VENTANA MEDICAL SYSTEMS PATHWAY ANTI-C-KIT (9.7) PRIMARY ANTIBODY
P020055 · Ventana Medical Systems, Inc. · NKF · Aug 11, 2004 · Hematology
Device Facts
Record ID
P020055
Device Name
VENTANA MEDICAL SYSTEMS PATHWAY ANTI-C-KIT (9.7) PRIMARY ANTIBODY
Applicant
Ventana Medical Systems, Inc.
Product Code
NKF · Hematology
Decision Date
Aug 11, 2004
Decision
APPR
Regulation
21 CFR 864.1860
Device Class
Class 3
Indications for Use
This antibody is intended for in vitro diagnostic (IVD) use. Ventana® Medical Systems' PATHWAY Anti-c-KIT (9.7) Primary Antibody is intended for laboratory use, via light microscopy, for the qualitative detection of KIT protein in formalin-fixed, paraffin-embedded gastrointestinal stromal tumors (GISTs) using either an automated immunohistochemistry staining system or a manual assay. It is indicated as an aid in the diagnosis of GIST within the context of the patient's clinical history, tumor morphology, and other diagnostic tests evaluated by a qualified pathologist. It may be used after the diagnosis of GIST as an aid in the selection of GIST patients who may qualify for imatinib mesylate (Gleevec®/Glivec®) therapy.
Device Story
Rabbit monoclonal antibody (clone 9.7) detects KIT protein in formalin-fixed, paraffin-embedded tissue sections; used in clinical laboratories by pathologists. Input: tissue sections on microscope slides. Process: immunohistochemistry (IHC) staining via automated systems (Ventana NexES, BenchMark, BenchMark XT) or manual assay; utilizes indirect avidin-biotin system with DAB chromogenic substrate. Output: light microscopy visualization of DAB precipitate; score of 1+ or greater indicates positive KIT expression. Clinical utility: aids GIST diagnosis and identifies patients eligible for imatinib mesylate therapy; results interpreted alongside clinical history and tumor morphology.
Clinical Evidence
Two clinical studies compared PATHWAY Anti-c-KIT (9.7) to an investigational immunocytochemical antibody (ICA) used in Gleevec clinical trials. Study 1 (129 GIST cases) and Study 2 (63 non-GIST sarcomas, 13 GIST cases) evaluated agreement across three independent readers. Overall agreement exceeded 75% in all reader cohorts. Preclinical testing included antibody specificity (normal tissue and tumor arrays), purity (Western Blot), and reproducibility (inter/intra-run, lot-to-lot, manual vs. automated).
Technological Characteristics
Rabbit monoclonal IgG antibody (clone 9.7) in Tris buffer (pH 7.5) with 0.09% sodium azide. Detection via indirect avidin-biotin system (iVIEW DAB). Compatible with Ventana automated slide stainers or manual staining. Requires antigen enhancement, endogenous peroxidase blocking, and hematoxylin counterstain. Form factor: 5 ml prediluted reagent dispenser.
Indications for Use
Indicated for qualitative detection of KIT protein in formalin-fixed, paraffin-embedded gastrointestinal stromal tumors (GISTs) to aid in diagnosis and selection of patients for imatinib mesylate therapy. No specific age or gender restrictions.
Regulatory Classification
Identification
Immunohistochemistry test systems (IHC's) are in vitro diagnostic devices consisting of polyclonal or monoclonal antibodies labeled with directions for use and performance claims, which may be packaged with ancillary reagents in kits. Their intended use is to identify, by immunological techniques, antigens in tissues or cytologic specimens. Similar devices intended for use with flow cytometry devices are not considered IHC's.
Special Controls
(2) Class II (special control, guidance document: “FDA Guidance for Submission of Immunohistochemistry Applications to the FDA,” Center for Devices and Radiologic Health, 1998). These IHC's are intended for the detection and/or measurement of certain target analytes in order to provide prognostic or predictive data that are not directly confirmed by routine histopathologic internal and external control specimens. These IHC's provide the pathologist with information that is ordinarily reported as independent diagnostic information to the ordering clinician, and the claims associated with these data are widely accepted and supported by valid scientific evidence. Examples of class II IHC's are those intended for semiquantitative measurement of an analyte, such as hormone receptors in breast cancer.
Related Devices
P040011 — C-KIT PHARMDX · Dako North America, Inc. · Jun 27, 2005
K941514 — VENTANA ANTI-S100 PRIMARY ANTIBODY · Ventana Medical Systems, Inc. · Apr 25, 1997
P140025 — VENTANA ALK (D5F3) CDX ASSAY · Ventana Medical Systems, Inc. · Jun 12, 2015
P230018 — VENTANA CLDN18 (43-14A) RxDx Assay · Ventana Medical Systems, Inc. · Oct 18, 2024
P030044 — DAKOCYTOMATION EGFR PHARMDX · Agilent Technologies, Inc. · Feb 12, 2004
Submission Summary (Full Text)
{0}
8
# SUMMARY OF SAFETY AND EFFECTIVENESS DATA
## I. GENERAL INFORMATION
| Device Generic Name: | Anti-c-KIT Rabbit
Monoclonal Antibody for
Immunocytochemical
Staining |
| --- | --- |
| Device Trade Name: | Ventana® Medical Systems'
PATHWAY Anti-c-KIT (9.7)
Primary Antibody |
| Applicant Name and Address: | Ventana Medical Systems, Inc.
1910 Innovation Park Drive
Tucson, AZ 85705 |
| Premarket Approval Application Number: | P020055 |
| Date of Panel Recommendation: | None |
| Date of Notice of Approval To the Applicant: | August 11, 2004 |
## II. INDICATIONS FOR USE
This antibody is intended for in vitro diagnostic (IVD) use. Ventana® Medical Systems' PATHWAY Anti-c-KIT (9.7) Primary Antibody is intended for laboratory use, via light microscopy, for the qualitative detection of KIT protein in formalin-fixed, paraffin-embedded gastrointestinal stromal tumors (GISTs) using either an automated immunohistochemistry staining system or a manual assay. It is indicated as an aid in the diagnosis of GIST within the context of the patient's clinical history, tumor morphology, and other diagnostic tests evaluated by a qualified pathologist. It may be used after the diagnosis of GIST as an aid in the selection of GIST patients who may qualify for imatinib mesylate (Gleevec®/Glivec®) therapy.
## III. CONTRAINDICATIONS
None
## IV. WARNINGS AND PRECAUTIONS
Warnings and Precautions for use of the device are stated in the product labeling.
## V. DEVICE DESCRIPTION
Reagent Provided
PATHWAY Anti-c-KIT (9.7) Primary Antibody (clone 9.7) consists of one dispenser of c-KIT primary antibody (rabbit monoclonal) and contains approximately 5 ml (50
{1}
Page 2 of 16 Summary of Safety and Effectiveness Data P020055
test) of prediluted reagent. The dispenser contains approximately 25 µg of antibody in Tris buffer, pH 7.5, with carrier protein, non-ionic detergent, and 0.09% sodium azide as a preservative.
In addition, the following materials and reagents are necessary to perform the assay:
- Negative tissue control slide (normal colon tissue)
- Positive tissue control slide (GIST tissue)
- Microtome
- Microscope slides, silanized or polylysine-coated
- Drying oven capable of maintaining a temperature of 70°C ± 5°C.
- Bar code labels for Automated Slide Stainer use (Ventana Medical Systems, Inc. Cat. No. 451-801 for negative control and any one of Cat. No. 451-001 through 451-451 for PATHWAY Anti-c-KIT (9.7) Primary Antibody)
- Xylene (histological grade)
- Ethanol or reagent alcohol (histological grade)
- Deionized/distilled water
- Ventana Automated Slide Stainer:
- Ventana NexES® Slide Staining System, or
- Ventana BenchMark™ Slide Staining System, or
- Ventana BenchMark™ XT Slide Staining System
- Ventana Medical Systems iVIEW DAB Detection Kit
- Ventana APK Wash Solution Concentrate (10X) (NexES IHC automated slide stainers)
- Ventana EZ Prep™ Solution Concentrate (10X)
- Ventana Cell Conditioning 1 (CC1) Solution Pre-dilute (BenchMark and BenchMark XT automated slide stainers)
- Ventana Low Temperature Liquid Coverslip™ Solution Pre-dilute (NexES IHC automated slide stainers, or Ventana High Temperature Liquid Coverslip Solution Pre-dilute (BenchMark and BenchMark XT automated slide stainers)
- CONFIRM™ Negative Control Rabbit Ig
- Mounting Medium: Permanent Mounting Medium for use with DAB
- Cover Glass
- Light microscope (20-80X)
- Staining jars or baths
- Timer (capable of 2-10 minute intervals)
- Wash bottles
- Absorbent wipes
- Ventana Hematoxylin or Nuclear Fast Red Counterstain
- Ventana Bluing Reagent
- Decloaking Chamber, Digital Pressure Cooker (Biocare Medical) (NexES IIIC automated slide stainers)
- Tissue Tek™ slide rack (Biocare Medical)
2
{2}
Page 3 of 16 Summary of Safety and Effectiveness Data P020055
# Principle of Device Methodology
PATHWAY Anti-c-KIT (9.7) Primary Antibody is a rabbit monoclonal antibody intended to qualitatively detect the expression of KIT protein in formalin-fixed, paraffin-embedded normal and neoplastic tissue, using either a Ventana automated immunohistochemistry staining system or a manual assay. PATHWAY Anti-c-KIT (9.7) Primary Antibody is a rabbit monoclonal antibody which binds a site on the internal domain of the KIT oncoprotein in paraffin-embedded tissue sections. This reagent is supplied as an IgG antibody from cell culture supernatant diluted in a tris buffer solution to a concentration of approximately $3\ \mu\mathrm{g/mL}$. Sodium azide is added at $0.09\%$ as a preservative.
The antibody may be used on an automated immunohistochemistry system or manually. The presence of bound antibody is detected by DAB precipitation from the avidin-biotin based iView binding reagent (secondary antibody conjugated to biotin) to the monoclonal antibody. The presence of staining with a score of $1+$ or greater is considered to be positive, while no or faint staining below $1+$ is considered negative. A positive score is used in conjunction with other clinical and histopathological findings to make a diagnosis of GIST.
The remainder of the reagents necessary to obtain results is recommended in the "Materials and Reagents Needed But Not Provided" section of the product labeling. The system is used with antigen enhancement reagent, endogenous peroxidase blocker, biotinylated goat anti-rabbit IgG, streptavidin-horseradish peroxidase conjugate, DAB chromogenic substrate, copper enhancing reagent, and hematoxylin counterstain. The additional reagents are not part of the PATHWAY Anti-c-KIT Primary Antibody device, but are purchased separately.
The specific antibody is localized by a biotin-conjugated secondary antibody formulation that recognizes rabbit and mouse immunoglobulins. This step is followed by the addition of a streptavidin-enzyme conjugate that binds to the biotin present on the secondary antibody. The specific antibody-secondary antibody-streptavidin-enzyme complex is then visualized with a precipitating enzyme reaction product. The visualization of the bound PATHWAY Anti-c-KIT (9.7) Primary Antibody in tissue specimens is facilitated using Ventana's iVIEW DAB Detection, an indirect avidin-biotin system coupled to an enzyme. Visualization occurs through the localization of a DAB precipitate. Each step is incubated for a precise time and temperature. At the end of each incubation step, sections are washed to stop the reaction and remove unbound material that would hinder the desired reaction in subsequent steps. Results are interpreted using a light microscope and aid in the differential diagnosis of pathophysiological processes, which may or may not be associated with a particular antigen. Detailed instructions in the package insert for use of control materials, reagents and how to interpret their results are provided to insure adequate assay control.
10
{3}
Page 4 of 16 Summary of Safety and Effectiveness Data P020055
# VI. ALTERNATIVE PRACTICES AND PROCEDURES
There are no FDA approved or cleared in vitro diagnostic immunohistochemical assays for detection of KIT protein in normal and neoplastic tissue.
Analysis of c-KIT status could be performed via KIT gene mutation analysis of genomic tumor DNA, as was done for 121 subjects enrolled in the CSTIB2222 clinical trial.10 KIT gene mutational analysis is most commonly performed using polymerase chain reaction (PCR), for which there are no commercially available assays.
Analyte Specific Reagents for c-KIT immunohistochemistry and flow cytometry and c-KIT assays for research are currently marketed by several companies, some of which are listed in Table 1.
Table 1. Commercially Available Assays for c-KIT
| Company | Application | Intended Use |
| --- | --- | --- |
| BioGenex | IHC | Analyte Specific Reagent |
| Cell Marque | IHC | Analyte Specific Reagent |
| DakoCytomation | IHC | Analyte Specific Reagent |
| Biocare Medical | IHC | Research Use Only |
| BioGenex | IHC | Research Use Only |
| Chemicon | IHC | Research Use Only |
| Lab Vision/NeoMarkers | IHC | Research Use Only |
| Zymed Laboratories | IHC | Research Use Only |
| Beckman Coulter | Flow Cytometry | Analyte Specific Reagent |
| Becton, Dickinson and Company | Flow Cytometry | Analyte Specific Reagent |
| Caltag Laboratories | Flow Cytometry | Analyte Specific Reagent |
# VII. MARKETING HISTORY
PATHWAY® Anti-c-KIT (9.7) Primary Antibody has not been marketed. Marketing of a Class I Exempt version of the device under the CONFIRM™ brand name began in June 2004 in the United States, Europe, and Japan. The device has not been withdrawn from any market for any reason of safety and effectiveness.
# VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Ventana® Medical Systems' PATHWAY Anti-c-KIT (9.7) Primary Antibody is indicated as an aid in the selection of GIST patients who may qualify for imatinib mesylate (Gleevec®) therapy. Patients falsely assigned as positive following assessment would be considered eligible for treatment. Because the design of the clinical studies did not include treatment of patients with negative assay results, the risks or benefits of treatment in this patient population are unknown. The risks of Gleevec treatment to c-KIT-positive patients included dermatologic reactions, fluid retention and edema, gastrointestinal irritation and bleeding, anemia, neutropenia, and
11
{4}
Page 5 of 16 Summary of Safety and Effectiveness Data P020055
thrombocytopenia. Liver and kidney toxicity and immunosuppression may result from long-term use. Women of childbearing age should be advised to avoid becoming pregnant because of the potential for teratogenic effects.
Patients falsely assigned as negative might be less likely to receive the benefits of therapy with Gleevec.
The labeling recommends that the testing laboratory employ positive and negative tissue controls to reduce the potential for an erroneous test result. The strong and weakly positive internal controls of mast cells and interstitial cells of Cajal (ICC), present in the patient sample also aid in the control of the immunohistochemistry assay and reduce the potential for erroneous results.
# IX. SUMMARY OF PRECLINICAL STUDIES
Preclinical testing of the Ventana® PATHWAY Anti-c-KIT (9.7) Primary Antibody included clinical and analytical specificity, purity of reactivity, reproducibility, and stability/stress studies.
# 1) Antibody Specificity
a) Clinical Specificity/Tour of Normal Tissues Throughout the Body To describe the specificity of positive staining in normal tissues, twenty-eight (28) normal tissue types (of 30 recommended in FDA's special control for immunocytochemistry devices' Guidance for Submission of Immunohistochemistry Applications to the FDA) in tissue arrays containing 81 specimens were stained using the Ventana BenchMark automated staining instrument using Ventana PATHWAY (9.7) Primary Antibody and the polyclonal c-KIT antibody submitted in the original submission to examine c-KIT expression in these tissues. Parathyroid, pituitary, and mesothelial tissues were not included since there is no published evidence that c-KIT would be detected there. Bone marrow was also not included for unspecified reasons. Slides were blinded to observers and read by three independent qualified readers.
Acceptance criteria were that specimens stained with PATHWAY 9.7 monoclonal antibody must demonstrate $\geq 90\%$ agreement with those stained by the original polyclonal antibody.
Results: Mast cell staining was identified in colon, esophagus, kidney, lung, prostate, skin, small intestine, stomach, tonsil and uterus. Membrane staining of ductal cells was observed (unexpectedly) in breast.
Agreement (specimens positive by both antibodies plus specimens negative by both antibodies divided by total specimens) was $100\%$, $95\%$ CI = 96.5-100%. Seventy-eight specimens were not stained by either antibody, and three specimens (breast) were stained by both antibodies.
{5}
Page 6 of 16 Summary of Safety and Effectiveness Data P020055
Conclusions: Ventana PATHWAY (9.7) Primary Antibody performed acceptably on normal tissue samples and the acceptance criteria were met. The results for the Ventana PATHWAY (9.7) Primary Antibody are published in its package insert so that users will know what to expect with regard to c-KIT staining in normal tissues in the background of tumor specimens.
## b) Clinical Specificity/Tour of Tumors
To describe the specificity of positive staining in cancer tissues, a tissue array containing 49 tumor specimens including 18 different neoplastic tumor types (at least two specimens of each tumor type and including tumors relevant to the differential diagnosis of GIST, leiomyoma and sarcoma) were stained with PATHWAY 9.7 monoclonal and the original polyclonal antibodies using the Ventana BenchMark automated staining instrument. Slides were blind-labeled and read by three independent qualified readers.
Acceptance criteria were that PATHWAY 9.7 monoclonal must show $\geq 90\%$ agreement with the original polyclonal antibody.
**Results:** Agreement (specimens positive by both antibodies plus specimens negative by both antibodies divided by total specimens) was $97.9\%$. The single discrepancy was a breast cancer with weak membrane staining with the original polyclonal.
FDA originally requested 60 different tumor tissues representing 20 tumor types to be tested. For the original submission, 50 tissues were tested. Ventana believes that the use of 49 specimens representing 18 tumor types is adequate and not significantly different from what was previously tested.
**Conclusions:** Acceptance criteria were met, and the Ventana PATHWAY (9.7) Primary Antibody performed equivalently to the original polyclonal antibody. Scientific Reviewer, Dr. Elizabeth Mansfield concurred that the smaller number of tissues tested was acceptable because this was a comparison study to demonstrate equivalence of antibodies raised against the same antigen. The staining of these neoplasms was stated in the Specificity section of the Ventana PATHWAY Anti-c-KIT (9.7) Primary Antibody package insert, so that users will know what staining to expect from this product.
## c) Purity of Staining of the Ventana PATHWAY (9.7) Primary Antibody
To examine the purity of staining of the Ventana PATHWAY (9.7) Primary Antibody Sodium Dodecyl Sulfate (SDS)-Polyacrylamide Gel Electrophoresis (PAGE) and Western Blotting were employed. GIST 822 cell line lysate was blotted and probed with the Ventana PATHWAY (9.7) Primary Antibody, and with antibody plus immunogenic peptide, and examined with chemiluminescent detection. The antibody was expected to detect a doublet at $140-145\mathrm{kDa}$, and the antibody plus peptide was expected to show no or faint
13
{6}
Page 7 of 16 Summary of Safety and Effectiveness Data P020055
bands due to competition for binding by the peptide.
Two bands (doublet) at 140-145 kDa were observed and no other bands were detected when the monoclonal antibody was used. When the monoclonal plus peptide were used no staining was detected, indicating complete competition for binding.
Conclusion: Dr. Mansfield concluded this was not a strict measure of purity but of specificity (purity of staining). It was her judgement that the antibody is sufficiently pure, given that it is a monoclonal antibody, that no other demonstration of purity should be requested.
## d) Proof of c-KIT specificity of PATHWAY 9.7 Monoclonal Antibody – Lack of Cross-Reactivity with Closely Related Types of Tyrosine Kinase Receptors
To demonstrate antibody specificity and lack of cross-reactivity with related tyrosine kinase receptors, cell line lysates containing Platelet Derived Growth Factor-Alpha (PDGFRa), Macrophage Colony Stimulating Factor Receptor (c-FMS), and FMS-like Tyrosine Kinase (Flt-3) in addition to a lysate containing c-KIT were run on SDS-PAGE and Western Blotted electrophoretically. One of the blots was probed with Ventana PATHWAY (9.7) Primary Antibody, and demonstrated the expected doublet at 140-145 kilo Daltons in c-Kit-expressing cell line lysate, and no bands in lysates with other related receptors. Other blots were probed with antibodies specific for PDGFRa, Flt-3, and c-FMS. For each related receptor type, the staining demonstrated the presence of the related receptor type in the appropriate lysate, and the absence of staining with Ventana PATHWAY (9.7) Primary Antibody in the same lysates. The same test was also performed in the presence of c-KIT immunizing peptide.
Results: No reaction of the Ventana PATHWAY (9.7) Primary Antibody with any of the related tyrosine kinase receptors was noted, while the reactivity of the Ventana c-KIT (9.7) Monoclonal Antibody was completely blocked by c-KIT immunizing peptide.
Conclusion: This is a convincing demonstration of the specificity and lack of cross-reactivity of the Ventana PATHWAY (9.7) Primary Antibody to other related receptor types. The blocking of the reactivity by the immunizing peptide demonstrated that the antibody is specific for its immunogen, and thus detects c-KIT. No additional data were recommended to be requested by Dr. Mansfield.
## e) Specificity of PATHWAY (9.7) Monoclonal Antibody on cells
Three lots of Ventana PATHWAY (9.7) Primary Antibody were tested for specific staining on neoplastic (GIST) and normal (colon) tissue sections. Ability of the immunizing peptide to block antibody binding on tissue sections was also examined. Slides were stained on the Ventana BenchMark
14
{7}
Page 8 of 16 Summary of Safety and Effectiveness Data P020055
automated staining instrument, blind-labeled and read by three independent qualified readers.
Acceptance criteria were that staining must be ≤ 1+ in the presence of blocking peptide, and ≥ 3+ in the absence of blocking peptide.
The antibody lots performed identically. Staining of GIST and normal colon sections was 4+ intensity. Mast cells and Interstitial Cells of Cajal (ICC) were detected in normal colon. Staining of GIST and colon sections was intensity = 0 with the addition of the immunizing peptide.
**Results:** Acceptance criteria were met, and the Ventana PATHWAY (9.7) Primary Antibody stained GIST and colon cells as expected.
**Conclusion:** The Ventana PATHWAY (9.7) Primary Antibody reactivity is completely blocked by the peptide used to immunize the rabbit to create it. It, therefore, gives expected reactivity and specificity.
## 2) Reproducibility
To determine the performance characteristic of reproducibility for declaration in the product labeling, FDA requested that the following reproducibility studies be repeated with the new PATHWAY (9.7) Monoclonal Antibody on the Ventana BenchMark automated staining instrument, and by manual staining.
### a) Inter- and intra-run reproducibility on automated platform
For the determination of inter-run reproducibility on the automated platforms, 5 GIST specimens and one leiomyosarcoma specimen were assessed on Ventana automated staining instruments on 3 days, using 3 different instruments (NexES, BenchMark, and BenchMark XT), and the manual version of the assay. The same specimens were tested in nine replicates using the same three different instruments to determine intra-run reproducibility. All instruments used the same protocol (NexES 9.1 paraffin recipe iView DAB detection protocol). Deparaffinization and antigen enhancement were performed manually for the NexES instruments, and the manual protocol was performed according to the Ventana PATHWAY (9.7) Monoclonal Antibody package insert. Slides were blind-labeled and scored by three independent qualified readers.
Acceptance criteria were that 90% of samples demonstrate specific staining intensity scores within 0.5 scoring grade across automated and manual staining platforms. Non-specific staining must be ≤ 0.5 scoring grade or ≤ staining intensity of negative control for all samples.
**Results:** Acceptance criteria were met for all platforms for inter-run and intra-run reproducibility.
**Conclusion:** This was not an extensive reproducibility study. However, because no samples deviated from expectations except for one falsely positive
IS
{8}
Page 9 of 16 . Summary of Safety and Effectiveness Data P020055
leiomyosarcoma sample that was found to be sectioned too thick (had high non-specific staining), it is reasonable to expect that the assay is reproducible. No further studies were requested. The results were summarized under the Performance characteristics section of the package insert, so that users will know what to expect with regard to reproducibility performance.
## b. Lot-to-lot reproducibility
To determine lot-to-lot reproducibility, 4 lots of Ventana PATHWAY (9.7) Primary Antibody were used to stain a tissue array containing 5 GIST specimens and an array containing 60 normal and neoplastic tissues. The study director confirmed the diagnoses and expected staining of the 60 tissues. The GIST specimens were stained in duplicate with the four lots of Ventana PATHWAY (9.7) Monoclonal Antibody, with the original polyclonal antibody, and in singlicate with negative control antibody. The other tissue specimens were stained in singlicate with Ventana PATHWAY (9.7) Primary Antibody, original polyclonal antibody, and negative control antibody. The staining took place on the Ventana BenchMark automated staining instrument using the PATHWAY protocol. The method used for the negative control was not specified. Slides were stained and blind labeled, then scored by three independent qualified readers.
Acceptance criteria were that ≥ 90% of specimens demonstrate specific staining intensity scores within 0.5 stain intensity across all lots of Ventana PATHWAY (9.7) Primary Antibody, and within 1.0 stain intensity of the original polyclonal antibody. Non-specific staining must be ≤ 0.5 scoring grade or ≤ intensity of the negative control.
