K092937 · Diagnostica Stago, Inc. · JIT · Aug 2, 2010 · Clinical Chemistry
Device Facts
Record ID
K092937
Device Name
STA-HYBRID HEP CALIBRATOR
Applicant
Diagnostica Stago, Inc.
Product Code
JIT · Clinical Chemistry
Decision Date
Aug 2, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Indications for Use
The STA® - Hybrid Hep Calibrator is a set of calibrator plasmas intended for use with analyzers of the STA® line suitable to these reagents, for the calibration of heparin (UFH and LMWH) activity assay by measuring the anti-Xa activity using the chromogenic method, STA® - Rotachrom® Heparin.
Device Story
STA® - Hybrid Hep Calibrator is a bundled kit of lyophilized human plasmas used to generate calibration curves for heparin (UFH and LMWH) activity assays. The kit contains five reagents: one heparin-free plasma, two plasmas with defined UFH concentrations, and two plasmas with defined LMWH concentrations. It is intended for use with the STA® line of IVD coagulation analyzers. The device operates by providing known reference values to calibrate the analyzer's chromogenic anti-Xa activity measurements. By establishing these curves, the analyzer can accurately quantify heparin levels in patient samples, assisting clinicians in monitoring anticoagulant therapy. The device is used in clinical laboratory settings by trained laboratory personnel.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Set of lyophilized human plasmas. Reagents include heparin-free plasma and plasmas with defined UFH or LMWH concentrations. Designed for use with STA® line IVD coagulation analyzers. Operates via chromogenic anti-Xa method. No software or electronic components.
Indications for Use
Indicated for use as a calibration set for heparin (UFH and LMWH) activity assays on STA® line analyzers using the chromogenic anti-Xa method. No specific patient population or contraindications are defined.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k092937
B. Purpose for Submission:
New device - combining two previously cleared devices into a single test kit.
C. Measurand:
Calibrator for heparin (unfractionated heparin and Low Molecular Weight Heparin) activity
D. Type of Test:
Quantitative
E. Applicant:
Diagnostica Stago, Inc.
F. Proprietary and Established Names:
STA® - Hybrid Hep Calibrator
G. Regulatory Information:
1. Regulation section:
21 CFR 864.7525 Heparin assay
21 CFR 862.1150, Calibrator
2. Classification:
Class II
3. Product code:
KFF, Assay, heparin
JIT, Calibrator, secondary
4. Panel:
81 Hematology
H. Intended Use:
1. Intended use(s):
The STA® - Hybrid Hep Calibrator is a set of calibrator plasmas intended for use with analyzers of the STA® line suitable to these reagents, for the calibration of heparin (UFH and LMWH) activity assay by measuring the anti-Xa activity using the chromogenic method, STA® - Rotachrom® Heparin.
2. Indication(s) for use:
Same as intended use
3. Special conditions for use statement(s):
Not applicable
4. Special instrument requirements:
Not applicable
I. Device Description:
STA® - Hybrid Hep Calibrator is a new device that bundles two of the company's previously FDA cleared devices, STA® - Hepanorm® H and STA® - Calibrator HBPM/LMWH into a single kit. The primary difference between the subject product and predicate devices is that it contains a set of calibration plasmas for calibration of assays of unfractionated heparin (UFH) and Low Molecular Weight Heparin
{1}
(LMWH) bundled together in a single kit.
J. Substantial Equivalence Information:
1. Predicate device name(s):
STA® - Calibrator HBPM/LMWH Kit
STA® - Hepanorm® H {formerly cleared as Hepanorm® Calibration Plasma Set}
2. Predicate 510(k) number(s):
k010350
k854762
3. Comparison with predicate:
| Similarities | | | |
| --- | --- | --- | --- |
| Item | Device | Predicate | Predicate |
| Device Name | STA® - Hybrid Hep Calibrator | STA® - Calibrator HBPM/LMWH | STA® - Hepanorm® H |
| Intended Use | The STA® - Hybrid Hep Calibrator is a set of calibrator plasmas intended for use with analyzers of the STA® line suitable with these reagents, for the calibration of heparin (UFH and LMWH) activity assay by measuring the anti-Xa activity using the chromogenic method, STA® - Rotachrom® Heparin. | The STA® - HBPM/LMWH Calibrator kit provides a set of calibrator plasmas intended for use by analyzers of the STA® brand name, for the calibration of low molecular weight heparin (LMWH) assays based on the anti-Xa principle. (STA® - Rotachrom® Heparin) | The STA® - Hepanorm® H kit provides a set of calibrator plasmas intended for use by analyzers of the STA® brand name, for the calibration of unfractionated heparins (UFH) assays based on the anti-Xa principle. (STA® - Rotachrom® Heparin, STA® - Staclot® Heparin). |
| Matrix | Lyophilized human plasmas reconstituted with distilled water, used in IVD analyzers. | Same | Same |
| Traceability of calibrator plasma | The LMWH levels are determined against the 01/608 International Standard for LMWH established in 2003. The UFH levels are determined against the 97/578 International Standard for UFH established in 1998. | Same | Same |
| Analytes | LMWH/ UFH | LMWH | UFH |
| Storage | 2-8 °C until Expiration date (24 months) | Same | Same |
| Format | Lyophilized | Same | Same |
| Open vial stability | 4 hours on board the analyzers of the STA® line | Same | Same |
There is no difference between the predicates and the new device.
K. Standard/Guidance Document Referenced (if applicable):
EP5-A2 Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition.
{2}
L. Test Principle:
The STA® - Hybrid Hep Calibrator is designed to operate utilizing the STA® product line of IVD coagulation analyzers for the purpose of creating calibration curves for assays of heparin by measuring the anti-Xa activity based on a chromogenic method.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Two samples containing different levels of heparin were tested using two different lots of STA® - Hybrid Hep Calibrator for the calibration. Each sample was tested for 22 operating days in two different laboratories. During each of the testing days, two separate runs were performed with alternatively two lots of STA® - Rotachrom® Heparin kit, on two analyzers, by a minimum of two operators.
Precision was estimated separately for each sample using the five components of precision described by the CLSI guidance: repeatability, between-run precision, within-day precision, between-day precision and within-laboratory precision (see results in tables 1-3 provided below).
Table 1 - Precision data for STA® - Hybrid Hep Calibrator in the Laboratory 1:
| Laboratory 1 | | | Sample 1 (UFH) | Sample 2 (LMWH) |
| --- | --- | --- | --- | --- |
| Theoretical concentration at which claim is made (spiked samples) | | IU/mL | 0.30 | 0.89 |
| Actual mean | | IU/mL | 0.27 | 0.89 |
| Point estimate of repeatability | SD | IU/mL | 0.01638 | 0.02459 |
| | CV | % | 6.1 | 2.8 |
| Point estimate of between-run | SD | IU/mL | 0.01264 | 0.03117 |
| | CV | % | 4.7 | 3.5 |
| Point estimate of within-day | SD | IU/mL | 0.01658 | 0.02174 |
| | CV | % | 6.1 | 2.4 |
| Point estimate of between-day | SD | IU/mL | 0.00013 | 0.00000 |
| | CV | % | 0.0 | 0.0 |
| Point estimate of within-laboratory precision | SD | IU/mL | 0.02358 | 0.03970 |
| | CV | % | 8.7 | 4.5 |
| Actual number of days involved in the experiment | | | 44 | 44 |
| Actual total number of runs | | | 88 | 88 |
| Total number of observations | | | 176 | 176 |
| Number of instruments used in evaluation | | | 2 | 2 |
{3}
Table 2 - Precision performance claims in Laboratory 2
| Laboratory 2 | | | Sample 1 (UFH) | Sample 2 (LMWH) |
| --- | --- | --- | --- | --- |
| Theoretical concentration at which claim is made (spiked samples) | | IU/mL | 0.30 | 0.89 |
| Actual mean | | IU/mL | 0.28 | 0.91 |
| Point estimate of repeatability | SD | IU/mL | 0.01344 | 0.02800 |
| | CV | % | 4.8 | 3.1 |
| Point estimate of between-run | SD | IU/mL | 0.01679 | 0.03322 |
| | CV | % | 6.0 | 3.7 |
| Point estimate of within-day | SD | IU/mL | 0.01525 | 0.02092 |
| | CV | % | 5.5 | 2.3 |
| Point estimate of between-day | SD | IU/mL | 0.00005 | 0.00000 |
| | CV | % | 0.0 | 0.0 |
| Point estimate of within-laboratory precision | SD | IU/mL | 0.02256 | 0.04345 |
| | CV | % | 8.1 | 4.8 |
| Actual number of days involved in the experiment | | | 44 | 44 |
| Actual total number of runs | | | 88 | 88 |
| Total number of observations | | | 176 | 176 |
| Number of instruments used in evaluation | | | 2 | 2 |
Table 3 - Site-to-site precision performance of STA - Hybrid Hep Calibrator
| Laboratory 1 + Laboratory 2 | | | Sample 1 (UFH) | Sample 2 (LMWH) |
| --- | --- | --- | --- | --- |
| Theoretical concentration at which claim is made (spiked samples) | | IU/mL | 0.30 | 0.89 |
| Actual mean | | IU/mL | 0.27 | 0.90 |
| Estimate of repeatability | SD | IU/mL | 0.01513 | 0.02638 |
| | CV | % | 5.5 | 2.9 |
| Estimate of between-run | SD | IU/mL | 0.01481 | 0.03223 |
| | CV | % | 5.4 | 3.6 |
| Estimate of within-day | SD | IU/mL | 0.01641 | 0.02305 |
| | CV | % | 6.0 | 2.6 |
| Estimate of between-day | SD | IU/mL | 0.00010 | 0.00000 |
| | CV | % | 0.0 | 0.0 |
| Inter-Laboratory variability | SD | IU/mL | 0.02347 | 0.04165 |
| | CV | % | 8.5 | 4.6 |
| Actual number of days involved in the experiment | | | 88 | 88 |
| Actual total number of runs | | | 176 | 176 |
| Total number of observations | | | 352 | 352 |
| Number of instruments used in evaluation | | | 4 | 4 |
{4}
b. Linearity/assay reportable range:
The entire range claimed for the assay kit is 0.1 to 2.0 IU/mL.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability: The UFH levels are determined against the 97/578 International Standard for UFH established in 1998. The LMWH levels are determined against the 01/608 International Standard for LMWH established in 2003.
