K200646 · Actuated Medical, Inc. · KNT · Dec 4, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K200646
Device Name
TubeClear System
Applicant
Actuated Medical, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Dec 4, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TubeClear System is intended to clear occlusions/clogs in Feeding and Decompression Tubes.
Device Story
The TubeClear System is a mechanical device used to restore patency to occluded gastro-jejunostomy (GJ) feeding tubes. It consists of a reusable Control Box and a single-use Clearing Stem (Model GJ-1422). The operator coats the Clearing Stem with coconut oil and manually inserts it into the patient's feeding tube. The Control Box actuates the Clearing Stem wire in a back-and-forth motion to mechanically break up clogs. This process is performed by clinicians to avoid the risks and costs associated with surgical or endoscopic tube replacement. By restoring tube patency, the device allows for the timely resumption of enteral nutrition, hydration, and medication delivery, benefiting critically ill patients who rely on these tubes for therapy.
Clinical Evidence
Bench testing only. Testing included benchtop efficacy in artificial clogs, clearing stem removal forces, feeding tube integrity (including kinked tubes in ex vivo porcine tissue), and flow rate testing to ensure no residue accumulation from coconut oil. USP monograph testing and a literature-based toxicological risk assessment were conducted to evaluate the safety of coconut oil usage. No clinical trials were performed.
Technological Characteristics
System comprises a reusable Control Box (Model 101) and a single-use Clearing Stem (Model GJ-1422). The stem consists of a wire enclosed in a sheath, a depth-limiting component, and a magnet for connection to the control box. Actuation is mechanical (back-and-forth motion). Requires coconut oil as a lubricant. Certified to ISO 13485:2016 and MDSAP.
Indications for Use
Indicated for clearing occlusions/clogs in adults with G-Jet®, MIC®, and MIC-KEY® Gastro-jejunostomy tubes, size 14-22 Fr, with a jejunal length of 15-45 cm (6-18 in).
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 4, 2020
Actuated Medical, Inc. Douglas Dillon Director. Ouality Assurance & Regulatory Affairs 310 Rolling Ridge Dr. Bellefonte, PA 16823
Re: K200646 Trade/Device Name: TubeClear System Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal Tube and accessories Regulatory Class: Class II Product Code: KNT Dated: March 11, 2020 Received: November 5, 2020
#### Dear Douglas Dillon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200646
Device Name TubeClear System
Indications for Use (Describe)
TubeClear Clearing Stem Model GJ-1422 is indicated for use ONLY and SOLELY in clearing occlusions/clogs in adults that have the following Tube type and size (French and length).
+ GJ-1422, for G-Jet®, MIC®, and MIC-KEY® Gastro-jejunostomy tubes that are size 14-22 Fr and have a jejunal length of 15-45 cm (6-18 in).
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Actuated Medical. The logo has the word "ACTUATED" in gray and "MEDICAL" in orange. Below the logo is the text "Innovative motion + Positive outcomes" in gray.
K200646 Page 1 of 6
310 Rolling Ridge Drive Bellefonte, PA 16823 + p (814) 355-0003 + f (814) 355–1532 ActuatedMedical.com
#### Certified ...
- + ISO 13485:2016
- + Medical Device Single Audit Program (MDSAP)
- + Women's Business Enterprise (WBE)
- + Women-Owned Small Business (WOSB)
