ACL TOP Family 50 Series (ACL TOP 750; ACL TOP 750 CTS; ACL TOP 750 LAS; ACL TOP 550 CTS; ACL TOP 350 CTS)
Applicant
Instrumentation Laboratory (IL) Co.
Product Code
GKP · Hematology
Decision Date
Aug 16, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 864.5400
Device Class
Class 2
Indications for Use
The ACL TOP Family 50 Series (ACL TOP 750; ACL TOP 750 CTS; ACL TOP 750 LAS; ACL TOP 550 CTS; ACL TOP 350 CTS) is intended for in vitro diagnostic use in clinical laboratories to perform coagulation and fibrinolysis testing on human citrated plasma samples.
Device Story
ACL TOP Family 50 Series: automated coagulation analyzers for clinical laboratories; process human citrated plasma samples to perform coagulation and fibrinolysis assays. Modification adds permission-based remote-control desktop sharing; enables remote delivery of software, OS patches, and test parameter updates. Device remains substantially equivalent to predicate K150877; fundamental scientific technology unchanged. System provides automated diagnostic results to clinicians for patient management.
Clinical Evidence
Bench testing only. Software verification and validation studies were conducted to confirm the safety and effectiveness of the new remote-access features and enhanced cybersecurity controls. No clinical performance data was required as the analytical performance remains unchanged from the predicate.
Technological Characteristics
Benchtop, fully automated, random access coagulation analyzer. Optical detection (405, 535, 671 nm). Fluidic handling includes aspiration, dispense, mixing, and temperature control. Connectivity: Cloud-based architecture for remote software/OS updates and desktop sharing. Software: Windows 10 (v6.5.3) with enhanced cybersecurity (MS BitLocker, Digital Signature, MS AppLocker).
Indications for Use
Indicated for in vitro diagnostic coagulation and fibrinolysis testing in clinical laboratories using human citrated plasma samples.
Regulatory Classification
Identification
A coagulation instrument is an automated or semiautomated device used to determine the onset of clot formation for in vitro coagulation studies.
Special Controls
*Classification.* Class II (special controls). A fibrometer or coagulation timer intended for use with a coagulation instrument is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
K231031 — ACL TOP Family 70 Series · Instrumentation Laboratory Company · Jun 21, 2023
K150877 — ACL TOP FAMILY 50 SERIES: ACL TOP 350 CTS; ACL TOP 550 CTS; ACL TOP 750; ACL TOP 750 CTS; ACL TOP 750 LAS · Instrumentation Laboratory CO · Dec 13, 2015
K160276 — ACL TOP (ACL TOP 700, ACL TOP 700 CTS, ACL TOP 700 LAS, ACL TOP 500 CTS, ACL TOP 300 CTS) · Instrumentation Laboratory CO · Mar 3, 2016
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# SPECIAL 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K242127
B Applicant
Instrumentation Laboratory (IL) Co.
C Proprietary and Established Names
ACL TOP Family 50 Series (ACL TOP 750; ACL TOP 750 CTS; ACL TOP 750 LAS; ACL TOP 550 CTS; ACL TOP 350 CTS)
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| GKP | Class II | 21 CFR 864.5400 - Coagulation Instrument | HE - Hematology |
## II Review Summary:
This 510(k) submission contains information/data on modifications made to the submitter's own CLASS II requiring 510(k). The following items are present and acceptable.
1. The name and 510(k) number of the SUBMITTER'S previously cleared device K150877.
2. Submitter's statement that the INDICATIONS FOR USE/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. The changes include:
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
- Added permission-based remote-control function for desktop sharing.
- Added functionality to remotely deliver software, OS updates (patches) and test parameter releases.
4. Comparison Information (i.e., similarities and differences) to the submitter's legally marketed predicate device including, labeling, intended use, and physical characteristics.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis.
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied.
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
K242127 - Page 2 of 2
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches