K982500 · Cook, Inc. · KNT · Jan 27, 1999 · Gastroenterology, Urology
Device Facts
Record ID
K982500
Device Name
CHAIT CECOSTOMY CATHETER
Applicant
Cook, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Jan 27, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Chait Cecostomy Catheter is used to instill fluids through a cecostomy into the colon to promote evacuation of the contents of the lower bowel through the anal opening and is intended to be an aid in the management of fecal incontinence. The opening and is and maintained in a percutaneously prepared opening such as a cecostomy .
Device Story
The Chait Cecostomy Catheter is a low-profile, 'trap door' catheter designed for antegrade colonic flushing. It is inserted into a pre-established percutaneous tract into the cecum. The device consists of a 10.2 French urethane catheter with five pigtail curves and three distal sideports, a trap door fitting that sits flush against the abdominal wall, and a metal access adapter cannula. When not in use, a plastic plug locks into the trap door fitting. Clinicians use the device to instill fluids into the colon to promote bowel evacuation. The device is supplied sterile for single-use. It benefits patients by providing a controlled access point for managing fecal incontinence.
Clinical Evidence
The submission references clinical trials, tensile strength tests, and biocompatibility tests to support the safety and performance of the device. No specific quantitative clinical metrics (e.g., p-values, sensitivity) are provided in the summary document.
Technological Characteristics
Materials: Urethane. Dimensions: 10.2 French diameter, 20cm length. Design: Single lumen, five pigtail curves, three distal sideports, trap door fitting with plastic plug, metal access adapter cannula. Sterilization: Sterile, single-use.
Indications for Use
Indicated for patients requiring management of fecal incontinence via antegrade colonic flushing through a percutaneously prepared cecostomy.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Ultrathane™ Biliary Drainage catheter with a COOK-Cope Type Loop (K851242)
Ultrathane™ Nephrostomy Catheter with a COOK-Cope Type Loop (K851242)
Submission Summary (Full Text)
{0}------------------------------------------------
Page 1 of 2 / K982500
113
510(k) Premarket Notification Chait Cecostomy Catheter COOK INCORPORATED
### Safety and Effectiveness Information
April Lavender, RAC Submitted By: Vice President, Regulatory Affairs COOK INCORPORATED 925 South Curry Pike P.O. Box 489 Bloomington, In 47402 (812) 339-2235 July 17, 1998
| Device: | Trade Name: Chait Cecostomy Catheter |
|-------------------------------|--------------------------------------|
| Proposed Classification Name: | Introduction, Drainage Catheter |
### Predicate Devices:
The Chait Cecostomy Catheter is similar in terms of intended use, materials of construction, and technological characteristics to the predicate devices reviewed, the Ultrathane™ Biliary Drainage catheter with a COOK-Cope Type Loop and the Ultrathane™ Nephrostomy Catheter with a COOK-Cope Type Loop.
## Device Description
The Chait Cecostomy Catheter is a "trap door", low profile catheter which facilitates antegrade colonic flushing. The device is supplied sterile and is intended for one time use. A tract must be established through the abdomen into the cecum prior to placement of the trap door catheter. Accessories used in conjunction with this device include, but are not limited to, a metal stiffener, a wire guide, a connecting tube, a syringe and saline solution.
The device consists of a cecostomy catheter with a trap door attachment and an access adapter. The 10.2 French catheter is made of urethane and is manufactured with five (total) pigtail curves. The distal end of the catheter includes three sideports which are evenly spaced inside the fifth curve. The endhole is 0.038 inches in diameter. The total length of the catheter is 20cm.
The trap door fitting is also made of urethane and lies flat against the abdomen wall. It incorporates a plastic plug which locks securely into place when the catheter is not being accessed.
The access adapter is a metal cannula. The distal end of the cannula is curved and can be
{1}------------------------------------------------
114
510(k) Premarket Notification Chait Cecostomy Catheter COOK INCORPORATED
attached to the trap door fitting as needed to fluids through the Chait Cecostomy Catheter and into the cecum. The proximal end of the cannula is designed with a large tapered fitting that will fit most taper tip syringes.
## Substantial Equivalence
COOK INCORPORATED currently markets two devices, the Ultrathane™ Nephrostomy Catheter with a COOK-Cope Type Loop and the Ultrathane™ Biliary Drainage Catheter with a COOK-Cope Type Loop, which are believed to be substantially equivalent to the catheter, subject of this submission. The Ultrathane Biliary Drainage Catheter is indicated for antegrade biliary drainage. The Ultrathane™ Nephrostomy Catheter is indicated for external urine drainage from the renal pelvis. Both devices are manufactured in either Ultrathane™ or polyurethane. These devices are classified as modified Preamendment devices and were also reviewed as substantially equivalent under K851242.
The Ultrathane™ Biliary Drainage Catheter is manufactured with a single lumen polyurethane tubing measuring 8.5, 10.2, 12.0, and 14.0 Fr. in outside diameter in a nominal 40cm length. The Ultrathane™ Nephrostomy Catheter is manufactured with a single lumen polyurethane tubing measuring 8.5, 10.2, 12.0, and 14.0 Fr. in outside diameter in a nominal 25cm length. The Chait Cecostomy Catheter will be manufactured with a single lumen urethane tubing measuring 10.2 Fr. in outside diameter and 20cm in length.
# Test Data
The Chait Cecostomy Catheter was subjected to the following tests to assure reliable design and performance under the specified testing parameters. These tests were comprised of:
- જુન Clinical Trials
- જુન Tensile Strength Tests
- જીન Biocompatibility Tests
The results of these tests provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its use as a catheter which facilitates antegrade colonic flushing.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or movement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 27 1999
COOK INCORPORATED c/o Ms. April Lavender, RAC Vice President Regulatory Affairs 925 South Curry Pike P.O Box 489 Bloomington, Indiana 47402
Re: K982500 Chait Cecostomy Catheter Set Dated: October 28, 1998 Received: October 29, 1998 Regulatory Class: II 21 CFR 876.5980/Procode: 78 KNT 21 CFR 876.5895/Procode: 78 EXD
Dear Ms. Lavender:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
.
Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear. Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
COOK INCORPORATED Response to Request for Additional Information K982500 Chait Cecostomy Catheter
510(k) Number (if known): K982500
Device Name:
Chait Cecostomy Catheter
Indications for Use:
The Chait Cecostomy Catheter is used to instill fluids through a cecostomy into the colon to promote evacuation of the contents of the lower bowel through the anal opening and is intended to be an aid in the management of fecal incontinence. The opening and is and maintained in a percutaneously prepared opening such as a cecostomy .
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
| | Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--|--------------------------------------------------------|--|
|--|--------------------------------------------------------|--|
| Prescription Use | <div> <svg height="15" width="15"> <path d="M4 10 L11 3" stroke="black" stroke-width="2"></path> </svg> </div> |
|-----------------------------|----------------------------------------------------------------------------------------------------------------|
| Use<br>(Per 21 CFR 801.109) | |
| | OR | |
|--|----|--|
|--|----|--|
| | Over-the-Counter |
|--|------------------|
|--|------------------|
| (Division Sign-Off) | |
|-----------------------------------------------------------------------|--|
| Division of Reproductive, Abdominal, ENT,<br>and Radiological Devices | |
| 510(k) Number | K982500/S001 |
|---------------|--------------|
|---------------|--------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.