## 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92.
#### 510(k) Number: K200646
#### Applicant Information
| Date Prepared: | November 4, 2020 |
|-------------------|-----------------------------------------------------------------------------------------------------------------------|
| Name and Address: | Actuated Medical, Inc.<br>310 Rolling Ridge Drive<br>Bellefonte, PA 16823<br>Ph: (814) 355-0003<br>Fx: (814) 355-1532 |
| Contact Person: | Douglas Dillon<br>Director, Quality Assurance & Regulatory Affairs<br>Ph: (814) 355-0003 x107<br>Fax: (814) 355-1523 |
Device Information
| Trade/Device Name: | TubeClear System |
|--------------------|----------------------------------------|
| Common Name: | In Patient Tube Clearing System |
| Regulatory Class: | Class II |
| Classification: | KNT |
| Regulation Number: | 21 C.F.R. 876.5980 |
| Regulation Name: | Gastrointestinal Tubes and Accessories |
douglas.dillon@actuatedmedical.com
Image /page/3/Picture/15 description: The image is an abstract design featuring a combination of geometric and organic elements. The background consists of a checkerboard pattern in varying shades of orange, creating a textured effect. Overlaid on this pattern are wavy lines, also in orange, that meander across the image, adding a sense of movement and fluidity to the composition. The overall effect is a visually engaging and dynamic design.
{4}------------------------------------------------
## Predicate Device
The legally marketed device to which substantial equivalence is being claimed is as follows:
| 510(k) Number | Trade Name | Manufacturer |
|---------------|------------------|------------------------|
| K163092 | TubeClear System | Actuated Medical, Inc. |
#### Device Description
The TubeClear System is comprised of a reusable Control Box and a single use Clearing Stem. Control Box Model 101 is used to actuate all Clearing Stem models. The Clearing Stem for the predicate device is TubeClear Clearing Stem Model TC-0608 (TC-0608). The Clearing Stem for the subject device is TubeClear Clearing Stem Model GJ-1422 (GJ-1422). GJ-1422 is connected to the Control Box. The Operator then coats GJ-1422 with Coconut Oil and manually inserts it into the gastro-jejunostomy (GJ) tube (Tube) and directs the progression of the Clearing Stem inside the Tube. The Control Box actuates the Clearing Stem Wire. The Clearing Stem Wire Tip mechanically clears the occlusion to restore Tube patency. Sixteen (16) Clearing Stem Models are currently marketed to accommodate different types, sizes, and materials of feeding tubes. This submission adds a seventeenth (17th) Clearing Stem model (GJ-1422) for use in a specific set of gastro-jejunostomy (GJ) tubes. Use of TubeClear Clearing Stem Model GJ-1422 requires coating the Clearing Stem with Coconut Oil prior to introduction into a patient's feeding tube.
## Intended Use
The TubeClear System is intended to clear occlusions/clogs in Feeding and Decompression Tubes.
## Indications for Use of GJ-1422
TubeClear Clearing Stem Model GJ-1422 is indicated for use ONLY and SOLELY in clearing occlusions/clogs in adults that have the following Tube type and size (French and length).
· GJ-1422, for G-Jet®, MIC®, and MIC-KEY® Gastro-jejunostomy tubes that are size 14-22 Fr and have a jejunal length of 15-45 cm (6-18 in).
Image /page/4/Picture/12 description: The image is an abstract design featuring a combination of geometric and organic elements. The background is predominantly orange, with a gradient effect that transitions from a lighter shade to a darker one. A pattern of squares in alternating shades of orange creates a checkerboard effect in the upper right portion of the image. A wavy line in a lighter shade of orange is overlaid on top of the checkerboard pattern, adding a sense of movement and fluidity to the design.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a close-up shot of an orange background with a pattern of squares and wavy lines. The squares are arranged in a checkerboard pattern, with alternating shades of orange. The wavy lines are drawn over the squares, creating a sense of movement and texture. The overall effect is abstract and visually interesting.
#### Technological Characteristics
K200646 Page 3 of 6
The technological characteristics of TubeClear Clearing Stem Model GJ-1422 are mostly identical to TubeClear Clearing Stem Model TC-0608. The differences are:
- Coconut Oil: GJ-1422 requires the application of Coconut Oil to the Clearing ● Stem prior to insertion, while applying a water-based medical lubricant is optional for TubeClear Clearing Stem Model TC-0608.