**Results:** A single reader on a single GIST specimen did not meet acceptance criteria, with a variation between lots of 1.0. Thus ~58/60 (96.7%) of cases conformed to acceptance criteria for GIST. For other normal and neoplastic tissues, all tissues were negative except normal breast (stained 1.75-2.0 with all lots), GIST, and seminoma, which are known to express c-KIT. Overall 98% of cases met acceptance criteria.
**Conclusion:** This study demonstrates adequate lot-to-lot reproducibility for the Ventana PATHWAY (9.7) Primary Antibody.
## c. Manual vs. automated reproducibility
To determine manual vs. automated reproducibility, a multi tissue block (MTB) containing 5 c-KIT positive GIST specimens, and a separate leiomyosarcoma specimen were used to compare staining reproducibility of three automated platforms (NexES, BenchMark, BenchMark XT) against manual staining. Single GIST MTB and leiomyosarcoma slides were stained with negative control reagent. The NexES protocol was used for both the Ventana PATHWAY (9.7) Primary Antibody and the original polyclonal c-KIT antibodies. Deparaffinization and antigen enhancement were performed manually for NexES and manual assays. The manual assay was as described
16
{9}
Page 10 of 16 Summary of Safety and Effectiveness Data P020055
in the package insert. Slides were blind-labeled and read by 3 independent qualified readers.
Acceptance criteria were that ≥ 90% of specimens have staining scores within 0.5 scoring grade across automated platforms and the manual assay. Non-specific staining must be ≤ 0.5 scoring grade or ≤ staining intensity of negative control.
**Results:** The acceptance criteria were met for all platforms and manual staining with both antibodies.
**Conclusion:** Platform-to-platform performance reproducibility is equivalent between the 3 different platforms and manual staining with Ventana PATHWAY (9.7) Primary Antibody.
## 3) Antibody Stability and Stress testing
To determine the stability of the Ventana PATHWAY Anti-c-KIT (9.7) Primary Antibody, 3 lots of product were stored at 2-8° C for 18 months. Testing intervals were day 0, week 1, and months 1, 2, 3, 5 and at appropriate intervals through month 18. In addition, to stress the product stability was analyzed after a mock shipping stress (2 days at 45° C followed by continuous storage at 2-8° C for 18 months) tested at day 3, week 1, and months 1, 2, 3, 6 and at appropriate intervals through month 18. Additional stress testing included tests performed at 37° C and 45° C continuous testing sampled at week 1, and months 1, 2, and 3, or until failure. Two freeze/thaw cycles were also tested from -20° C to room temperature at month 2. On-board open-bottles were tested stored at room temperature for one week after opening. Intended storage is at 2-8° C.
The tissue used for stability studies was paraffin-embedded GIST and normal colon tissue. Normal stomach tissue will be incorporated into the study in the future to increase the probability that the 2+ interstitial cells of Cajal (ICC) will be present. Testing of Ventana® Medical Systems' PATHWAY Anti-c-KIT (9.7) Primary Antibody included antigen enhancement, stained on the Ventana BenchMark® automated staining instrument and was visualized using Ventana's iView DAB detection kit. The slides were stained in triplicate for evaluation. Cells expected to be negative in the GIST and normal tissue slides are used as a negative tissue control to look for unintended cross-reactivity.
The test parameters measured were staining intensity of GIST tumors, mast cells and ICC at the time points indicated. Slides were masked and scored in comparison with rated reference slides of intensity levels of 0, 1, 2, 3, and 4+. Passing criteria were set as maintaining intensity when compared to day 0 slides for each lot. Should the sample time point fail, the assay shall be repeated once. A second failure constitutes a product failure.
17
{10}
Page 11 of 16 Summary of Safety and Effectiveness Data P020055
Results and Conclusions: Dating for this product has been set at 5 months from manufacture when stored at 2-8° C. Real-time stability and stress testing are ongoing.
4) Overall Conclusions for Preclinical Studies: The results of the preclinical testing performed on Ventana PATHWAY Anti-c-KIT (9.7) Primary Antibody demonstrate that this product is reproducible and is specific to c-KIT expression with preclinical performance characteristics appropriate to aid in the assessment of patients considered for Gleevec® (imatinib mesylate) therapy.
## X. SUMMARY OF CLINICAL STUDIES
Two studies were conducted with clinical samples to determine the agreement between PATHWAY Anti-c-KIT (9.7) Primary Antibody vs. the investigational immunocytochemical antibody (ICA) used in the Gleevec clinical trial using clinical cases involved in the differential diagnosis of GIST with known diagnoses.
## THE OBJECTIVE OF THE CLINICAL STUDIES
The objective of the clinical studies was to demonstrate acceptable agreement between the PATHWAY Anti-c-KIT (9.7) Primary Antibody and the investigational immunocytochemical antibody (ICA) used in the Gleevec clinical trial using clinical cases with known diagnoses that would be appropriate for the differential diagnosis of GIST. The acceptance criterion was 75% overall agreement defined as the total number of specimens that agreed with the ICA results divided by the total number of samples tested.
All clinical testing was done in-house at Ventana.
## STUDY POPULATIONS
Study #1. A multi-tissue array of 129 United States GIST cases obtained from Diomeda in Maryland. Thirty-five of the specimens were collected during the Gleevec clinical trial.
