Benchtop hematology analyzer; microprocessor-controlled. Electrical safety per IEC 1010-1; Electromagnetic Compatibility per EN 61326. Analyzes whole blood specimens for CBC and 3-part differential. Configuration-dependent parameter output (5-18 parameters).

Indications for Use

Indicated for in vitro diagnostic testing of whole blood specimens or blood cell concentrates in clinical laboratory environments to perform complete blood counts (CBC) and 3-part differential leukocyte counts.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a logo for ABX Horiba Diagnostics. The logo is black and white and features the letters "ABX" in a stylized font. Below the letters, the word "HORIBA" is written in a smaller, sans-serif font. Below that, the word "AGNOSTICS" is written in a smaller, sans-serif font. There is a black circle to the left of the letters. # Premarket Notification [510(k)] Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 The assigned 510(k) number is : H030799 Company: ABX Diagnostics Parc Euromédecine Rue du Caducée – BP 7290 34184 Montpellier cedex 4 FRANCE + (33) 4 67 14 73 20 Telephone: + (33) 4 67 14 15 17 Fax: Contact Person: Tim Lawton (tlawton@fr.abx.fr) Date Prepared: 07 March, 2003 #### Device Name: | Trade/Proprietary Names: | ABX MICROS 60 Hematology Analyzer | |--------------------------|-------------------------------------------------------------------------------------------| | Common or Usual Name: | Automated cell counter and<br>Automated differential cell counter | | Device Class | Class II : Special Controls Guidance Document | | Classification Name: | Automated cell counter (§864.5200) and<br>Automated differential cell counter (§864.5220) | | Product Code: | GKZ | {1}------------------------------------------------ ### Substantial Equivalence: The ABX MICROS 60 is based on the same fundamental technology as the predicate devices cleared to market under K014203. The use of a predicate device to show the clinical capability was ABBOTT CD 4000 (K961439). ### Description: The ABX MICROS 60 device is a benchtop, clinical laboratory instruments that analyze in-vitro samples of whole blood specimens. The device operates in a complete blood count (CBC) and 3 part DIFF (LMG). As either a closed tube / closed system (CT/CS) or open tube/open system (OT/OS). The ABX MICROS 60 can evaluate 5,8, 16 or 18 hematology parameters depending on its internal configuration. ## Intended Use : The ABX MICROS 60 is a fully automated (microprocessor controlled) multi-parameter hematology analyzer intended for in in-vitro diagnostic use in the clinical laboratory environment. Compared to the previous 510k submissions, there is no change to the intended use # Determination of substantial equivalence : The ABX MICROS 60 the released software version 1.6 is substantially equivalent to the already cleared device MICROS 60 v1.4 (K992511) with respect to the indications for use, the hematological parameters for complete blood count and 3 part differential leucocyte count, and the principles of operation (fundamental scientific technology). # Discussion of Performance Data: This submission is carried out in accordance with appropriate indications given by the FDA guidelines. The data presented in this 510K Pre-market Notification demonstrate good precision in accordance with EPS-A (NCCLS guidelines) and is entirely acceptable for all parameters. No change to the claimed limits is being made in this submission. The linearity claim for the parameters WBC (0-100 x 103/ mm3), RBC (0 - 8.0 x 109/ mm"). HGB(0 - 26g/dl). HCT (0 - 80%). PLT with Hgb>2g/dl (0 - 2,200 x 10 / mm ) {2}------------------------------------------------ and PLT with Hgb<2g/dl (0 – 4000 x 103/ mm3) are entirely supported by the clinical data provided in this submission. Accuracy (Inter-procedural Correlation) demonstrated no evidence of significant bias between the MICROS 60 and the Abbott CD 4000 provided good correlation of R>0.95 for all parameters. No change to the claimed limits is being made in this submission. The clinical studies conclude that the safety and effectiveness of the devices is not compromised. #### Conclusions for non clinical and clinical tests : The clinical studies tests conclude that the safety and effectiveness of the device is not compromised. Clinical testing met all acceptance criteria. The device meets with the IEC 1010-1 standard of the International Electro-technical Commission on electrical equipment for measurement, control, and laboratory use. As well as the EN 61326 standard for Electromagnetic Compatibility. All clinical and non clinical tests show appropriate levels of safety and effectiveness. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling three stylized human profiles facing to the right, with flowing lines beneath them. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 APR 0 8 2003 Mr. Tim Lawton Regulatory Affairs Manager ABX Diagnostics Parc Euromedecine Rue du Caducee - BP 7290 34184 Montpellier cedex 4 FRANCE Re: k030799 > Trade/Device Name: ABX Micros 60 Hematology Analyzer Regulation Number: 21 CFR 864.5220 Regulation Name: Automated differential cell counter Regulatory Class: Class II Product Code: GKZ Dated: March 7, 2003 Received: March 13, 2003 Dear Mr. Lawton: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ Page 2 -- This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Butman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows a logo for ABX Horiba Diagnostics. The logo is a black square with the letters "ABX" in white, stacked on top of the word "HORIBA" in smaller white letters. Below the square, the word "DIAGNOSTICS" is written in a smaller font. The logo is simple and modern, and the use of black and white gives it a clean and professional look. # MICROS 60 Special 510(k): Device Modification #### INDICATIONS FOR USE STATEMENT 510(k) Number (if known): ABX MICROS 60 Hematology Analyzer Device Name: Indications For Use: The ABX MICROS 60 Hematology Analyzer is a fully automated (microprocessor controlled) hematology analyzer used for the in vitro diagnostic testing of whole blood specimens or blood cell concentrates. It operates in complete blood count (CBC) mode, (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CRFR 801.109) OR Over-The-Counter Use (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number K 030799 ABX Diagnostics (Horiba Group)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...