- Indications: GJ-1422 is indicated for a specific set of gastro-iejunostomy (GJ) . tubes for which TC-0608 is not indicated.
- Stem Label: GJ-1422 has a blue Stem Label to visually differentiate it from TC-0608, which has a white Stem Label. This does not affect the safety nor the functionality of the device, as it is not a patient-contacting component and is only used to convey intellectual property information and to differentiate the models.
The design and function of all remaining components are identical. Both models consist of a Wire enclosed in a Sheath, a Depth Limiting Component, and a Magnet which connects to the reusable Control Box. Once the Clearing Stem is inserted into the Tube, the Control Box actuates the Clearing Stem in a backward and forward movement to physically break up clog contents in a Tube.
## Non-Clinical Performance Data
AMI reviewed the testing performed in support of the predicate model and repeated those tests where risk analysis suggested potential risk due to the use of the Coconut Oil and the different Indications for Use. Clearing Stem Removal Forces, Benchtop Efficacy, Feeding Tube Integrity and Flow Rate testing (following use of Clearing Stem with Coconut Oil) was performed. In addition, USP monograph testing (USP 42) and a literature-based toxicological risk assessment was conducted to provide evidence that the addition of the Coconut Oil does not present unacceptable adverse risks to the patients.
The following brief discussion of these tests and the conclusions drawn demonstrate the substantial equivalence of GJ-1422 to TC-0608 by establishing that the proposed Clearing Stem is substantially equivalent to TC-0608.
## Clearing Stem Removal Forces and Benchtop Efficacy:
Benchtop Efficacy: Worst-case artificial clogs were placed in series of indicated Tubes including 14Fr-18Fr, silicone, G-Jet® Button Low profile Gastric-Jejunal (GJ) enteral feeding tubes, 14Fr-22Fr, silicone, MIC-KEY* Gastric-Jejunal (GJ) enteral feeding tubes (low profile), and 16Fr-22Fr, silicone, MIC* Gastric-Jejunal (GJ) enteral feeding tubes
{6}------------------------------------------------
K200646
(conventional). In each case, the GJ-1422, when used in accordance with the Instructions Page 4 of 6 for Use (IFU) (i.e., Operator's Manual), effectively cleared the required occluded Tubes in accordance with the a priori acceptance criteria. The acceptance criteria were the same as that used for the legally marketed TC-0608 model, indicating that GJ-1422 is substantially equivalent to TC-0608 in clearing occluded Tubes.
Clearing Stem Removal Forces: In order to ensure that the different material and geometry of the Tubes indicated for GJ-1422 do not lead to unnecessarily high removal forces that might dislodge the Tube or exceed the component strength of GJJ-1422, the force required to remove the Clearing Stem from an artificially created clogged Tube was measured. In each case, the GJ-1422, when used in accordance with IFU, had lower forces than the a priori acceptance criteria. The acceptance criteria were the same as that used for the leqally marketed TC-0608 model, indicating that GJ-1422 is substantially equivalent to TC-0608.
## Feeding Tube Integrity:
Feeding tube integrity was tested against the different geometry and materials used in indicated Tubes for GJ-1422. The GJ tubes were folded and placed into ex vivo porcine tissue model forming a small radius curved Tube (i.e., kink). Using GJ-1422 as per the IFU, a simulated clog was first cleared, then GJ-1422 was advanced until the distal end of the Clearing Stem contacted the kink. The Operator moved the Clearing Stem backward and forward while in contact with the kink for 70 minutes. After the test, the Tubes were examined under a microscope for signs of damage. No signs of scratching or marring was observed. The Tube Integrity Tests support that GJ-1422 is substantially equivalent to TC-0608.
