K023309 · Dade Behring, Inc. · GIZ · Nov 1, 2002 · Hematology
Device Facts
Record ID
K023309
Device Name
CONTROL PLASMA N
Applicant
Dade Behring, Inc.
Product Code
GIZ · Hematology
Decision Date
Nov 1, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5425
Device Class
Class 2
Indications for Use
Control Plasma N is assayed for use as an accuracy control for use of the following parameters in the normal range: Prothrombin time (PT); Activated partial thromboplastin time (aPTT); Thrombin time (TT)*; Batroxobin time*; Fibrinogen; Coagulation factors II, V, VII, VIII, VWfF IX, X, XI, XII and XIII **; Inhibitors: Antithrombin III, Protein C, Protein S, as -antiplasmin, C, inhibitor**: Plasminogen; Total complement activity**, Lupus anticoagulants. (* Not for BFT ™II analyzer, ** Not available in the U.S.)
Device Story
Control Plasma N is a lyophilized, HEPES-buffered human plasma control; used in clinical laboratories to monitor accuracy and precision of coagulation and fibrinolysis assays. Device serves as a quality control material; processed by mechanical or photo-optical coagulation analyzers. Healthcare providers compare analyzer results against assigned control values to validate system performance. Benefits include ensuring reliable diagnostic testing for coagulation disorders.
Clinical Evidence
Bench testing only. Stability studies demonstrated that reconstituted Control Plasma N maintains assigned values for 4 hours at +15 to +25°C and for 4 weeks when stored at -20 to -30°C (following a 2-hour thaw at +15 to +25°C).
Technological Characteristics
Lyophilized pooled human plasma; stabilized with HEPES buffer. Formulated for use with mechanical and photo-optical coagulation systems. No electronic components, software, or energy sources.
Indications for Use
Indicated for use as an accuracy control for coagulation and fibrinolysis tests in the normal range. Intended for professional laboratory use to monitor precision and accuracy of coagulation parameters including PT, aPTT, TT, Batroxobin time, Fibrinogen, coagulation factors, inhibitors, and Lupus anticoagulants.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
K023312 — CONTROL PLASMA P · Dade Behring, Inc. · Nov 1, 2002
K062306 — PLASMACON N, PLASMACON L-1, PLASMACON L-2 · R2 Diagnostics, Inc. · Mar 16, 2007
K981248 — SIGMA DIAGNOSTICS ACCUCLOT CONTROL I · Sigma Diagnostics, Inc. · Apr 29, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Dade Behring Inc. Control Plasma N 510(k) Modification
:
·
K023309
NOV 1 2002
# ATTACHMENT 5
510(k) Summary
{1}------------------------------------------------
# 510(k) Summary Control Plasma N
#### 1. Manufacturer's Name, Address, Telephone, and Contact Person, Date of Preparation:
| Manufacturer: | Dade Behring Marburg GmbH<br>Emil-von-Behring Str. 76<br>Marburg/Germany |
|----------------------|-----------------------------------------------------------------------------------------------------------------------|
| Contact Information: | Dade Behring Inc.<br>Glasgow Site<br>P.O. Box 6101<br>Newark, Delaware 19714<br>Attn: Donna Wolf<br>Tel: 302-631-0384 |
| Preparation date: | October 2, 2002 |
#### 2. Device Name/ Classification:
Control Plasma N / Multipurpose system for in vitro coagulation studies, Class II (864.5425)
#### 3. Identification of the Legally Marketed Device:
Control Plasma N (K00125)
#### 4. Device Description:
Control Plasma N is a Iyophilized control prepared from pooled human plasma. stabilized with HEPES buffer solution. It is an assayed control intended to monitor and evaluate the precision and accuracy of coagulation and fibrinolysis tests in the normal range.
#### 5. Device Intended Use:
Control Plasma N is assayed for use as an accuracy control for use of the following parameters in the normal range: Prothrombin time (PT); Activated partial thromboplastin time (aPTT); Thrombin time (TT)*; Batroxobin time*; Fibrinogen; Coagulation factors II, V, VII, VIII, VWfF IX, X, XI, XII and XIII **; Inhibitors: Antithrombin III, Protein C, Protein S, as -antiplasmin, C, inhibitor**: Plasminogen; Total complement activity**, Lupus anticoagulants. (* Not for BFT ™II analyzer, ** Not available in the U.S.)
### 6. Medical device to which equivalence is claimed and comparison information:
Control Plasma N (modified) is substantially equivalent in intended use to the Control Plasma N (K001256) currently marketed. Control Plasma N (modified). like the current Control Plasma N is intended for use as an assayed quality control material to monitor the accuracy and precision of various coagulation and fibrinolysis tests in the normal range using a variety of mechanical and photooptical coaqulation systems.
### 7. Device Performance Characteristics: Stability:
In duplicate determinations, reconstituted stability data met the acceptance criteria of recovering within the assigned values for the following claims: 4 hours at +15 to +25°C; 4 weeks at -20 to -30°C (2 hours after thawing at +15 to +25°C, stored for 4 weeks at -20 to -30°C).
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is centered in the image. The font is a serif font, and the text is black. The background is white.
od and Drug Administra 98 Gaither Road ockville MD 20850
Ms. Donna A. Wolf Senior Regulatory Affairs Specialist Dade Behring Inc. P.O. Box 6101 Newark, DE 19714
k023309 Re: Trade/Device Name: Control Plasma N Regulation Number: 21 CFR 864.5425 Regulation Name: Multipurpose System for In Vitro Coagulation Studies Regulatory Class: Class II Product Code: GIZ Dated: October 2, 2002 Received: October 3, 2002
Dear Ms. Wolf:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Control Plasma N. K023309 Control Plasma P. K023312 Response to Questions
# Indications for Use Statement
Device Name:
Control Plasma N
.
# Indications for Use:
Control Plasma N is assayed for use as an accuracy control of the following parameters in the normal range:
- Prothrombin time (PT) · 1.
- Activated partial thromboplastin time (aPTT) 2.
- Thrombin time (TT)* 3.
- 4. Batroxobin time*
- ട്. Fibrinogen
- Coagulation factors II, V, VII, VIII, VWF, IX, X, XI, and XII ర్.
- 7. Inhibitors: Antithrombin III, Protein C, Protein S, az-antiplasmin,
- Plasminogen 8.
- മ. Lupus anticoagulants
* Not for BFT™II analyzer
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
EDRH, Office of Device Evaluation (ODE) rence · 11 11
(Division)Sian-Off Division of Clinical Laboratory Devices
510(k) Number K 023309
Prescription Use v (Per 21 CFR 801.109)
Over-The-Counter-Use (Optional Format 1-2-96)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.