Stability: Closed vial: In the lyophilized state the STA® - Hybrid Hep Calibrator plasmas remain stable until the expiration date printed on the box when stored at 2-8°C. This corresponds to 24 months after the date of manufacture.
Open vial: After reconstitution with distilled water, STA® - Hybrid Hep Calibrator reagents are stable for 4 hours on board the STA Satellite® (k082248), STA Compact® (k961579) and STA-R® (k983460) analyzers. The acceptance criteria for the 5 calibrator reagents are as follows:
Reagent 1: ≤0.10 IU/mL
Reagent 2: 0.20 – 0.35 IU/mL
Reagent 1: 0.40 – 0.60 IU/mL
Reagent 1: 0.70 - 1.10 anti-Xa IU/mL
Reagent 1: 1.60 – 2.00 anti-Xa IU/mL
## Calibrator Value Assignment:
For each level of calibrator the heparin (UFH or LMWH) value corresponds to the mean of a minimum of four determinations performed with four instruments, using at least two (2) different lots of STA®-Rotachrom® Heparin (K010455). The determinations are carried out by a minimum of two (2) laboratory technicians. The assay calibration is carried out with a pool of plasmas supplemented with the International standard for UFH or LMWH and/or other lots of STA® - Hybrid Hep Calibrator.
Table 4 - Determination of heparin level for a given lot of STA® - Hybrid Hep Calibrator.
| Laboratory technicians | Analyzer | STA® - Rotachrom® Heparin Lot | Heparin level (IU/ml) | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Reagent 1 | Reagent 2 | Reagent 3 | Reagent 4 | Reagent 5 |
| 1 | 1 | 061772 | 0.01 | 0.28 | 0.54 | 1.01 | 1.95 |
| 2 | 2 | 060871 | 0.01 | 0.27 | 0.49 | 0.97 | 1.96 |
| 1 | 1 | 061772 | 0.00 | 0.34 | 0.58 | 0.98 | 1.91 |
| 2 | 2 | 060871 | 0.00 | 0.30 | 0.56 | 0.93 | 1.88 |
| 3 | 3 | 062553 | / | / | / | 0.97 | 2.13 |
| 4 | 4 | 100973 | / | / | / | 0.98 | 2.00 |
| 3 | 3 | 062553 | / | / | / | 1.01 | 2.07 |
| 4 | 4 | 100973 | / | / | / | 0.95 | 1.97 |
| 4 | 5 | 100973 | / | 0.30 | 0.48 | / | / |
| 1 | 6 | 061772 | / | 0.29 | 0.48 | / | / |
| 4 | 5 | 100973 | -0.02 | 0.34 | 0.54 | / | / |
| 1 | 6 | 061172 | 0.00 | 0.33 | 0.57 | / | / |
| X | | | 0.00 | 0.31 | 0.53 | 0.97 | 1.98 |
{5}
Table 5 - After verification of the consistency of the results obtained, the heparin values assigned to the STA® -Hybrid Hep Calibrator are as follows:
| Reagent 1 - lyophilized human plasma free of heparin | Reagent 1: 0.00 IU/ml |
| --- | --- |
| Reagent 2 - lyophilized human plasma containing a well defined quantity of UFH | Reagent 2: 0.31 IU/ml |
| Reagent 3 - lyophilized human plasma containing a well defined quantity of UFH that is greater than that of reagent 2 | Reagent 3: 0.53 IU/ml |
| Reagent 4 - lyophilized human plasma containing a well defined quantity of LMWH | Reagent 4: 0.97 anti-Xa IU/ml |
| Reagent 5 - lyophilized human plasma containing a well defined quantity of LMWH that is greater than that of reagent 4 | Reagent 5: 1.98 anti-Xa IU/ml |
STA® - Hybrid Hep Calibrator lot used as an example is 100773
d. Detection limit: Not applicable
e. Analytical specificity: Not applicable
f. Assay cut-off: Not applicable
2. Comparison studies:
a. Method comparison with predicate device: Not applicable
b. Matrix comparison: Not applicable
3. Clinical studies:
a. Clinical Sensitivity: Not applicable
b. Clinical specificity: Not applicable
c. Other clinical supportive data (when a. and b. are not applicable): Not applicable
4. Clinical cut-off: Not applicable
5. Expected values/Reference range: The Expected Values are designated in the Assay Value Insert provided for each respective lot.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.