Study #2. Two micro arrays comprised of 63 non-GIST sarcoma cases and 13 GIST cases present in duplicate or triplicate obtained from Dr. Ray Tubbs at the Cleveland Clinic Foundation of Cleveland, Ohio.
## STUDY SUMMARIES
STUDY #1
A total of 129 GIST cases assembled into a tissue array were obtained from Diomeda (Maryland). Sections of each tissue array were stained with both of the antibodies (PATHWAY 9.7 and ICA Polyclonal), and with a negative control rabbit IgG antibody. The Ventana antibody and the control were used with the BenchMark instrument, and the ICA antibody and its control were run on the DAKO Autostainer. The PATHWAY and the original antibody were run using the same protocol on the
18
{11}
Page 12 of 16 Summary of Safety and Effectiveness Data P020055
BenchMark instrument. The ICA antibody was run on the DAKO Autostainer with the protocol recommended for the DAKO (ICA) antibody. The slides were blinded and read by three independent qualified readers.
## STUDY #2
A cohort of 63 non-GIST sarcoma cases and 13 GIST cases assembled into tissue arrays was obtained from Dr. Ray Tubbs of the Cleveland Clinic. The samples were obtained with IRB approval and the diagnostic information was provided by Dr. Tubbs. The data was reviewed by the study director to confirm the presence of cases representing non-GIST sarcomas and GIST. Sections of each tissue array were stained with both of the antibodies (PATHWAY 9.7 and ICA Polyclonal), and with a negative control rabbit IgG antibody. The Ventana antibody and the control were used with the BenchMark instrument, and the ICA antibody and the control were run on the DAKO Autostainer. The PATHWAY and the original antibody were run using the same protocol on the BenchMark instrument. The ICA antibody was run on the DAKO Autostainer with the protocol recommended for the DAKO (ICA) antibody. The slides were blinded and read by three independent qualified readers.
## RESULTS OF THE CLINICAL STUDIES
## STUDY #1. PATHWAY Anti-c-KIT (9.7) Primary Antibody vs. ICA Using a GIST Tissue Micro array with cases of known mutational status
A total of 129 cases was a part of the original GIST micro array. However, about one-fourth of the cores were variably lost to each of the three qualified readers. Since one of the reader's results was considerably different than the other two, the results are presented separately by qualified reader.
Qualified Reader #1
| | ICA Polyclonal | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| PATHWAY
Anti-c-KIT
(9.7) | Positive | 86 | 0 | 86 |
| | Negative | 1 | 9 | 10 |
| | Total | 87 | 9 | 96 |
Positive Agreement = 99% (86/87) with 95% Clopper-Pearson Lower Confidence Bound (LCB) = 95%
Negative Agreement = 100% (9/9) with 95% LCB = 72%
Overall Agreement = 99% (95/96)
19
{12}
Page 13 of 16 Summary of Safety and Effectiveness Data P020055
## Qualified Reader #2
| | ICA Polyclonal | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| PATHWAY
Anti-c-KIT
(9.7) | Positive | 67 | 0 | 67 |
| | Negative | 15 | 9 | 24 |
| | Total | 82 | 9 | 91 |
Positive Agreement = 82% (67/82) with 95% LCB = 73%
Negative Agreement = 100% (9/9) with 95% LCB = 72%
Overall Agreement = 83.5% (76/91)
## Qualified Reader #3
| | ICA Polyclonal | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| PATHWAY
Anti-c-KIT
(9.7) | Positive | 85 | 0 | 85 |
| | Negative | 1 | 9 | 10 |
| | Total | 86 | 9 | 95 |
Positive Agreement = 99% (85/86) with 95% LCB = 95%
Negative Agreement = 100% (9/9) with 95% LCB = 72%
Overall Agreement = 99% (94/95)
## Results and Conclusions:
Ten (10) of the 95 cases with a core evaluated by at least 2 of the qualified readers, did not demonstrate over-expression of c-KIT protein. These results demonstrate that the absence of c-KIT over expression alone cannot be used to eliminate a diagnosis of GIST, and reinforce the importance of considering the patient’s complete histopathologic profile and clinical history before making a diagnosis.
All of the qualified readers met study criterion of the ≥ 75% overall agreement and thus demonstrated acceptable agreement with the ICA. This data demonstrates that the PATHWAY Anti-c-KIT (9.7) Primary Antibody may be used to select GIST patients for therapy with the drug, Gleevec.
## STUDY 2. PATHWAY Anti-c-KIT (9.7) Primary Antibody vs. ICA Using GIST/Non-GIST Sarcoma Tissue Micro array
Micro arrays comprised of 63 non-GIST sarcoma cases and 13 GIST cases were also studied by 3 qualified readers. However, about one-fifth of cores was variably lost to the readers or contained no tumor. Two GIST and two non-GIST tumors had no evaluable cores. These results are presented separately by pathologist, also.