## Flow Rate:
In order to confirm that the repeated use of Coconut Oil does not lead to accumulated residue that neqatively affects the patency of the Tube, AMI completed worst-case Flow Rate testing demonstrating that repeated introduction of GJ-1422 coated in Coconut Oil into a feeding tube of the smallest French size and maximal length does not negatively alter the Tube's subsequent flow rate. The Tubes that were used with Coconut Oil showed a statistically significant higher amount of water collected, demonstrating that the Coconut Oil does not lead to accumulation of residue, nor does it reduce patency of the Tube. The Flow Rate Tests support that GJ-1422 is substantially equivalent to TC-0608.
## USP Monograph Testing and Toxicological Assessment:
GMP compliant USP method verification and release testing per USP 42 for a representative sample of the Coconut Oil was performed by an accredited test laboratory. Nineteen (19) of the twenty-one (21) analytes met the USP specification. At 0.7%, the fatty acid composition, carbon chain length 18:2 was below the specific of 1.0-3.0%.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image is an abstract design featuring a pattern of orange squares and squiggly lines. The squares are arranged in a grid-like fashion, with alternating shades of orange creating a subtle contrast. The squiggly lines are drawn in a lighter shade of orange and meander across the squares, adding a sense of movement and fluidity to the design. The overall effect is visually appealing and creates a sense of depth and texture.
K200646 Arsenic testing could not be verified and was therefore not tested as part of the release P&G 5 of 6 testing.
A literature review found eight (8) clinical trials exploring the effects of consuming coconut oil. For GJ-1422, in a worst-case scenario, ≤1 mL of Coconut Oil could be administered per week. In the reviewed studies, between 140 and 378 times the weekly dose of coconut oil was administered compared to the GJ-1422 worst-case scenario. In the reviewed studies, even with the higher dose regimen, no negative effects from coconut oil consumption was reported.
A risk-based assessment of the USP monograph testing and literature-based toxicological risk assessment in totality - especially considering the disparity in volumes between the clinical studies in the literature and that potentially administered through worst-case use with GJ-1422 - indicates that GJ-1422 does not present unacceptable adverse risks to the patient.
# Risk-benefit analysis:
For patients experiencing a clogged GJ tube, their prescribed enteral therapy (i.e., medication, hydration, and/or nutrition) is interrupted. Utilization of GJ-1422 provides critically ill patients with an effective alternative for clearing their clogged GJ tubes compared to the traditional Tube replacement. GJ tubes are placed either endoscopically, using topical analgesics and/or sedation, or surgically, under general anesthesia. Avoiding a Tube replacement can significantly reduce unnecessary risks for patients, whom may already be exhibiting gastrointestinal comorbidities. Successful clearings would render scheduled GJ tube replacement procedures due to clogging unnecessary, thus reducing costs for both the healthcare institution and patients. And would quickly restore their enteral therapy which is critical to the patient's health outcomes.
# Conclusions
Benchtop testing in artificially clogged GJ tubes demonstrate that GJ-1422 is substantially equivalent to the predicate device (TC-0608), which is currently marketed for the same intended use.
This submission claims substantial equivalence of GJ-1422 to TC-0608 based on repeated benchtop testing of Clearing Stem Removal Forces, Benchtop Efficacy, and Feeding Tube Integrity in clogged GJ Tubes. The methodology was largely identical between the two Clearing Stem types, and the differences in methodology do not affect the validity or applicability of the results. Both Clearing Stem models TC-0608 and GJ-1422 passed their respective a priori acceptance criteria for each test, which reflects clinically acceptable success rates. USP monograph testing and a literature-based toxicological risk assessment indicates that the use of Coconut Oil with GJ-1422 does not
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows an orange patterned background with the text "K200646" at the bottom. The pattern consists of orange squares and squiggly lines. The text is in black and is located on a white background.
present unacceptable adverse risks to the patient. A benefit risk analysis reveals that the Page 6 of 6 benefits of GJ-1422 greatly outweigh the risks, which are low in severity and infrequent in occurrence. For these reasons TubeClear Clearing Stem Model GJ-1422 is substantially equivalent to the TubeClear Clearing Stem Model TC-0608 (K163092).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.