## Qualified Reader #1
| | ICA Polyclonal | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| PATHWAY
Anti-c-KIT | Positive | 11 | 0 | 11 |
| | Negative | 0 | 61 | 61 |
13
20
{13}
Page 14 of 16 Summary of Safety and Effectiveness Data P020055
| (9.7) | Total | 11 | 61 | 72 |
| --- | --- | --- | --- | --- |
Positive Agreement = 100% (11/11) with 95% LCB = 76%
Negative Agreement = 100% (61/61) with 95% LCB = 95%
Overall Agreement = 100% (72/72)
Qualified Reader #2
| | ICA Polyclonal | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| PATHWAY
Anti-c-KIT
(9.7) | Positive | 10 | 0 | 10 |
| | Negative | 4 | 54 | 58 |
| | Total | 14 | 54 | 68 |
Positive Agreement = 71% (10/14) with LCB = 46%
Negative Agreement = 100% (54/54) with LCB = 95%
Overall Agreement = 94% (64/68)
Qualified Reader #3
| | ICA Polyclonal | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| PATHWAY
Anti-c-KIT
(9.7) | Positive | 11 | 0 | 11 |
| | Negative | 2 | 58 | 60 |
| | Total | 13 | 58 | 71 |
Positive Agreement = 85% (11/13) with 95% LCB = 59%
Negative agreement = 100% (58/58) with 95% LCB = 95%
Overall Agreement = 97.2% (69/71)
## Results and Conclusions:
Two of the concordant positive specimens consisted of one spindle cell melanoma specimen and one synovial sarcoma specimen. C-KIT expression has been reported in approximately 20% of malignant melanomas (18). There is controversy in the literature regarding c-KIT expression in synovial sarcoma (32).
Four cases, all GISTs, had discrepant results according to at least one reader. For two of these cases, the discrepancy was noted with two of the readers. One reader had no discrepant results. These results demonstrate comparable performance between the PATHWAY Anti-c-KIT (9.7) Primary Antibody and the ICA and between the three different qualified readers.
All of the qualified readers met study criterion of the ≥ 75% overall agreement and thus demonstrated acceptable agreement with the ICA. This data demonstrates that the PATHWAY Anti-c-KIT (9.7) Primary Antibody may be used to select GIST patients for therapy with the drug, Gleevec.
21
{14}
Page 15 of 16 Summary of Safety and Effectiveness Data P020055
## Conclusion of the Clinical Studies
By identifying c-KIT positive GIST tissue, the Ventana PATHWAY Anti-c-KIT (9.7) Primary Antibody was demonstrated to be an effective aid in the assessment of patients being considered for therapy with Gleevec
## XI. OVERALL CONCLUSIONS DRAWN FROM ALL OF THE STUDIES
The results of the preclinical and clinical testing performed on Ventana PATHWAY Anti-c-KIT (9.7) Primary Antibody demonstrate that this product is reproducible and is specific to c-KIT expression with analytical and clinical performance characteristics appropriate to aid in the assessment of patients considered for Gleevec® (imatinib mesylate) therapy.
## Risk Benefit Analysis
The testing performed on Ventana PATHWAY Anti-c-KIT (9.7) Primary Antibody indicates that the assay performs consistently and that the assay results are clinically relevant in the assessment of patients considered for c-KIT targeted therapy.
Patients falsely assigned as c-KIT-positive following assessment would be considered eligible for treatment. Because the design of the clinical studies did not include treatment of patients with negative assay results, the risks or benefits of treatment in this patient population are unknown. The risks of Gleevec treatment to c-KIT-positive patients included dermatologic reactions, fluid retention and edema, gastrointestinal irritation and bleeding, anemia, neutropenia, and thrombocytopenia. Liver and kidney toxicity and immunosuppression may result from long-term use. Women of childbearing age should be advised to avoid becoming pregnant because of the potential for teratogenic effects.
False negative test results would potentially exclude the patient from treatment.
Based on the information in the studies provided, the FDA has concluded that the benefits of using the Ventana PATHWAY Anti-c-KIT (9.7) Primary Antibody for its intended use outweigh the risks associated with using it.
## Safety
The Ventana PATHWAY Anti-c-KIT (9.7) Primary Antibody is an in vitro diagnostic test and does not contact the patient. Instructions for the safe use of the product are included in the package insert.
## Effectiveness
The results of testing performed with the Ventana PATHWAY Anti-c-KIT (9.7) Primary Antibody indicated that the assay is effective in aiding in the assessment of patients considered for c-KIT targeted therapy.
CDRH has, therefore, concluded that the device is safe and effective for the stated indication.
15
{15}
Page 16 of 16 Summary of Safety and Effectiveness Data P020055
## X. PANEL RECOMMENDATIONS
In accordance with the provisions of section 515(c)(2) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Immunology Advisory Panel, an FDA advisory committee, for review and recommendation because the information in the PMA substantially duplicates information previously reviewed by this panel.
## XI. CDRH DECISION
CDRH issued an approval order for the applicant’s PATHWAY Anti-c-KIT (9.7) Primary Antibody on August 11, 2004.
The applicant’s manufacturing facility was inspected in June 4, 2004 and found to be in compliance with the Quality System Regulation (21 CFR 820).
## XII. APPROVAL SPECIFICATIONS
Directions for use: See labeling
Hazards to Health from Use of the Device: See Indications, Warnings and precautions in the labeling.
Postapproval Requirements and Restrictions: See approval order.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.